Linear Accelerator QA

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/25

encourage image

There's no tags or description

Looks like no tags are added yet.

Last updated 1:53 PM on 10/13/26
Name
Mastery
Learn
Test
Matching
Spaced
Call with Kai
Chat

No analytics yet

Send a link to your students to track their progress

26 Terms

1
New cards

acceptance testing

•performed upon equipment installation

•performed by physicist and vendor representative

2
New cards

commissioning

•performed by a qualified physicist

•establishes baselines

3
New cards

quality assurance testing

•performed by physicist

•equipment calibrated by Accredited Dosimetry Calibration Laboratory (ADL)

4
New cards

spot checks

•performed by therapists

•equipment less precise

•use array detectors

•multiple tests in 1 exposure (output, symmetry, constancy checks)

5
New cards

daily QA dosimetry

•xray output constancy: 3%

•electron output constancy: 3% (all energies at least 2x weekly)

6
New cards

daily QA mechanical

•lasers: 2 mm (non IMRT), 1.5 mm (IMRT), 1mm (SBRT)

•ODI: 2mm

•collimator size: 2 mm (non IMRT/IMRT), 1 mm (SBRT)

7
New cards

daily QA safety

•door interlock: functional

•audiovisual monitor: functional

8
New cards

daily QA checks

•water level/temp/pressure

•gas pressure (SF6)

•general conditions

•tx devices

•supplies

9
New cards

monthly QA check dosimetry

•xray output constancy: 2%

•xray beam profile constancy: 1%

•electron output constancy: 2%

•electron beam profile constancy: 1%

•backup monitor: 2%

•dose rate constancy: 2%

10
New cards

monthly QA check safety

•emergency off: functional

•wedge interlocks: functional

11
New cards

monthly QA check mechanical

•lasers: 2mm (non IMRT), 1mm (IMRT), 1mm (SRS/SBRT)

•light vs xray: 2mm/1mm (asym)

•FS indicators: 2mm/1mm (asym)

•tx couch position: 2mm/1 degree

•gantry/collimator: 1 degree

•cross hair center: 1 mm

•light field intensity: functional

•tray/wedge latches: functional

12
New cards

temperature and pressure correction

•most ion chambers unsealed

•chamber sensitivity changes with temp and pressure

•increased temp, reading goes up

•increased pressure, reading goes down

=(reading x temp K x 760 mmHg x Cal factor)/(exp time x 295 K x pressure in mmHg)

13
New cards

monthly QA respiratory gating

•beam output constancy: 2%

•phase, amplitude beam control: functional

•in room respiratory monitoring system: functional

•gating interlock: functional

14
New cards

annual QA check dosimetry

•xray flatness/symmetry: 1%

•electron flatness/symmetry: 1%

15
New cards

annual QA check safety

•per manufacturer

16
New cards

annual QA check mechanical

•collimator rotation iso: 1 mm dia

•coincidence of radiation and mechanical iso: 2 mm dia

•table top sag: 2 mm

•table angle: 1 degree

•electron applicator interlocks: functional

17
New cards

stability of isocenter

•stable: convergence is at iso

•wobble: nonconvergence at iso

•split field test: used to QA POP fields

18
New cards

split field test

•used to QA POP fields

•materials: film, plastic tray on 4" feet, 2 shielding blocks

19
New cards

split field test technique

•gantry @ 0, film on tray

•shielding block on top of film

•shielding block under tray, opposite side/jaw

•expose film at 0 and 180

20
New cards

couch isocenter

•technique: gantry at 0, reference mark with table at iso height, rotate thru various angles

•not to exceed 1mm

•other sources of error: bad reference mark, collimator, light crosshair alignment

21
New cards

machine dose rate constancy

•constancy: machine output, cGy/MU

•spot check confirms baseline parameters

•often combined with flatness, symmetry, beam penetration

•possible reasons: magnetron current, electron gun current, change in monitor ion chamber

•instruments: ion chamber, diodes, film , TLD

22
New cards

flatness

•maximum difference from CA dose over 80% field dimension

•defined at a depth of 10cm

•Max= (Ymax-Yo)/(Yo)x100%

•Min= (Ymin-Yo)/(Yo)x100%

23
New cards

symmetry

•difference in dose rate between any 2 points within 80% of the field dimension

=(Y1-Y2)/Yo x 100%

24
New cards

flattening filter free

•flattening filter is removed in beam to attenuate photons

25
New cards

beam penetrative quality

•percentage depth dose/HVL (established upon commissioning using precision instruments)

•spot check (standardized SSD and FS)

•materials: tissue equivalent, ion chamber/diodes

•technique: 2 measurements, compare ratio to established baseline usually at dmax and some depth

26
New cards

imagers QA kV and MV

•daily: collision/door interlock, warning lights/sounds, laser/image/tx iso coincidence (2mm), phantom localization and couch shift

•monthly: kV/MV/laser alignement (1mm), high contrast spatial resolution, CT# accuracy

•annual: imaging dose and xray generator performance