Doppler is used to

detect motion, primarily of blood flow

2
New cards

Types of Doppler

  • Continuous Wave Doppler

  • Pulsed Wave Doppler

  • Color Doppler

  • Power Doppler


3
New cards

does ultrasound physics apply to Doppler?

yes

4
New cards

Doppler Shift is Doppler Frequency

describes the apparent change in pitch due to a change in position (movement) between the sound source and receiver

5
New cards

the frequency of a sound changes when

the sound source and receive move closer together or farther apart

6
New cards

if a wave reflects from a moving object….

the reflected wave will have a different wavelength than the original value

7
New cards

the faster the moving object, the ____

higher the frequency

8
New cards

source of the sound wave =

transducer

9
New cards

reflector of sound wave =

red blood cells

10
New cards

Positive Doppler Shift

reflectors moving TOWARD the transducer

  • will create a shorter wavelength than the original

  • makes a higher frequency


11
New cards

Negative Doppler Shift

Reflectors moving AWAY from the transducer

  • will create a longer wavelength than the original

  • makes a lower frequency


12
New cards

Motion that is exactly perpendicular (90 degrees) will

not show any shift in frequency

13
New cards

as RBC’s move faster…

the reflected frequency is higher

14
New cards

Demodulation

signal processing that happens within the ultrasound system

  • looks at Doppler only

  • gets rid of gray scale


15
New cards

Doppler Shift =

Doppler Shift = received frequency - transmitted frequency

16
New cards

transmitted and reflected frequencies are in what unit

MHz

17
New cards

Doppler shift falls into what range

audible range

  • 2-20kHz or 20-20,000 Hz


18
New cards

Positive Doppler Shifts are assigned as

RED on Color Doppler

appears ABOVE the baseline in spectral Doppler

<p>RED on Color Doppler</p><p>appears ABOVE the baseline in spectral Doppler</p>
19
New cards

Negative Doppler Shifts are assigned as

BLUE on color doppler
appears BELOW the baseline of spectral Doppler

<p>BLUE on color doppler<br>appears BELOW the baseline of spectral Doppler</p>
20
New cards

Speed gives info about

magnitude only (how fast is something traveling?)

21
New cards

Velocity gives info about

magnitude AND direction (how fast and in what direction is something traveling?)

22
New cards

The Doppler Equation

Doppler Shift = (2 x velocity of blood x transducer frequency x cosine) / propagation speed (1.54 mm/us)

23
New cards

Desired angle

60 degrees or less

<p>60 degrees or less </p>
24
New cards

how to angle

we must angle correct and make the cursor parallel to the vessel wall that allow the machine to make accurate velocity calculations

<p>we must angle correct and make the cursor parallel to the vessel wall that allow the machine to make accurate velocity calculations</p>
25
New cards

Nondirectional

  • like a fetal doptone

  • measures the PRESENCE of RBC’s and make audio with single speaker to hear it

  • CANNOT determine direction


26
New cards

Bidirectional

  • uses Doppler Shift to determine direction of flow

    • positive and negative shift determines toward or away from the transducer


27
New cards

Bidirectional has…

2 speakers

  • one toward and one away

    • headphones (one ear piece hears towards, the other hears away)


28
New cards

Bidirectional allows for what display of information?

visual/graphic display of information

  • forward flow is above the baseline

  • away flow is below the baseline


29
New cards

Phase Quadrature is AKA

Quadrature Detection

30
New cards

Phase Quadrature definition

signal processing that separates the echos into negative and positive echoes (allows directional information)

31
New cards

Phase Quadrature obtains _____ information

directional information

32
New cards

Demodulation vs. Quadrature Detection

  • both are signal processing techniques

  • Demodulation — separates signals to extract Doppler frequency from transmitted frequency

    • does NOT give info about direction

  • Quadrature detection —— separates echoes into negative and positive


33
New cards

What information does the signal processing known as Phase Quadrature detection decipher?

directional information

34
New cards

What does the signal processing known as demodulation provide for information?

Doppler frequency shift itself

35
New cards

Continuous Wave Doppler

two crystals required in the transducer

  • one constantly transmits

  • one constantly receives


<p>two crystals required in the transducer</p><ul><li><p>one constantly transmits</p></li><li><p>one constantly receives</p></li></ul><p></p>
36
New cards

continuous waves are ______ accurate and measures very ______ velocities

very, high

<p>very, high</p>
37
New cards

Can a continuous wave doppler determine the exact location of the signal?

NO

<p>NO</p>
38
New cards

continuous wave Doppler listens….

along the ENTIRE LENGTH OF THE BEAM PATH

<p>along the ENTIRE LENGTH OF THE BEAM PATH</p>
39
New cards

does continuous wave doppler create an image?

no

40
New cards

Advantages of Continuous Wave Doppler

accurately measures very high velocity blood flow

<p>accurately measures very high velocity blood flow</p>
41
New cards

Disadvantages of Continuous Wave Doppler

  • exact location of RBC’s cannot be determined

  • RANGE AMBIGUITY—> signals arise from all blood cells in the region of overlap between the transmit and receive beams


<ul><li><p>exact location of RBC’s cannot be determined</p></li><li><p>RANGE AMBIGUITY—&gt; signals arise from all blood cells in the region of overlap between the transmit and receive beams</p></li></ul><p></p>
42
New cards

Pulsed Wave Doppler

one ONE crystal

  • crystal alternates between transmitting and receiving

  • 5-30 cycles per pulses with pulsed wave

    • 2D is more like 2-3 cycles


43
New cards

Can the location of Pulsed Wave Doppler be determined by the sonographer?

yes

44
New cards

Changing Location on Pulsed Wave Doppler

  • huge advantage over continuous wave

  • range resolution

  • range specificity

  • freedom from range ambiguity artifact


45
New cards

Sample volume or gate

the small marker that can be adjusted by the sonographer to determine the location to sample to Doppler signal

<p><span>the small marker that can be adjusted by the sonographer to determine the location to sample to Doppler signal</span></p>
46
New cards

Advantages of Pulsed Wave Doppler

  • ability to select the exact location where velocity is measured

    • called range resolution, range specificity, or freedom from range ambiguity artifact


47
New cards

Disadvantages of Pulsed Wave Doppler

  • inaccurate measurement of high velocity signals

  • high velocity flow in one direction is incorrectly displayed as traveling in the opposite direction

    • Aliasing


48
New cards

Aliasing is the…

most common error associated with Doppler ultrasound

49
New cards

Aliasing definition

  • a false depiction of very high velocities incorrectly displayed as going the opposite direction

  • When flow is very fast and goes all the way to the top of the spectral image (forward flow), it then wraps around incorrectly and reappears at the very bottom of the display


<ul><li><p><span>a false depiction of very high velocities incorrectly displayed as going the opposite direction</span></p></li><li><p><span>When flow is very fast and goes all the way to the top of the spectral image (forward flow), it then wraps around incorrectly and reappears at the very bottom of the display</span></p></li></ul><p></p>
50
New cards

Nyquist Limit

the very top of the display with the maximum velocity that can be displayed before aliasing occurs

  • or the opposite

    • the very bottom of the display is the maximum away velocity that can be displayed before aliasing wraps around to the top of the screen


<p>the very top of the display with the maximum velocity that can be displayed before aliasing occurs</p><ul><li><p>or the opposite</p><ul><li><p><span>the very bottom of the display is the maximum away velocity that can be displayed before aliasing wraps around to the top of the screen</span></p></li></ul></li></ul><p></p>
51
New cards

Aliasing occurs with ____ velocities because…

high velocities because the velocity needs to be measured at such a rapid rate

52
New cards

Aliasing occurs with PW doppler due to

pulses

53
New cards

does aliasing happen with CW doppler?

NO, able to measure high velocities, continuous sampling

54
New cards

PRF =

number of pulses per second

55
New cards

as depth decreases, PRF _______

increases

  • the system only waits a short time for the return and can create new pulses more rapidly


56
New cards

as depth increases, PRF ______

decreases

  • the system waits for a longer time for the return which affects the frequency of creating new pulses


57
New cards

Nyquist Limit =

Nyquist Limit = ½ the PRF or PRF/2

  • Nyquist Limit and PRF are directly related


58
New cards

if the PRF is too low, the blood flow…

cannot be sampled fast enough

59
New cards

Shallow Sample

  • HIGH PRF (rapidly moving blood is measured frequently)

  • HIGH Nyquist Limit (speed limit)

  • HIGH velocities measured accurately


60
New cards

Deep Sample

  • LOW PRF (rapidly moving blood is measured infrequently)

  • LOW Nyquist limit (speed limit)

  • System is unable to measure high velocity flow

  • Prone to aliasing


61
New cards

Aliasing occurs when….

Doppler Shift (speed) exceeds the Nyquist Limit (speed limit)

<p>Doppler Shift (speed) exceeds the Nyquist Limit (speed limit)</p>
62
New cards

Avoid Aliasing by

raising the Nyquist limit (speed limit) or reduce the Doppler shift

<p>raising the Nyquist limit (speed limit) or reduce the Doppler shift</p>
63
New cards

5 Ways to fix/avoid Aliasing

  1. Adjust the scale to its maximum

  2. Find a different view with a shallower sample volume

  3. Use a lower frequency transducer

  4. Shift the baseline

  5. Use CW doppler


64
New cards

Gray Shades of a Spectrum

amplitude of the reflected signal

OR

Number of blood cells creating the reflection

65
New cards

Brighter areas on the spectrum (waveform) mean that…

there are MORE RBC’s at that particular point, creating the Doppler shift

66
New cards

x axis =

time

67
New cards

y axis =

velocity

68
New cards

ABOVE the baseline is a _____

positive shift

69
New cards

BELOW the baseline is a ____

negative shift

70
New cards

Color Doppler

2D doppler

  • not in a single location (PW)

  • Not along a single transmit line (CW)


71
New cards

in Color Doppler, velocity info is

assigned a color and superimposed on a 2D gray scale anatomic image

72
New cards

black and white =

anatomy

73
New cards

color =

blood flow

74
New cards

Color Flow Doppler gives info on

location so it HAS to be a pulsed ultrasound technique

direction (basically is a color depiction of the Doppler shift)

75
New cards

Color Flow Doppler has

range resolution (or range specificity)

76
New cards

is Color Flow Doppler subject to aliasing?

yes

77
New cards

Color Flow Doppler is _________ in terms of angles

semiquantitive

angles are less important than with PW


78
New cards

Semi Quantitative

  • provides mean (average) velocities

  • CW and PW give peak velocities


79
New cards

Color Maps

convert measured velocity information into color that appears in gray scale image

<p>convert measured velocity information into color that appears in gray scale image</p>
80
New cards

the dictionary or look up table that is used for this conversion for color amps is displayed as

a vertical bar is various colors

  • black at center


<p>a vertical bar is various colors </p><ul><li><p>black at center</p></li></ul><p></p>
81
New cards

2 common dictionaries

  1. velocity mode

  2. variance mode


<ol><li><p>velocity mode</p></li><li><p>variance mode</p></li></ol><p></p>
82
New cards

To figure out if flow is moving toward or away, we need to understand a

“home” for the transducer

<p>“home” for the transducer</p>
83
New cards

How to find a “home” for the transducer

  • find the corner of the color box where you can draw a straight line down and not hit the colored part of the vessel (this is home)

    • if the vessel is BLUE/BELOW baseline on the color map, the blood is moving AWAY from the transducer and vise versa for red


<ul><li><p>find the corner of the color box where you can draw a straight line down and not hit the colored part of the vessel (this is home)</p><ul><li><p>if the vessel is BLUE/BELOW baseline on the color map, the blood is moving AWAY from the transducer and vise versa for red</p></li></ul></li></ul><p></p>
84
New cards

Power Doppler is AKA

energy mode or color angio

85
New cards

Power Doppler identifies

PRESENCE of a Doppler Shift

86
New cards

Power Doppler does NOT evaluate…

speed or direction

  • nondirectional color doppler


87
New cards

do all vessels have the same color in Power Doppler?

yes

88
New cards

in Power Doppler, strength (amplitude) is processed…

without deciphering direction or speed

89
New cards

Amplitude is ____ related to

DIRECTLY related to the number of moving blood cells

  • larger amplitude signals = brighter

  • weaker amplitude signals = darker


90
New cards

Advantages of Power Doppler

  • increased sensitivity to low flow/velocity

    • venous flow

    • small vessels

  • Unaffected by Doppler angles

    • except exactly 90 degrees

  • No aliasing

    • velocity information is ignored


91
New cards

Disadvantages of Power Doppler

  • no measurement of velocity

  • lower frame rates

  • flash artifact or bursts of color can occur from motion of the transducer, patient, or soft tissue


92
New cards

Ghosting (doppler artifact)

low frequencies that appear on Color Doppler

93
New cards

Clutter (doppler artifact)

low frequencies that appear on Spectral Doppler

94
New cards

Crosstalk (doppler artifact)

mirror image of spectral display with baseline being the mirror

  • fix with decreasing wave gain or changing angle


95
New cards

other doppler artifact

aliasing

96
New cards

Spectral Analysis

not all blood cells travel at the same speed or direction

  • breaks the signals into individual signals that make up the entire Doppler signal


97
New cards

2 methods for Spectral Analysis

  1. Fast Fourier Transformation (FFT)

  2. Autocorrelation (color)