Digital image quality is

the fidelity with which the anatomy under investigation is rendered on the IR

2
New cards

spatial resolution is

ability to image small objects that have high subject contrast

3
New cards

as pixel size decreases, spatial resolution

increases

4
New cards

Bennett’s definition of spatial resolution

ability of a system to resolve spatial frequency

5
New cards

spatial frequency is

the area you take up or exist in so it requires a pixel inside and outside an area

6
New cards

contrast resolution is

ability to distinguish anatomical structures of similar subject contrast

7
New cards

Bennett’s definition of contrast resolution

ability of system to show it to me

8
New cards

noise is

random fluctuation of x-ray interaction on the IR

9
New cards

Quantum mottle decreases as

mAs increases, kVp decreases, and focal spot size increases

10
New cards

Why causes quantum mottle to decrease overall

more photons

11
New cards

Bennett’s definition of noise

inability to accurately draw the edges

12
New cards

A Fast IR has

high noise and low contrast resolution

13
New cards

A slow IR has

low noise and high contrast resolution

14
New cards

Spatial resolution is limited to

pixel size

15
New cards

DeL stands for

detector elements

16
New cards

What are the radiographic quality Image receptor factors

pixel size

dynamic range

intensity response

signal to noise

post processing

17
New cards

what are the radiographic quality geometric factors

magnification

distortion

focal spot blur

18
New cards

how often does magnification occur

every exam

19
New cards

why does magnification occur

because the beam diverges from a point source

20
New cards

equation for magnifcation factor

MF= Img size/Obj size

21
New cards

Magnification facotr alternative equation

MF= SID/SOD

22
New cards

The numerator of the magnification factor equation is always

positive and higher

23
New cards

For every inch of OID, it requires how much of an increase in SID

7”

24
New cards

If the part is angled, what distortion occurs

foreshortening

25
New cards

if the tube is angled, what distortion occurs

elongation

26
New cards

if the IR is angled against the tube, what distortion occurs

elongation

27
New cards

If the IR is angled to the part, what distortion occurs

foreshortening

28
New cards

distortion depends on what three things

Object thickness(OID)

Object position

Object shape

29
New cards

are thick objects more distorted than thin

yes

30
New cards

when outside the center of central ray what happens to distortion

increases

31
New cards

focal spot blurring occurs more toward what side of the tube

cathode

32
New cards

what are the subject factors

subject contrast

part thickness

33
New cards

what help display subject contrast

capabilites of the IR, the system, and the monitor

34
New cards

as part thickness increases

scatter increases and technique increases

35
New cards

what is the effective atomic number

collective atominc numbers of stuff

36
New cards

when the effective atomic number is high, subject contrast is

high

37
New cards

Iodine Z#

53

38
New cards

Barium Z#

56

39
New cards

the more the shape coincides with the x-ray beam shape

contrast increases

40
New cards

best solution to involuntary motion

short exposure time

41
New cards

best solution to voluntary motion

good pt instructions

42
New cards

if good pt instructions isnt an option you choose

short exposure time

43
New cards

exit skin exposure is how much of the entrance skin exposure

10%

44
New cards

pixel size affects pt

relative dose

45
New cards

the smaller the pixel size, it rquires what

increase in pt dose so more photons hit the pixels

46
New cards

increases in Pt dose are due to increases in

pt thickness

field size

mAs

Tiem

kVp

Voltage ripple

47
New cards

increase in magnification is due to increases in

pt thickness

OID

48
New cards

increase in focal spot blur is due to increases in

pt thickness

focal spot size

OID

49
New cards

increase in motion blur is due to increases in

pt thickness

OID

Time

Voltage ripple

50
New cards

increase in absorption blur is due to increases in

pt thickness

51
New cards

three main factors in scatter radiation are

kVp

Field size

Pt thickness

52
New cards

2 main tools for controlling scatter

beam restriction

grids

53
New cards

scatter does what to contrast

decreases

54
New cards

can scatter change the shape of an object

yes

55
New cards

kVp effects the levels of

scatter production

56
New cards

why is low kVp not the solution to scater

it would increase pt dose due to an increase in mAs

57
New cards

collimation has what affect on quality

decreases pt dose and increases contrast

58
New cards

as field size decreaes, scatter

decreases

59
New cards

as field size decreases, contrast

increases

60
New cards

as pt thickness increases, scatter

increases

61
New cards

with an increased pt thickness, more x-rays undergo multiple scattering which in turn causes an increase in

average angle of scatter

62
New cards

putting a pt prone increases

compression

63
New cards

beam restrictors are

aperture diaphragm

cones & cylinders

variable aperture collimator

64
New cards

What are the grid types

parallel

crosses

focused

moving

65
New cards

what are the 2 reasons to use beam restriction

only tissue examined is exposed

large field sizes produce more scatter and decrease contrast

66
New cards

radiolysis of water is

x-rays scission water molecules into hydrogen peroxide

67
New cards

the aperture diaphragm is designed for the opening to be 1cm ______ on all sides of film

smaller

68
New cards

what uses aperture diaphragm

fixed chest rooms

trauma machines

dental

69
New cards

why is aperture diaphragm least used

most restrictive and simplest

70
New cards

cones and cylinders are modifications to what type of beam restriction

aperture diaphragm

71
New cards

the position and size of the distal end of the cone or cylinder determines

field size

72
New cards

cones have a fixed

SID

73
New cards

when does cone cutting occur

when the center of central ray is not in center of cone or cylinder

74
New cards

collimator is what type of beam restriction

variable aperture diaphragm

75
New cards

what is the most common beam restriction

collimator box

76
New cards

how many field sizes does a collimator box have

infinite

77
New cards

does beam restriction prevent scatter production

yes

78
New cards

the first stage entrance shutters control

off focus radiation

79
New cards

positive beam limitation was instated by the FDA why

techs wouldnt collimate

80
New cards

PBL was mandated in what year

1974

81
New cards

PBL was resended in what year

1994

82
New cards

do grids prevent scatter production

no

83
New cards

grid ratio equation

H/D

84
New cards

in the grid ratio equation, H stands fpr

height of the grid strip

85
New cards

in the grid ratio equation, D stands for

how far the strips are from one another

86
New cards

a 5:1 grid has how much cleanup

87%

87
New cards

a 16:1 grid has how much cleanup

97%

88
New cards

Grid frequency is

the number of grid strips per cm or inch

89
New cards

High frequency grids show less distinct what

grid strips

90
New cards

how is grid frequency increased

adding lead strips and thinning interspace materials

91
New cards

Does grid strip size ever change

no

92
New cards

what are the interspace materials made of

aluminum or plastic/carbon fiber

93
New cards

aluminum is perferred why

less visible grid lines and nonhygroscopic

94
New cards

what is the disadvantage of aluminum interspace material

absorption of primary beam which is compensated with an increase in mAs

95
New cards

The contrast improvement factor equation is

Image contrast w grid/Image contrast w/o grid

96
New cards

the numerator of the CIF equation is always

greater than 1

97
New cards

the range for the CIF for most grids is usually

1.5-2.5

98
New cards

The bucky factor determines

the amount of technique required to penetrate (?)

99
New cards

what is the ascending order of grid ratios

0,5:1,6:1,8:1,10:1,12:1,16:1

100
New cards

what is the ascending order of bucky factors

1,2,3,4,5,5,6,