Photon output constancy daily tolerance
3%
Electron output constancy daily tolerance
3%
Localizing lasers should be tested ___ and within ___ of standard?
Daily, 2mm
Optical Distance Indicator (ODI) frequency & tolerance
Daily, 2mm
Jaw position indicators frequency & tolerance
Daily, 2mm
The door interlock should be checked:
Daily, functional
Audio/visual should be checked:
Daily, functional
Beam-on indicator should be checked:
Daily, functional
Electron output monthly tolerance:
2%
Photon output monthly tolerance:
2%
Light field/radiation (symmetric jaws) frequency & tolerance
monthly, 2 mm or 1% on a side
Light field/radiation (asymmetric jaws) frequency & tolerance
monthly, 1 mm or 1% on a side
Gantry/collimator angle indicators frequency & tolerance
monthly, 1 degree
Emergency off buttons should be checked:
monthly, functional
Table sag should be checked:
annually
Couch rotation to isocenter frequency & tolerance
annually, 2mm diameter
Gantry rotation to isocenter frequency & tolerance
annually, 2mm diameter
Photon/electron MU linearity is checked:
annually
CT number for water QA frequency & tolerance
daily; +/- 5 HU
MLC performance frequency & tolerance
weekly; OD<20%. Picket Fence Test
What is test used for MLC QA?
Picket Fence Test
Imaging QA frequency & tolerance
daily; 2mm for non SRS/SBRT; 1mm for SRS/SBRT
Laser guard lock should be checked:
monthly
Respiratory gating should be checked:
monthly
What is used to check light and radiation field agreement?
Ready pack film; done monthly
Mechanical isocenter (collimator) frequency & tolerance
annually, 2mm
Mechanical isocenter (Gantry) frequency & tolerance
annually; 1mm
Radiation isocenter frequency & tolerance
Star pattern test; checks gantry, collimator, and couch. annually, 2mm diameter