What is biophotonics?
The technology that focuses on the interaction of biological materials with light and other forms of radiant energy whose quantum unit is the photon
Define radiation in the context of biophotonics
Energy that comes from a source and can travel through material or space
Four main biomolecules that compose tissues and cells
DNA, proteins, lipids, and carbohydrates
Four ways light can interact with biomolecules
Reflection, absorption, transmission, and light scattering
What physical principle is the Bradford assay based on?
The absorption of light as a function of wavelength
What gives a material its color in terms of light interaction?
The removal (absorption) of specific wavelengths from visible light
Absorption wavelength and color of the Coomassie dye alone (without protein)
Absorbs blue light at 470 nm, making the dye a reddish-brown color (cationic state)
Absorption wavelength and color of the protein-dye complex
Absorbs yellow wavelengths at 595 nm, making the protein-dye complex blue (unprotonated state)
Main components of a spectrophotometer
A light source that generates specific wavelengths, a sample cuvette path, and a detection sensor that reads light emission/absorption
Peak absorbance wavelength set on the spectrophotometer for the Bradford assay
595 nm
Purpose of colorimetric assays using standard curves
To determine the concentration of unknown samples by measuring absorbances of solutions of known concentrations
Three colorimetric methods for determining total protein content mentioned in the text
Biuret, Lowry, and Bradford
Characteristics of the Biuret protein assay
The oldest method, commonly used in high school labs, involves chelation and redox reactions, and is the least sensitive of the three
Characteristics of the Lowry protein assay
Involves two redox reactions, is more sensitive than Biuret, but is affected by interference from many common laboratory reagents and chemicals
Characteristics of the Bradford protein assay
The most sensitive of the three methods, uses Coomassie Brilliant Blue G-250, and is relatively unaffected by many common laboratory reagents.
Who first described the Bradford protein assay and in what year?
M. Bradford in 1976
Dye used in the Bradford protein assay
Coomassie Brilliant Blue G-250
Chemical form and color of Coomassie G-250 in the Bradford solution before binding
Cationic state, reddish-brown color, with a peak absorption at 470 nm
Chemical form and color of Coomassie G-250 after binding to proteins
Stable unprotonated blue form, with absorption maximum shifting to 595 nm
Primary molecular interaction between Coomassie G-250 and proteins
Electrostatic interactions between negatively charged sulfate groups on the dye and arginine (a very basic amino acid).
Secondary (weaker) dye-protein interaction involving aromatic rings
Electron stacking interactions between the aromatic rings of Coomassie G-250 and aromatic amino acids like tryptophan
Tertiary (weaker) dye-protein interaction with polar amino acids
Weak hydrophobic interactions with polar amino acids having hydrophobic R-groups, such as tyrosine (and phenylalanine)
Four main steps of the Bradford assay procedure
(1) Prepare dilution series of known standards and unknowns
(2) Add Bradford dye and incubate for 5–60 min
(3) Measure A595 in a spectrophotometer
(4) Compile data into a standard curve to determine unknown concentrations
Incubation time requirements for the Bradford assay
At least 5 minutes, but not to exceed 60 minutes
What protein is most abundant in milk?
Casein
Total amino acid count and molecular mass of Casein
224 amino acids with a molecular mass of 24,967 daltons
Number and types of strong Coomassie-binding amino acids in Casein
13 total: 4 arginines (R), 1 tryptophan (W), 4 tyrosines (Y), and 4 histidines (H).
Molecular mass comparison between Coomassie dye and a typical amino acid
Coomassie is ~854 daltons, compared to an average of ~110 daltons per amino acid.
How Coomassie dye interacts with a typical protein in solution based on size
Because Coomassie is much larger than amino acids, a few dye molecules bind and "coat" the protein in solution
What graph type is used to plot raw absorbance data vs. concentration for the standard curve?
Linear graph paper (or graphing utilities like Microsoft Excel)
What does a high correlation coefficient (R2 > 0.9) indicate?
A high degree of linearity in the data, meaning the standard curve fits well and can accurately estimate unknown values
What are the axes assignments when generating a Bradford standard curve in this lab?
Measured absorbance (A595) on the y-axis plotted against known protein concentration (mg/mL) on the x-axis.
Initial preparation step for testing milk samples in Laboratory Activity 1
Prepare a 1:50 dilution of the milk samples using 1x PBS (e.g., 4uL milk added to 196uL 1x PBS)
What solution is used as the "blank" in the Bradford cuvettes?
1x PBS (with Bradford dye reagent added)
Volume of Bradford dye reagent added to each cuvette
1 mL (of 1x Bradford dye reagent).
Volume of standards, blank, or diluted samples added to each cuvette
20uL
What material is used to cover each cuvette before mixing?
Parafilm (small pieces to seal cuvettes)
Mixing technique used after sealing cuvettes with Parafilm
Invert each cuvette 3 times to mix
Spectrophotometer instrument used for quantitative evaluation in Lab Activity 2
DeNovix Spectrophotometer (DS-11x).
Proper orientation when loading a cuvette into the DeNovix spectrophotometer
Make sure the clear window of the cuvette is placed in the direction of the arrow
Critical warning regarding re-blanking the instrument after running the standard curve
DO NOT REBLANK THE INSTRUMENT!
Why must an optional dilution factor be entered into the DeNovix software?
To automatically multiply and correct the calculated concentration if the sample was diluted prior to the assay (e.g., the 1:50 milk dilution).
What does the Reports tab show on the DeNovix after data collection?
A list of sample names, concentrations (mg/mL), dilution factors, recalculated concentrations, assay type, absorbance, and R2 values.
Formula used to calculate standard serial dilution volumes (V1) in Laboratory Activity 3
C1V1 = C2V2 solved as V1 = (C2V2)/C1
How is the volume of 1x PBS calculated when making standard serial dilutions from a stock?
Subtract the calculated V1 volume from the total target volume (e.g., 50uL - V1 = PBS volume)
Mathematical definition of the coefficient of determination (R2) given in the text
R2 = 1 - (SSRES/SSTOT) = 1 - Ei(yi - y^i)2 / Ei(yi - y-)2
What does an R2 value of 1.0 mean?
All the variation in the y values is perfectly accounted for by the x values (zero residual error from the regression line).
What does an R2 value of 0 mean?
None of the variation in the y values is accounted for by the x values (no linear relationship).
What does a low R2 value indicate about pipetting in this lab?
High variability and poor precision/accuracy in pipetting skills, leading to inaccurate concentration readings of unknown samples
When a MedTech gets a low R2 discrepancy in clinical sample runs, what is typically investigated first?
Human technique errors (the technician's pipetting skills), as humans have a greater propensity for variability than certified instruments and controls