Doppler is used to
detect motion, primarily of blood flow
Types of Doppler
Continuous Wave Doppler
Pulsed Wave Doppler
Color Doppler
Power Doppler
does ultrasound physics apply to Doppler?
yes
Doppler Shift is Doppler Frequency
describes the apparent change in pitch due to a change in position (movement) between the sound source and receiver
the frequency of a sound changes when
the sound source and receive move closer together or farther apart
if a wave reflects from a moving object….
the reflected wave will have a different wavelength than the original value
the faster the moving object, the ____
higher the frequency
source of the sound wave =
transducer
reflector of sound wave =
red blood cells
Positive Doppler Shift
reflectors moving TOWARD the transducer
will create a shorter wavelength than the original
makes a higher frequency
Negative Doppler Shift
Reflectors moving AWAY from the transducer
will create a longer wavelength than the original
makes a lower frequency
Motion that is exactly perpendicular (90 degrees) will
not show any shift in frequency
as RBC’s move faster…
the reflected frequency is higher
Demodulation
signal processing that happens within the ultrasound system
looks at Doppler only
gets rid of gray scale
Doppler Shift =
Doppler Shift = received frequency - transmitted frequency
transmitted and reflected frequencies are in what unit
MHz
Doppler shift falls into what range
audible range
2-20kHz or 20-20,000 Hz
Positive Doppler Shifts are assigned as
RED on Color Doppler
appears ABOVE the baseline in spectral Doppler

Negative Doppler Shifts are assigned as
BLUE on color doppler
appears BELOW the baseline of spectral Doppler

Speed gives info about
magnitude only (how fast is something traveling?)
Velocity gives info about
magnitude AND direction (how fast and in what direction is something traveling?)
The Doppler Equation
Doppler Shift = (2 x velocity of blood x transducer frequency x cosine) / propagation speed (1.54 mm/us)
Desired angle
60 degrees or less

how to angle
we must angle correct and make the cursor parallel to the vessel wall that allow the machine to make accurate velocity calculations

Nondirectional
like a fetal doptone
measures the PRESENCE of RBC’s and make audio with single speaker to hear it
CANNOT determine direction
Bidirectional
uses Doppler Shift to determine direction of flow
positive and negative shift determines toward or away from the transducer
Bidirectional has…
2 speakers
one toward and one away
headphones (one ear piece hears towards, the other hears away)
Bidirectional allows for what display of information?
visual/graphic display of information
forward flow is above the baseline
away flow is below the baseline
Phase Quadrature is AKA
Quadrature Detection
Phase Quadrature definition
signal processing that separates the echos into negative and positive echoes (allows directional information)
Phase Quadrature obtains _____ information
directional information
Demodulation vs. Quadrature Detection
both are signal processing techniques
Demodulation — separates signals to extract Doppler frequency from transmitted frequency
does NOT give info about direction
Quadrature detection —— separates echoes into negative and positive
What information does the signal processing known as Phase Quadrature detection decipher?
directional information
What does the signal processing known as demodulation provide for information?
Doppler frequency shift itself
Continuous Wave Doppler
two crystals required in the transducer
one constantly transmits
one constantly receives

continuous waves are ______ accurate and measures very ______ velocities
very, high

Can a continuous wave doppler determine the exact location of the signal?
NO

continuous wave Doppler listens….
along the ENTIRE LENGTH OF THE BEAM PATH

does continuous wave doppler create an image?
no
Advantages of Continuous Wave Doppler
accurately measures very high velocity blood flow

Disadvantages of Continuous Wave Doppler
exact location of RBC’s cannot be determined
RANGE AMBIGUITY—> signals arise from all blood cells in the region of overlap between the transmit and receive beams

Pulsed Wave Doppler
one ONE crystal
crystal alternates between transmitting and receiving
5-30 cycles per pulses with pulsed wave
2D is more like 2-3 cycles
Can the location of Pulsed Wave Doppler be determined by the sonographer?
yes
Changing Location on Pulsed Wave Doppler
huge advantage over continuous wave
range resolution
range specificity
freedom from range ambiguity artifact
Sample volume or gate
the small marker that can be adjusted by the sonographer to determine the location to sample to Doppler signal

Advantages of Pulsed Wave Doppler
ability to select the exact location where velocity is measured
called range resolution, range specificity, or freedom from range ambiguity artifact
Disadvantages of Pulsed Wave Doppler
inaccurate measurement of high velocity signals
high velocity flow in one direction is incorrectly displayed as traveling in the opposite direction
Aliasing
Aliasing is the…
most common error associated with Doppler ultrasound
Aliasing definition
a false depiction of very high velocities incorrectly displayed as going the opposite direction
When flow is very fast and goes all the way to the top of the spectral image (forward flow), it then wraps around incorrectly and reappears at the very bottom of the display

Nyquist Limit
the very top of the display with the maximum velocity that can be displayed before aliasing occurs
or the opposite
the very bottom of the display is the maximum away velocity that can be displayed before aliasing wraps around to the top of the screen

Aliasing occurs with ____ velocities because…
high velocities because the velocity needs to be measured at such a rapid rate
Aliasing occurs with PW doppler due to
pulses
does aliasing happen with CW doppler?
NO, able to measure high velocities, continuous sampling
PRF =
number of pulses per second
as depth decreases, PRF _______
increases
the system only waits a short time for the return and can create new pulses more rapidly
as depth increases, PRF ______
decreases
the system waits for a longer time for the return which affects the frequency of creating new pulses
Nyquist Limit =
Nyquist Limit = ½ the PRF or PRF/2
Nyquist Limit and PRF are directly related
if the PRF is too low, the blood flow…
cannot be sampled fast enough
Shallow Sample
HIGH PRF (rapidly moving blood is measured frequently)
HIGH Nyquist Limit (speed limit)
HIGH velocities measured accurately
Deep Sample
LOW PRF (rapidly moving blood is measured infrequently)
LOW Nyquist limit (speed limit)
System is unable to measure high velocity flow
Prone to aliasing
Aliasing occurs when….
Doppler Shift (speed) exceeds the Nyquist Limit (speed limit)

Avoid Aliasing by
raising the Nyquist limit (speed limit) or reduce the Doppler shift

5 Ways to fix/avoid Aliasing
Adjust the scale to its maximum
Find a different view with a shallower sample volume
Use a lower frequency transducer
Shift the baseline
Use CW doppler
Gray Shades of a Spectrum
amplitude of the reflected signal
OR
Number of blood cells creating the reflection
Brighter areas on the spectrum (waveform) mean that…
there are MORE RBC’s at that particular point, creating the Doppler shift
x axis =
time
y axis =
velocity
ABOVE the baseline is a _____
positive shift
BELOW the baseline is a ____
negative shift
Color Doppler
2D doppler
not in a single location (PW)
Not along a single transmit line (CW)
in Color Doppler, velocity info is
assigned a color and superimposed on a 2D gray scale anatomic image
black and white =
anatomy
color =
blood flow
Color Flow Doppler gives info on
location so it HAS to be a pulsed ultrasound technique
direction (basically is a color depiction of the Doppler shift)
Color Flow Doppler has
range resolution (or range specificity)
is Color Flow Doppler subject to aliasing?
yes
Color Flow Doppler is _________ in terms of angles
semiquantitive
angles are less important than with PW
Semi Quantitative
provides mean (average) velocities
CW and PW give peak velocities
Color Maps
convert measured velocity information into color that appears in gray scale image

the dictionary or look up table that is used for this conversion for color amps is displayed as
a vertical bar is various colors
black at center

2 common dictionaries
velocity mode
variance mode

To figure out if flow is moving toward or away, we need to understand a
“home” for the transducer

How to find a “home” for the transducer
find the corner of the color box where you can draw a straight line down and not hit the colored part of the vessel (this is home)
if the vessel is BLUE/BELOW baseline on the color map, the blood is moving AWAY from the transducer and vise versa for red

Power Doppler is AKA
energy mode or color angio
Power Doppler identifies
PRESENCE of a Doppler Shift
Power Doppler does NOT evaluate…
speed or direction
nondirectional color doppler
do all vessels have the same color in Power Doppler?
yes
in Power Doppler, strength (amplitude) is processed…
without deciphering direction or speed
Amplitude is ____ related to
DIRECTLY related to the number of moving blood cells
larger amplitude signals = brighter
weaker amplitude signals = darker
Advantages of Power Doppler
increased sensitivity to low flow/velocity
venous flow
small vessels
Unaffected by Doppler angles
except exactly 90 degrees
No aliasing
velocity information is ignored
Disadvantages of Power Doppler
no measurement of velocity
lower frame rates
flash artifact or bursts of color can occur from motion of the transducer, patient, or soft tissue
Ghosting (doppler artifact)
low frequencies that appear on Color Doppler
Clutter (doppler artifact)
low frequencies that appear on Spectral Doppler
Crosstalk (doppler artifact)
mirror image of spectral display with baseline being the mirror
fix with decreasing wave gain or changing angle
other doppler artifact
aliasing
Spectral Analysis
not all blood cells travel at the same speed or direction
breaks the signals into individual signals that make up the entire Doppler signal
2 methods for Spectral Analysis
Fast Fourier Transformation (FFT)
Autocorrelation (color)