Dynamic range
the number of signal values the detector can capter
the greater the number of signal values that a receptro can capture the greater the receptors dynamic range
gray scale
the range of pixel values present in a displayed image
the greater the dynamic range, the longer the gray scale in the iimage
more shades of gray
exposure latitude
the range of exposure factors that will create an acceptable image
look up table
for each naatomical type of procedure an ideal dynamic rage and average gray level (brightness) is preprogrammed into the system in a look up table
image analysis consist of the following steps
recognizing the image
partitioned patter recognition (aka segmentation)
exposure field recognition (EFR)
histogram constructuion
histogram analysis
rescaling (adjusting image)
what can cause the image to be displayed improperly
desnitits (noise) outside the exposure areas from scatter radiation, fog or electrical interference
exposure field recognition (EFR)
both CR and DrR use to distinguish between dark structures within the anatomy and dark structures outside the anatomy
EFR goal is to keep only the useful anatomy data and ignore everything else
exposure field recognition is used
when only one exposure is make on the image receptor
Process of exsposure field recognition
during the process of recognition, the computer identifies the exposure field, the edges of the image and all exposure data outside of this field is excluded from the histogram
ideally all four edges of a collimated field are recognized
if at three edges are not identified then all data including raw exposure or scatter outside of the field may be included in the histogram, resulting in a histogram analysis error
this is done by recognizing landmarks in the image histogram, so it is technically part of histogram analysis
histogram characteristics
graphical computation of pixel brightness values
arranges signals values from minimum to maximum values
shape of histogram varies depending upon anatomy and exposure
used to identify pixel values of interest and exposure index
what is a histogram
a bar graph
represents the number of digital pixel values versus the relative prevalence of those values in the image
x axis on histogram
represents the amount of exposure
y axis on histogram
represents the incidence of pixels for each exposure level
sampling frequiences
different sampling frequencies represented by the number of bars that make up each histogram
higher sampling frequency =
more bars packed cose together
higher the sampling rate =
more accurate the histogram
how many lobes does the most common type of histogram have?
2
the main lobe =
tissues with anatomy
Smin =
start
Smax =
stop
the tail lobe =
“raw” background exposure
right of Smax, must be eleiminated from averaging calculations used for the exposure indicator and for rescaling
SMAX is not used on exams when?
there is not bakckground air
examples:
abdomen
lumbar spine
body parts that cover whole IR
what must be eliminated from averaging calculations for the exposure indicator?
both raw background densities and large barium areas that are nearly black
voxels of interest (VOI)
information the computer wants to keep
this tells the computer which information is important and which cna be ignored whne copmleting the rescaling calculations
goal of defining voi
keep anatomy of interest
ignore scatter, background exposure, unwated data
how does the computer scan histogram data
inward from both the left and right ends of the histogram
from each end the first bin containing non zero pixel values is identified, the 1st point at th eleft is called Smin
how is Smax identified
from within the data curve
represents the darkest density within the body part
Save
the average densitiy within the anatomy
used to calculate the exposure indicator
what is histogram analysis
the comparison of exposure data in an imaging system to standardized data in a look up table
important to choose correct anantomic region on the menu bef exposing the patient
raw data used to form the histgram are compared with a “normal” histogram of the same body part by the computer
resclaing
modifing the numbers in the histogram to make it match the standard
done if the numbers in the histogram are higher or lower than the standard
rescaling process
automatically changes the brightness and gray scale of the digital image so that it appears like a normal radiogrpah
rescaling is simply a process of algebraically “re-labeling” incoming data
can the computer align the average brightness of the image?
yes
the computer finds the difference in average brightness between the reference histogram (permanent LUT) and the new exposure, then subtracts this from all values
can the computer force the histogram to match a perfect histogram by changing how many pixels exist in each brightness category
moving pixels around image
creating new anatoym infor
deleting anatomy info
no
computers cannot invent info that was never captures