1/83
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
Rate control medication
slow down the fast rates of abnormal rhythms
Rhythm control medication
suppress abnormal rhythms from occurring
Rate Control Medications:
beta blockers: metoprolol, atenolol they slow impulses passing through AV node
calcium channel blockers: verapamil, diltiazem
Rhythm Control Medications:
antiarrhythmic medications
potassium channel blockers: aminodarone, sotalol, tikosyn, dofetilide
sodium channel blockers: flecainide, rythmol, aka propafenone
Permanent Pacemaker (PPM)
pacemaker therapy is required when there is a problem with the heart’s conduction system that results in slow heartbeats
sick sinus syndrome
heart block
pacemakers work on demand
monitors intrinsic rhythm, paces when rate falls below programmed pacing rate
ICD Therapy
an ICD, or implantable cardioverter defibrillator is for people who are at risk of having a life threatening heart rhythm such as VT or VF
rapid irregular beats originate in the lower chamber of the heart and can lead to sudden cardiac arrest
indicated for patients who have a weak heart, ejection fraction <40% from a heart attack or muscle injury, at risk for having VT and VF
if the heart beats too slow, it will act as a pacemaker and pace the heart
if the device detects a ventricular arrhythmia, it will deliver pacing and shock therapy to convert the heart to a normal rhythm
CRT Therapy
cardiac resynchronization therapy or CRT delivers pacing stimuli to resynchronize the pumping function of the heart
left bundle branch block slows electrical conduction to left ventricle resulting in an out of sync heartbeat
pacing lead placed through coronary sinus into venous anatomy
CRT-D and CRT-P Devices
CRT-D provides CRT pacing and defibrillation
CRT-P provides CRT pacing only
Class I
benefits outweigh risk. procedure or treatment should be performed
Class IIa
benefits still outweigh risk, considered reasonable to perform procedure
Class IIb
benefits equal to risk. May consider performing procedure or treatment. Additional studies needed
Class III
risk is greater than or equal to the benefit. Procedure should not be performed and could be harmful to the patient
Voltage
is the force or the push that causes an electron to move through a circuit
is measured in volts, represented with the letter V
called pacing amplitude in the context of external stimuli delivered to the heart
Current
is the flow of electrons through a completed circuit
is measured in milliamps (mA) represented with the letter I
Impedance
is the opposition or the resistance to current
is measured in ohms, represented with the letter R
often referred to as resistance
Ohm’s Lam
represents the relationship between voltage, current, and impedance
V = IR formula states the voltage is equal to current in amps multiplied by resistance in ohms
if we know any two of these variables, we can solve for the third
Cathode
a negatively charged electrode that is the source of electrons entering an electrical device
tip electrode of pacing lead (-)
an electrode that is in contact with cardiac tissue
delivers electrical impulses in attempt to depolarize cardiac tissue
negatively charged when electrical current is flowing
Anode
a positively charged electrode by which electrons leave an electrical device
ring electrode of pacing lead (+)
an electrode that receives electrical impulses after depolarization of cardiac tissue
positively charged when electrical current is flowing
Conduction Pathway
electrons flow out the lead tip, through the blood and the tissue of the heart, and then return to the anode
Pacing Spike
a pacing spike is the graphical representation on the ECG of a delivered pacing impulse
delivered energy should cause myocardial stimulation or capture heart muscle
Myocardial Capture
Capture is a function of:
Amplitude:
the amplitude of the impulse must be large enough to cause depolarization (capture)
the amplitude of the impulse must be sufficient to provide an appropriate pacing safety margin (2:1) based on pacing threshold
Pulse width:
the pulse width must be long enough for depolarization to disperse to the surrounding tissue
Amplitude
the strength of the impulse expressed in volts
Pacing Threshold
the minimum amount of energy needed to consistently capture cardiac tissue
voltage over time for calculation
Pulse Width
the duration of the current flow expressed in milliseconds
Supernormal period
when a cell will respond to a weaker than normal stimulus
on the ECG, this correlates to the end of the downslope on the T wave up until the return to resting state
Effective refractory period
correlates to the absolute refractory period but also includes a small segment of phase three
stimulus may cause the cell to depolarize minimally, but not result in a propagated action potential
may see movement at the cellular level from an external stimulus, such as pacing impulse, but that impulse will not cause widespread depolarization of cardiac tissue
Capture
enough energy
tissue is not refractory
Loss of Capture
not enough energy
tissue is not refractory
Functional Non Capture
enough energy
tissue is refractory
Sensing
is the identifying of cardiac depolarization from an intracardiac electrogram
each signal represents cardiac depolarization occurring at a particular point in the heart
sensing can also be described as the difference in electrical potential between two points: the cathode and anode
Nearfield
are close range signals and have a sharp looking morphology
Farfield
are farther range signals and typically have a rounded morphology
Slew rate
is a measurement of an intrinsic signal’s slope or change over time
measured in volts per second
Frequency
is the amount an event occurs over time
measured in hertz
Sensitivity
is the ability for a device to sense an intrinsic electrical signal
is also a value that can be programmed based on a sensing measurement
measured in mV
more sensitive the device programming, the more it can see and the lower that programmed value will be
Unipolar Pacing System
cathode: tip electrode on lead
anode: pulse generator (can)
large antenna
Bipolar Pacing System
cathode: tip electrode on lead
anode: ring electrode on lead
small antenna
Electricity in Cardiac Pacing - Battery
a device that produces electricity; may have several primary or secondary cells arranged in parallel or series
Electricity in Cardiac Pacing - Capacitor
an electrical device characterized by its ability to store an electrical charge
Lead Breakage - Insulation Breach
decreases the resistance in the lead, which causes an increase in flow
Lead Breakage - Conductor or Coil Damage
increase in resistance in the lead, which cause a decrease in flow
Programmer
interfaces with the device, collects patient and device information, used for testing and programming
Lead
implanted through a patient’s venous anatomy and placed in either or both the right atrium and right ventricle; LV lead placed through CS into coronary venous anatomy for CRT devices
Pulse generator
metal device encased in titanium
the header is where the leads connect to the pulse generator
Circuit board
capacitors, resistors, accelerometers, reed switch, crystal oscillators, telemetry coil
Pacemaker Header
is part of device where the electrical connection is made between the lead and the generator
International Standard - 1 (IS-1)
lead enters the connecter block and is then held in place by a set screw
Passive fixation
leads utilize tines which embeds into the trabeculae
Active fixation
leads utilize a helix or a screw
extendable / retractable active fixation most common lead; fixation tool rotated clockwise to extend
Endocardial Leads
also known as transvenous leads
are placed in the heart through venous access
Epicardial Leads
also known as myocardial leads
are placed on the epicardium surgically
may be placed during open heart surgery or valve surgery, pediatric cases, and for patients with comprised venous access
Steroid Eluting Leads
today all pacing leads are steroid eluting
steroid collar built into lead
leads to decrease inflammation, which in turn decreases thresholds, and increases the device longevity
Lead Insulation - Silicone
reliable, flexible, and pliable
stickier and more difficult to navigate, prone to abrasions
Lead Insulation - Polyurethane
resistance to low friction in the blood, high abrasion resistance
stiffer, less pliable, history of failure
55D most current
Lead Insulation - Optim
combination of both insulation types thought to create a strong, lubricious, and abrasion resistant lead that is also flexible
Porous Tip Electrode
tip electrode is porous
porous design increases the overall surface area of the electrode, which increases contact with tissue and facilitates better sensing
keeps electrode small for pacing stimulation
Coaxial design
one insulated coil is surrounded by a second insulated coil
damage to the other coil would still provide an opportunity to reprogram the lead to unipolar
Coradial design
the coils are wrapped around each other and separately insulated
tend to be thinner
lead failure are more likely to involve a fracture of the cathode wire
Terminal pin
portion of the lead that connects to the header on the pulse generator
IS-1
5/6 mm unipolar
3.2 mm low profile
Selective Capture
if capturing only the specific part of the conduction system
also known as PURE capture
Non selective Capture
if capturing the conduction system and some of the surrounding tissue
also known as Fusion
SSI
single chamber pacing
single chamber sensing
inhibition in response to sensing
program to AAI or VVI
AAI
atrial pacing
atrial sensing
inhibition in response to sensing
no ventricular support
not commonly implanted in the US
AOO
atrial pacing
no sensing
asynchronous mode
magnet application to AAI pacemaker
VVI
ventricular pacing
ventricular sensing
inhibition in response to sensing
longstanding persistent or permanent AF with SVR
VOO
ventricular pacing
no sensing
asynchronous mode
magent application to VVI pacemaker
Single chamber Modes
AAI (R)
AOO
VVI (R)
VOO
DDD
sensing and pacing in both chambers, inhibition and triggering in response to sensing
intrinsic p wave: atrial inhibition
intrinsic r wave: ventricular inhibition
intrinsic p wave with no intrinsic r wave: ventricular triggering
DDI
sensing and pacing in both chambers, inhibition in response to sensing
no atrial tacking
DOO
dual chamber pacing, no sensing
asynchronous mode
magnet application to DDD pacemaker
VDD
pacing in the ventricle, sensing in both chambers, inhibition and tracking in response to sensing
single VDD lead
atrial electrode floating blood
VDI
pacing in the ventricle, sensing in both chambers, inhibition in response to sensing
no atrial tracking
Dual Chamber Modes
DDD (R)
DDI (R)
DOO
VDD (R)
VDI (R)
Refractory Periods
programmable
begins after every sensed or paced beat
device will not respond to intrinsic signals
typically 250 ms
Alert Period
not directly programmable
begins when refractory period expires
device will inhibit if it senses an intrinsic signal
Automatic intervals
pacing intervals initiated by paced beat
measured from pacing spike to pacing spike
Escape intervals
pacing interval initiated by sensed beat
measured from intrinsic beat to pacing spike
Rate Hysteresis
separately programmed escape interval
program hysteresis rate slower than programmed rate
Search Hysteresis
device periodically extends the pacing interval to search for intrinsic beats
no intrinsic beat: switches back to normal pacing interval
intrinsic beat: switches to hysteresis interval
Oversensing
device senses non-cardiac signals that it should ignore
sensitivity should set low enough to detect cardiac signals, but not low enough to detect extra cardiac signals
oversensing leads to underpacing - inappropriate inhibition from oversensing leads to underpacing
gaps in the paced rhythm: measure backwards to find where device over sensed
Undersensing
device ignores cardiac signals that should sense
set high enough to ignore extra cardiac signals, but not high enough to ignore cardiac signals
inappropriate inhibition from oversensing leads to overpacing
Loss of Capture and Causes
pacing impulses does not lead to cardiac depolarization
every pacing spike should lead to immediate depolarization
patient’s intrinsic rhythm will take over at slower rate
lead failure
lead dislodgment
perforation
most common: device output set too low
Loss of Output
no pacing impulses at all
most common cause: connector pin out of header
patient may be very symptomatic
requires invasive intervention
Fusion
pacemaker’s output pulse combines with intrinsic beat to create hybrid depolarization
ventricular lead in apex
intrinsic depolarization starting in His bundle
intrinsic signal starts just before pacing interval expires
device paces and signals meet halfway
mix between paced and intrinsic beat
Pseudofusion
pacemaker paces but has no effect on depolarization
ventricular lead in apex
intrinsic depolarization starting His bundle
intrinsic signal starts just before pacing interval expires
device paces after surrounding tissue has already depolarized
intrinsic morphology with pacing spike