daily qc
PSP not used in 48 hrs are erased
Artifacts
Clean CR/DR detectors
Inspect for damage
Verify interface operation
Record image retakes
Check laser printers (brightness & contrast w no artifacts)
monthly qc
Erase PSP plates
Evaluate for dark noise
Acquires QC phantom images
Calibrate monitors
quarterly test
clean & inspect plates
preform tests: resolution, contrast, noisse, laser jitter, EI
annual qc
acceptance
review test data & service reports
pt exp records & dose creep
reports & improvements
what does laser printer test eval for
density contrast
what does Image erasure test do
confirms PSP is erased by white light within plate reader's erasure area
dark noise definition
residual signals from bg rad on unexposed cassette
what does dark noise do to an image
lowers contrast & brightness
linearity response def
measure pixel values produced by 3 diff exp at diff lvls of rad
relationship of linearity response
Should demon linear relationship w changes in lvls of exp to receptor
what is EI calibration for
monitoring pt exp for psp receptors
why is EI calibration needed
to receive accurate pt dose
what does laser jitter test
Transport system of PSP plate reader evaluated AND how laser beam interacts w PSP plates
what does laser jitter test display
wavy band perpendicular to the direction of travel
what does laser jitter test use
metal ruler/T-bar processed
what is measured in brightness uniformity & what is the variance
pixel values; 10%
what does high contrast res test use & why
Resolution grid used to see small lines & spat res at a high contrast lvl
what is used for distortion assessment
wire mesh screen; spacing used to determine distortion
what is anti-aliasing
When laser sampling freq of PSP plate reader & anti scatter grid freq match, & creates a moire artifact
how is linearity response shown
plotted pixel values to compare exp lvl of exp
how does image retention happen in DR
when DEL doesn't return to 0 after an exp
what is used to test blurring/stitching artifacts
fine wire mesh used
how is noise/low contrast res tested in DR
eval images of leeds low contrast test phantom
what does blurring/stitching artifacts assess
localize blurring/stitching issues w individual DELs
what qualities can be assessed when using the leeds phantom test