Impregnation (Infiltration)
is the process whereby the clearing agent is completely removed from the tissue and replaced by a medium that will completely fill all the tissue cavities and give a firm consistency to the specimen.
This allows easier handling and cutting of suitably thin sections without any damage or distortion to the tissue and its cellular components.
Embedding
is the process by which the impregnated tissue is placed into a precisely arranged position in a mold containing a medium which is then allowed to solidify.
Casting or Blocking
Embedding is also termed as?
Soluble in processing fluids
suitable for sectioning and ribboning
molten between 30C and 60C
Translucent or transparent; colorless
stable
homogenous
capable of flattening after ribboning
non-toxic
odorless
easy to handle
inexpensive
An infiltrating medium should be?
Four types of impregnation and embedding medium
Paraffin wax
Celloidin (Collodion)
Gelatin
Plastic
Paraffin wax
Is the simplest, most common and best embedding medium used for routine tissue processing.
At its melting point, it tends to be slightly viscous, but this decreases as the temperature is increased.
Wax hardness (Viscosity)
What depends on the molecular weight of the components and the ambient temperature
60C to 65C
To decrease viscosity and improve infiltration of the tissue, technologists often increase the temperature to above
39C to 68C
High molecular weight mixtures melt at higher temperatures than waxes comprised of lower molecular weight fractions.
Paraffin wax is traditionally marketed by its melting points which range from?
Paraffin
Solid at room temperature but melts at temperatures up to about 65C or 70C.
It can be purchased with melting points at different temperatures
56C to 58C
The most common degree Celsius for histological use of paraffin wax is about what?
Tissue wax adhesion
______ depends upon the crystal morphology of the embedding medium.
Paraffin wax
_____ is a polycrystalline mixture of solid hydrocarbons produced during the refining of coal and mineral oils.
close packing
Small, uniform sized crystals provide better physical support for specimens through _____
Crystalline morphology
What can be altered by incorporating additives which result in a less brittle, more homogeneous wax with good cutting characteristics.
There is consequently less deformation during thin sectioning. Setting temperature does not appreciably affect crystal size
Thin individual serial sections may be cut with ease from the majority of tissues without distortion
The process is very rapid, allowing sections to be prepared within 24 hours.
Tissue blocks and unstained mounted sections may be stored in paraffin for an indefinite period of time after impregnation without considerable tissue destruction.
Because formalin fixed, paraffin-embedded tissues may be stored indefinitely at room temperature and nucleic acids (Both RNA And DNA) may be recovered from them decades after fixation, they are an important resource for historical studies in medicine
many staining procedures are permitted with good results
What are the advantages of paraffin wax impregnation
Overheated paraffin makes the specimen brittle
Prolonged impregnation will cause excessive tissue shrinkage and hardening, making the cutting of sections difficult.
Inadequate impregnation will promote retention of the clearing agent. Tissues become soft and shrunken, and tissue blocks crumble when sectioned and break up when floated out in a water bath.
Tissues that are difficult to infiltrate, e.g. bones, teeth, brains and eyes, need long immersion for proper support; otherwise they will crumble on sectioning. Prolonged immersion in paraffin, on the other hand, is not advisable.
Paraffin processing is not recommended for fatty tissues. The dehydrating and clearing agents used in the process dissolve and remove fat from the tissues
What are the disadvantages of paraffin wax impregnation
After being completely cleared, the tissue is submerged in two or more changes of melted paraffin wax, either in a paraffin oven or in an incubator which has been regulated at ___ - ___°C.
Size and type of tissues: Longer time is required for thicker tissues.
Use of vacuum imbedding: Vacuum reduces the time required for complete impregnation.
Clearing agent employed
The duration and number of changes required for thorough impregnation of tissue depends on:
45°C, 52°C, 56°C and 58°C
Common waxes have melting points of
20-24°C, 54-58°C
In a laboratory with temperature ranging from ___°C - ___°C, paraffin wax with a melting point of ___-___ is indicated.
15-18°C, 50 and 54°C
If the laboratory temperature is between ______, the melting point of wax to be used should be between _____.
Manual Processing
Automatic processing
vacuum embedding
Three ways impregnation and embedding is performed?
Manual Processing
At least four changes of wax are required at 15 minutes intervals in order to insure complete removal of the clearing agent from the tissue. The specimen is then immersed in another fresh solution of melted paraffin for approximately 3 hours to insure complete embedding or casting of tissue.
Manual prcessingns
What process is shown here?

Automatic Processing
This method makes use of an automatic tissue processing machine (i.e., Autotechnicon) which fixes, dehydrates, clears and infiltrates tissues, thereby decreasing the time and labor needed during the processing of tissues. This results in a more rapid diagnosis with less technicality. Usually, only 2- 3 changes of wax are required to remove the clearing agent and properly impregnate the specimen. This is made possible due to constant tissue agitation which accelerates and improves tissue penetration giving rise to more consistent results.
Example Elliot bench type processor
Give an example of an automatic tissue processing machine
Transfer arm
it is controlled by electrical current moves the tissues from one processing reagent to another (by clock schedules).
It can be removed by raising a spring-loaded plunger in the center of the cover plate, thereby allowing the tissue to be arranged manually anytime during the processing.
Agitation of fluid
is accompanied by a continuous vertical movement or rotation of the specimen carrier by a mechanism connected to the transfer arm.
Elliott Bench-Type Processor.
This machine is mounted on rollers to permit the turning of platforms and easy access to beakers and wax baths. It makes use of 12 individual processing steps, with ten 1-liter capacity glass beakers and two thermostatically controlled wax baths with a safety device cut-out switch to protect the wax against over-heating.
Electrical clock
connected to a metal disc notched in positions of 15 minutes or more, serves to control the time needed for each processing step.
rotates and sets the transfer arm and mechanism into motion, moving the tissue to the next position. A delay mechanism is provided in instances where processing time may exceed 24 hours
Precautions with Automatic Processing
The frequency with which fluids are changed depends on the number and sizes of the tissues processed. The presence of any odor in the clearing agent during final paraffin wax bath indicates that the paraffin wax needs to be changed. What kind of activity is being done here?
Dehydrating fluids
should be changed frequently since _____ is the most critical stage of tissue processing and inadequate _____ is difficult to correct once the tissue is in paraffin.
once
The first 100% ethanol bath should be discarded, and the others moved down, so that the final bath has fresh 100% ethanol after two complete processing runs of loads of at least three-quarters capacity. The clearing agent and the dilute ethanols should be changed at least ___ a week.
To avoid spillage, 3 degrees
_____ _______, fluid and wax containers must be filled to the appropriate level and correctly located in the machine. Wax accumulating on any surface or beaker leads must be removed and any spillage should be wiped away. Wax bath thermostats should be set at least _ _______ above the melting point of the wax, and timing should be checked when loading the machine, especially if the machine is equipped with a delay mechanism.
Due to the viscosity of molten paraffin wax, Due to the viscosity of molten paraffin wax
___ _____ _______, some form of gentle agitation is highly desirable.
If the ______ _____ is to be run overnight it should be programmed to hold on the first ethanol bath and not finish until the next morning so the specimens do not sit in hot paraffin longer than the time indicated.
Vacuum Embedding
involves wax impregnation under negative atmospheric pressure inside an embedding oven. It reduces the time when tissues are subjected to high temperatures thus minimizing heat-induced tissue hardening.
the time required for complete impregnation is reduced by 25% -75% of the normal time required for tissue processing. The tissue is not over-exposed to heat; brittleness, shrinkage and hardening of tissues consequent to overheating is therefore prevented. The tissue can also be transferred after clearing to a heated bath of paraffin wax from which air can be evacuated.
Vacuum
_____ hastens the removal of air bubbles and clearing agent from the tissue block, thereby promoting a more rapid wax penetration of the tissue.
Vacuum infiltration, vacuum infiltrator
requires a _____ or embedding oven, consisting of wax baths, fluid trap and vacuum gauge, to which a vacuum of up to 760 mm Hg is applied using a water or mechanical pump.
Vacuum Embedding
It facilitates complete removal of transition solvents, and prolongs the life of wax by reducing solvent contamination.
Modern tissue processors
____ _____ ____ are equipped to deliver vacuum, or vacuum and pressure, to all or some reagent stations during the processing cycle.
vacuum embedding oven
consists of a flat-bottomed heavy brass chamber covered with a heavy glass lid resting on a wide and thick rubber valve which produces an airtight seal when the chamber is being used.
Vacuum chamber, 2-4C. 500
is enclosed in a thermostatically controlled water-jacket, usually maintained at a temperature of _- _°C above the melting point of the wax.
The degree of this should not exceed _—- mm.
Vacuum Embedding
This technique is particularly recommended for urgent biopsies, for delicate tissues such as lung, brain, connective tissues, decalcified bones, eyes, spleen and central nervous system.
Stopcock
What is provided to prevent water from being sucked back into the trap bottle and vacuum chamber when the water or suction pump is closed
Embedding Procedure
1. Clear in two changes of xylene, for 1 hour each.
2. Place the tissue in molten wax, in vacuum chamber and make the oven
airtight. Exhaust the air slowly by means of a vacuum pump or
Venture suction pump until there is a negative pressure of 400 to 500
mm. mercury.
3. Leave for 15 minutes, then slowly readmit air until normal atmospheric
pressure is reached.
4. Place the tissue in fresh wax.
5. Repeat steps 3 and 4.
6. Place the tissue into fresh wax.
7. Repeat step 3 and leave for 30 to 45 minutes.
8. Bring to normal atmospheric pressure and embed the tissue.
NOTE: The exhaustion and readmission of air must be gradual or
the specimen may be ruined
Benzene and xylene, chloroform and cedarwood oil
What are easily removed from the tissues? and what is difficult and requires more frequent wax changes?
Addition of benzene
What may hasten displacement of cedarwood oil with less tissue shrinkage.
4 hours, above 60°C), 2 to 5°C
Since prolonged treatment in melted paraffin causes shrinkage and hardening of tissues, making cutting difficult, the tissue should not be left in the paraffin oven for more than _ _____. The shorter the time in the hot oven with adequate paraffin impregnation and evaporation of clearing agent, the better it is for the tissue. Tissues become increasingly harder and more brittle as they are heated. Infiltration in overheated paraffin (above _____) will also produce shrinkage and hardening of tissues and destroy lymphoid tissues completely. To avoid this, the paraffin oven must be maintained at a temperature _ to _ C above the melting point of paraffin to be used for impregnation.
Paraffin wax
______ ___ must be pure, i.e. free from dust, water droplets and other foreign matter. Fresh wax should be filtered before use in a wax oven at a temperature 2°C higher than its melting point.
Green's No. 904
Wax that has been trimmed away from the impregnated tissue may be melted and filtered for future use, with a coarse filter paper, e.g. Green's No. 904.
Water, 100 -105°C
When wax has been reused, some amount of ____ inevitably is mixed with it. If excessive, this may impair the impregnating capacity of the medium and prevent formation of a good tissue block. Water must therefore be removed by heating the wax to ____ ____, thereby raising its melting point.
Paraffin wax
____ ___ may be used only twice, after which, fresh wax must be utilized. When using an automatic tissue processing machine, wax usually becomes admixed with the clearing agent, especially in the first beaker; hence, water must be discarded.
fixed knife microtomes, heavier microtome
For ___ ____ ____, a relatively hard wax with a higher melting point is recommended.
____ _____ knives require harder paraffin wax than lighter ones
Paraplast
is a mixture of highly purified paraffin and synthetic plastic polymers, with a melting point of 56-57°C. It is more elastic and resilient than paraffin wax thereby permitting large dense tissue blocks such as bones and brain to be cut easily with the same result as in double embedding.
Blocks obtained are more uniform than with any other medium, with better ribboning of sections. Serial sections may be cut with ease, without cooling the tissue block, thereby preventing the formation of ice crystal artefacts. No deposit is left on the slide after staining, and no special processing schedule is required. It is soluble in common clearing agents and follows the same time schedule for paraffin impregnation, and does not tend to crack like other paraffin wax substitutes.
Double embedding
is done in histology to provide extra structural support and prevent the distortion or loss of tiny, delicate, friable, or multiple fragmented tissue specimens during processing and microtome sectioning. The tissue is first infiltrated with an initial supportive medium (like agar, gelatin, celloidin, or agarose) and then embedded a second time in a final medium like paraffin wax.
56 to 58oC, 54 to 56oC, 60 to 63oC
Generally, Paraplast with a melting point of __ is recommended. During the winter, _ ____ Paraplast may be used if the tissue is cut in a cool room. During the summer it may be necessary to use ___ ____, although this is to be avoided if possible in order to not to "cook" the tissue. "Cooked" tissue does not section well or, if it does, it does not stain well and most details are destroyed.
Embeddol
is synthetic wax substitute similar to Paraplast with a melting point of 56-58°C. It is less brittle and less compressible than Paraplast.
Bio/aid
is a semisynthetic wax recommended for embedding eyes.
Tissue Mat
is a product of paraffin, containing rubber, with the same property as Paraplast.
Ester Wax
has a lower melting point (46-48°C), but it is harder than paraffin.
It is not soluble in water, but is soluble in 95% Ethyl Alcohol and other clearing agents; hence, it can be used for impregnation without prior clearing of the tissue.
Cellosolve (ethylene glycol monoethyl ether) or xylene may be used as clearing agents, if indicated. In such instances, removal of the clearing agent must be gradual; that is, the tissue must be placed in a solution containing equal proportion of clearing agent and this wax for 3- 6 hours before finally transferring it to pure wax. Three to four changes of wax are required to ensure complete tissue impregnation.
Sectioning of these wax’s tissues should be done on a heavy duty microtome (e.g. sliding or sledge type microtome) due to the relative hardness of the wax.
Water Soluble Waxes, 38-42 or 25-56
are plastic polymers, mostly polyethylene glycols with melting points of —- ——C.
Carbowax
It is a polyethylene glycol containing 18 or more carbon atoms, which appears solid at room temperature. It is soluble in and miscible with water; hence does not require dehydration and clearing of the tissue. The tissues are fixed, washed out and transferred directly into the melted _____
Deeper for Carbowax
Processing time is reduced with the special advantage that harmful effects produced by ordinary dehydrating agents are consequently avoided.
It does not remove neutral fats and lipids which are soluble in reagents used for routine processing with paraffin, hence, allowing these substances to be demonstrated in thin sections.
Tissues are not exposed to too much heat so that excessive hardening, shrinkage and brittleness of tissue is avoided; hence, making this technique suitable for many enzyme histochemical studies. Cytologic details are excellently preserved
Routine processing of carbowax
four changes of Carbowax, one each in 70% and 90% and 2 times in I 00% concentration, at a temperature of 56°C are used, at 30 minutes, 45 minutes and 1 hour (with agitation), respectively. Specimens are then embedded in fresh Carbowax at 50°C and rapidly cooled in a refrigerator.
Due to its hygroscopic nature
Why is this?
Carbowax is very easily dissolved in water. Hence care must be taken to avoid contact of the block with water or ice. Tissue sections are very difficult to float out and mount due to its extreme solubility in water, dehydrating and clearing agents. Adding soap to water or using 10% Polyethylene Glycol 900 in water will reduce tissue distortion and promote flattening and "floating out" of sections.
by adding soap to water or using 10% polyethylene glycol 900 in water
How can you reduce tissue distortion and promote flattening and floating out of sections in carbowax?
Dimethyl sulphoxide (DMSO)
It is Added to proprietary blends of plastic polymer paraffin waxes reduces infiltration times and facilitates thin sectioning.
It scavenges residual transition solvent and probably alters tissue permeability by substituting for or removing bound water thus improving infiltration.
Some individuals who handle this wax may experience an unpleasant and annoying oyster or garlic taste probably caused by this wax’s metabolites.
Possible health risks associated with the use of this wax are minimal if correct laboratory hygiene is observed
Celloidin (Collodion)
It is a purified form of nitrocellulose soluble in many solvents, suitable for specimens with large hollow cavities which tend to collapse, for hard and dense tissues such as bones and teeth and for large tissue sections of the whole embryo.
It is supplied in thin (2%), medium (4%) or thick (8%) solutions of cellulose dissolved in equal parts of ether and alcohol.
This is used mainly for preparing soft tissue sections of mixed consistency such as eyes and brain. No heat is required, and the resultant block has a rubbery consistency which gives good support to the tissues.
Disadvantages include inability to cut thin sections, storage of blocks in alcohol and speed of technique (which can take several weeks or months).
1. It permits cutting of tissue sections which are thicker than in paraffin wax, and is recommended for processing of neurological tissues.
2. Its rubbery consistency allows tissue blocks that are either very hard or of varying consistency, to be cut without undue distortion.
3. Dense tissues which are hard to infiltrate (e.g. bones and brain) and specimens which tend to collapse easily due to air spaces (e.g. eyes) are supported better, thereby avoiding the crumbling of tissues during sectioning. When eye sections are embedded by the paraffin method, the retina may be detached from the harder tissues (e.g. sclera and choroid) that encircle it. The cedarwood oil used in the dry celloidin technique helps to soften the brittle layers.
4. It does not require heat during processing; hence, producing minimum shrinkage and tissue distortion especially for cutting large bone sections. It is, therefore, recommended in cases when minimum shrinkage is required and when frozen section technique cannot be done.
Advantages of Celloidin impregnation
The impregnation is very slow (lasting for several days or weeks).
Very thin sections (less than I 0 µ) are difficult to cut.
Serial sections are difficult to prepare.
Vapor of the ether solvent is very flammable; hence, it should never be used near an open flame.
Photomicrographs are difficult to obtain.
It is very volatile and therefore must be kept in bottles with ground glass stoppers to prevent evaporation.
Disadvantages of celloidin impregnation
Wet Celloidin Method
It is recommended for bones, teeth, large brain sections and whole organs. After the usual fixation and dehydration of the tissue, it is placed in equal parts of ether and alcohol for 12-24 hours. The tissue is then placed in thin celloidin (2-4%) for 5-7 days, transferred to medium celloidin (4-6%) for another 5-7 days, drained off and poured with thick celloidin (8-12%) until the specimen has become impregnated, usually between 3-5 days.
The specimen is removed from the celloidin, transferred to an embedding medium containing freshly poured thick celloidin and kept in a tightly covered jar or dessicator in order to evaporate the alcohol-ether solvent. The dessicator top is removed for a few seconds, time and again, to admit fresh air and harden the tissue block. Evaporation must be gradual to achieve a consistent, uniform degree of hardness throughout the block and prevent the formation of air bubbles.
When the ball of the finger leaves no mark on the surface of the tissue block, evaporation and consequently, embedding, is considered to be complete. The tissue block is then stored in 70-80% alcohol until ready for cutting. This is done to avoid dehydration and shrinkage of tissues.
Dry Celloidin Method
is preferred for processing of whole eye sections. The principle and procedure of this method is similar to wet celloidin method, except that 70% alcohol is not used for storage before cutting. Instead, Gilson's mixture, made up of equal parts of chloroform and cedarwood oil, is added to the celloidin block before hardening, to make the tissue transparent.
This method does not make use of alcohol due to the presence of cedarwood oil in the block
Gilson’s mixture
In the Dry celloidin method what is added to the celloidin block before hardening to make the tissue transparent
Low Viscosity Nitrocellulose (L.V.N.)
is another form of celloidin soluble in equal concentration of ether and alcohol, with a lower viscosity, allowing it to be used in higher concentrations and still penetrate tissues rapidly. Because of this, many workers prefer ___ __ ___ to the ordinary celloidin for impregnation and embedding. It forms a harder tissue block and makes cutting of thinner sections possible.
The tendency of tissues to crack may be prevented by adding plasticizers (e.g. oleum ricini or castor oil) when embedding chrome-mordanted tissues.
Low viscosity nitrocellulose is more explosive than celloidin and should therefore be handled with care.
When dry, striking or dropping the container will cause the substance to explode.
It is usually marketed while wet with alcohol. The container must be kept tightly covered and protected from sunlight to avoid evaporation of alcohol.
When no longer needed for future use, this should be carefully destroyed, since the material becomes increasingly dangerous as the alcohol continues to evaporate.
GELATIN IMPREGNATION
It is rarely used except when dehydration is to be avoided and when tissues are to be subjected to histochemical and enzyme studies. It is used as an embedding medium for delicate specimens and frozen tissue sections because it prevents fragmentation of tough and friable tissues when frozen sections are cut.
It is water-soluble, and does not require dehydration and clearing, although fixatives (such as 10% formalin) should still be washed out by running water whenever indicated. It has a low melting point and does not cause over-hardening of tissues by heating.
After the fixative has been completely washed out, the tissue is placed in 10% gelatin with 1% phenol for 24 hours, transferred to 20% gelatin with 1% phenol for the next 12 hours, and finally to another fresh solution of 20% gelatin with 1% phenol which is then allowed to cool in a refrigerator until impregnation and embedding are completed.
Gelatin-embedded tissues are then transferred to 10% formalin for 12-24 hours in order to harden the tissue. Tissues should not be more than 2-3 mm. thick since gelatin-embedded specimens are harder to freeze than non-impregnated tissues.
1 % phenol
The ______ serves to prevent the growth of molds. Excess gelatin may be removed by floating the sections oh to paper and trimming them with scissors. The volume of the impregnating medium should be at least 25 times the volume of the tissue.
EMBEDDING
After impregnation, the tissue is placed into a mold containing the embedding medium and this medium is allowed to solidify. Ideally the embedding medium should match the tissue type in strength and hardness.
If the embedding medium is too soft for the material, the tissue will not be supported and sections will be torn or shredded.
If the medium is too hard for the tissue, sections will be brittle and will shatter.
To infiltrate the tissues with supporting embedding medium, tissues must be free of all water (since usually embedding medium is not miscible with water).
Paraffin embedded tissues are arranged at the bottom of the mold together with their proper labels and immersed in melted paraffin at a temperature between 5-10°C above its melting point and then cooled rapidly in a refrigerator at -5°C or immersed in cold water to solidify. This will allow hardening of tissues, giving them a firmer consistency and better support, thereby facilitating the cutting of sections.
Orientation
The process by which a tissue is arranged in precise positions in the mold during embedding, on the microtome before cutting, and on the slide before staining, is known as? Generally speaking, the surface of the section to be cut should be placed parallel to the bottom of the mold in which it is oriented.
Leuckhart’s Embedding Mold
Consists of two L-shaped strips of heavy brass or metal arranged on a flat metal plate and which can be moved to adjust the size of the mold to the size of the specimen. Blocks produced are even, with parallel sides, and with a fairly shaped initial setting of the wax. The mold is adjustable to give a wide variety of sizes to fit the size of the tissue block for casting. It is recommended for routine use, although, too slow and cumbersome for use in a busy laboratory.
Compound Embedding Unit
is made up of a series of interlocking plates resting on a flat metal base, forming several compartments. It has the advantage of embedding more specimens at a time, thereby reducing the time needed for blocking.
Plastic Embedding Rings and Base Mold
consist of a special stainless steel base mold fitted with a plastic embedding ring, which later serves as the block holder during cutting.
Tissue Tek
is equipped with a warm plate to manage the impregnated specimen, and a cold plate at -5°C for rapid solidification of the block. It consists of a white plastic cassette mold with detachable, perforated stainless steel hinge and Snap-On lid, used to hold the tissue specimen through-out fixation, dehydration, clearing and wax impregnation.
With the ____ ___ system, the specimen is placed on the base mold, the plastic embedding ring is placed in position, filled up with wax, and then placed on a small cool area to allow the wax in the base of the mold to semi-harden. This will allow easy orientation of the block.
Once the tissue has been properly oriented, the base of the cassette is placed on top and together, they are placed on the cold plate so that the paraffin wax can cool and harden quickly. After the paraffin wax has solidified (usually 5 minutes), the block is taken out together with the embedding ring and is immediately ready for cutting without having to undergo trimming or mounting, thereby saving time and effort.
The advantage of Tissue tek
include ease of use, less paraffin wax needed, faster embedding, firmly attached tissue and holder, and permanent identification. It produces easier orientation when re sectioning of tissue is required, and blocks can be filed immediately after sectioning.
Disposable Embedding Molds
These three are
a. Peel-away
b. plastic ice trays
c. Paper boats
Peel-Away
disposable thin plastic embedding molds, available in 3 different sizes, are simply peeled off one at a time, as soon as the wax has solidified, giving perfect even block without trimming. It may be placed directly in the chuck or block holder of the microtome.
Plastic Ice Trays
-such as those used in ordinary refrigerators may be recommended for busy routine laboratories. Each compartment may be utilized for embedding one tissue block, which may then be removed by bending the plastic tray once the wax has solidified or by smearing the inner mold with glycerin or liquid paraffin before embedding
Paper Boats
are normally utilized for embedding celloidin blocks but are equally useful for paraffin wax blocks. They have the advantage of being cheap and easy to make. They provide easy and accurate identification of specimen, thereby avoiding confusion and interchange of tissue blocks. Rapid embedding of small or large volume of individual specimen is possible, since paper molds can be made to suit any size of tissue.
The basic method of embedding
(READ AND READ)
1. Open cassette to view tissue sample and choose a mold that best corresponds to the size of the tissue. A margin of at least 2 mm of paraffin surrounding all sides of the tissue gives best cutting support. Discard cassette lid.
2. Put small amount of molten paraffin in mold, dispensing from paraffin reservoir.
3. Using warm forceps, transfer tissue into mold, placing cut side down, as it was placed in the cassette.
4. When the tissue is in the desired orientation, add the labeled tissue cassette on top of the mold as a backing. Press firmly.
5. Hot paraffin is added to the mold from the paraffin dispenser. Be sure there is enough paraffin to cover the face of the plastic cassette.
6. Cool the top surface of the Paraplast by blowing gently on it. Tissues at this stage are very brittle and should be handled with care. If necessary, fill cassette with paraffin while cooling, keeping the mold full until solid.
7. Cool thoroughly in cold running tap water. If you use ice water for the final cooling, you may split the block owing to too rapid shrinkage. Paraplast naturally splits in the line of least resistance-right through the tissue
8. Paraffin should solidify in 30 minutes. When the wax is completely cooled and hardened (30 minutes) the paraffin block can be easily popped out of the mold; the wax blocks should not stick. If the wax cracks or the tissues are not aligned well, simply melt them again and start over. 9. If you use plastic cups, the Paraplast block can be removed as soon as it is cooled. The stainless steel mold should slip off easily when cool and can be used again.
ORIENTATION
_____ of tissues in the Paraplast block is important for tissues such as an artery or fallopian tube so that they can be properly placed in a predetermined plane, such as cross sections. Also, trimming is excessively difficult in a block embedded with two or more tissues if they are not carefully lined up before the Paraplast is cooled.
Celloidin or Nitrocellulose Embedding Method
Used to be recommended for embedding hard tissues such as bones and teeth, and for large sections of whole organs like the eye, since the delicate layers of the eyeball are difficult to keep intact when other media are used. Tissues are embedded in shallow tins of enamel pans which are covered by sheets of weighted glass. Bell jars can be used to control the rate or evaporation of the solvent. The use of celloidin is discouraged now because of the special requirements needed for processing and the limited use of these types of sections in neuropathology
Double-Embedding
is the process by which tissues are first embedded or fully infiltrated with a supporting medium such as agar or nitrocellulose, then infiltrated a second time with paraffin wax in which they are subsequently embedded. This is used to facilitate cutting of large blocks of dense firm tissues like the brain. They are also recommended for making small sections of celloidin blocks. A shortcoming of using agar as the pre-embedding media is that certain tissues shrink during the embedding process, and the agar-based pre-embedding media limits tissue expansion during slide mounting, resulting in difficulties with the tissue sample adhering to the microscope slide. The availability of paraffin waxes containing different types of resins has made this technique obsolete
PLASTIC (RESIN) EMBEDDING
Has provided superior results for light microscopic studies, particularly in hard tissues such as undecalcified bone and for high resolution light microscopy of tissue sections thinner than the usual 4-6 µm, such as renal biopsies and bone marrow biopsies.
Epoxy
_____ embedding plastics are made up of a carefully balanced mixture of ____ plastic, catalysts and accelerators.
bisphenol A (Araldite),
or glycerol (Epon),
or cyclohexene dioxide (Spurr).
Three types of epoxy plastics used in microscopy
Araldite
Infiltration by _____ is slow, partly because the epoxy plastic itself is a large molecule.
glycerol-based epoxy plastics (Epon)
The _______ plastics (____) have a lower viscosity but are often sold as mixtures of isomers
Cyclohexene dioxide-based plastics
A epoxy plastic embedding type that can be obtained pure, have very low viscosity, and infiltrate fastest.
Is a Low Viscosity mixture which provides rapid infiltration of tissues. It's easy to prepare and mixes rapidly. This resin is compatible with ethanol so no change to propylene oxide is needed prior to infiltration. Polymerization at 60°C is recommended.
Epoxy plastics Disadvantages
They are hydrophobic and subsequent oxidation by peroxide to correct this may produce tissue damage.
Epoxide groups may reduce antigenicity of embedded tissue, and may compromise the result of immunohistochemical staining.
More importantly, epoxy resins may cause sensitization if absorbed by skin or inhalation.
The components of many epoxy plastics are toxic and one of its components, vinyl cyclohexane dioxide (VCD) is known to be carcinogenic.
For protection, gloves should always be worn when handling these plastics, and adequate facilities including an operational fume hood must be provided to remove the toxic vapors and properly dispose of toxic waste.