scatter
A change in direction of an xray photon after interacting with matter
Primary radiation
radiation emitted from the xray tube
Remnant radiation
unabsorbed and scattered xrays that exit the patient
Imaging forming xrays
photons that interact with the IR
INC scatter = ?
INC scatter = INC density/exposure
INC scatter = ? contrast
INC scatter = DEC contrast
compton scatter vs coherent/classical scatter
compton scatter = ionization takes place, interacts w valence/outer electron shell
coherent/classical scatter = nonionizing
biggest source of scatter for tech
patient
4 Factors that affect the amount of scatter reaching the IR
field size, thickness of tissue or body habitus, tube potential, anatomy
INC field size = ? scatter
INC field size = INC scatter (direct relationship)
INC thickness/body habitus = ? scatter
INC thickness/body habitus = INC scatter (direct relationship)
INC mas or kv = ? scatter
INC mas or kv = INC scatter (direct relationship)
Specific body part itself ? scatter
Specific body part itself (foot versus abdomen) (more dense body parts have more scatter)
beam limitation
same as collimation
beam limitation = ? pt exposure
DEC Patient exposure
beam limitation = ? image quality and how?
beam limitation = INC image quality by reducing scatter
beam limitation = ? visibility of detail
beam limitation = INC visibility of detail
beam limitation = ? image contrast
beam limitation IMPROVES image contrast
types of beam limiting devices
aperture diaphragms (similar to collimator)
cone
automatic collimator
aperture diaphragms
similar to collimator
simplest, lease expensive
set size, not adjustable (piece of cut out lead)
large amount of blur
aperture diaphragms = ? penumbra
aperture diaphragms = INC penumbra
aperture diaphragm dedicated units
dental units, chest rooms
2 types of cones
flared cone
cylinder cone
flared cone
doesnt really restrict beam, matches divergence of beam
flare is usually larger than the beam
cylinder cone
blur is reduced
circular field
excessive radiation
cone disadvantages (2)
image comes out round and IR is rectangular
if the image fills the entire IR, then parts of the patient may be unnecessarily irradiated
variable aperture
this is what we use
collimator
beam limiting device
positive beam limiting device PBL or automatic collimator
limits the radiation field size so it cannot be larger than the IR in the bucky
automatic collimator
allows for infinite number of field sizes
provides a light field
scale for field sizes
adds to ADDED filtration (1 mm Al collimator box)
upper/entrance shutters control stem radiation
how many number of field sizes are allowed with automatic collimator?
infinite
automatic collimator adds to what type of filtration?
added filtration (1 mm Al from collimator box)
which shutters of automatic collimator controls stem radiation?
upper/entrance shutters controls stem radiation
which shutters of automatic collimator reduce blur?
lower shutters/lead leaf shutters reduce blur
stem radiation
AKA off focus radiation OR extrafocal radiation
xr photons that are not produced at focal spot
can be produced in tube housing or evaporated tungsten on glass envelope
which shutters do we control of automatic collimator?
lower shutters AKA lead leaf shutters
what does lower shutter of automatic collimator affect?
lower shutters / lead leaf shutter REDUCE blur
what does upper shutter of automatic collimator affect?
controls stem radiation AKA off focus radiation AKA extrafocal radiation