1/267
Looks like no tags are added yet.
venous line, suckers, vents
venous return components
reservoir below source, fluid filled lines
passive venous drainage requirements
CVP, height difference, resistance
passive venous drainage variables
air lock, venodilation with drugs, hypovolemic condition, tubing kinking during rewarming, manipulation of heart
causes of low venous return with siphon drainage
nitroglycerin, inhalation anesthetics
drugs that cause venodilation
kinetic-assisted venous drainage
centrifugal or roller pump between venous cannula and venous reservoir
hemolysis, venous line chatter, micro-air aspiration from venous cannula insertion sites
potential problems with active venous drainage
reduced venous cannula size, no reservoir restrictions
KAVD advantages
cost, complicates circuit, additional drive console, coordination of two pumps, chance of depriming
KAVD disadvantages
reduced circuit size, reduced cannula size, low cost, decreased prime volume
VAVD advantages
reservoir restrictions (hardshell only)
VAVD disadvantages
vents and suckers must be occlusive, arterial roller head must have proper occlusion, centrifugal heads are non-occlusive
technical considerations of VAVD
utilize one-way valve, maintain adequate RPMs, utilize flow probe
additional considerations for centrifugal head in VAVD
(mL of volume x 60)/mL per min of flow
reaction time =
safety (difficult to pump air, controlled cardiotomy return), multiple manufacturer use, isovolumetric control of venous drainage
softshell pros
circuit complication (cardiotomy reservoir), cost, monitoring difficulties, purging, limited access ports
softshell cons
VAVD option, increased inlet ports, simplification of circuit, reservoir/oxy combo, variable priming volumes, volume measurement markings, decreased venous line resistance, no separate cardiotomy reservoir
hardshell pros
positive pressure, vortexing, cascading/cavitation, capacity, breakthrough
hardshell cons
1/4 in
cardiotomy reservoir inlet ports usual size
1/4 or 3/8 in
cardiotomy reservoir rapid prime port usual size
3/8 in
cardiotomy reservoir outlet port size
-200 mmHg
most reservoirs can accommodate max pressure of
interception process (polyester)
cardiotomy reservoir screen filtration
impaction process (polyurethane or packed polyester fibers)
cardiotomy reservoir depth filtration
dimethylpolysiloxane
cardiotomy reservoir antifoam
20-40 microns
cardiotomy reservoir filtered size
70-200 microns
cardiotomy reservoir unfiltered pore size
2-4 microns
platelet diameter
megakaryocytes in bone marrow
platelet production cells and site
platelets
most common hemostatic defect following CPB
collagen, glycoprotein molecules in subendothelial tissue, ADP, thrombin
triggers of activation of platelets
von willebrand factor
initial attachment of platelets is mediated by platelet glycoprotein with the
alpha-granules
contain fibrinogen, thrombospondin, factor V, von willebrande factor, beta-thromboglobulin, platelet factor 4
dense bodies
contain calcium, serotonin, adenine, nucleosides
thromboxane A2
what part of platelet activation does aspirin override
GP1b/IX
mediates platelet adhesion to subendothelial collagen via vWF
GP1a
mediates platelet adhesion via binding to collagen
GPIIb/IIIa
binding site for adhesive molecules (fibrinogen, vWF, etc)
dense tubular system
serves as pool for intraplatelet calcium, also compartment for thromboxane synthesis
open canulicular system
surround organelle zone, invaginations of the platelet’s plasma membrane, offers additional membrane capacity during activation
alpha granules
approximately 40 per platelet, contain PDGF, TGFbeta, fibroblast growth factor, fibrinogen, thrombospondin, factor V, vWF, beta-thromboglobulins, platelet factor 4, PDEGF, PDAF, PAF, IGF
PDGF
sentinel growth factor that initiates nearly all wound healing, stimulates cell replication of healing capable stem cells
dense granules
contain calcium, serotonin, histamine, epinephrine, ADP (stimulates aggregation)
platelet poor plasma
which has high native fibrinogen, platelet rich or poor plasma?
PDGF
increases the rate of epithelial and granulation tissue production, enhances osteogenesis and angiogenesis, promotes epithelial and collagen synthesis resulting in accelerated soft tissue wound healing and bone osteogenesis
5 times
increase in level of native platelets in platelet gel
inflammatory, proliferative, remodeling
3 phases of wound healing
hypofibrinogenemia, anemia, hypersensitivity to bovine thrombin, hemodynamic instability, platelet dysfunction syndrome, critical thrombocytopenia, pregnancy
contraindications to platelet gels
DeBakey and Liotta 1966
first LVAD implanted
1990
first LVAD implanted patient discharged from hospital with FDA approval
2017
FDA approval of Thoratec HeartMate 3
Abbott HeartMate 3, Medtronic Heartware, Syncardia Total Artificial Heart
long-term VADs
Johnson and Johnson Impella, Abbott Thoratec Centrimag, TandemHeart pVAD, IABP
short-term VADs
centrifugal, pulsatile, 9000 RPMs, 10 L/min, 200 g, 17 hours
Abbott HeartMate 3 characteristics
centrifugal, smallest, 3200 RPMs, 8.7 L/min, 160 g, 4-6 hours
Medtronic HeartWare (HVAD) characteristics
pneumatically driven, pulsatile, 70 or 50 cc
Syncardia Total Artificial Heart characteristics
aortic stenosis
contraindication for impella
centrifugal, 9.9 L/min, 6 hours
Thoratec Centrimag characteristics
LA to femoral artery
Tandem Heart pVAD blood movement
RA to PA
Tandem Life ProTek Duo blood movement
<5%
acceptable rate of mortality
protamine reaction, hypoperfusion, oxygenator failure, blood clotting, line separation, gross contamination, transfusion errors, drug errors, gas embolism, electrical failure
most common perfusion problems
biomed engineering department
who keeps perfusion equipment record
patient demographics, equipment used (lot numbers), drugs/fluids used, perfusion parameters, key events, pre-bypass checklist
perfusion case record included information
stop arterial pump and cardioplegia, isolate the patient and surgical field, try to evacuate air back to reservoir, try to evacuate air out of lure-lock port on arterial cannula, re-prime/circulate, find and address the source, resume flow
response steps for arterial line air
evacuate residual air in arterial line, cool patient and place into trendelenburg position, clamp distal aorta, establish retrograde cerebral perfusion with aortic root venting
response steps for air passed to patient via arterial line
allogeneic blood transfusion
the transfusion of blood from one member of a species to another member of the same species (homologous)
direct donor transfusion
an allogeneic blood transfusion where blood is collected from a chosen donor to be transfused into a specific recipient
autologous blood transfusion
the collection and reinfusion of the donor’s own blood
pre-deposit autologous donation
patient donates their own blood on a non-emergent basis prior to an operation where there is high potential for necessary transfusion
severe AS, unstable angina, CHF, recent MI
heart conditions that may force exclusion from pre-deposit autologous donation
33%
hematocrit must be this or greater for pre-deposit autologous donation
acetaminophen, aspirin, alcohol
should be avoided 48 hours prior to pre-deposit autologous donation
acute normovolemic hemodilution
controlled removal of whole blood from patient after under anesthesia but before start of the operation
8 hours
how long is ANH blood good
anemia, thrombocytopenia, coagulopathy, preload dependency, renal dysfunction, operation time greater than 8 hours
ANH contraindications
intraoperative cell salvage
the collection and reinfusion of shed blood during surgery
discontinuous autotransfusion system
intermittent filling and washing cycles, consists of specific bowl sizes, requires larger volumes of blood to process
continuous autotransfusion system
can process any amount of blood continuously, primarily for pediatric use
ACD/CPD, 30,000 units heparin in 1L NS
intraoperative cell salvage anticoagulation
20-120 microns
intraoperative cell salvage collection reservoir pore size
40 micron pall filter, leukocyte-depleting filter
filters used for cell saver product prior to reinfusion
60%
hematocrit of cell saver blood
12 units/hr
processing capability of cell saver
two
how many units of cell saver blood make it cost effective
750-1500 mL
blood loss volume that is indication for intraoperative cell salvage
procoagulant materials (collagen) in operative field, neoplastic cells in operative field, infection in operative field, aspiration of other body fluids
intraoperative cell salvage contraindications
hemolysis, sepsis, air embolism, clot embolism, fat embolism, excessive bleeding from loss of platelets
intraoperative cell salvage complications
postoperative cell salvage
blood is collected from wound drains then reinfused either unwashed and filtered or washed and filtered
low hct, high cytokines, tissue factor and activated clotting factors, debris
characteristics of sanguineous fluids
sanguineous fluid
first drainage a wound produces
4-6 hours
time from initiation of collection when reinfusion of shed blood stored at room temperature must be completed
500 mL
at what volume of blood should reinfusion be started with postoperative cell salvage
80-90%
how much does diastolic arrest decrease oxygen demand
97%
reduction in myocardial oxygen demand achieved by diastolic arrest and hypothermia
superior buffering capacity, antioxidant enzymes act as free radical scavengers, o2 delivery, protein increases oncotic pressure
blood cardioplegia advantages
del nido, custodial, HTK
one-shot dose cardioplegias
custodial, HTK
heart transplant cardioplegias
50-100 mmHg
normal antegrade cardioplegia aortic root pressure
150-200 mmHg
normal antegrade cardioplegia line pressure