1/98
Looks like no tags are added yet.
Perfix
Metallic orthopedic implant; Screw used for ACL reconstruction that is screwed firmly into bone; Safe; Ferromagnetic; But not in danger of dislodgement because of the way it is implanted
Revo Surescan
Cardiac pacemaker; Treat bradycardia with low energy electrical impulses and monitor heart rhythm; Conditional; 1st FDA approved MR conditional cardiac pacemaker; Most pacemakers are strictly contraindicated for standard MRIs (several patients have died); Must be put in asynchronous mode for MRI
Advantio, advisa, ensure, enrhythm, ingenio, promri
Cardiac pacemaker; Treat bradycardia with low energy electrical impulses and monitor heart rhythm; Conditional
Kora
Cardiac pacemaker; Treat bradycardia ith low energy electrical impulses and monitor heart rhythm; Conditional; Detects magnetic field and automatically switches to asynchronous pacing
Evera surescan
Defibrillator; Provide defibrillation therapy for tachycardia when needed; Conditional
Activa
Deep brain neurostimulator (DBS); Treats Parkinson's disease, tremor, dystonia; Conditional
RestoreSensor Surescan
Spinal cord neurostimulator (SCS); Treats chronic, intractable (severe) pain of neurological origin; Conditional
Argus II
"Second sight" retinal prosthesis neurostimulator; Electrical stimulation of retina to induce visual perception in blind patients; Conditional:; 3T or less; External equipment (VPU and glasses) is unsafe
Enterra
GES neurostimulator; Treats chronic, intractable nausea from gastroparesis; Unsafe
Interstim
Sacral nerve neurostimulator; Treats urinary incontinence or bowel control; Conditional
Actiflo
Indwelling bowel catheter system; "Zassi bowel management system"; Conditional:; 3T field and 720 G/cm gradient or less
Cyberonics
Vagus nerve neurostimulator; Treats epilepsy, depression, anxiety, alzheimer's, morbid obesity, migranes; Conditional
Bone/spinal fusion stimulator (SpF)
Spinal nerve stimulator; Adjunct therapy for spinal fusion procedures; Conditional
Accurx constant flow implantable pump
Drug pump; Continuous medication storage and dispenser; Conditional; Made of a titanium shell and diaphragm
DuraCath
Drug pump; Long term intraspinal (epidural or intrathecal) medication; dispenser; Conditional; Mad of silicone and steel
Actipatch
Pain reduction compression sleeve; Drug free compression sleeve that accelerates healing of soft tissue injuries with EM frequency therapy; Unsafe; Battery operated microchip and EM frequency can cause burns
Alsius coolgard 3000
Intravascular temperature management; Cools blood; Unsafe; Catheters must be disconnected
Martenistic stainless steel aneurysm clip
Aneurysm clip; Manage intracranial aneurysms and arteriovenous malformations (AVMs); Unsafe; Ferromagnetic stainless steel; Martenistic =; never safe
Vari-angle aneurysm clip
Aneurysm clip; Manage intracranial aneurysms and arteriovenous malformations (AVMs); Unsafe; 1 reported aneurysm clip fatality
Yasargil aneurysm clip
Aneurysm clip; Manage intracranial aneurysms and arteriovenous malformations (AVMs); Conditional:; 1.5T or less
Austenitic stainless steel aneurysm clip
Aneurysm clip; Manage intracranial aneurysms and arteriovenous malformations (AVMs); Conditional:; 1.5T or less; Austenitic = ok
Phynox aneurysm clip
Aneurysm clip; Manage intracranial aneurysms and arteriovenous malformations (AVMs); Conditional:; 1.5T or less
Elgiloy aneurysm clip
Aneurysm clip; Manage intracranial aneurysms and arteriovenous malformations (AVMs); Conditional:; 1.5T or less
Titanium alloy aneurysm clip
Aneurysm clip; Manage intracranial aneurysms and arteriovenous malformations (AVMs); Conditional:; 1.5T or less; Gives smallest artifact out of all types of aneurysm clips
Baha
Bone conduction implant; Hearing aid that works via bone (skull and jaw) conduction instead of outer/middle ear; Conditional; Outer sound processor must be removed but implant itself is conditional; Uses bone conduction rather than sound waves
Bravo
Gastric monitoring system; PH monitoring system of gastric reflux inserted endoscopically; Unsafe; No MRIs within 30 days of procedure
Breast tissue expanders
Tissue expander; Used for breast reconstruction, deformity corrections, cosmetic reasons; Generally unsafe; Some expanders have magnetic ports that are significantly attracted to MR field, produce large artifacts, and can injure patients
Breast implants
Breast implant; Typically silicone or saline; Safe
Cardiovascular catheter
Catheter; Used to measure cardiac pressure, output, saturation, etc; Conditional or unsafe (when made with conducting material); Used in iMRI
Poppen Blaylock carotid artery vascular clamp
Vascular clamp; Unsafe; Only contraindicated carotid clamp because of ferromagnetism; All other carotid clamps are conditional
Celsius control system (innercool)
Temperature control system; Catheter, console, and disposables; Conditional; "Head only" scanning; MRIs on all other body parts is unsafe
Codman Hakim programmable valve
Cerebrospinal fluid (CSF) shunt valve; Treat hydrocephalus by bypassing fluid; Conditional; Valve pressure can be changed by magnetic field
Polaris adjustable pressure valve
Cerebrospinal fluid (CSF) shunt valve; Treat hydrocephalus by bypassing fluid; Conditional; Despite magnetic components, it can be scanned at 3T or less
Delta shunt assembly, pulsar valve, Sophy mini monopressure valve
Cerebrospinal fluid (CSF) shunt valve; Treat hydrocephalus by bypassing fluid; Conditional
Cochlear implants
Hearing implant; Electrically activated device for hearing; Conditional at very low fields or unsafe
Codman microcsensor
Intracranial pressure (ICP) monitor; Monitors ICP; Conditional
Coils
Cardiac implant; Stainless steel implant that slows blood flow; Conditional: wait 6 weeks after implantation for device to be secure in vessel wall if it contains ferromagnetic material; 1.5T
Gianturco (Cook) stent
Cardiac implant; Stainless steel implant that prevents vessel closure; Conditional: wait 6 weeks after implantation for device to be secure in vessel wall if it contains ferromagnetic material; 1.5T
Coronary artery stent
Cardiac implant; Implant that prevents vessel closure; Conditional:; SAR 2W/kg; 3T or less; 15min max per pulse sequence; There are no ferromagnetic coronary stents so patients can undergo MRIs immediately after placement
Greenfield filter
IVC filter; Traps blood clots, prevents pulmonary embolism; Conditional: wait 6 weeks after implantation for device to be secure in vessel wall if it contains ferromagnetic material; 1.5T
Cozmo pump
Insulin pump; Manages diabetes through insulin infusion; Conditional; Must be removed before MRI
Craniofix
Bone flap fixation device; Secures bone flaps after craniotomy; Conditional:; 1.5T or less
Copper T, Mirena, Etonogestriel
IUDs; Contraceptive; Conditional:; 1.5T; Copper creates only small artifacts
Aquacel Ag, Nyenhuis, Duan, Mepilex Ag
Silver containing wound dressings; Antimicrobial dressing containing silver; Safe
Ambit
Epidural pump; Infuse pain management drugs; Not recommended to be scanned but not tested much
Essure
Contraceptive; Metal implant for birth control near fallopian tube; Safe
Esteem
Hearing aid; Sound processor with sensor and drivers, replicates ossicular chain; Conditional
Hoffmann II, TenXor
External fixation device; Orthopedic device; Conditional; Heating is the primary concern
Fiber optic pacing leads
Pacemaker leads; Temporary pacing; Conditional:; SAR 1.5W/kg
Foley catheter with temperature sensor
Urinary catheter; Sense temperature of urine; Conditional:; Temp sensor removed; Temperature sensor must be disconnected for MRI
Freehand system
Functional neurostimulator (FNS); Neuroprosthesis that stimulates peripheral nerves for hand grasping; Conditional:; 1.5T only
Nonmetallic shunt tubes
Glaucoma drainage shunt tube; Creates drainage pathway to control intraocular pressure for glaucoma patients, made with nonmetallic materials; Safe
Optonol LTD
Glaucoma drainage shunt tube; Creates drainage pathway to control intraocular pressure for glaucoma patients, made with stainless steel; Unsafe
Halo vest
Cervical fixation device; Used to immobilize neck after injury; Conditional; Potential induced current (Faraday's law) in ring and eddy currents which could lead to burns
Hearing aid
Hearing aid; Used to enhance hearing; Must remove prior to entering MR room; Magnetic fields can damage device
Annuloplasty ring
Heart valve prostheses; Reinforces shape of heart valve so it can open and close properly; Conditional; Minimal magnetic attraction at 3T
Hemostatic vascular clips
Vascular clip; Stops bleeding; Conditional:; 1.5T; Made of nonferromagnetic steel and tantalum; Nonmagnetic clips can be scanned immediately after placement
Insound XT lyric hearing device
Hearing device; Aids in hearing; Conditional; Must be removed prior to MRI
Isomed constant flow
Infusion pump; Medication administration; Conditional:; 1.5T; Image distortion only directly around site of implant, worsens with GRE, reduced with SE; Potential PNS
Licox
Oxygen monitoring and temperature probe system; Monitors oxygen and temperature; Conditional
Magec
Magnetic growing rods; For early onset scoliosis (EOS) with an external remote controller (ERC); Conditional
Medivance
Nasogastric sump tube with temperature sensor; Conditional:; 3T or less; 720 G/cm gradient or less; Max of 3W/kg SAR
Medstream
Programmable infusion system; Medication administration; Conditional
Medtronic minimed
Insulin pump; Medication administration; Must be removed prior to MRI
Barostim neo
Neuromodulator; Treats heart failure; Unsafe or conditional; IPG model 2101: unsafe; Lead models: unsafe; IPG model 2102: conditional at 1.5T
Endostim LES
Lower esophagus stimulation (LES) neurostimulator; Stimulates dysfunctional LES muscle; Highly specific conditional
Pulsante SPG (sphenopalatine ganglion)
Neurostimulator; Treats cluster headaches; Conditional
Senza SCS
Spinal cord stimulator; Manages pain; Conditional
RNS neuropace
Neurostimulator; Treats refractory partial epilepsy; Unsafe; RNS stimulator and leads are both unsafe
Stimwave
Spinal cord and peripheral nerve stimulators; Treats pain; Conditional
Stimrouter
Neurostimulator; Treats pain of peripheral origin; Conditional
Neurocybernetic prosthesis (NCP)
Vagus nerve stimulator; Treats pain, epilepsy, depression, etc; Conditional
Fatio eyelid spring
Occular clip; Treats scleral buckling; Unsafe
Retinal tack
Occular clip; Treats scleral buckling; Unsafe
Troutman magnetic occular implant
Occular clip; Treats scleral buckling; Unsafe
Non-magnetic occular clips
Occular clip; Treats scleral buckling; Safe
McGee stapedectomy piston prosthesis
Otologic implant; Aids in hearing; Unsafe; Made of ferromagnetic platinum and stainless steel alloy; Recalled by manufacturer
Tuebingen type ventilton tubes with wire
Otologic implant; Aids in hearing; Unsafe
Minmal type ventilation tube with wire (Kurz)
Otologic implant; Aids in hearing; Unsafe
MP35N occluder
Cardiac occluder; Treates PDA, ASD, VSD, PFO; Conditional:; 1.5T; Nonferromagnetic; Can be scanned immediately after placement
304V stainless steel occluder
Cardiac occluder; Treats PDA, ASD, VSD, PFO; Conditional; Weakly ferromagnetic; Wait 6 weeks after placement to scan
Palatal system implant assembly
Palatal implant; Treats sleep disordered breathing; Conditional
Duraphase
Penile implant; Penile dysfunction; Inadvisable
Omniphase
Penile implant; Penile dysfunction; Inadvisable
Pessary
Vaginal or rectal ring shaped device; Nonsurgical way to treat pelvic support defects; Safe; Metallic components cause significant artifacts
Pillcam
Capsule endoscopic device; Ingested capsule camera for visualization of the GI tract, moved through system by peristalsis, used to diagnose various small intestine diseases; Unsafe
Prometra programmable pump
Drug pump; Intrathetcal infusion of morphine; Conditional
RF bion microstimulator
Peripheral nervous system stimulator; Stimulates PNS; Conditional:; Wait 6-8 weeks post op to scan; 1.5T; Heating risk; Only certain pulse sequences allowed; Check for function post MRI
Reveal plus insertable loop recorder
Loop recorder; Single use programmable ECG recorder for patients at risk for cardiac arhythmias; Conditional; Contains no leads
St Jude implantable cardiac monitor (SJM ICM)
Cardiac monitor; Records ECG for patients with increased risk of arrhythmias and transient syndromes; Unsafe
SAM sling
Pelvic belt; Stabilizes open pelvic fractures; Conditional:; 3T or less; Contains ferromagnetic springs that must stay firmly in place during MRI
Bioresorbable scaffolds
Vascular support device; Provides support for coronary arteries; Safe; Same benefits as metal stents without using metallic materials
Tantalum scleral buckles
Occular implant; Repairs retinal detachments; Conditional
Alpha magnetic implant
Bone conduction hearing system; Aids in hearing; Conditional implant; Unsafe spacer, processor, headband
Synchrommed
Drug infusion pump; Infuses medication; Not recommended; MRI temporarily stops rotor of pump
Theraseed
Radioactive seed implant; Delivers low level radiation to prostate gland; Conditional; 1.5T
Deponit (nitro system)
Transdermal patch; Medication patch on skin; Should be removed; Caused a second degree burn from aluminum component
Verichip microtranspoder
Mini implanted radio frequency identification device (RFID); Implanted passive identification device; Conditional:; 7T or less
Neurocontrol vocare
Sacral nerve stimulator/ neuroprosthesis; Bladder control stimulator with both internal and external component; Conditional:; 1.5T