High Flow Nasal Cannula (HFNC)
An oxygen delivery system delivering flows typically from 30L/min to 60L/min with an FiO2 range from room air (21%) to 100%, incorporating active or cartridge-based warming and humidification.
Three Required Components of an HFNC System
The essential structural parts of any high flow nasal cannula setup: a patient interface (e.g., cannula or prongs), a gas delivery device that regulates FiO2 (e.g., a blender or two flow meters), and a humidification system.
PEEP Generation in HFNC
The generation of positive end-expiratory pressure at a rate of approximately 1cmH2O of pressure for every 10L/min of gas delivered, which stents small airways open during inspiration and expiration.
Carbon Dioxide Washout
The flushing of carbon dioxide from the anatomical dead space by high gas flows, which helps lower PaCO2 in patients with hypercapnic respiratory failure.
Vapotherm
A brand of high flow nasal cannula device that utilizes a paper transfer cartridge to provide warmth and humidification to the delivered breathing gas.
Indications for High Flow Nasal Cannula
Clinical situations appropriate for HFNC, including hypoxemic respiratory failure (e.g., pneumonia), hypercapnic respiratory failure (e.g., COPD exacerbation), reducing the risk of intubation, palliative comfort for DNI or hospice patients, and device intolerance (such as inability to wear a CPAP mask).
Manual Gas Blending Formula
The mathematical method used to mix compressed air and oxygen manually: O2 Flow=79Total Flow×(%O2−21) and Air Flow=Total Flow−O2 Flow.
Oxygen Blender
A precision device connected to 50psi compressed air and oxygen high-pressure wall sources (yellow and green hoses) that combines the two gases to allow direct dialing of a specific FiO2.
Oxygen Blender Quality Control
The safety check required when a blender is first placed on a patient and at least once per shift, performed using a calibrated oxygen analyzer placed at the patient outlet to confirm FiO2 accuracy.
Oxygen Hood (Oxy Hood)
An infant oxygen enclosure covering only the head that requires flow rates greater than 7L/min to flush out carbon dioxide, capable of delivering heated, humidified gas at an FiO2 from room air to 100%.
Noise Amplification in Oxygen Hoods
A known clinical side effect occurring because the high flow from the nebulizer or humidification system generates noise that is amplified inside the enclosed head chamber.
Isolette
A temperature-controlled infant enclosure designed for stable newborns to maintain a neutral thermal environment, capable of administering heated, humidified oxygen via an internal 50psi gas source, an oxygen bleed line, or free-flowing gas from an anesthesia bag.
Oxygen Tent
A pediatric enclosure system requiring gas flows of 12L/min to 15L/min to flush carbon dioxide, delivering a variable FiO2 of 40% to 50% that fluctuates significantly due to frequent opening and closing.
The Three P's of Device Selection
The clinical decision-making framework for selecting oxygen equipment: Purpose (correct hypoxemia, reduce work of breathing), Patient (age, condition, airway, minute volume), and Performance (device precision and FiO2 stability).
Emergent Oxygen Therapy
The clinical approach for trauma, cardiac/respiratory emergencies, or suspected tissue hypoxia, calling for the highest possible FiO2 (100% if available), typically initiated using a non-rebreather mask.
Critically Ill Adult Oxygen Therapy
The oxygen strategy for patients presenting with moderate to severe hypoxemia, requiring a reservoir or high flow system providing an FiO2 of at least 60%.
Acute-on-Chronic Hypoxemia Management
Oxygen therapy tailored for chronic lung disease patients to prevent suppressing their hypoxic drive, targeting a PaO2 between 50mmHg and 70mmHg (saturation between 85% and 92%) using low flows of 1L/min to 2L/min or an air-entrainment mask at 24% to 28%.
Desired FiO2 Formula
The mathematical formula used to calculate a target oxygen fraction: Known FiO2Known PaO2=Desired FiO2Desired PaO2.
Protocol-Based Oxygen Therapy
A clinical framework that authorizes respiratory therapists to initiate, modify, and discontinue oxygen therapy within their scope of practice based on systematic patient evaluations rather than requiring a physician order for each change.
Oxygen Titration
The process of adjusting oxygen delivery upward or downward in increments of 1L/min to 2L/min or small percentage increments to maintain target arterial oxygen saturation (SpO2≥92%).
Post-Titration Recheck Times
The time required to reassess a patient following an oxygen adjustment: 10minutes for patients with no pulmonary history, and 20minutes for patients with a pulmonary history.
Oxygen Discontinuation Criterion
The guideline indicating oxygen therapy should be discontinued when the measured PaO2 on supplemental oxygen is greater than 5 times the set FiO2.
Textbook Weaning Frequency Recommendation
The textbook rule suggesting that a patient's oxygen concentration should be decreased no more frequently than once per shift.