Color Doppler
Used to assess perfusion of tissue and flow within blood vessels
Demonstrates average flow velocities
Flow direction displayed according to color map
Mosaic patterns indicate turbulence which can be caused by normal anatomy abnormalities or incorrect machine settings
Persistence is also called
Temporal compounding
Persistence
Color sensitivity
Uses average of previous frames to create the displayed image
Used to highlight low velocity flow
Improves signal to noise ratio on color image
Wall filter is also known as
High pass filter
Always use less that __ degrees angle on PW
60
RI equation
PVS - EDV divided by PSV
RI
Compares systolic and diastolic peak flow and represents the difference between them
PI equation
PSV - EDV divided by PSV + EDV
Pulsatility index
Describes the variation of arterial flow during a single cardiac cycle
Which of the following is true regarding color Doppler system controls?
A smaller sample size is preferred
Lower persistence settings is preferred for venous evaluations
Lower wall filter settings prevent artifactual color display
Low priority settings are preferred for slower blood flow
A smaller sample size is preferred
Which of the following is true regarding color Doppler system controls
Larger color packet size is preferred
Increasing wall filter can reduce color ghosting/clutter
Lower wall filter settings prevent artifactual color display
Lower priority settings are preferred for slower blood flow
Increasing the wall filter can reduce color ghosting/clutter
Lower threshold settings display _____ color (lower priority)
Less
Higher threshold settings display _____ color (higher priority)
More
What is preferred for evaluating slow flow in a vessel
Smaller color sample size
Higher persistence, wall filter and priority settings
2D imaging requires __________ of the US beam to reflectors, while other color and PW Doppler require _________ of the US beam to blood flow
Perpendicular incidence; parallel incidence
Primary advantage of Power Doppler over color Doppler
More sensitive to low flow velocities and less angle dependent
Larger PW Doppler sample sizes are use to evaluate?
Ovarian flow
What 3 waveform measurements do not require the use of an angle on the Doppler cursor?
Pulsatile index
Resistive index
S/D ratio
What causes a higher RI value
High PSV and Low EDV
S/D ratio is typically used to evaluate the:
Umbilical artery