what are the 3 main part of a typical fluoroscopic unit?
1. tube and generator
2. Image intensifier OR Flat panel detector
3. video/image monitoring system
True or false: fluoroscopic tubes are similar to diagnostic tubes except that they are designed to operate for LONGER periods of time at a much LOWER mA.
true
typical diagnostic tubes operate at _____ mA
50-1,200
fluoroscopic ma rang is _________mA
0.5-5
RF systems do both _______
radiography and fluoroscopy
when the tube is under the patient, backscatter is directed ______
downward
when the tube is above the patient, scattered radiation is going towards the operators _____ and _______
eyes; thyroid
the image intensifier was designed to electronically amplify the ______ of an image
brightness
what part of the image intensifier is a vacuum tube that allows the free flow of electrons from the photocathode to the anode?
glass envelope
what is the input phosphor made of?
cesium iodide
what part of the image intensifier is made of cesium iodide?
input phosphor
what part of the image intensifier converts the x-ray energy into light energy?
input phosphor
what part of the image intensifier is connected to the photocathode
input phosphor
what part of the image intensifier is a thin sheet of metal made of cesium and antimony?
photocathode
what part of the image intensifier converts light energy into electron energy via photoemission?
photocathode
what is the photocathode made of?
cesium and antimony
what part of the image intensifier are positively charged metal plates that focus and accelerate the electrons?
electrostatic focusing lenses
true or false: the electrostatic focusing lenses help magnify the image during "mag mode" in multifield image intensifiers?
true
what part of the image intensifier is a positive terminal to attract the electrons from the photocathode toward the output phosphor?
accelerating anode
what is the reason that electrons accelerate to the anode in an image intensifier?
a 25-30 kV potential difference
what part of the image intensifier is made up of glass and emits green light when struck by electrons?
output phosphor
what part of the image intensifier is made of zinc cadmium sulfide?
output phosphor
what is the output phosphor made of?
zinc cadmium sulfide
what part of the image intensifier converts electron energy into light energy?
output phosphor
what is a setting of the image intensifier that increases the voltage to the electrostatic focusing lenses?
magnification mode
when "mag mode" is used, field of view _______
decreases
what "mag mode" is used, spatial resolution ______
increases
when "mag mode" is used, patient dose _____
increases
the _______the voltage applied to the electrostatic lenses, the GREATER the acceleration and the CLOSER the focal point moves to the input phosphor.
greater
Magnification will increase patient dose because the _______ will increase the tube output to compensate for the loss of electrons with in image intensifier
ABC (automatic brightness control)
what is the equation for magnification?
input screen diameter/diameter of input being used during magnification
when magnification mode is used during fluoroscopy, which of the following occurs?
patient dose increases
what part of the fluoroscopic unit maintains a consistent overall appearance of the image by automatically adjusting the kVp and/or mAs?
automatic brightness control
Automatic brightness control: the beam is regulated by varying the _____
kVp, mA and pulse time
True or false: minification gain is an increase in brightness or intensity.
true
why does magnification gain make the image appear brighter?
because the same number of electrons are being concentrated on a smaller surface area
what is the equation for minification gain?
input screen diameter^2/output screen diameter^2
what is defined as the number of photons generated at the output phosphor for every photon generated at the input phosphor?
flux gain
what represents the tube's conversion efficiency?
flux gain
what is a measurement of the increase in image intensity achieved by an image intensifier tube?
total brightness gain
what determines the total brightness gain?
minification gain and flux gain
what is the equation for brightness gain?
minification gain x flux gain
all of these are true about the grids used in fluoroscopic units EXCEPT for?
a non-focused type of grid is used
what are some quality factors of the fluoroscopic unit?
contrast
conventional fluoro resolution
distortion
quantum mottle
contrast is controlled by ______
increasing the amplitude of the video signal
true or false: digital fluoro system can also use post-processing functions to improve image contrast.
true
Image Intensifier fluoro contrast is affected by:
scatter
penumbral light scatter in the input and output screens
light scatter within the image intensifier tube itself
size distortion in fluoro is caused by?
OID
shape distortion in fluoro is caused primarily by?
the way the tube is shaped
what kind of distortion distorts appearance by increasing magnification towards the center?
barrel distortion
what kind of distortion distorts appearance by increasing magnification towards the periphery?
pincushion distortion
what kind of distortion is a decrease in image brightness at the periphery, compromising up to 8-10% of the image area?
vignetting
what kind of distortion is a warping of the image along the S-shaped axis due to a strong magnetic field?
S distortion
what is a blotchy or grainy appearance caused by insufficient radiation to create the image?
quantum mottle
Flat-panel fluoroscopy designs convert ______ to _____ then to _______
x-rays; light; electrical charge
what are the components of a flat-panel fluoroscopy tube?
scintillation layer
photodiode (photocathode)
thin flat panel transistor (TFT)
True or false: flat-panel fluoroscopic tubes go through these steps
xrays-> xray scintillator (creating light) -> photocathode ->thin film transistor (TFT)-> analog signal -> analog-to-digital converter -> digital data
true

in a flat-panel fluoro design, the scintillation layer is made of?
cesium iodide or gadolinium
FLAT PANEL: the scintillation layer converts _____ into _____
xray photons; light photons
FLAT PANEL FLUORO: the photodiode is made of ?
amorphous silicon
FLAT PANEL FLUORO: the photodiode converts ____into ______ that goes to the TFT
light photons; electrical charge
what is a matrix of detector elements (DEL) called?
Thin Flat Panel Transistor (TFT)
in a DEL, what captures information that becomes a pixel in the display image?
sensing area
in a DEL, what stores the charge?
storage capacitor
in a DEL, what is an electrical switch that controls when to read out the charge?
thin film transistor (TFT)
true or false: with a flat panel detector, the same image magnification is achieved with no increase in radiation.
true
what is the most important benefit of the Flat panel designs when it comes to image quality?
does not suffer from the many inherent limitations of an image intensifier (vignetting and other shape distortions)
which of the following statements are accurate regarding fluoroscopic systems that use flat panel detectors for fluoroscopy?
flat panel detectors replace the image intensifier tube
what is the sequence of energy conversions that occur in a flat panel detector during a fluoroscopic procedure?
x-rays to light photons to electrical signals
what function of the fluoroscopic unit maintains the last real-time image until it is replaced by the unit being activated again?
last image hold
how does last image hold achieve a still image during a fluoroscopic procedure?
last image hold enables the last live image to be displayed when the exposure is terminated
Federal mandate for the exposure rate for conventional fluoroscopy is ____
100 mGy/min
with pulsed fluoroscopy, beam width is determined by:
beam on time
in pulsed fluoroscopy, pulse height is determined by:
the determined mA, usually set by the ABC
In pulsed fluoroscopy, pulse interval is determined by:
beam off time
which of the following fluoroscopy modes is characterized by a continuous x-ray beam while the exposure switch is activated?
conventional
which fluoroscopic mode can be compared to a strobe light in terms of how it operates?
pulsed
in pulsed fluoroscopy, the fluoroscopic mA value is expressed as which of the following?
pulse height
what is the primary difference between pulsed and intermittent fluoroscopy?
intermittent fluoro has the beam turned on and off at non-regular intervals that are controlled by the operator
what are some essential features to a mobile x-ray unit?
6-foot exposure cord
dead-man switch
clear display of exposure factors
why are mobile exams difficult to accomplish?
exam conditions can vary widely
if a chest radiograph requires a mAs of 4 at 72", how would the mAs need to be adjusted to maintain IR exposure if the SID is changed to 56"?
decrease the mAs
which of the following is NOT an essential feature of portable machines?
automatic exposure control
how many lead aprons should be carried to a mobile examination?
two, one for the patient and one for the technologist
as you prepare to complete a portable chest x-ray on a patient in the CVICU, you notice several tubes and lines going into the patient. what should you do prior to the exposure?
ask a nurse to move the lines from the area being imaged