Flashcards covering the analytical method validation parameters, experimental results, and conformity conclusions documented in the SH-FORM 43 medical biology evaluation sheet.

7 Terms

1
New cards

SH-FORM 43

A standardized form (fiche type) used for the verification (scope A / portée A) or validation (scope B / portée B) of a medical biology method.

2
New cards

Repeatability (Répétabilité) Evaluation

A performance test conducted on Lyphochek N (n=30n = 30) yielding a mean of 2.072.07, standard deviation of 0.040.04, and a coefficient of variation of 1.93%1.93\%, which was concluded compliant (conforme) because the obtained CV was lower than the supplier CV (2.02%2.02\%).

3
New cards

Intermediate Precision (Fidélité Intermédiaire) Evaluation

A performance assessment on Lyphochek N and Lyphochek P (n=30n = 30 each) concluded non-compliant (non conforme) because the obtained CVs (20.4%20.4\% and 78.3%78.3\%) exceeded the laboratory's retained CV limit of <2.3%< 2.3\%.

4
New cards

Trueness (Justesse) Evaluation

An assessment performed using normalized internal quality controls (CIQ) where Level 1 was compliant (bias=1.36%\text{bias} = 1.36\% vs. limit <5%< 5\%) but Level 2 was non-compliant (bias=7.3%\text{bias} = 7.3\% vs. limit <5%< 5\%), preventing the adoption of the technique.

5
New cards

Lyphochek N

The control sample (échantillon contrôle) tested with n=30n = 30 values, yielding a repeatability CV of 1.93%1.93\% and an intermediate precision CV of 20.4%20.4\%.

6
New cards

Lyphochek P

The control sample evaluated in intermediate precision testing (n=30n = 30) that produced a mean of 3.093.09, standard deviation of 2.422.42, and an unacceptable CV of 78.3%78.3\%.

7
New cards

CIQ (Contrôle Interne de Qualité)

Normalized internal quality controls evaluated across two levels to verify method trueness against an established limit of <5%< 5\% bias.