1
New cards

daily qc

PSP not used in 48 hrs are erased

Artifacts

Clean CR/DR detectors

Inspect for damage

Verify interface operation

Record image retakes

Check laser printers (brightness & contrast w no artifacts)

2
New cards

monthly qc

Erase PSP plates

Evaluate for dark noise

Acquires QC phantom images

Calibrate monitors

3
New cards

quarterly test

clean & inspect plates

preform tests: resolution, contrast, noisse, laser jitter, EI

4
New cards

annual qc

acceptance

review test data & service reports

pt exp records & dose creep

reports & improvements

5
New cards

what does laser printer test eval for

density contrast

6
New cards

what does Image erasure test do

confirms PSP is erased by white light within plate reader's erasure area

7
New cards

dark noise definition

residual signals from bg rad on unexposed cassette

8
New cards

what does dark noise do to an image

lowers contrast & brightness

9
New cards

linearity response def

measure pixel values produced by 3 diff exp at diff lvls of rad

10
New cards

relationship of linearity response

Should demon linear relationship w changes in lvls of exp to receptor

11
New cards

what is EI calibration for

monitoring pt exp for psp receptors

12
New cards

why is EI calibration needed

to receive accurate pt dose

13
New cards

what does laser jitter test

Transport system of PSP plate reader evaluated AND how laser beam interacts w PSP plates

14
New cards

what does laser jitter test display

wavy band perpendicular to the direction of travel

15
New cards

what does laser jitter test use

metal ruler/T-bar processed

16
New cards

what is measured in brightness uniformity & what is the variance

pixel values; 10%

17
New cards

what does high contrast res test use & why

Resolution grid used to see small lines & spat res at a high contrast lvl

18
New cards

what is used for distortion assessment

wire mesh screen; spacing used to determine distortion

19
New cards

what is anti-aliasing

When laser sampling freq of PSP plate reader & anti scatter grid freq match, & creates a moire artifact

20
New cards

how is linearity response shown

plotted pixel values to compare exp lvl of exp

21
New cards

how does image retention happen in DR

when DEL doesn't return to 0 after an exp

22
New cards

what is used to test blurring/stitching artifacts

fine wire mesh used

23
New cards

how is noise/low contrast res tested in DR

eval images of leeds low contrast test phantom

24
New cards

what does blurring/stitching artifacts assess

localize blurring/stitching issues w individual DELs

25
New cards

what qualities can be assessed when using the leeds phantom test

noise & low contrast res