1/116
Lecture
assay
process of determining the amount of analyte in a sample
analyte
the chemical substance being measured
qualitative analysis
correlate signal to what is in a sample
quantitative analysis
correlate signal to how much is in a sample
voltage
electrical potential energy
measured as the electric potential difference between two points
currect
the rate of flow of charge past a point in a circuit
transducer
device that converts an input stimulus into an electrical output
signal
a measured quantity that is correlated to the amount of analyte
contains info about the chemical species of interest
noise
unwanted variation in a measured quantity
random fluctuations in background
σBkg
signal to noise ratio (S/N)
magnitude of the signal divided by the magnitude of the noise
background
approximately constant signal (μBkg) measured in the absence of analyte
detection limit
amount of analyte that corresponds to a signal just greater than the mean of the background plus 3 std. dev. of its noise
minimum = (+)
maximum = (-)
sample matrix
all components of a sample except the analyte
blank
man-made sample that lacks analyte but contains solvent and reagents
positive control
standard sample that contains known quantity of analyte of interest
prevents false negative results
negative control
standard sample that does not contain any analyte
prevents fast positive results
interference
a specific chemical substance in a sample matrix that causes a systematic error in a measured quantity
act on analyte
act on reagent
be a source of a large background signal
cause negative or positive bias
cause absolute or proportional errors
selectivity
the extent to which other substances interfere with the determination of an analyte
good selectivity = minimal interferences
poor selectivity = prone to certain interferences
masking agent
a reagent that prevents one or more components in a sample matrix from interfering with an analysis
attempts to improve analytical sensitivity and selectivity
is NOT a source of large background signal
does NOT reduce personal errors
accuracy
measurement denotes the closeness of an experimental value to the true value
right on target
precision
set of measurements refers to the agreement among the results
std. dev.
all in one spot, may not be target
the reproducibility between replicate measurements
absolute error
difference between measured and true value
can be positive or negative
relative error
error in a measurement expressed as a percentage
replicates
samples from the same source, run using the same method, under the same conditions, expected to give the same result in the absence of error
random / indeterminate errors
introduce uncertainty
symmetric about the true value
treat with statistics
affects precision

systematic / determinate errors
introduce bias
measured value is always higher or lower than its true value
eg. rxn that did not go to completion
affects accuracy

types of systematic error: instrument
minimize with careful, regular calibration
voltage drift
can usually be corrected
types of systematic error: method
chemistry does not behave as expected
most difficult to identify and correct
incomplete reactions
interference from non-analytes
false positive/negative results
types of systematic error: personal
incorrect recording of data
deviations from an established method
types of systematic error: proportional
error is proportional to samples
always measuring % higher than true value
effect is independent of the magnitude of the measurement
types of systematic error: constant
error is always same in each sample
always measuring x lower than true value
is small if measuring small but large if measuring large
gaussian distribution
used to describe data clustered about a mean value
continuous and symmetric distribution
bell curve shape
peak position determined by mean
width and heigh determined by std. dev.
total area under curve is 1
std. dev. of gaussian distribution
1 std. dev. = 68.3%
2 std. dev. = 95.4%
3 std. dev. = 99.7%
population
all possible measurements that are of interest
sample
limited number of measurements that are representative of the population
mean
average
deviation
difference between a measured value and the mean value of all measurements
difference between true and average
std. dev. and RSD
measure of the uncertainty and precision associated with a measurement
propagation of error
find uncertainty after calculations
purpose of t-statistic
permits use of sample data to test hypotheses about unknown population means without knowledge of the population std. dev.
purpose of significance test
to determine whether the difference between 2 or more values is too large to be explained by indeterminate error
null hypothesis
assume true until proven wrong
postulates the 2 observed quantities are the same
assume 2 results are the same then apply test to see if null can be statistically rejected
t test: case 1
compare measured value to actual value
t exp < t table = no significant difference
null true
t exp > t table = significant difference
reject null
standard reference materials
highest-quality materials
certified and traceable
carefully analyzed by the supplier
accurately known composition
use and sold for a variety of analytical techniques
prepared and sold by both companies and organizations
p value
probability value of results
lower = stronger evidence = data is unlikely under null
statistically significant
higher = weak evidence = could easily happen by random chance
cannot reject null
t-test: case 2
comparing 2 data sets
std. dev. is pooled
compare sample means (replicate measurements)
F-test
compare variances by calculating the ratio of the 2 variances (or std. dev.)
used to compared the precision of two sets of measurements
t-test: case 3
paired t-test
uses paired/matched measurement data
eg. comparing single measurement made with 2 methods on several different samples
eg. comparing measurements before vs after
compare methods of single measurements on several different samples
same number of measurements
Grubbs Test
used to exclude an outlier from a data set
significant figures
reflection of a measurement’s magnitude and uncertainty
significant figures: zeros
in decimals:
leading = not significant
training = significant
significant figures: pH
right of decimal = significant
eg. pH = 2.45 → [H] = 2.8E-2
significant figures: exact number
infinite
significant figures: addition/subtraction
round answer to the last decimal place that is significant for each measurement in calculation
# with least amount of digits/decimal places
135.621 + 97.33 + 21.2163 = 254.17
significant figures: scientific notation
convert to common exponent

significant figures: multiplication/division
round to the measurement with the fewest significant figures
standard solutions
has known concentrations
sample solution
has unknown concentration
linear regression of straight line calibration curve equation
y = mx + b
assumptions of calibration curve
signal to concentration is linear
any difference between experimental data and calculated regression line is result of indeterminate errors affecting y only
indeterminate errors affecting y are normally distributed
indeterminate errors in y are independent of x value
calibration curve
plot of a measured signal vs a known quantity
calibration curve: sensitivity
slope (m) of calibration curve
high sensitivity = small change in [] yields large change
calibration curve: dynamic range
the concentration range over which the calibration curve is analytically useful
high much [] change will give change in y signal
concentration limit of detection
smallest signal detectable
calibration curve: selectivity
for linear calibration plots, the selectivity of a method for compound 1 vs compound 2 is reflected by the ratio of slopes m1/m2
bigger ratio = better
matrix effect
combined effect of all non-analyte components in a sample on the quantitative measurement of the analyte
used when specific interference is not known
eg. some component in the sample:
generates signal similar to analyte
has a chemical interaction with analyte
is co-isolated with the analyte
standard addition
means of accounting for matrix effect
most precise results are obtained when the amount of standard added is comparable in magnitude to original quantity of analyte
time consuming
extra opportunity for dilution error
added standard should not overwhelm any interferences
may require large quantities of sample
internal standard
intentionally added substance of known quantity that is not expected to be found in the sample but is expected to behave similarly
different than the analyte
is analyzed with the analyte to account for losses during sample processing or fluctuations in instrument signals
reference the analyte signal to the internal standard signal
chemically similar to analyte so systematic effects of matrix that increase/decrease analyte signal will have similar effect on signal of internal standard
used with both external standard calibration and standard addition methods
sampling
process by which a sample population is reduced to a size suitable for laboratory analysis
identify population
develop sampling strategy
collect and preserve a gross sample
reduce gross sample to a lab sample
replicate analyses of the samples
problems with real samples: defining the problem
expected concentration rage of analyte?
accuracy and precision required?
where, when, how many samples to take, what size?
problems with real samples: storing and preserving sample prior to analysis
protect from light, control temperature and pH
store most samples in glass but store metal solutions in plastic
problems with real samples: homogenization
may require crushing and grinding, drying, digesting, decomposing
problems with real samples: overwhelming matrix effects
requires extraction, precipitation, distillation, or other separations of analyte from interferences
problems with real samples: analyte too dilute for reliable analysis
requires pre-concentration steps such as extraction of evaporation
sensitivity, specificity, ROC curve
sensitivity = true positive rate
rule out when gives negative result
specificity = true negative rate
rule in when gives positive result
too see if method is working
ROC curve
receiver operating characteristic
too see if method is working
plots sensitivity against 1-specificity
x = false positive (1-specificity)
y = sensitivity
area under curve = 1 = perfect
AUC < 0.5 = useless
AUC > 0.8 = method is somewhat effective
analog signal
takes on any value and the signal is continuous
equilibrium constant
K: ratio of [products] and [reactants] when there is no net change in system
products over reactants
bigger is better and determines if solution becomes acidic or basic
strong acid
ionize completely in water
strong base
ionize completely in water
weak acid
ionize partially in water
weak base
ionize partially in water
Kw
Kw = Ka * Kb = 1.00E-14
polyprotic acids
have 2 or more ionizable protons
amphiprotic
act as either proton donor (acid) or acceptor (base)
acid HA with a negative charge
charge balance equation
sum of positive charges in solution equals sum of negative charges
include water dissociation
if has numbered, charge, go infront of species
eg. HPO4 ^(-2) = 2[HPOr ^(-2)]
mass balance
sum of the amounts of species in a solution containing a particular atom must equal the amount of that atom delivered to solution

idk bro
isoelectric point
point that has no charge
solubility
can be dissolves in a solvent up to a certain point
Ksp
larger = more soluble compound
precipiate will form
complexation equilibria
binding between metal and ligand
denticity
number of coordinate bonds to a central metal ion
chelating agent
non-cyclic, multidentate ligand
cryptand
polycyclic, multidentate ligands
EDTA
has 6 pKas
4 basic
2 acidic
titration has 3 components
determine quantity of substance A by adding measured increments of substance B
substance B reacts with substance A
some means of indicating the endpoint of which all of A has reacted
titration examples
acid-base
precipitation
complexometric
oxidation-reduction
advantages of volumetric titrations
rapid
convenient
economical
accurate
easily automated
disadvantages of volumetric titrations
manual titrations are tedious
sensitivity and selectivity can be limited
titrant
solution of known composition and concentration dispensed from buret