1/44
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
Final digital image quality
is essentially a function of total exposure to the detector
Milliampere-second (mAs)
Digital image receptors can detect a wide range of radiation intensities (wide dynamic range)
Exposure errors can adversely affect the quality of the digital image
Di >+3
Excessive radiation exposure
Repeat only if relevant anatomy is clipped or "burned out"
Require immediate management follow-up
+1 to +3.0
Overexposure
Repeat only if relevant anatomy is clipped or "burned out"
-0.5 to +0.5
Target range
Less than -1.0
Underexposure
Consult radiologist for repeat
Less than -3.0
repeat
total amount of exposure
mas does not affect image brightness, but what does matter is the ______________ incident upon the detector
Kilovoltage-peak (kUp)
It affects the variation in radiation intensities exiting the patient and image contrast
should be lower compared to film-screen radiography
50-60 kVp
infant extremities
65-75 kVp
adult extremities
75-90 kVp
Bucky extremities
85-95 kVp
AP spine
85-100 kVp
Lateral Spine: cervical, Thoracic, Lumbar
110-130 kVp
chest
80-90 kVp
skull
higher kVp
-Smaller signal difference
-Narrower data range
-Display more anatomical data
-Lower dose
reduction in entrance skin exposure, and lower mAs
Higher kVp values are now recommended with digital detectors because ________________________
fixed kVp systems
The optimal kVp is the maximum kVp level that will produce images with appropriate contrast that are consistently within acceptance limits
double or halve mAs for every 5 cm of thickness
rule for fixed kvp systems
Variable kVp systems
The rule is 2 kVp is adjusted per cm of subject thickness, requiring a base of 30 kVp (40-50 kVp can be used) to be used
Fixed kVp systems
are better suited for use with digital image receptor
Collimation
Changes in beam restriction alter the amount of tissue irradiated, affecting the amount of exposure to the image receptor
imaging large anatomic areas
performing examinations without a grid]
high kVp
The effect of collimation is greater when:
Source-to-Image Receptor Distance (SID)
It is the distance between the source of the radiation and the image receptor, and it affects the amount of radiation reaching the patient
Divergence of the x-ray beam and the intensity of the radiation varies at different distances
Object-to-Image Receptor Distance (OID)
It is the distance between the object radiographed and the image receptor
decreased beam intensity
The presence of OID may result in
zero
Optimal recorded detail is achieved when the OID is
Grid
It is a device placed between the anatomic area and the image receptor to absorb scatter radiation exiting the patient
Thicker phosphors
have faster speed and absorbs more radiation
thinner phosphors
have slower speed and absorbs less radiation, producing sharper image quality
Fog
It is produced as a result of scatter radiation reaching the image receptor, and can be visualized on both a digital and a film image
Scatter radiation
can add unwanted exposure to the radiographic image as a result of compton interactions
thick, thin
A ____ part absorbs more radiation, whereas a _____ part transmits more radiation
increased
Maintaining the exposure to the image receptor when imaging thicker part requires an_____ mAs
increases. increases, decreases
As thickness ______ , scatter radiation_______, and radiographic contrast _______
Filtration
It is the removal of low-energy x-rays from the x-ray beam spectrum which does not contribute to image quality
The x-ray beam has increased energy and can increase the amount of scatter radiation reaching the image receptor
Automatic Exposure Control (AEC)
It provides x-rays with consistent receptor exposure, reducing chances of overexposure
It is used to control the amount of radiation reaching the receptors by terminating the length of exposure
ionization chambers
measures receptor exposure
Not necessarily applied in digital imaging but is essential to ALARA compliance and optimum image quality
kVp, mAs
increased collimation, decreased field size, decreased patient dose, decreases scatter radiation
collimation
increased SID, decreased intensity = increased mAs
SID
increased OID, increased magnification, decreased spatial resolution
OID
increased grid, decreased scatter radiation = increased mAs
Grid
increased thickness, increased scatter radiation, decreased contrast
thickness