What is an orientation notch?
A marker on the transducer that corresponds to the orientation marker displayed on the ultrasound screen.
What is a preset?
A group of preselected machine settings designed for a specific examination type.
What does the transducer icon identify?
The probe being used for the examination.
What does frame rate measure?
The number of ultrasound images displayed per second.
What happens when frame rate increases?
Motion is displayed more smoothly, although other imaging parameters may need to be adjusted.
What does overall gain control?
The brightness of the entire ultrasound image.
What does dynamic range control?
The range of echo amplitudes displayed as shades of gray.
What happens when dynamic range increases?
More shades of gray are displayed, producing a smoother image with lower contrast.
What happens when dynamic range decreases?
Fewer shades of gray are displayed, producing higher contrast.
What does Harmonics/Fusion represent?
Imaging features that combine or process ultrasound frequencies to improve image quality.
What is the compression bar?
A display indicator associated with compression or grayscale mapping.
What is a depth marker?
A screen marker that identifies depth within the ultrasound image.
What does the TGC curve represent?
The time gain compensation settings at different depths.
What does the focus marker indicate?
The depth at which the ultrasound beam is focused.
What is maximum depth?
The greatest depth displayed in the ultrasound image.
What does the thermal index (TI) indicate?
An estimate related to the potential for tissue heating.
What does the mechanical index (MI) indicate?
An estimate related to the potential for mechanical bioeffects.
Why are TI and MI important?
They help monitor ultrasound output and support safe scanning practices.
What is a sample gate?
The selected location where pulsed-wave Doppler measures blood-flow velocities.
What is a Color Doppler box?
The region of the ultrasound image where color flow information is displayed.
What does the Nyquist limit represent?
The maximum Doppler shift that can be measured without aliasing, equal to half the PRF.
What is the Color bar?
The scale that indicates the colors assigned to blood-flow direction and velocity.
What is the Color baseline?
The reference line separating positive and negative Doppler shifts on the color or spectral display.
What is Color gain?
The amplification of Doppler signals in the Color Doppler image.
What is pulse gain?
The amplification of the pulsed Doppler signal.
What is the spectral scale?
The scale displaying Doppler velocity measurements.
What are time markers?
Markers used to identify time along the Doppler waveform display.
What happens if Color gain is too high?
Color noise and color blooming can occur.
What happens if Color gain is too low?
Blood flow may be underrepresented or difficult to detect.
What happens when the Nyquist limit is exceeded?
Aliasing occurs.
Which Doppler mode uses a sample gate?
Pulsed Wave Doppler.
Which Doppler mode continuously transmits and receives signals?
Continuous Wave Doppler.
What is the difference between the Color Doppler box and sample gate?
The Color Doppler box identifies the region displaying color flow, while the sample gate identifies the specific location sampled by PW Doppler.