Chromatography Midterm 2 - HPLC

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/37

flashcard set

Earn XP

Description and Tags

38 Terms

1
New cards

Pump

sets flow and gradient composition.

2
New cards

Injector

adds a narrow, reproducible sample plug.

3
New cards

Column

creates selectivity and efficiency.

4
New cards

Detector

converts eluting concentration into signal.

5
New cards

5.0 µm particle

Low pressure

lower efficiency

longer run.

6
New cards

3.0 µm particle

Higher efficiency and speed

clear pressure cost.

7
New cards

1.5 µm particle

fast and efficient

only a high-pressure system can operate

8
New cards

Symptom: Pressure rises and falls with each stroke

First check: the inlet and outlet check valves.

9
New cards

symptom: solvent appears near piston

First check: the piston seal and wash system.

10
New cards

symptom: pressure collapses and erratic flow

First check: solvent supply and trapped air.

11
New cards

Dual piston HPLC pump

A delivers, B refills

Then B delivers and A refills

12
New cards

Degassing prevents what failure

Removes dissolved gas before it forms bubbles in pump or detector.

13
New cards

Filtration prevents what failure

Removes particles before they reach narrow tubing, valves, frits, and the packed bed

14
New cards

If sharp baseline spikes:

Inspect degassing and the detector flow cell first.

15
New cards

if pressure rises over several runs

Inspect sample and solvent filtration first.

16
New cards

if pump loses prime…

Inspect: solvent inlet, degassing

17
New cards

Steps to loading injection valve:

Draw sample → fill loop → switch valve → carry to column

18
New cards

5 µL x 1 concentration

Small volume and low mass

inlet can focus sample, and stationary phase remains within working capacity.

19
New cards

25 µL x 1 concentration

Large volume, moderate mass


Mass overload unlikely

but large solvent plug may cause broadening or splitting

20
New cards

5 µL x 8 concentration

Small volume, high mass

The solvent plug small, but stationary phase may overload.

Fronting becomes more likely

21
New cards

25 µL x 8 concentration

Large volume and high mass


Both mechanisms act together.

Reduce concentration or volume, then test one change at a time

22
New cards

If carryover appears:

Inject concentrated sample

Inject blank

Run needle wash

Inject another blank

23
New cards

Matched or weaker diluent

The analyte focuses near the column head before separation begins

so band stays narrow

24
New cards

Stronger diluent than the starting mobile phase

The sample plug begins eluting the analyte before the inlet can focus it. Broadening, fronting, or splitting can result.

25
New cards

Isocratic elution

Composition stays constant.

Choose when solvent strength resolves sample in reasonable time.

26
New cards

Gradient elution

Elution strength increases during run.

Use it when retention spans a wide range or late compounds elute too slowly.

27
New cards

Changing solvent strength: isocratic

Use a fixed composition when the analytes have a modest retention range.

It is simple to transfer and needs little composition recovery between runs.

28
New cards

Changing solvent strength: linear gradient

Increase solvent strength steadily.

This is a strong default for mixtures with a wide polarity range because late compounds elute sooner and often more sharply.

29
New cards

Changing solvent strength: step gradient

Jump to a stronger composition for a fast wash or a defined fraction.

Abrupt changes can compress or distort bands, so step timing requires care.

30
New cards

Simple routine wants what kind of elution

Isocratic

31
New cards

Mobile phase

32
New cards
33
New cards
34
New cards
35
New cards
36
New cards
37
New cards
38
New cards