1/37
Pump
sets flow and gradient composition.
Injector
adds a narrow, reproducible sample plug.
Column
creates selectivity and efficiency.
Detector
converts eluting concentration into signal.
5.0 µm particle
Low pressure
lower efficiency
longer run.
3.0 µm particle
Higher efficiency and speed
clear pressure cost.
1.5 µm particle
fast and efficient
only a high-pressure system can operate
Symptom: Pressure rises and falls with each stroke
First check: the inlet and outlet check valves.
symptom: solvent appears near piston
First check: the piston seal and wash system.
symptom: pressure collapses and erratic flow
First check: solvent supply and trapped air.
Dual piston HPLC pump
A delivers, B refills
Then B delivers and A refills
Degassing prevents what failure
Removes dissolved gas before it forms bubbles in pump or detector.
Filtration prevents what failure
Removes particles before they reach narrow tubing, valves, frits, and the packed bed
If sharp baseline spikes:
Inspect degassing and the detector flow cell first.
if pressure rises over several runs
Inspect sample and solvent filtration first.
if pump loses prime…
Inspect: solvent inlet, degassing
Steps to loading injection valve:
Draw sample → fill loop → switch valve → carry to column
5 µL x 1 concentration
Small volume and low mass
inlet can focus sample, and stationary phase remains within working capacity.
25 µL x 1 concentration
Large volume, moderate mass
Mass overload unlikely
but large solvent plug may cause broadening or splitting
5 µL x 8 concentration
Small volume, high mass
The solvent plug small, but stationary phase may overload.
Fronting becomes more likely
25 µL x 8 concentration
Large volume and high mass
Both mechanisms act together.
Reduce concentration or volume, then test one change at a time
If carryover appears:
Inject concentrated sample
Inject blank
Run needle wash
Inject another blank
Matched or weaker diluent
The analyte focuses near the column head before separation begins
so band stays narrow
Stronger diluent than the starting mobile phase
The sample plug begins eluting the analyte before the inlet can focus it. Broadening, fronting, or splitting can result.
Isocratic elution
Composition stays constant.
Choose when solvent strength resolves sample in reasonable time.
Gradient elution
Elution strength increases during run.
Use it when retention spans a wide range or late compounds elute too slowly.
Changing solvent strength: isocratic
Use a fixed composition when the analytes have a modest retention range.
It is simple to transfer and needs little composition recovery between runs.
Changing solvent strength: linear gradient
Increase solvent strength steadily.
This is a strong default for mixtures with a wide polarity range because late compounds elute sooner and often more sharply.
Changing solvent strength: step gradient
Jump to a stronger composition for a fast wash or a defined fraction.
Abrupt changes can compress or distort bands, so step timing requires care.
Simple routine wants what kind of elution
Isocratic
Mobile phase