Microfluidic cell-counting chip
whole blood → lysing buffer lyses RBCs →quenching buffer neutralizes hemoglobin from lysed cells → counting of lymphocytes and other white blood cells → capure chamber → exit counter
Used for HIV/AIDS CD4+ counting
Microcantilever biosensor
DNA oligonucleotic probes immobilized on a cantilever, target DNA hybridization causing bending of the cantilever, which can be measured to detect specific DNA sequences.
Nanopore biosensors
quantify DNA translocation through nanoscale pores by measuring ionic current changes as DNA molecules pass through.
Viral Load on a Chip
blood → metering → cell lysing → RT-LAMP amplification kit → droplets on dried primers → cell-phone fluorescence detection
higher viral loads = shorter threshold time
Microfluidic CD4+ capture chip
whole blood introduced → CD4+ cells captured via controlled flow → listed under microscope
sensitive to ~20 cells/µL in a 10 µL device volume; governed by shear stress at the chamber wall: τₗ = 6μQ / (h²w)
Biosensore from Lab to Market (Daktari-style workflow)
prick finger → apply blood drop to disposable cartridge → insert cartridge into reader unit → washing, capture, measurement → CD4 count displayed