Haemodialyser Design and performance

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/11

encourage image

There's no tags or description

Looks like no tags are added yet.

12 Terms

1
New cards

Design requirements

Efficiency in removal of waste solutes

small internal blood volume

Minimal blood retention

Moderate resistance to blood flow

Compatibility of blood contacting surfaces

Safety, ease of use, sterility and cheapness

2
New cards

Small internal blood volune

40-140ml (depends on device)

3
New cards

Minimal blood retention after use

2-7ml (device and washback technique dependant)

Synthetic EPO important for maintaining RBC levels

4
New cards

Moderate resistance to blood flow

Ultrafiltration most effective at low resistances

5
New cards

The 2 main geometries used

Parallel Plate (rarely used, <10%)

Hollow tibe (>90% of dialysers)

6
New cards

Parallel Plate design mechanism

Blood passes through 2 sheets of membrane, dialysate flow is in opposite direction to bloodflow, high membrane area is required by stacking multiple plates.

7
New cards

Hollow fibre design

5-20,000 fine bore (200 micrometres) membrane tubes.

End of fibres are embedded in potting resin

Fibre bundle is encased in a plastic shell

Entry and exit points allow dialysate fluid to be pumped around outside surface of the hollow fibres

8
New cards

Mass transfer rate of solute crossing membrane

m=QbiCbi-QboCbo

9
New cards

Solute clearece (subbing in m)

Cl= (QbiCbi-QboCbo)/Cbi

10
New cards

Cl

Solute clearence (can be thought of as blood flow rate)

11
New cards

m

Mass transfer rate

12
New cards