1/44
Looks like no tags are added yet.
Gray top
(sodium fluoride / potassium oxalate)
Suitable for glucose in the lab exercise
-if can’t test in a timely manner, it stops glycolysis
use the validated tube for BUN, triglycerides and total cholesterol
Red top glucose specimen
let blood clot, centrifuge and separate serum from cells promptly
the course exercise specifies separation within 1 hour
4 hour delay risk?
a 4 hour delay before separation puts glucose, bilirubin, and electrolyte at risk
cells consume glucose and light can lower bilirubin
when to collect iron?
Collect iron studies in the morning when specified
serum iron varies with time of day
Hemolysis
releases intracellular potassium (K) and may increase measured K
a moderately hemolyzed red top specimen makes a high potassium suspect
Free hemoglobin
can also affect AST, LDH and photometric assays
use the assay specific hemolysis index and recollection criteria
Icterus
reflects increased bilirubin and gives serum or plasma a yellow-brown appearance
Bilirubin can interfere with some photometric reactions
Lipemia
often reflects high triglyceride rich lipoproteins
turbidity can affect photometry, and severe lipemia may produce falsely low sodium by indirect ISE
Errors across the testing process
Pre-examination
bilirubin specimen exposed to light during handling
(specimen too long or improper tube)
Errors across the testing process
Examination
A sample cup placed in the wrong analyzer position
(wrong position, reagent expired, improper dilution)
Errors across the testing process
Post-examination
Indirect bilirubin calculated incorrectly from the total and direct results
(manually calculated incorrectly, typed in wrong number when reporting results)
How to get indirect bilirubin
Total bilirubin - Direct bilirubin = indirect bilirubin
Glucose by hexokinase
The Axel hexokinase reaction forms glucose-6-phosphate from glucose and ATP
Glucose-6-phosphate dehydrogenase then reduces NAD+ to NADH
the increase in absorbance at 340 nm tracks glucose concentration
Course reagent blank limit is <0.500 absorbance
A blank of 0.850 exceeds the limit and is consistent with NADH already present in the reagent
resolve the failed blank and verify QC before releasing patient glucose results
Fasting plasma glucose (>8 h)
Normal: <100 mg/dL
Pre-diabetic: 100-125 mg/dL
Diabetes: >126 mg/dL
2 hour plasma glucose after a 75-g OGTT
Normal: <140 mg/dL
Pre-diabetic: 140-199 mg/dL
Diabetes: >200 mg/dL
HbA1c
Normal: <5.7%
Pre-diabetic: 5.7-6.4%
Diabetes: >6.5%
Random plasma glucose
>200 mg/dL supports diagnosis when classic symptoms or hyperglycemic crisis are present
There is no random glucose pre-diabtes cutoff
OGTT reportability
Check patient and specimen IDs, fasting status, glucose dose and actual collection times
ex.) Fasting glucose 112 mg/dL and 2-hour glucose 128 md/dL
the fasting result is in the pre-diabetes range and the 2 hour result is normal
fasting and 2-hour results measure different aspects of glucose regulation
review unexpected pairs for collection or labeling error
If specimen and analytical checks pass, a difference in diagnostic categories can be valid
Total cholesterol
desirable: <200
boderline high: 200-239
High: >240 mg/dL
LDL-C
Optimal: <100
Near or above optimal: 100-129
boderline high: 130-159
Very high: >190 md/dL
HDL-C
Coronary risk factor: <40 mg/dL
Protective in the ATP III risk factor count: >60 mg/dL
Triglycerides
Normal: <150 mg/dL
Boderline high: 150-199 mg/dL
High: 200-499 mg/dL
Very high: >500 mg/dL
Calculated LDL
Total cholesterol - HDL-C - (triglyceride ÷ 5)
at triglyceride >400 mg/dL withhold the Friedewald LDL-C
report valid measured components, including triglyceride
HDL measurement
The Axel direct HDL method suppresses reactions from non HDL lipoproteins, including LDL and VLDL, so the signal reflects HDL cholesterol
Ion selective electrodes
ISE
Measure a change in electrical potential relative to a reference electrode
The response depends on ion activity
Ion selective electrodes
The potassium selective membrane contains valinomycin, an ionophore that preferentially binds K+
Sodium and chloride often change together with changes in water balance
true acid-base disturbances can make them diverge
If multiple results show an unexpected electrolyte pattern, review QC, calibration and the relevant electrode
Anion gap
AG = Na+ - (Cl + HCO3)
ex.) Na 140, Cl 104, HCO3 24 mmol/L gives 140 - (104 + 24) = 12
mmol/L
if the sodium electrode reads 5 mmol/L too high, the same sample gives an anion gap of 17 mmol/L
an upward shift across many patients can indicate positive sodium bias
check the sodium electrode, QC and calibration
Reportability and critical values
A mildly abnormal result can be reportable when the specimen, QC and analytical checks pass
Total protein includes albumin and globulins
(an albumin results higher than total protein calls for investigation)
→ repeat a critical result to confirm (our lab)
→ follow the lab policy and communicate promptly
ex.) glucose ≤50 or ≥400 mg/dL, K⁺ ≤2.5 or ≥6.0 mmol/L, Na⁺ ≤120 or ≥160 mmol/L.
total calcium ≤6.5 or ≥13.0 mg/dL and magnesium ≤1.0 or ≥9.0 mg/dL
Creatinine by kinetic Jaffé
In alkaline solution, creatinine reacts with picrate to produce a colored complex.
The Axcel kinetic method measures the early change in absorbance. The lab exercise describes an early window of about 80 seconds.
The early read reduces the contribution of slower-reacting non-creatinine chromogens.
Residual chemical and spectral interference is still possible
Azotemia and glomerulonephritis
Pre-renal
(reduced perfusion): BUN rises disproportionately to creatinine.
BUN:creatinine is often ≥20:1, for example 48:2 = 24:1
Azotemia and glomerulonephritis
Renal (intrinsic renal injury)
Damage with kidney
BUN and creatinine rise together
A ratio around 10–15:1 is typical
Azotemia and glomerulonephritis
Post renal
obstruction to urinary tract obstruction
⬆ creatinine
⬆BUN
ratio varies with timing and severity
Acute glomerulonephritis
Creatinine and BUN may rise
Urinalysis can show hemauria, dysmorphic RBCs, RBC casts and proteinuria
24 hour creatinine clearance
Use the same concentration units for U and P
enter urine volume in mL, time in minutes and BSA in m2
Course reference intervals (mL/min/1.73 m²): males 85–125, females 75–115

Bilirubin and enzyme methods
Axcel uses DMSO, which acts as an accelerator in the total bilirubin diazo assay. It makes unconjugated bilirubin accessible to the aqueous diazo reagent.
Albumin on the Axcel is measured by binding to bromocresol green (BCG). The color signal tracks albumin concentration.
Enzyme activity assays use excess substrate to establish zero-order kinetics with respect to substrate.
Under those conditions, the initial reaction rate reflects enzyme activity and is independent of small changes in substrate concentration
Jendrassik grof solubilizing agents
Caffeine benzoate solution
(caffeine, sodium benzoate, and sodium acetate)
Evelyn malloy solubilizing agents
Methanol
(methyl alcohol)
DMSO solubilizing agents
Dimethyl sulfoxide
Jaundice patterns: Prehepatic
0.2 - 0.8 mg/dL
Indirect bilirubin ↑
Due to excess RBC destruction (hemolysis or ineffective erythropoiesis)
-urine bilirubin is generally absent
Ex.) hemolytic disease, hereditary spherocytosis, sickle cell anemia
Jaundice patterns: Hepatic
0.2 - 0.8 mg/dL (Indirect)
0.0 - 0.2 mg/dL (Direct
Direct and/or indirect ↑
Due to impaired uptake, conjugation, or secretion
-hepatocellular injury such as viral hepatitis
⬆ALT
⬆AST
Ex.) hepatitis, cirrhosis
Impaired: Gilbert (lack enzyme to conjugate bilirubin effectively)
Conjugation: Crigler - Najjar (liver cant conjugate bilirubin)
Secretion: Dubin johnson (liver can’t pump conjuagted out of cells)
Jaundice patterns: Post hepatic
Direct bilirubin ↑
Due to blockage preventing elimination of conjugated bilirubin
⬆ALP
⬆⬆⬆GGT
-dark urine and pale stool
Ex.) gallstones, tumors, cholestasis
Conjugated bilirubin
Can appear in urine due to being soluble
Low AST
Course scenario: total bilirubin, direct bilirubin and ALT are markedly elevated, while AST is low or normal.
Consider substrate depletion as one analytical explanation. With rapid substrate consumption, the late reaction rate can appear falsely low.
Review the reaction trace and analyzer flags. Check AST QC and reagent status, and rerun or dilute as the method directs.
A discordant patient panel alone cannot establish reagent failure
Thyroid specimen
Use approved
red-top serum
Serum separator tube
lithium heparin specimen
❌EDTA