what are the 3 main part of a typical fluoroscopic unit?

1. tube and generator

2. Image intensifier OR Flat panel detector

3. video/image monitoring system

2
New cards

True or false: fluoroscopic tubes are similar to diagnostic tubes except that they are designed to operate for LONGER periods of time at a much LOWER mA.

true

3
New cards

typical diagnostic tubes operate at _____ mA

50-1,200

4
New cards

fluoroscopic ma rang is _________mA

0.5-5

5
New cards

RF systems do both _______

radiography and fluoroscopy

6
New cards

when the tube is under the patient, backscatter is directed ______

downward

7
New cards

when the tube is above the patient, scattered radiation is going towards the operators _____ and _______

eyes; thyroid

8
New cards

the image intensifier was designed to electronically amplify the ______ of an image

brightness

9
New cards

what part of the image intensifier is a vacuum tube that allows the free flow of electrons from the photocathode to the anode?

glass envelope

10
New cards

what is the input phosphor made of?

cesium iodide

11
New cards

what part of the image intensifier is made of cesium iodide?

input phosphor

12
New cards

what part of the image intensifier converts the x-ray energy into light energy?

input phosphor

13
New cards

what part of the image intensifier is connected to the photocathode

input phosphor

14
New cards

what part of the image intensifier is a thin sheet of metal made of cesium and antimony?

photocathode

15
New cards

what part of the image intensifier converts light energy into electron energy via photoemission?

photocathode

16
New cards

what is the photocathode made of?

cesium and antimony

17
New cards

what part of the image intensifier are positively charged metal plates that focus and accelerate the electrons?

electrostatic focusing lenses

18
New cards

true or false: the electrostatic focusing lenses help magnify the image during "mag mode" in multifield image intensifiers?

true

19
New cards

what part of the image intensifier is a positive terminal to attract the electrons from the photocathode toward the output phosphor?

accelerating anode

20
New cards

what is the reason that electrons accelerate to the anode in an image intensifier?

a 25-30 kV potential difference

21
New cards

what part of the image intensifier is made up of glass and emits green light when struck by electrons?

output phosphor

22
New cards

what part of the image intensifier is made of zinc cadmium sulfide?

output phosphor

23
New cards

what is the output phosphor made of?

zinc cadmium sulfide

24
New cards

what part of the image intensifier converts electron energy into light energy?

output phosphor

25
New cards

what is a setting of the image intensifier that increases the voltage to the electrostatic focusing lenses?

magnification mode

26
New cards

when "mag mode" is used, field of view _______

decreases

27
New cards

what "mag mode" is used, spatial resolution ______

increases

28
New cards

when "mag mode" is used, patient dose _____

increases

29
New cards

the _______the voltage applied to the electrostatic lenses, the GREATER the acceleration and the CLOSER the focal point moves to the input phosphor.

greater

30
New cards

Magnification will increase patient dose because the _______ will increase the tube output to compensate for the loss of electrons with in image intensifier

ABC (automatic brightness control)

31
New cards

what is the equation for magnification?

input screen diameter/diameter of input being used during magnification

32
New cards

when magnification mode is used during fluoroscopy, which of the following occurs?

patient dose increases

33
New cards

what part of the fluoroscopic unit maintains a consistent overall appearance of the image by automatically adjusting the kVp and/or mAs?

automatic brightness control

34
New cards

Automatic brightness control: the beam is regulated by varying the _____

kVp, mA and pulse time

35
New cards

True or false: minification gain is an increase in brightness or intensity.

true

36
New cards

why does magnification gain make the image appear brighter?

because the same number of electrons are being concentrated on a smaller surface area

37
New cards

what is the equation for minification gain?

input screen diameter^2/output screen diameter^2

38
New cards

what is defined as the number of photons generated at the output phosphor for every photon generated at the input phosphor?

flux gain

39
New cards

what represents the tube's conversion efficiency?

flux gain

40
New cards

what is a measurement of the increase in image intensity achieved by an image intensifier tube?

total brightness gain

41
New cards

what determines the total brightness gain?

minification gain and flux gain

42
New cards

what is the equation for brightness gain?

minification gain x flux gain

43
New cards

all of these are true about the grids used in fluoroscopic units EXCEPT for?

a non-focused type of grid is used

44
New cards

what are some quality factors of the fluoroscopic unit?

contrast

conventional fluoro resolution

distortion

quantum mottle

45
New cards

contrast is controlled by ______

increasing the amplitude of the video signal

46
New cards

true or false: digital fluoro system can also use post-processing functions to improve image contrast.

true

47
New cards

Image Intensifier fluoro contrast is affected by:

scatter

penumbral light scatter in the input and output screens

light scatter within the image intensifier tube itself

48
New cards

size distortion in fluoro is caused by?

OID

49
New cards

shape distortion in fluoro is caused primarily by?

the way the tube is shaped

50
New cards

what kind of distortion distorts appearance by increasing magnification towards the center?

barrel distortion

51
New cards

what kind of distortion distorts appearance by increasing magnification towards the periphery?

pincushion distortion

52
New cards

what kind of distortion is a decrease in image brightness at the periphery, compromising up to 8-10% of the image area?

vignetting

53
New cards

what kind of distortion is a warping of the image along the S-shaped axis due to a strong magnetic field?

S distortion

54
New cards

what is a blotchy or grainy appearance caused by insufficient radiation to create the image?

quantum mottle

55
New cards

Flat-panel fluoroscopy designs convert ______ to _____ then to _______

x-rays; light; electrical charge

56
New cards

what are the components of a flat-panel fluoroscopy tube?

scintillation layer

photodiode (photocathode)

thin flat panel transistor (TFT)

57
New cards

True or false: flat-panel fluoroscopic tubes go through these steps

xrays-> xray scintillator (creating light) -> photocathode ->thin film transistor (TFT)-> analog signal -> analog-to-digital converter -> digital data

true

<p>true</p>
58
New cards

in a flat-panel fluoro design, the scintillation layer is made of?

cesium iodide or gadolinium

59
New cards

FLAT PANEL: the scintillation layer converts _____ into _____

xray photons; light photons

60
New cards

FLAT PANEL FLUORO: the photodiode is made of ?

amorphous silicon

61
New cards

FLAT PANEL FLUORO: the photodiode converts ____into ______ that goes to the TFT

light photons; electrical charge

62
New cards

what is a matrix of detector elements (DEL) called?

Thin Flat Panel Transistor (TFT)

63
New cards

in a DEL, what captures information that becomes a pixel in the display image?

sensing area

64
New cards

in a DEL, what stores the charge?

storage capacitor

65
New cards

in a DEL, what is an electrical switch that controls when to read out the charge?

thin film transistor (TFT)

66
New cards

true or false: with a flat panel detector, the same image magnification is achieved with no increase in radiation.

true

67
New cards

what is the most important benefit of the Flat panel designs when it comes to image quality?

does not suffer from the many inherent limitations of an image intensifier (vignetting and other shape distortions)

68
New cards

which of the following statements are accurate regarding fluoroscopic systems that use flat panel detectors for fluoroscopy?

flat panel detectors replace the image intensifier tube

69
New cards

what is the sequence of energy conversions that occur in a flat panel detector during a fluoroscopic procedure?

x-rays to light photons to electrical signals

70
New cards

what function of the fluoroscopic unit maintains the last real-time image until it is replaced by the unit being activated again?

last image hold

71
New cards

how does last image hold achieve a still image during a fluoroscopic procedure?

last image hold enables the last live image to be displayed when the exposure is terminated

72
New cards

Federal mandate for the exposure rate for conventional fluoroscopy is ____

100 mGy/min

73
New cards

with pulsed fluoroscopy, beam width is determined by:

beam on time

74
New cards

in pulsed fluoroscopy, pulse height is determined by:

the determined mA, usually set by the ABC

75
New cards

In pulsed fluoroscopy, pulse interval is determined by:

beam off time

76
New cards

which of the following fluoroscopy modes is characterized by a continuous x-ray beam while the exposure switch is activated?

conventional

77
New cards

which fluoroscopic mode can be compared to a strobe light in terms of how it operates?

pulsed

78
New cards

in pulsed fluoroscopy, the fluoroscopic mA value is expressed as which of the following?

pulse height

79
New cards

what is the primary difference between pulsed and intermittent fluoroscopy?

intermittent fluoro has the beam turned on and off at non-regular intervals that are controlled by the operator

80
New cards

what are some essential features to a mobile x-ray unit?

6-foot exposure cord

dead-man switch

clear display of exposure factors

81
New cards

why are mobile exams difficult to accomplish?

exam conditions can vary widely

82
New cards

if a chest radiograph requires a mAs of 4 at 72", how would the mAs need to be adjusted to maintain IR exposure if the SID is changed to 56"?

decrease the mAs

83
New cards

which of the following is NOT an essential feature of portable machines?

automatic exposure control

84
New cards

how many lead aprons should be carried to a mobile examination?

two, one for the patient and one for the technologist

85
New cards

as you prepare to complete a portable chest x-ray on a patient in the CVICU, you notice several tubes and lines going into the patient. what should you do prior to the exposure?

ask a nurse to move the lines from the area being imaged