treatment planning system (TPS)
uses reconstructed CT data set as foundation of treatment planning
presented w/ series of CT image slices
uses dataset to reconstruct images into sagittal + coronal planes
1 slice contains BB reference marks placed during CT sim
if not all BBs visible → may be misaligned
BBs visible on CT image = known location within CT dataset
identifies right/left lateral + anterior BBs within CT dataset
intersection point of BBs = reference centre (user origin)
acts as known location → reference point for other structures
reference centre (user origin)
known location within CT dataset set as (0,0) → where planning begins
intersection point of BBs
internal starting point within CT dataset
zero position along all 3 coordinate axes (X, Y, Z)
coordinates of other points described relative to this point
6MV
less penetrating → dmax = 1.5cm
more rapid dose fall off
more suitable for shallower targets (3DCRT)
15MV
more penetrating → dmax = 3.5cm
gradual relative dose fall off
more suitable for deeper targets (3DCRT)
CT coordinate system
CT dataset 3D → use 3D coordinate system to navigate
user origin = (0,0) = reference point
each axis represents different anatomical direction + CT view/plane
identify location of any point within CT dataset
represents direction as we scroll through images
corresponding coordinate value changes → in relation to (0,0)
change in coordinate value = distance moved in mm
transverse CT images → distance b/w consecutive slices = slice thickness
3 axes:
X = sagittal
➕ → left
➖ → right
Y = coronal
➕ → posterior
➖ → anterior
Z = transverse (axial)
➕ → superior
➖ → inferior
contouring
process of outlining important anatomical structures on CT images
structures outlined directly on CT images
outlines define boundaries of each structure
outlines allow planning system to recognize individual structures
goal:
to create treatment plan, need to identify where to deliver radiation + which healthy structures to avoid
to identify + outline these areas within CT dataset for TPS to recognize
software can’t differentiate b/w organs → only shades of grey
structures outlined:
OARs
body - external surface
tumour volume → identified + done by radiation oncologist
process:
done slice-by-slice throughout CT dataset
structure identified on each CT slice where it is visible
structure boundary outlined on every slice
individual contours combine = 3D volume
organs at risk (OARs)
normal tissue sensitive to radiation
should receive as little dose as reasonably achievable
can restrict treatment margins to avoid
those that are close enough to tumour to potentially receive radiation are contoured
common structures for tumour in:
breast
heart
lung
axilla
spinal cord
esophagus
contralateral breast + lung
brachial plexus
prostate + vagina
bladder
rectum
femoral heads
small intestine
uterus → for vagina
oral cavity
optic nerves + chasm
brain stem
spinal cord
esophagus
eyes + lenses
target delineation
process done by radiation oncologist
outlining tumor volumes + surrounding healthy organs on CT dataset to plan precise radiation treatment
3 target volumes contoured
gross target volume
clinical target volume
planning target volume
gross target volume (GTV)
visible tumour identified on CT dataset
visible/palpable or macroscopic tumour
smallest contour of tumour
doesn’t account for
microscopic disease in neighbouring tissue
variable patient positioning
changing tumour size

clinical target volume (CTV)
includes gross tumour volume (GTV) + margin that may contain microscopic disease
cancer cells can extend beyond visible tumour
microscopic disease may not be detectable on imaging
includes tissue at risk of containing microscopic disease

planning target volume (PTV)
includes clinical target volume (CTV) + margin to account for treatment setup uncertainties + internal organ motion
position of target may not be exactly same every day of treatment
several contributing factors
patient positioning: small differences in position each day
internal organ motion: organ movement, ie. breathing, other physiological processes
changes in organ filling: differences in bladder/rectal filling can affect target position
eg. lung vs. prostate
margins larger for lung → moves more than prostate
need to account for organ stability

margins
distance b/w target volumes
sizes vary depending on several factors
tumour type + location
different cancers = different patterns of microscopic spread
internal organ motion
breathing + changes in bladder/rectal filling
patient positioning + immobilization
reproducibility of patient positioning each day
clinical protocols
evidence-based, tumour-specific guidelines help determine appropriate margins
not always equal in every direction
typical margins
GTV → CTV = 5-20mm
CTV → PTV = 3-10mm

isotropic margin
equal expansion in all directions of target volume
eg. 5mm around entire CTV
most commonly used across radiation oncology

anisotropic margin
unequal expansion in different directions of target volume
eg. 5mm anteriorly + 3mm posteriorly
may be used to reduce radiation exposure to nearby (sensitive) organs
eg. spinal cord, parotid glands, rectum

prescription dose
total radiation dose to be delivered to PTV + over how many treatments (fractions)
determined by radiation oncologist
eg. gynae tumour → 45Gy/25#
total dose = 45 Gy
# of treatment sessions = 25
1.8 Gy per treatment
common daily dose for many treatment sites for radical treatments
delivered over multiple fractions
delivering too much radiation in 1 day can critically harm OARs
allows healthy tissue time to recover b/w treatments + still deliver full prescribed dose to tumour
gray (Gy)
measure of absorbed dose of ionizing radiation
defines amount of energy deposited by radiation into material
ie. absorption of radiation by body tissue
dose
measured as absorbed energy → energy causes physical damage
radiation passing through body w/o hitting anything does no harm
damage to tissue happens when radiation crashes into atoms → energy deposited
certain organs can only receive max amount of dose before significant side effects present
deposits energy by interacting w/ atoms along its path
photon interacts w/ atom → electrons knocked loose
electrons deposit energy into nearby tissue = biological damage
energy not deposited evenly → changes w/ depth
palliative dose
focus on symptom relief
lower total dose
higher daily dose
less damage to OARs
planning priorities
efficiency
greater dose variation may be acceptable
less emphasis on conformity - may be tightly conformed or not
adequate dose coverage
radical dose
focus on cure, long-term tumour control or symptom relief, depending on intent
higher total dose
lower daily dose
more concerned about OARs
planning priorities
greater emphasis on dose homogeneity + conformity
adequate dose coverage
treatment field
radiation delivered from chosen gantry angle, eg. gantry angle at 180° to treat T5
delivers photon beam
continuous stream of photons produced by LINAC
deposits radiation into tissues along its path as it travels through body
build up region
beam first hits skin → few electrons set in motion
relatively little energy deposited
photon continues through body → more electrons move
more energy deposited = dose increase → dose builds up
dose maximum (Dmax)
max dose delivered by treatment field
point at which beam reaches max dose
point below surface of skin where radiation dose = highest
# of electrons set in motion = # of electrons stopping
electron equilibrium
dose fall off
gradual decrease in dose as beam travels deeper into body
fewer photons remain to create new electrons
less electrons = less energy deposited into tissue
dose gradually decreases
more gradual as photon energy increases
6MV
more rapid
10MV
more gradual
15MV
gradual
percentage depth dose (PDD)
curve showing dose distribution w/ depth - expressed as %
illustrates
dose build-up
maximum dose (Dmax)
dose fall-off
describes how radiation dose changes w/ depth as it travels through body
how photon beam behaves as it travels deeper into tissue
photon energy
LINAC can produce beams w/ different energies
6/10/15/18 MV
MV used to describe nominal energy of photon beam
can change beam energy by altering kinetic energy gained by electrons as they accelerate through waveguide
⬆ KE = ⬆ energy photons at tungsten target
⬆ photon energy = deeper beam penetration = deeper dmax
beam energy influences where depth of maximum dose (dmax) occurs for that treatment field
⬇ energy beams = max dose closer to skin
⬆ energy beams = max deeper within body
can change dose fall off
⬆ = more rapid dose fall off
dmax
refers to depth at which maximum dose (Dmax) occurs
influenced by photon energy
⬆ photon energy = greater depth
3D conformal radiation therapy (3DCRT)
treatment field shaped to match target (tumour) via jaws + MLCs
gantry, collimator, MLCs = stationary during beam delivery
1+ treatment fields can be used
single field treatment
uses 1 treatment field from single direction to deliver prescribed dose per fraction
not used very often → only for specific cases
pros
only 1 treatment field used = quick + efficient treatment delivery
cons
non-uniform dose distribution
dose decreases beyond Dmax as depth increases
shallower region of PTV receives higher dose than deeper region
dose homogeneity
does not deliver homogenous dose throughout target volume
uneven distribution of radiation across target
dose conformity
distribution does not closely follow shape of PTV
shallower region receives higher dose than deeper region
applications
can be used if treatment is low-total-dose palliative treatment → focus on relieving symptoms quickly
depends prescribed dose + planning objectives
doses received by surrounding healthy tissues remain within acceptable limits
adequate target coverage is still ensured
for relatively shallow targets where adequate target coverage can be achieve
eg. vertebrae, pedicles, laminate, spinous processes
dose gradient
change in radiation dose over distance within tissue
commonly, dose decreases as depth increases
occurs in single field treatments
homogenous dose (uniform dose)
even distribution of radiation dose across target volume
not always necessary
acceptable dose distribution depends on treatment intent + clinical planning objectives
determined by ROs
simple palliative treatment = great dose variation may be acceptable
single field is quick → efficient treatment = achieve pain relief in shorter period of time
BUT still need to ensure adequate dose coverage + protect surrounding healthy tissues
homogeneity
how evenly dose is distributed throughout PTV → distribution

conformity
how closely high-dose region follows shape of target volume → shape

target coverage
how much of PTV receives prescribed radiation dose
objective may specify what % of PTV should receive particular dose
can use Dose Volume Histogram (DVH) in Eclipse to access this coverage
dose volume histogram (DVH)
shows % of PTV receiving at least particular radiation dose
x-axis: dose cGy + relative dose %
y-axis: % of PTV volume
treatment planning objectives
describes dose requirements aim to achieve when creating treatment plan
established by ROs
uses clinical protocols or national guidelines
may specify what % of PTV should receive particular dose
eg. PTV receive at least 85% of prescribed dose of 8Gy = 6.8Gy
hotspots to be less than 124% of prescription
hotspot
area within patient that receives significantly higher radiation dose than prescribed dose
anywhere that is receiving 107%+ of prescription
treatment plan considerations
homogeneity
how evenly dose is distributed throughout PTV
conformity
how closely high-dose region follows shape of PTV
target coverage
whether PTV received required radiation dose
hotspots
whether areas receive excessively high doses = >107% of prescription
dose profile tool
allows us to examine calculated dose distribution through patient in Eclipse
dose displayed as % of prescribed dose
x-axis = distance (cm)
y-axis = dose (%)
parallel opposed fields (POP)
treatment technique that uses 2 treatment fields positioned 180° apart
each treatment field delivers portion of prescribed dose → combined dose = total prescribed dose to PTV
beam weighting can be equal or unequal
pros
prescription delivered from 2 opposite directions → both contribute dose to PTV
although does make plan hotter overall
dose combines to produce more uniform dose distribution throughout PTV = good homogeneity
cons
high dose not confined to PTV = poor conformity
high dose delivered throughout entire treatment field
applications
useful for deeper targets (eg. high BMI patients, bone metastases, whole brain, prostate)
single treatment field would result in very high dose near dmax
common in selected palliative treatments (eg. lung)
relieve symptoms rather than cure disease
often used for lower total-dose prescriptions
limit radiation exposure to surrounding healthy tissues + OARs
considerations
different tissue density affects how dose is distributed
patient anatomy → changes in patient thickness + shape influence dose delivery by each field
beam energy → high = deeper targets + low = shallower targets
POP dose profile
red + green lines
dose delivered by each individual treatment field
each field follows similar pattern to PDD curve
Dmax is halved for each field
prescription dose distributed b/w two
blue line
combined dose from both treatment fields
as dose from one field ⬆ → opposing field contributes dose from other direction
get ~100% throughout treatment field
healthy tissue also receiving 100% of prescription
amount OARs are receiving are still acceptable (low dose)
together = more uniform dose distribution = dose homogeneity
total dose relatively lo
unequal beam weighting
when one field contributes more dose than the other
used when target is not equally distanced from surface
field that has to travel further distance will deliver lower dose
field that is closer to target will deliver higher dose
for more shallow targets
opposing fields can lessen hotspot intensity
eg. in POP technique: 20Gy/#5 → 4Gy per fraction
field 1 = 2Gy per fraction
field 2 = 2Gy per fraction = 4Gy
three-field technique
3DCRT technique uses 3 treatment fields directed towards target from different angles
each field delivers portion of prescribed dose = combines to deliver total prescribed dose to PTV
beam weighting can be equal or unequal
pros
reduces dose to surrounding healthy tissues + OARs while maintaining good PTV coverage
dose delivered from 3 different directions = spread across different areas
each field contributes dose to PTV = lower dose from each field
provides better dose conformity
reduces volume of healthy tissues receiving high dose (compared to POP)
applications
may be used in palliative treatment in long-term tumour control cases
especially for higher total prescription dose
eg. rectum, pelvis
may be used for radical treatments
3 field dose profile
blue + green
dose contributions from 2 opposing fields
red
dose contribution from 3rd field
only where it intersects profile line
purple
combined dose from all 3 fields = higher dose region
where all 3 fields overlap
ICRU (International Commission on Radiation Units and Measurements)
provides international recommendations for prescribing, recording, reporting radiation therapy doses
recommended dose range within PTV = 95-107% of prescribed dose
aims to achieve adequate PTV coverage while maintaining relatively homogenous dose distribution
guides stricter planning objectives → important for 3DCRT
more conformal 3DCRT techniques
aim to achieve more uniform PTV coverage while limiting dose to surrounding healthy tissues
(for single-field + POP → palliative treatments = not as strict)
may use lower total prescription doses
allows for greater flexibility in dose distribution + wider acceptable dose ranges
isocenter
point in space where rotation axis of gantry, collimator, treatment couch intersect
Eclipse automatically positions this point within PTV target using SAD technique
commonly positioned near PTV centre
provides common reference point around which multiple fields can be arranged
as gantry rotates → SAD remains constant at 100cm
different fields can be delivered w/o repositioning patient b/w fields
isocentre positioning
start at user origin
CT sim = BB mark intersection → user origin (reference centre) established
LINAC room lasers intersect at isocentre
treatment delivery = position patient (using tattoo marks) + align w/ LINAC room lasers
places user origin at LINAC isocentre
Eclipse determines how to get to planned isocentre
get to planned isocentre within/near PTV established during treatment planning
Eclipse provides X/Y/Z coordinate difference between/w user origin + planned isocentre
move user origin to treatment isocentre
use differences to determine how far/which direction to move couch → bring planned isocentre into alignment w/ LINAC isocentre
SSD technique to double-check patient position
during planning, Eclipse calculates expected SSD for each treatment field
use ODI to to measure SSD + compare w/ expected SSD measurements in patient setup instructions
NOTE: using SAD technique but reading SSD as additional safety check to ensure correct patient distance from radiation source