Mass Spectrometry

The analysis of a protonated molecule by it’s mass, normally fragmented in some way

2
New cards

First step for mass spectrometry

The production of gas-phase ions, fragmentation, and separation by abundance

3
New cards

Why are ions required for mass spectrometry?

So they can be manipulated with electric al or magnetic fields which is required for detection

4
New cards

Molecular ion

Positive radical ion, formed in EI through the removal of an electron from the molecule

5
New cards

mass-to-charge ratio (m/z)

a unitless measurement of an ion based on it’s relative mass and charge

6
New cards

Isotopes

Elements that differ in atomic weight due to differing amounts of neutrons present in the molecule (denoted as a superscript before the element letter)

7
New cards

Nominal mass

the mass of the predominant isotope of each element rounded to the nearest integer value that corresponds to the mass number

8
New cards

monoisotpic mass

calculated using the exact mass of the most abundant isotope for each element

9
New cards

Where is the molecular ion most commonly found in a mass spectrum?

It is almost always represented by the peak at the highest m/z value (though adducts, isotopes, and background ions can all create higher m/z values as well)

10
New cards

Base Peak

The most abundant peak in the mass spectrum, and is related to the stability of the ion

11
New cards

Why is the base peak important for understanding the presence of other peaks in the spectrum?

Mass spectra are displayed in relative ion abundance, so all other peaks and there amounts are based on how much of the base peak there is

12
New cards

Fragment Ions

observed at smaller m/z values than the molecular ion (fragments of your original molecule)

13
New cards

Fragment Ion rules

Ions sequentially fragment, no fragment will have an overall mass gain, and no fragment will contain a larger number of atoms than the molecular ion

14
New cards

Why are isotopes important in mass spectrometry?

Provide confirmation of actual peak rather than just noise, the also provide information about the elemental composition of a fragment, also fragment ions will not show isotopic patterns for elements not in the molecular ion.

15
New cards

Why is a vacuum necessary for a MS?

They allow ions to reach the detector without colliding with other neutral molecules in the gas phase (which would complicate a spectrum)

16
New cards

What is the sample inlet?

Where the sample is introduced in the vacuum for MSes

17
New cards

Important considerations for ionization sources

Internal energy transferred during ionization and the physics-chemical properties of the analytes

18
New cards

Ways ionization sources produce ions

Electron ejection/capture, de/protonation, adduct formation, transfer of charged species from condensed phase to gas phase

19
New cards

Hardest electron ionization source

Electron Ionization

20
New cards

Softest electron ionization source

Electron spray ionization

21
New cards

How does EI work?

Electrons are emitted from a hot tungsten rhenium filament, e- are accelerated with a 70-eV potential, e- collide with neutral analytes which forms a +1 ion

22
New cards

Why is the 70-eV potential standardized?

To increase reproducibility and allow for the creation of spectral libraries

23
New cards

Explain the mechanism of EI

when the electron collides with the neutral molecule, it excites the molecule rotational-vibrationally, which leads to the ejection of an electron due to electrostatic repulsion

24
New cards

Advantages of EI

Provides fragment-rich MS (good for library searches), useful for structural elucidation, easily coupled with GC separation

25
New cards

Disadvantages of EI

Extensive fragmentation can result in the absence of the molecular ion, requires the introduction of neutral molecules under vacuum, requires thermal stables and volatile analytes

26
New cards

Is chemical ionization a hard or soft ionization technique?

Hard, but softer than EI

27
New cards

Explain the mechanism of chemical ionization

ions are produced via collision with primary ions present in the source under pressure (to increase collisions). The primary ions are the reagent gas that is ionized with EI before being pumped into the source. Differing ions in the reagent gas will lead to a variety of fragments formed from the analyte of interest

28
New cards

What property of the reagent gas in CI effects its fragmentation ability?

Proton affinity difference between reagent ion and the analyze (methane produces the most fragmentation, followed by isobutane then ammonia)

29
New cards

Electrospray ionization

soft ionization technique that occurs at atmospheric pressure and works well for large, non volatile, chargeable molecules (typically in biology)

30
New cards

Advantages of CI

Preserves molecular species (protonated) and has a tunable degree of fragmentation based on reagent gas

31
New cards

Disadvantages of CI

Requires EI source modification, requires additional reagent gas, and generation of even-electron and odd-electron molecular species is possible

32
New cards

ESIs mechanism

Solution is nebulizer in the presence of a strong electric field and heated drying gas, and there is an applied electric field between he nebulizer and the capillary for MS. Taylor cones form when the coulomb regulation overcomes the surface tension and the drops are desolvated from the heated drying gas. The charged droplets are then manipulated by the electrostatic field

33
New cards

Advantages of ESI

Extended MS application to biological and polar drug metabolites, preserves the molecular species, and is easily interfaced with LC techniques

34
New cards

Disadvantages of ESI

Produces limited fragmentation and is only useful for molecular weight information unless combined with an activation technique

35
New cards

Atmospheric Pressure chemical ionization

Gas-phase chemical ionization process that is the atmospheric pressure equivalent to CI

36
New cards

APCI mechanism

Solution is passed through a nebulizing needle, the spray is then fully vaporized through a heated ceramic tube (400-500C). Reagent gas ions formed through interaction with discharge from corona (a nearby needle) discharge. Everything must be in the gas phase for this to work

37
New cards

Advantages of APCI

Very high ionization efficiency because of high pressure, non polar and slightly polar analytes can be analyzed, works better at higher flow rates.

38
New cards

Disadvantages of APCI

Analyze must by thermally stable, and it is not suitable for compounds with a molecular weight greater than 1,000 Da

39
New cards

Different types of mass analyzers

single quadruple, triple quadruple, time of flight, ion trap

40
New cards

How does a single quadrupole work?

Four parallel rods are set in a square formation, pairs of opposite rods held as the same potential which oscillates between the rods. Only stable ions will traverse the quadrupole to reach the detector

41
New cards

Full scan mode

Collection of all m/z information over a given range (scanned low to high mass)

42
New cards

Selected Ion Monitoring (SIM)

Increases sensitivity compared to full scan mode, monitoring for only a single or few ions at a time

43
New cards

Advantages of Single quadrupole

Lightweight, compact, robust, inexpensive, high scan speeds for chromatographic peaks

44
New cards

Disadvantages of single quadrupole

Low resolution, lower sensitivity in full scan mode

45
New cards

Tandem Mass Spectrometry

can be done in space (QQQ) or time (IT), combination of mass analyzers to increase the opportunity for additional fragmentation of a desired ion. Typically includes collision induced dissociation with a neutral gas

46
New cards

Triple quadrupole MS parts

Q1 used to isolate desired ion (precursor ion), Q2 where CID occurs (RF only), Q3 used to scan product ions to the detector

47
New cards

Triple quadratole process

Isolated ions in Q1 are accelerated into the collision cell (Q2). Kinetic energy of precursor ions is converted into internal energy through neutral collision gas. Energy redistribution leads to the formation of fragment ions and neutral losses (not detected). Product ions emerge from collision cell and are transported to Q3

48
New cards

QQQ scan modes

Product ion scan, precursor ion scan, neutral loss scan, multiple reaction monitoring

49
New cards

Product Ion scanning

Desired ion is isolated in Q1, isolated precursor ion is fragmented by CID in Q2, product ions are scanned to detector in Q3

50
New cards

Precursor Ion Scanning

Q1 is scanned across the desired mass range, ions are fragmented in the collision cell, only a single ion is isolated and transmitted to the detector in Q3. Useful for class-specific identifications of unknown analytes

51
New cards

Neutral Loss Scanning

Q1 and Q3 are scanned with a constant offset. Isolated ions in Q1 are fragmented in Q2, therefore, ions which loses a neutral molecule of the defined mass will be transmitted and detected. Useful to identify characteristics neutral losses

52
New cards

Multiple reaction monitoring

Q1 is set to a specific precursor ion, isolated precursor ion is fragmented in Q2, and Q3 is set to allow only a specific product ion to reach the detector. Reduces noise and increases sensitivity since it works as a double mass filter. Most useful for quantitative analysis.

53
New cards

Advantages of QQQ over single quadrupoles

Higher selectivity, better S/N and lower LOD, more reliable identification of detected analytes, wider linear range of quantitation, better accuracy and reproducibility (especially at lower concentrations), Different scan modes available

54
New cards

Exact Mass

theoretical value calculated by summing the exact masses of the individual isotopes of the molecule

55
New cards

Accurate Mass

Experimentally determined mass that allows the elemental composition to be determined by comparison with the calculated exact mass

56
New cards

Time of Flight MS

Separation of ions based on the time it takes to travel a known distance upon uniform acceleration, requires very low background pressure

57
New cards

Advantages of Quadrupole Time of Flight

High resolution MS, elemental formula determinations, simulataneous spectral acquisition, increased sensitivity from improved ion transmission, extended mass range.

58
New cards

Disadvantages of Quadrupole Time of Flight

Very expensive, requires space, requires tight control of lab environment

59
New cards

Electron Multiplier Detector

Ions are accelerated at a high velocity where they hit a dynode (has the oposite charge of the ion) which causes emission of secondary particles (cations or e-) secondary particles are amplified in the electron multiplier which creates a current

60
New cards

Tuning

The adjustment of parameters to achieve the best analytical response based on the analysis of a standard reference compound. Poor peak shapes are often due to this not being done well, which will result in incorrect identification of peak

61
New cards

Calibration

The process of assigning m/z values based on the known m/z values of standard compounds

62
New cards

Optimization

the adjustment of source-dependent parameters and the compound-dependent parameters to maximize the instrument response for a sample or compound

63
New cards

What is the main advantage of ambient ionization?

No sample prep or training, which saves money and time

64
New cards

What samples can DART analyze?

all phases

65
New cards

DART mechanism

Gas enters DART source (HE, nitrogen, Ar) where a high voltage is applied and generates an excited state species. Electrodes draw out these species (metastable are the only ones that pass through) which enters a heated region. Grid electrode prevents recombination and the insulator protects from high voltages. Reagent ions are formed from atmospheric compounds.

66
New cards

What gas creates the most reagent gases in DART MS?

He because it carries the most energy

67
New cards

What factors determine if an adduct or molecular ion is formed in DART MS?

Proton affinity and ionization energy

68
New cards

What is the tradeoff for using nitrogen or Argon instead of He in DART MS?

More selective but less sensitive

69
New cards

What are the two sampling techniques in DART MS?

Scanning and pulsing

70
New cards

Vapur Interface Advantages

makes DART compatible with different mass spectrometers and improves reproducibility by reducing turbulence

71
New cards

Vapur Interface disadvantages

Ion-molecule reactions can occur over longer transport distance, loss of non polar or reactive compounds, increased adduct formation, and carryover due to absorption onto ceramic tubing

72
New cards

Chromatography

Separation of components in a mixture based on different interactions between two phases

73
New cards

How is the partition coefficient defined?

concentration of sample in stationary phase over concentration of sample in mobile phase

74
New cards

How is retention time (tr) defined?

Time spent in stationary phase (ts) + dead time, essentially how long it takes a sample to elute, including the time it takes the solvent to work its way through the column

75
New cards

What is the retention factor?

Describes how well an analyze is retained by the stationary phase, and the higher the value the stronger the retention

76
New cards

What is the selectivity factor?

alpha, the ration of retention factors for two adjacent peaks (more strongly retained/less strongly retained) and a larger value indicates better separation

77
New cards

What are the optimal conditions for chromatography?

Short run, good separation, complete baseline separation, symmetry in peak shape, reproducible retention times

78
New cards

What can cause peak fronting?

large injection volume, poor solubility of sample, degradation of stationary phase, column saturation

79
New cards

What can cause peak tailing?

Secondary interactions, column overload, contaminants

80
New cards

What factors describe column efficiency?

Plate height (H), number of plates (N) and length of the column (N=L/H)

81
New cards

What does A represent in the van Deemter equation?

Multiple paths (eddy diffusion)

82
New cards

What does B represent in the van Deemter equation?

longitudinal diffusion, diffusion through the column

83
New cards

What does u represent in the van Deemter equation?

average mobile phase velocity

84
New cards

What does Cs represent in the van Deemter equation?

resistance to mass transfer of the stationary phase

85
New cards

What does Cm represent in the van Deemter equation?

Resistance ti mass transfer of the mobile phase

86
New cards

What effects Cs and Cm?

for Cs film thickness and diffusion coefficient and for Cm it’s the particle or column diameter

87
New cards

What specific number is considered baseline resolution?

1.5

88
New cards

What phases can be analyzed with GC?

All phases with proper preparation

89
New cards

What are ideal characteristics for GC samples?

Boiling point up to 400C, non polar or slightly polar, and must be thermally stable

90
New cards

What are some characteristics that might make a sample difficult to analyze on GC?

Does not vaporize at GC operation conditions, highly reactive, thermally unstable or strongly polar

91
New cards

Derivatization

Process of chemically modifying a compound to produce a new molecule suitable for GC analysis

92
New cards

Types of derivatization

Silylation, acylation, alkylation

93
New cards

Characteristics of carrier gases in GC

Must be inert, and lead to a small H, cost and safety are also important

94
New cards

List the carrier gases in order from best to worst resolution

hydrogen, He, nitrogen

95
New cards

Why does hydrogen and helium have a better resolution than nitrogen in GC?

solutes are able to diffuse rapidly through these gases due to decreased resistance to mass transfer

96
New cards

How is oxygen prevented from getting into the GC column?

Pure carrier gas and oxygen moisture traps or filters

97
New cards

How should the sample be introduced to the column in GC?

as a “plug of vapor”

98
New cards

Describe the steps for manual injection in GC?

Rinse syringe with solvent, rinse with sample, draw up desired volume, carefully insert syringe into septum, let needle heat up, then depress the plunger and pull the syringe out

99
New cards

What are some advantages of auto sampler injection?

More robust and reproducible, instantaneous injection

100
New cards

What is the inlet liner in GC?

In the injection port and enhances vaporization, guides the needle, and limits sample degredation