line voltage input
Main incoming power supply (220 V AC).
Line monitor
Measures incoming line voltage to ensure proper
system operating voltage
line compensator
Adjusts line voltage tap to keep autotransformer
voltage precisely at 220 V.
autotransformer
Operates on self-induction; acts as the primary kVp
selector across single coil.
major kVp selector
Taps across autotransformer turns in 10-kVp step
increments.
minor kVp selector
Taps across autotransformer turns in 1-kVp fine
increments.
kVp meter
Voltmeter connected across autotransformer output;
indicates set kVp prior to exposure
timing, circuit and selector
Electronic timer that terminates exposure after set
duration.
mA selector
Bank of precision rheostats (variable resistors)
controlling filament heating current.
High voltage step up transformer
Mutual-induction transformer multiplying primary
voltage to high kilovoltage (kV).
rectifiers(diodes)
AC from transformer secondary into DC.
ground
Safety ground connected at secondary coil center tap
mA meter
Measures actual x-ray tube current in milliamperes
local spot selector/switch
Directs filament current to either small or large
tungsten filament coil.
filament step down transformer
Decreases voltage to ~10–12 V and increases current
to ~3–6 A to heat filament.
X-ray tube
Evacuated glass/metal envelope housing Cathode (-)
filament and Anode (+) target.