[10a] Name the 4 main non-blood diagnostic samples listed
Urine, bone marrow, tissue samples (biopsy), plus emerging ones (saliva, nasopharyngeal swab, breath, tears, stool)
[10a] Urine is used to diagnose
Prostate cancer, urinary tract infection (UTI), drug monitoring
[10a] Bone marrow is
Semi-solid tissue; primary site of new blood cell production. Used for childhood cancers, myeloid and lymphoid leukemia
[10a] Tissue samples are collected by
Biopsy; used for cancer diagnostics
[10a] Saliva is used for
Oral health and respiratory infections
[10a] Nasopharyngeal swab is used for
Respiratory infections (COVID-19, influenza), whooping cough, diphtheria
[10a] Breath samples are used for
Cancer biomarkers, e.g. lung cancer
[10a] Tears are used for
Eye health diagnostics
[10a] Stool is used for
Microbiome and cancer diagnostics
[10a] Whole blood is collected mainly by
Venipuncture or finger prick
[10a] List whole blood biomarkers
Blood cells, circulating proteins, cytokines, miRNA, cfRNA, cfDNA, pathogens, exosomes, circulating tumor cells (CTCs), antibodies, other molecules (lactate, glucose, drugs)
[10a] cfDNA stands for
Circulating free DNA
[10a] cfRNA stands for
Circulating free RNA
[10a] CTC stands for
Circulating tumor cell
[10b] Why do we collect blood?
Diagnostics, therapy monitoring, blood transfusion
[10b] Venipuncture is
Process of obtaining intravenous access to collect venous blood
[10b] Vacutainer tubes are coated with
Additives (anticoagulants) to prevent clotting, e.g. EDTA and heparin
[10b] Anticoagulant examples
EDTA and heparin
[10b] Purpose of anticoagulants in collection tubes
Prevent blood clotting
[10b] Rotisserie is used to
Mix blood samples after collection in a vacutainer tube
[10b] New venipuncture technology
Automated robotic arm with multiple degrees of freedom (DOF) to control the needle, plus near-infrared (NIR) imaging to find veins
[10b] NIR imaging in automated venipuncture is used to
Find/locate veins
[10b] Finger prick is
Pricking a finger with a lancet to get a small quantity of capillary blood (also called finger stick)
[10b] Best fingers for finger prick
3rd and 4th fingers
[10b] Where NOT to prick the finger
Do not use the tip or center; avoid the side of the finger
[10b] Avoid pricking a finger that is
Cold or blue, swollen, scarred, or covered with a rash
[10b] How to warm a cold finger before pricking
Wrap a warm cloth or rub the finger with your hand
[10b] Finger prick site is cleaned with
Alcohol swab; let it air dry completely before pricking
[10b] After the finger prick, what do you do with the first drop?
Wipe it off with dry clean cotton wool
[10b] Finger prick requires what kind of lancet
Sterile lancet with full skin penetration for adequate blood flow
[10b] Neonatal heel prick volume
About 300-500 uL using multiple blood drops
[10b] Main way to collect blood from neonates
Heel prick (others: venous or arterial needle sticks)
[10b] Finger prick vs venous: parameter with the highest correlation listed (PCT)
PCT (plateletcrit) r = 0.999; MCV also very high (0.996); total WBC 0.991
[10b] Finger prick vs venous: basophil percentage correlation
Essentially none (r = 0.002)
[10b] Finger prick vs venous CD4 study
110 HIV-infected Malawian adults; bias 6.6 cells/uL (limits -50.7 to 63.7) for absolute counts; 0.71% bias for %CD4/lymphocyte
[10b] Whole blood composition: plasma fraction
55% of whole blood
[10b] Whole blood: platelets per uL and size
3 x 10^5 per uL; 1-2 um
[10b] Whole blood: RBC per uL and size (per slide)
5 x 10^6 per uL; 4-5 um
[10b] Whole blood: WBC per uL and size
7 x 10^3 per uL; 7-14 um
[10b] Whole blood: granulocytes + monocytes per uL and size
4 x 10^3 per uL; 9-14 um
[10b] Whole blood: lymphocytes per uL and size
3 x 10^3 per uL; 7-8 um
[10b] Most abundant cell in blood
RBCs (about 5 x 10^6 per uL)
[10b] Where are proteins, cytokines, antibodies, cfDNA, cfRNA, microRNA found?
Blood plasma
[10b] Two main divisions of whole blood biomarkers
Cell biomarkers and plasma biomarkers
[10b] Cell biomarkers include
Cell counts, surface protein expression, morphological changes, mechanical properties, cellular contents (e.g. pro-viral load), circulating tumor cells
[10b] Plasma biomarkers include
Proteins, cytokines, antibodies, cfDNA, cfRNA, microRNAs and others
[10b] If you want to count RBCs, you need
Blood dilution (RBCs are highly abundant)
[10b] If you want WBCs and their subtypes, you need
RBC lysis (to remove highly abundant RBCs)
[10b] If you want plasma biomarkers, you need
Plasma separation
[10b] To quantify both cell and plasma biomarkers
Integrate plasma separation + RBC lysis modules (and possibly blood dilution)
[10b] Hemolysis
Rupturing (lysis) of RBCs and release of their contents into surrounding fluid
[10b] Osmosis
Spontaneous net movement of solvent through a selectively permeable membrane into a region of HIGHER solute concentration
[10b] Isotonic solution effect on RBC
Cells keep normal size/shape (no net osmosis)
[10b] Hypertonic solution effect on RBC
Cells lose water and shrink (crenate)
[10b] Hypotonic solution effect on RBC
Cells take on water, swell and burst (lyse)
[10b] RBC lysis by osmotic swelling works by
Making the solution hypotonic: lower solute concentration outside the cell creates inward osmotic pressure, cell swells and bursts
[10b] ACK lysing buffer: blood to buffer ratio
1 mL EDTA whole blood into 10-20 mL ACK buffer at room temp
[10b] ACK lysing incubation time
3-5 minutes at room temperature
[10b] ACK protocol centrifugation
300 x g for 5 minutes (room temp first, then 2-8 C after PBS wash)
[10b] ACK protocol: leave how much supernatant
About 50 uL to avoid disturbing the pellet
[10b] ACK protocol: wash solution
5 mL cold phosphate buffered saline (PBS)
[10b] ACK protocol: blood collection tube type
EDTA-treated tubes (venipuncture)
[10b] Microfluidic RBC lysis: lysis solution
0.12% (v/v) formic acid + 0.05% (w/v) saponin in DI water
[10b] Microfluidic RBC lysis: quenching solution
0.6% (w/v) sodium carbonate + 3% (w/v) sodium chloride in DI water
[10b] Why use a quenching solution?
Stops lysis so WBCs are not damaged/permeabilized
[10b] Microfluidic lysis time scale
Seconds (RBC lysis seen over about 0.5 to 5.5 s) vs minutes for traditional ACK
[10b] Microfluidic lysis: 3 s vs 8 s
At 3 s cells are not permeabilized; at 8 s lymphocytes start to permeabilize and shift left on the plot
[10b] Why optimize lysis time?
To prevent WBCs from becoming permeabilized
[10b] Impedance at low frequency (500 kHz) measures
Cell size
[10b] Impedance at high frequency (1.7 MHz) measures
Cell morphology, granularity and permeabilization
[10b] Opacity vs output voltage plot axes
Opacity (y) vs output voltage at low frequency (x)
[10b] Traditional plasma separation
Centrifugation of whole blood
[10b] After centrifugation: layers
Plasma (about 55%) on top, buffy coat (leukocytes + platelets, under 1%) in the middle, erythrocytes (about 45%) at the bottom
[10b] Buffy coat contains
Leukocytes and platelets (<1% of whole blood)
[10b] Plasma composition
About 91% water, about 7% proteins, plus traces of bacteria, viruses, metabolites, circulating nucleic acids
[10b] Parameters to consider in microfluidic plasma separation
Initial blood dilution, rate of plasma collection, overall assay time, fraction of plasma separated, possible hemolysis
[10b] Microfluidic plasma separation principle
Cells migrate toward channel walls, leaving a low-cell region near the axial center where plasma is collected
[10b] Microfluidic plasma separation device geometry
Serpentine channels; plasma collected at the middle outlet branch, cells at the side branches
[10b] Microfluidic plasma separation result
With 25x dilution in 1x PBS, over 90% plasma purity
[14a] Hemocytometer is
Thick glass slide with a rectangular indentation (counting chamber) with a known gridded area and known depth, used to count cells
[14a] Hemocytometer uses
Counting cells, live/dead analysis, efficacy of cell culture
[14a] Hemocytometer chamber depth
0.1 mm
[14a] Hemocytometer: 1 x 1 mm square volume
1 mm^2 area, 100 nL at 0.1 mm depth
[14a] Hemocytometer: 0.25 x 0.25 mm square volume
0.0625 mm^2, 6.25 nL
[14a] Hemocytometer: 0.25 x 0.20 mm square volume
0.05 mm^2, 5 nL
[14a] Hemocytometer: 0.20 x 0.20 mm square volume
0.04 mm^2, 4 nL
[14a] Hemocytometer: 0.05 x 0.05 mm square volume
0.0025 mm^2, 0.25 nL
[14a] How many 1 x 1 mm squares in the gridded area?
Nine
[14a] Central 1 mm square is further subdivided into
0.05 x 0.05 mm squares (0.0025 mm^2)
[14a] Hemocytometer counting steps
Pre-process sample (lysis, dilution) > pipette onto hemocytometer > image under microscope > count cells in specific regions > calculate concentration (cells per unit volume)
[14a] Concentration from hemocytometer count
cells counted divided by volume of counted region (count / (area x depth)), corrected for dilution
[14a] Cell viability formula
Live cells / (live + dead cells)
[14a] Dye used for viability on a hemocytometer
Trypan blue
[14a] Viable cells with trypan blue
Intact membranes, do NOT take up dye, appear bright/clear
[14a] Dead cells with trypan blue
Damaged membranes take up dye, appear blue
[14b] Blood cell biomarkers (flow cytometry)
Size, numbers and types, surface proteins (CD4, CD8, CD64, CD11b), expression of surface proteins
[14b] Flow cytometry is the ___ in health-care settings
Current standard; optical based with fluorescence detection
[14b] Flow cytometry identifies cell types by
Tagging with conjugated fluorescent antibodies
[14b] Flow cytometry drawbacks
Expensive ($50k to over $125k), about $10k/year maintenance contracts, trained technicians necessary
[14b] Four parts of a flow cytometer
Fluidics system, lasers, optics, detectors, plus electronics and computer system