Mass Spectrometry
The analysis of a protonated molecule by it’s mass, normally fragmented in some way
First step for mass spectrometry
The production of gas-phase ions, fragmentation, and separation by abundance
Why are ions required for mass spectrometry?
So they can be manipulated with electric al or magnetic fields which is required for detection
Molecular ion
Positive radical ion, formed in EI through the removal of an electron from the molecule
mass-to-charge ratio (m/z)
a unitless measurement of an ion based on it’s relative mass and charge
Isotopes
Elements that differ in atomic weight due to differing amounts of neutrons present in the molecule (denoted as a superscript before the element letter)
Nominal mass
the mass of the predominant isotope of each element rounded to the nearest integer value that corresponds to the mass number
monoisotpic mass
calculated using the exact mass of the most abundant isotope for each element
Where is the molecular ion most commonly found in a mass spectrum?
It is almost always represented by the peak at the highest m/z value (though adducts, isotopes, and background ions can all create higher m/z values as well)
Base Peak
The most abundant peak in the mass spectrum, and is related to the stability of the ion
Why is the base peak important for understanding the presence of other peaks in the spectrum?
Mass spectra are displayed in relative ion abundance, so all other peaks and there amounts are based on how much of the base peak there is
Fragment Ions
observed at smaller m/z values than the molecular ion (fragments of your original molecule)
Fragment Ion rules
Ions sequentially fragment, no fragment will have an overall mass gain, and no fragment will contain a larger number of atoms than the molecular ion
Why are isotopes important in mass spectrometry?
Provide confirmation of actual peak rather than just noise, the also provide information about the elemental composition of a fragment, also fragment ions will not show isotopic patterns for elements not in the molecular ion.
Why is a vacuum necessary for a MS?
They allow ions to reach the detector without colliding with other neutral molecules in the gas phase (which would complicate a spectrum)
What is the sample inlet?
Where the sample is introduced in the vacuum for MSes
Important considerations for ionization sources
Internal energy transferred during ionization and the physics-chemical properties of the analytes
Ways ionization sources produce ions
Electron ejection/capture, de/protonation, adduct formation, transfer of charged species from condensed phase to gas phase
Hardest electron ionization source
Electron Ionization
Softest electron ionization source
Electron spray ionization
How does EI work?
Electrons are emitted from a hot tungsten rhenium filament, e- are accelerated with a 70-eV potential, e- collide with neutral analytes which forms a +1 ion
Why is the 70-eV potential standardized?
To increase reproducibility and allow for the creation of spectral libraries
Explain the mechanism of EI
when the electron collides with the neutral molecule, it excites the molecule rotational-vibrationally, which leads to the ejection of an electron due to electrostatic repulsion
Advantages of EI
Provides fragment-rich MS (good for library searches), useful for structural elucidation, easily coupled with GC separation
Disadvantages of EI
Extensive fragmentation can result in the absence of the molecular ion, requires the introduction of neutral molecules under vacuum, requires thermal stables and volatile analytes
Is chemical ionization a hard or soft ionization technique?
Hard, but softer than EI
Explain the mechanism of chemical ionization
ions are produced via collision with primary ions present in the source under pressure (to increase collisions). The primary ions are the reagent gas that is ionized with EI before being pumped into the source. Differing ions in the reagent gas will lead to a variety of fragments formed from the analyte of interest
What property of the reagent gas in CI effects its fragmentation ability?
Proton affinity difference between reagent ion and the analyze (methane produces the most fragmentation, followed by isobutane then ammonia)
Electrospray ionization
soft ionization technique that occurs at atmospheric pressure and works well for large, non volatile, chargeable molecules (typically in biology)
Advantages of CI
Preserves molecular species (protonated) and has a tunable degree of fragmentation based on reagent gas
Disadvantages of CI
Requires EI source modification, requires additional reagent gas, and generation of even-electron and odd-electron molecular species is possible
ESIs mechanism
Solution is nebulizer in the presence of a strong electric field and heated drying gas, and there is an applied electric field between he nebulizer and the capillary for MS. Taylor cones form when the coulomb regulation overcomes the surface tension and the drops are desolvated from the heated drying gas. The charged droplets are then manipulated by the electrostatic field
Advantages of ESI
Extended MS application to biological and polar drug metabolites, preserves the molecular species, and is easily interfaced with LC techniques
Disadvantages of ESI
Produces limited fragmentation and is only useful for molecular weight information unless combined with an activation technique
Atmospheric Pressure chemical ionization
Gas-phase chemical ionization process that is the atmospheric pressure equivalent to CI
APCI mechanism
Solution is passed through a nebulizing needle, the spray is then fully vaporized through a heated ceramic tube (400-500C). Reagent gas ions formed through interaction with discharge from corona (a nearby needle) discharge. Everything must be in the gas phase for this to work
Advantages of APCI
Very high ionization efficiency because of high pressure, non polar and slightly polar analytes can be analyzed, works better at higher flow rates.
Disadvantages of APCI
Analyze must by thermally stable, and it is not suitable for compounds with a molecular weight greater than 1,000 Da
Different types of mass analyzers
single quadruple, triple quadruple, time of flight, ion trap
How does a single quadrupole work?
Four parallel rods are set in a square formation, pairs of opposite rods held as the same potential which oscillates between the rods. Only stable ions will traverse the quadrupole to reach the detector
Full scan mode
Collection of all m/z information over a given range (scanned low to high mass)
Selected Ion Monitoring (SIM)
Increases sensitivity compared to full scan mode, monitoring for only a single or few ions at a time
Advantages of Single quadrupole
Lightweight, compact, robust, inexpensive, high scan speeds for chromatographic peaks
Disadvantages of single quadrupole
Low resolution, lower sensitivity in full scan mode
Tandem Mass Spectrometry
can be done in space (QQQ) or time (IT), combination of mass analyzers to increase the opportunity for additional fragmentation of a desired ion. Typically includes collision induced dissociation with a neutral gas
Triple quadrupole MS parts
Q1 used to isolate desired ion (precursor ion), Q2 where CID occurs (RF only), Q3 used to scan product ions to the detector
Triple quadratole process
Isolated ions in Q1 are accelerated into the collision cell (Q2). Kinetic energy of precursor ions is converted into internal energy through neutral collision gas. Energy redistribution leads to the formation of fragment ions and neutral losses (not detected). Product ions emerge from collision cell and are transported to Q3
QQQ scan modes
Product ion scan, precursor ion scan, neutral loss scan, multiple reaction monitoring
Product Ion scanning
Desired ion is isolated in Q1, isolated precursor ion is fragmented by CID in Q2, product ions are scanned to detector in Q3
Precursor Ion Scanning
Q1 is scanned across the desired mass range, ions are fragmented in the collision cell, only a single ion is isolated and transmitted to the detector in Q3. Useful for class-specific identifications of unknown analytes
Neutral Loss Scanning
Q1 and Q3 are scanned with a constant offset. Isolated ions in Q1 are fragmented in Q2, therefore, ions which loses a neutral molecule of the defined mass will be transmitted and detected. Useful to identify characteristics neutral losses
Multiple reaction monitoring
Q1 is set to a specific precursor ion, isolated precursor ion is fragmented in Q2, and Q3 is set to allow only a specific product ion to reach the detector. Reduces noise and increases sensitivity since it works as a double mass filter. Most useful for quantitative analysis.
Advantages of QQQ over single quadrupoles
Higher selectivity, better S/N and lower LOD, more reliable identification of detected analytes, wider linear range of quantitation, better accuracy and reproducibility (especially at lower concentrations), Different scan modes available
Exact Mass
theoretical value calculated by summing the exact masses of the individual isotopes of the molecule
Accurate Mass
Experimentally determined mass that allows the elemental composition to be determined by comparison with the calculated exact mass
Time of Flight MS
Separation of ions based on the time it takes to travel a known distance upon uniform acceleration, requires very low background pressure
Advantages of Quadrupole Time of Flight
High resolution MS, elemental formula determinations, simulataneous spectral acquisition, increased sensitivity from improved ion transmission, extended mass range.
Disadvantages of Quadrupole Time of Flight
Very expensive, requires space, requires tight control of lab environment
Electron Multiplier Detector
Ions are accelerated at a high velocity where they hit a dynode (has the oposite charge of the ion) which causes emission of secondary particles (cations or e-) secondary particles are amplified in the electron multiplier which creates a current
Tuning
The adjustment of parameters to achieve the best analytical response based on the analysis of a standard reference compound. Poor peak shapes are often due to this not being done well, which will result in incorrect identification of peak
Calibration
The process of assigning m/z values based on the known m/z values of standard compounds
Optimization
the adjustment of source-dependent parameters and the compound-dependent parameters to maximize the instrument response for a sample or compound
What is the main advantage of ambient ionization?
No sample prep or training, which saves money and time
What samples can DART analyze?
all phases
DART mechanism
Gas enters DART source (HE, nitrogen, Ar) where a high voltage is applied and generates an excited state species. Electrodes draw out these species (metastable are the only ones that pass through) which enters a heated region. Grid electrode prevents recombination and the insulator protects from high voltages. Reagent ions are formed from atmospheric compounds.
What gas creates the most reagent gases in DART MS?
He because it carries the most energy
What factors determine if an adduct or molecular ion is formed in DART MS?
Proton affinity and ionization energy
What is the tradeoff for using nitrogen or Argon instead of He in DART MS?
More selective but less sensitive
What are the two sampling techniques in DART MS?
Scanning and pulsing
Vapur Interface Advantages
makes DART compatible with different mass spectrometers and improves reproducibility by reducing turbulence
Vapur Interface disadvantages
Ion-molecule reactions can occur over longer transport distance, loss of non polar or reactive compounds, increased adduct formation, and carryover due to absorption onto ceramic tubing
Chromatography
Separation of components in a mixture based on different interactions between two phases
How is the partition coefficient defined?
concentration of sample in stationary phase over concentration of sample in mobile phase
How is retention time (tr) defined?
Time spent in stationary phase (ts) + dead time, essentially how long it takes a sample to elute, including the time it takes the solvent to work its way through the column
What is the retention factor?
Describes how well an analyze is retained by the stationary phase, and the higher the value the stronger the retention
What is the selectivity factor?
alpha, the ration of retention factors for two adjacent peaks (more strongly retained/less strongly retained) and a larger value indicates better separation
What are the optimal conditions for chromatography?
Short run, good separation, complete baseline separation, symmetry in peak shape, reproducible retention times
What can cause peak fronting?
large injection volume, poor solubility of sample, degradation of stationary phase, column saturation
What can cause peak tailing?
Secondary interactions, column overload, contaminants
What factors describe column efficiency?
Plate height (H), number of plates (N) and length of the column (N=L/H)
What does A represent in the van Deemter equation?
Multiple paths (eddy diffusion)
What does B represent in the van Deemter equation?
longitudinal diffusion, diffusion through the column
What does u represent in the van Deemter equation?
average mobile phase velocity
What does Cs represent in the van Deemter equation?
resistance to mass transfer of the stationary phase
What does Cm represent in the van Deemter equation?
Resistance ti mass transfer of the mobile phase
What effects Cs and Cm?
for Cs film thickness and diffusion coefficient and for Cm it’s the particle or column diameter
What specific number is considered baseline resolution?
1.5
What phases can be analyzed with GC?
All phases with proper preparation
What are ideal characteristics for GC samples?
Boiling point up to 400C, non polar or slightly polar, and must be thermally stable
What are some characteristics that might make a sample difficult to analyze on GC?
Does not vaporize at GC operation conditions, highly reactive, thermally unstable or strongly polar
Derivatization
Process of chemically modifying a compound to produce a new molecule suitable for GC analysis
Types of derivatization
Silylation, acylation, alkylation
Characteristics of carrier gases in GC
Must be inert, and lead to a small H, cost and safety are also important
List the carrier gases in order from best to worst resolution
hydrogen, He, nitrogen
Why does hydrogen and helium have a better resolution than nitrogen in GC?
solutes are able to diffuse rapidly through these gases due to decreased resistance to mass transfer
How is oxygen prevented from getting into the GC column?
Pure carrier gas and oxygen moisture traps or filters
How should the sample be introduced to the column in GC?
as a “plug of vapor”
Describe the steps for manual injection in GC?
Rinse syringe with solvent, rinse with sample, draw up desired volume, carefully insert syringe into septum, let needle heat up, then depress the plunger and pull the syringe out
What are some advantages of auto sampler injection?
More robust and reproducible, instantaneous injection
What is the inlet liner in GC?
In the injection port and enhances vaporization, guides the needle, and limits sample degredation