types of technique selection
Manual
Automatic exposure control
Automatically programmed techniques
Technique chart
Automatic exposure control
Ensure quality radiographic image is produced
Control consistently the amount of radiation reaching the IR by terminating the length of exposure
Predetermined amount of radiation reaches detector
Exposure terminates
measure the amount of x-rays exiting the patient with some type of detector; and
end the x-ray exposure as soon as a pre-set amount of radiation has been measured;
produces a radiograph with a pre-set amount of exposure in the area of the image where the detector was located
Manual technique selection technologist determines
Kvp
mA
Time
AEC technologist determines
Kvp
mA (sometimes)
Detector selection
Two types of ARC devices
Phototimers
Ionization chamber system
Phototimers
Light paddles, coated with a fluorescent material act as the detector
Radiation interacts with the paddles, producing visible light
This light is transmitted to photomultiplier tubes or photodiodes that convert this light into electricity
The electric signal trips the timer
The radiographic exposure is terminated when a sufficiently large charge has been received
Aka: light paddles, together with photomultiplier tubes, measure radiation exposure AFTER it passes through the image receptor
Ionization chamber system
A hollow cell that contains air and is connected to the timer circuit via an electrical wire
Ionization chamber is exposed to exit radiation
The air inside the chamber becomes ionized, creating an electrical charge
Charge travels along the wire to the timer circuit
The timer is tripped and the radiographic exposure is terminated when a sufficiently large charge has been received
AKA: ionization chamber measures radiation exposure before it reaches the image receptor
AEC system features
mAs readout
Minimum response time
Backup time
Density controls
mAs readout
Actual mAs used for the exam will be displayed following the exposure
Provide critical information needed before attempting any repeat exposures or using a manual technique for other projections
Minimum response time
Shortest exposure time a system can produce
You can’t set this
Backup time
mAs
Maximum allowed time that x-ray can stay on before exposure is terminated
Law requires max of 600 mAs when kVp is >50
May be able to set this depending on the department and machines
Density adjustment
Density settings (example -1,0,+1)
Adjusts the exposure time by some predetermined amount of increment expressed as a percentage
Alignment and positioning considerations
Select the detector that will be superimposed by the anatomic structures area of interest
The anatomic area of interest must be centered properly over the selected detectors
Patient anatomic area of interest must adequate cover the detector combination
Improper collimation may affect termination of the exposure time
Collimation
Too little collimation can cause excessive scatter to terminate exposure too quickly
Too much collimation can cause the detector to extend the length of exposure
Patient considerations
The composition of anatomic area centered over the detector will influence the exposure time
Increasing mA
Exposure time decreased
Decreasing kvp
Exposure time increased
Increasing OID
Exposure time increased
Decreasing SID
Exposure time decreased
Anatomic part has excessive gas
Exposure time decreased
Excessive collimation
Exposure time increased
Increased density control
Exposure time increased
changing mA and/or kVp
maintains exposure to IR at pre-set level
changing the SID
maintains exposure to IR pre-set level
patient thickness
maintains exposure to IR at pre-set level
selecting different detectors
maintains exposure to IR at pre-set level
what des ARC provide
AEC devices provide sufficient exposure only for structures positioned directly above the ionization chamber
how can Photocell(s) selection can affect image quality
Area of interest must cover selected photocell adequately
Underexposure and Quantum Noise
Overexposure and Loss of Detail
Contrast and Dynamic Range Shifts
False Triggering and Shadow Artifacts
Photocell vs. Phototransistor Characteristics:
Photocell (LDR) uses variable resistance and a slower response time, making it suitable for slow ambient tracking but prone to lag or blurring if light shifts rapidly.
Phototransistor offers amplified current output and faster response, delivering tighter precision for dynamic or high-speed optical capture
back up mAs should be what to produce sufficient exposure
Back-up mAs should be slightly greater than the mAs required to produce sufficient exposure
The actual mAs used for an exam will provide critical information needed before attempting an repeat exposures
why Back-up mAs should be slightly greater than the mAs required to produce sufficient exposure
typically 150% to 200% (or about 1.5 to 2 times)
Allow Normal AEC Operation: Provides enough exposure time for the AEC to reach the required image receptor exposure. Setting the back-up mAs too close to or lower than the required exposure will cause the system to shut off prematurely, resulting in an underexposed, noisy image that must be repeated.
Protect the Patient and X-Ray Tube: If the AEC ionization chambers fail, or if the wrong chamber/anatomy is selected (e.g., shooting through air or a prosthesis), the AEC will never reach its target density. The back-up mAs forces the X-ray exposure to terminate automatically before the patient receives a massive, harmful overdose or the X-ray tube overheats.
Account for Anatomical and Pathological Variations: A safety margin of roughly 150% (up to 200% for certain setups) accommodates minor variations in patient size, tissue density, and positioning without triggering a false-positive safety abort.
Anatomically programmed techniques allow techs to do what
Allows techs to select on the control panel that represents an anatomic area
Preprogrammed set of exposure factors is displayed and selected for use
Once an anatomic part and projection or position has been selected, the radiographer can adjust the exposure factors that are displayed
Exposure technique charts
Provide a method of consistency in the quality of the image production
Standardize the selection of exposure factors for the typical patient to improve quality and consistency
Designed for the average or typical patient
Accurate measurement of part critical
Variable kVp fixed mAs
kVp can be increased as the anatomic part size increases
kVp is increased by 2 for every 1 cm increase in part thickness
Kvp used for a particular projection is varied depending on the measured body part thickness
Fixed kVp variable mAs
kVp held constant for a given range of tissues types and thicknesses while the mAs is varied to achieve an appropriate exposure to IR
General rule: every 4-5 cm (1.6 -2 inch) changes in part thickness requires a change in mAs by a factor of 2
Variable kVp fixed mAs Advantages Disadvantages
Advantages
Easy to construct and simple to use
At low kVp levels, small changes in kVp more effective
Disadvantages
Anatomic part penetration not guaranteed
Changing kVp less effective for range of procedures
Less accurate for part size extremes
Fixed kVp variable mAs Advantages Disadvantages
Advantage
Part measurement less critical
Increased accuracy with range of procedures
More consistent image contrast and overall quality
Part penetration assured
Disadvantage
Optimal kVp values not standardized
Part thickness affects
Beam attenuation
Exposure reaching image receptor
Scatter production and image contrast
SPECIAL CONSIDERATIONS Geriatric patients
May require decreased exposure techniques, lower kVp values, and increased mA to decrease exposure time
SPECIAL CONSIDERATIONS Pediatric patients
Small size may require a reduction in exposure.
Quicker exposure times may be necessary
SPECIAL CONSIDERATIONS Bariatric patients
Need increase in kVp and mAs, require a grid, may need to be imaged in quadrants.
SPECIAL CONSIDERATIONS Projections and positions
May require changes in exposure technique due to changes in tissue thickness.
SPECIAL CONSIDERATIONS Casts and splints
Plaster casts require increase in exposure factors
Splints may require adjustment in exposure factors due to type of material if in the path of the x-ray beam
SPECIAL CONSIDERATIONS Contrast media
Used to increase radiographic contrast when imaging areas of low subject contrast.
Positive agents versus negative agents.