69 Terms

1
New cards
Direct digital radiography (DR)
Converts x-ray photons directly into electrical charge without first producing light.
2
New cards
Direct DR detector material
Amorphous selenium (a-Se).
3
New cards
Direct DR sequence
X-rays โ†’ a-Se โ†’ electrical charge โ†’ TFT readout โ†’ ADC โ†’ digital image data.
4
New cards
Indirect digital radiography (DR)
Converts x-ray photons into light and then converts the light into electrical charge.
5
New cards
Indirect DR detector materials
A scintillator converts x-rays to light, and amorphous silicon (a-Si) photodiodes convert the light to electrical charge.
6
New cards
Scintillator
Converts x-ray photons into visible light.
7
New cards
Photodiode
Converts light produced by the scintillator into electrical charge.
8
New cards
Thin-film transistor (TFT)
Functions as an electronic switch that releases stored charge from each detector element for readout.
9
New cards
Analog-to-digital converter (ADC)
Converts the analog electrical signal into digital numerical data.
10
New cards
Indirect DR sequence
X-rays โ†’ scintillator โ†’ a-Si photodiode โ†’ electrical charge โ†’ TFT readout โ†’ ADC.
11
New cards
Computed radiography (CR)
Uses a photostimulable phosphor (PSP) imaging plate to store the latent image.
12
New cards
Photostimulable phosphor (PSP) plate
Stores energy from the x-ray exposure until stimulated by a laser in the CR reader.
13
New cards
CR reader laser
Stimulates the PSP plate, causing stored energy to be released as light.
14
New cards
Photomultiplier tube (PMT)
Detects and amplifies light released from the PSP plate and converts it into an electrical signal.
15
New cards
CR readout sequence
PSP plate โ†’ laser stimulation โ†’ light emission โ†’ PMT โ†’ electrical signal โ†’ ADC โ†’ digital image.
16
New cards
Pixel
The smallest picture element in a digital image.
17
New cards
Matrix
An arrangement of pixels in rows and columns that forms the digital image.
18
New cards
Field of view (FOV)
The physical area of anatomy represented in the digital image.
19
New cards

Pixel size = FOV รท matrix size.

Pixel size formula

20
New cards

Pixel size decreases and spatial resolution improves.

What happens when matrix size increases while FOV remains constant?

21
New cards

Spatial resolution improves.

What happens when pixel size decreases?

22
New cards
Spatial resolution
The ability of an imaging system to display small, closely spaced structures as separate objects.
23
New cards
Spatial frequency
Describes the size and spacing of image detail and is commonly expressed in line pairs per millimeter (lp/mm).
24
New cards
High spatial frequency
Represents small, closely spaced structures or fine detail.
25
New cards
Low spatial frequency
Represents larger structures or coarse detail.
26
New cards

The largest lp/mm value.

Which lp/mm value represents the highest spatial frequency?

27
New cards

Modulation transfer function (MTF)

Measures how accurately an imaging system transfers object detail and contrast into the recorded image.

28
New cards
High MTF
Indicates better preservation of image detail and better spatial resolution.
29
New cards
MTF of 1.0
Represents perfect transfer of object contrast and detail.
30
New cards

MTF generally decreases.

What happens to MTF as spatial frequency increases?

31
New cards
Detective quantum efficiency (DQE)
Measures how efficiently an image receptor converts incoming x-ray information into a useful image signal.
32
New cards
High DQE
Indicates greater detector efficiency and the ability to maintain image quality with less receptor exposure.
33
New cards

DQE generally decreases.

What happens to DQE as spatial frequency increases?

34
New cards
Signal-to-noise ratio (SNR)
Compares useful image signal with unwanted image noise.
35
New cards
High SNR
Indicates strong useful signal compared with noise and produces better image quality.
36
New cards
Quantum noise (quantum mottle)
Grainy image appearance caused by an insufficient number of x-ray photons reaching the detector.
37
New cards

Increase receptor exposure appropriately, most directly by increasing mAs.

How can quantum noise be reduced?

38
New cards

SNR increases because more photons reach the detector and quantum noise decreases.

What happens to SNR when mAs increases?

39
New cards
Scatter radiation
X-ray photons that change direction after interacting with matter and may reach the image receptor as unwanted exposure.
40
New cards
Scatter versus quantum noise
Scatter is unwanted redirected radiation; quantum noise results from too few photons reaching the detector.
41
New cards
Dynamic range
The range of exposure values that an image receptor is capable of detecting and recording.
42
New cards
Exposure latitude
The range of exposure techniques that can still produce an acceptable diagnostic image.
43
New cards
Dynamic range versus exposure latitude
Dynamic range describes what the detector can record; exposure latitude describes how much exposure can vary and still produce a usable image.
44
New cards
Exposure indicator (EI)
A numerical value that reflects the amount of radiation reaching the image receptor.
45
New cards
Deviation index (DI)
Indicates how far the actual detector exposure differs from the target exposure.
46
New cards
DI of 0
Actual detector exposure equals the target exposure.
47
New cards
Positive DI
Detector exposure is greater than the target exposure.
48
New cards
Negative DI
Detector exposure is less than the target exposure.
49
New cards
Histogram
A graph showing the distribution of pixel or exposure values within a digital image.
50
New cards
Histogram X-axis
Represents pixel values or signal intensity values from low to high.
51
New cards
Histogram Y-axis
Represents the frequency, or number of pixels, at each value.
52
New cards
Histogram analysis
The computer evaluates the distribution of pixel values to identify the useful exposure data for image processing.
53
New cards
Automatic rescaling
Adjusts acquired pixel values so the displayed image has appropriate brightness despite variations in receptor exposure.
54
New cards
Look-up table (LUT)
Assigns processed pixel values to displayed gray shades and determines the appearance of image contrast.
55
New cards
Histogram versus rescaling versus LUT
Histogram analyzes the pixel values; rescaling adjusts the values; the LUT determines how those values are displayed as gray shades.
56
New cards
Digital image brightness
Primarily controlled by image processing and display settings rather than directly by mAs.
57
New cards
mAs in digital imaging
Primarily affects photon quantity, detector exposure, SNR, quantum noise, and patient dose.
58
New cards

mAs.

A digital image appears grainy because too few photons reached the receptor. Which technical factor should be increased?

59
New cards

DQE.

Which detector characteristic describes how efficiently x-ray photons are converted into useful image information?

60
New cards

Spatial resolution.

Which image-quality characteristic describes the ability to display small structures separately?

61
New cards

MTF.

Which image-quality characteristic describes how well object detail is transferred to the image?

62
New cards

Dynamic range.

Which term describes the range of exposure values a detector can record?

63
New cards

Exposure latitude.

Which term describes the range of techniques that still produces an acceptable image?

64
New cards

Y-axis.

A histogram question asks how many pixels contain a specific signal intensity. Which axis represents this?

65
New cards

X-axis.

A histogram question asks for the range of signal intensity values. Which axis represents this?

66
New cards

LUT.

Which processing function maps pixel values to displayed gray shades?

67
New cards

Automatic rescaling.

Which processing function helps maintain relatively consistent displayed brightness despite exposure differences?

68
New cards

Detector underexposure relative to the target.

A negative deviation index indicates what?

69
New cards

Detector overexposure relative to the target.

A positive deviation index indicates what?