Why is it important to leave space between items when loading steam sterilizer carts
to permit air removal from and steam penetration into packs
2
New cards
How often should biological monitoring of the hydrogen peroxide-ozone sterilizer process be performed
at least daily
3
New cards
Which microorganism is used to biologically monitor the hydrogen peroxide-ozone sterilizer
geobacillus stearothermophilus
4
New cards
Which of the following is compatible with hydrogen peroxide-ozone sterilization
A. natural rubber and latex
B. liquids
C. woven textiles
D. stainless steel instruments
D. stainless steel instruments
5
New cards
What are the parameters for low-temperature gas plasma sterilization
time, temperature, and hydrogen peroxide gas plasma
6
New cards
Which of the following should not be sterilized in a low-temperature gas plasma sterilizer
A. single-channel endoscopes
B. gauze sponges
C. cameras
D. light cords
B. gauze sponges
7
New cards
the Association for the Advancement of Medical Instrumentation recommends that rigid sterilization containers be biologically tested before purchase and annually to verify that
the sterilizer is able to sterilize the contents
8
New cards
unless stated by the sterilizer manufacturer, where should the ethylene oxide process challenge pack (biological test pack) be placed
geometric center of the load
9
New cards
when a positive biological indicator demonstrates a sterilization process failure, and the failure is a result of an operator error what should be done
correct the error and reprocess the load
10
New cards
All devices processed by low-temperature gas plasma must be
clean and dry
11
New cards
Which sterilizer requires only a special electrical outlet or installation
low-temperature gas plasma
12
New cards
which microorganism is most resistant to the steam sterilization process
geobacillus stearothermophilus
13
New cards
which sterilization method requires aeration after processing
ethylene oxide
14
New cards
define super-heating of steam
heat added to dry steam in the absence of water
15
New cards
a loaner instrument has been contaminated during the procedure and the OR wants to sterilize it in the immediate-use sterilizer. what's the first step to take
consult the manufacturer's written instructions for approved cleaning and sterilization methods and parameters
16
New cards
you should unload items from the steam sterilizer with
clean hands
17
New cards
when are sterile packs are most vulnerable to contamination
when the packages are still hot
18
New cards
where do you place biological indicators for testing rigid sterilization containers
as recommended by the manufacturer of the container
19
New cards
all loads containing implantable devices should be
monitored with a biological indicator
20
New cards
routine biological testing of steam sterilizers should be performed no less than
weekly
21
New cards
unless otherwise indicated by the biological indicator manufacturer, if a biological test is performed daily, how long is it necessary to incubate
a control biological indicator
once a day and if a new lot number is used
22
New cards
what is the recommended sterilizer setting for the Bowie-Dick
3 1/2 to 4 minutes at 270*F (132*C) dynamic-air-removal
23
New cards
what should be done if a Bowie-Dick test fails the first time
report to the supervisor for further instructions
24
New cards
type (class) 6 indicators are also called
emulating indicators
25
New cards
a Bowie-Dick test is considered which type of monitoring process
chemical
26
New cards
what is the correct action to take if an item needs to be corrected or removed from the sterilization load record
draw a line through the item and initial
27
New cards
a type (class) 5 indicator is called a(n)
integrating indicator
28
New cards
which monitoring method verifies the conditions within the load were adequate to kill bacterial spores
biological
29
New cards
one reason for adding chemicals to the boiler waterfeed is to
minimize contaminates
30
New cards
what is used to detect air leaks or insufficient air removal(vacuum)
Bowie-Dick test
31
New cards
stacking of rigid sterilization containers in a steam sterilizer is permitted when
the manufacturer of the container provides technical data showing that it is acceptable
32
New cards
how should basin sets be positioned for steam sterilization
tilted on their sides
33
New cards
where should textile packs be placed in a mixed load of textile and metal goods
on the top rack
34
New cards
if packages are moist at the end of the steam cycle what can occur
microorganisms can wick into the package causing possible contamination
35
New cards
for steam sterilization to occur steam must
make direct contact with all surfaces
36
New cards
which best describes a "wet load"
A. wet sets placed inside the sterilizer
B. a single package with moisture droplets
C. multiple packages in the same load found with excess moisture
D. excess moisture collecting in the drain line
C. multiple packages in the same load found with excess moisture
37
New cards
when might immediate-use steam sterilization be indicated
when a one-of-a-kind instrument is contaminated
38
New cards
what is the minimum exposure time for wrapped goods in a dynamic-air-removal steam cycle at 275*F (135*C) and 28 to 30 pounds per square inch (psi)
3 minutes
39
New cards
in a non-tabletop sterilizer, what does the gauge on the main steam line (usually located on the back of the sterilizer) tell you
the steam pressure as steam enters the jacket
40
New cards
define sterilization
the process by which all forms of microbial life, including spore, are destroyed