Flashcards covering the analytical method validation parameters, experimental results, and conformity conclusions documented in the SH-FORM 43 medical biology evaluation sheet.
SH-FORM 43
A standardized form (fiche type) used for the verification (scope A / portée A) or validation (scope B / portée B) of a medical biology method.
Repeatability (Répétabilité) Evaluation
A performance test conducted on Lyphochek N (n=30) yielding a mean of 2.07, standard deviation of 0.04, and a coefficient of variation of 1.93%, which was concluded compliant (conforme) because the obtained CV was lower than the supplier CV (2.02%).
Intermediate Precision (Fidélité Intermédiaire) Evaluation
A performance assessment on Lyphochek N and Lyphochek P (n=30 each) concluded non-compliant (non conforme) because the obtained CVs (20.4% and 78.3%) exceeded the laboratory's retained CV limit of <2.3%.
Trueness (Justesse) Evaluation
An assessment performed using normalized internal quality controls (CIQ) where Level 1 was compliant (bias=1.36% vs. limit <5%) but Level 2 was non-compliant (bias=7.3% vs. limit <5%), preventing the adoption of the technique.
Lyphochek N
The control sample (échantillon contrôle) tested with n=30 values, yielding a repeatability CV of 1.93% and an intermediate precision CV of 20.4%.
Lyphochek P
The control sample evaluated in intermediate precision testing (n=30) that produced a mean of 3.09, standard deviation of 2.42, and an unacceptable CV of 78.3%.
CIQ (Contrôle Interne de Qualité)
Normalized internal quality controls evaluated across two levels to verify method trueness against an established limit of <5% bias.