1/22
Looks like no tags are added yet.
I
Highly resistant borosilicate;
buffered or non-buffered aqueous parenterals
Powdered glass and surface glass
II
Treated soda-lime;
acidic and neutral aqueous parenterals
Water attack test
III
Soda-lime;
non-aqueous or dry-solid parenterals
Powdered glass and surface glass
Old powdered-glass test:
crushed glass;
titration with 0.02 N H2SO4.
Listed limits:
Type I 1.0 mL;
Type III 8.5 mL.
Water-attack test
: inner surface of Type II glass; 0.02 N H2Surface-glass test: inner surface of Type I and III; 0.1 N HCl.S
Surface-glass test:
inner surface of Type I and III;
0.1 N HCl.
NMT 10% at any wavelength 290–450 nm.
Colored glass light transmission:
NMT 0.1 µg/g (0.1 ppm).
Arsenic in Type I/II glass:
albino mice; .
In vivo:
systemic injection—parenteral containers,
In vivo:
albino rabbit;
intracutaneous and eye-irritation tests
In vivo:
rabbit
implantation—material in direct tissue contact,
In vitro:
agar diffusion,
direct contact, and
elution tests;
sample includes elastomeric material; cell/tissue culture.
Optical microscopy:
individual particles seen;
reliable but tedious and provides a 2D image.
projected-area circle diameter
Diameter descriptors:
Feret
Martin
(longest distance between tangents),
(bisects particle)
Sieving:
official and fastest;
sieve number = square openings per linear inch;
higher sieve number means smaller particle size.
Sieving limitations
: sample ≥25 g;
not suitable for oily or cohesive materials.
Dry mechanical sieving for >75 µm;
air entrainment for smaller particles.
Endpoint:
weight change NMT 5%;
f <5% of sample weight is on a sieve, % weight change NMT 20%.
mass / untapped volume
mass / tapped volume
Bulk density =
Tapped density
Tapped volume procedure:
tap 10, 500, then 1,250 times;
if V500 − V1250 ≤ 2 mL, use V1250.
Angle of repose
: tan θ = h/r;
measured with fixed funnel.
Smaller angle indicates better flow
Flow through an orifice:
mass/time;
for free-flowing powders only.
(CI) = [(Vi − Vf)/Vi] × 100 or
[(ρtapped − ρbulk)/ρtapped] × 100.
Carr’s index
(HR) = Vi/Vf or ρtapped/ρbulk.
Hausner ratio