1/31
Looks like no tags are added yet.
What is resolution?
How accurately an ultrasound system can show and separate structures. In addition to axial and lateral, slice thickness or elevational resolution contributes to image quality.
What are the three types of resolution in ultrasound imaging?
Axial (shallow-to-deep).
Lateral (side-to-side).
Elevational (above-to-below the imaging plane).
What is another name for elevational resolution?
Slice thickness resolution.
What does elevational resolution help determine?
Answers the question: are the reflections on the image created from structures directly inside the imaging plane or from structures that lie above or beneath the imaging plane?
In what direction is slice thickness measured?
Perpendicular to the imaging plane (above-to-below).
Is the ultrasound imaging plane truly razor-thin?
No, the beam has a measurable thickness that can vary with depth.
What happens when the ultrasound beam has a thick slice?
Structures above and beneath the “assumed razor thin plane” create reflections that appear in the image.

What shape of active element creates the best elevational resolution? Which transducer types have the best elevational resolution?
Active elements that are disc shaped provide the thinnest ultrasound slices and the best elevational resolution within the focal zone. Disc-shaped crystals are found in mechanical and annular phased array transducers.

What is the cross-sectional shape of a sound beam produced by a disc-shaped crystal?
the cross section of a sound beam is a circle.
What does the width of the circular beam represent?
indicates the beam diameter (which determines lateral resolution)
What does the height of the circular beam represent?
Beam thickness, which determines elevational resolution.

With disc-shaped active elements, how are lateral and elevational resolution related?
They are identical because the beam has a circular cross-section.
What determines elevational resolution in phased, linear, and convex array transducers?
The thickness of the sound beam, which is controlled by an acoustic lens.

What shape are the active elements in these three types of arrays?
These array probes have rectangular-shaped active elements placed side by side.
When these 3 arrays use phasing, in which direction is the sound beam narrowed?
With phasing, the beam is focused or narrowed only side to side in the imaging plane, but not above and below. This narrowing improves lateral resolution but does not affect slice thickness.
What is used to make the ultrasound slice thinner in these 3 arrays?
an acoustic lens is used to create a thinner slice. Recall that lenses have a fixed focal depth.
At the transducer, what determines the initial thickness of the sound beam?
At the transducer, the beam thickness is equal to the height of the active element. At the focal depth, the beam thins to its minimum value. At greater depths, the beam thickens substantially. At all depths, the beam is much thicker than it is wide
What is the order of resolution from best to worst in phased, linear, and convex arrays?
Axial (best)
Lateral
Elevational (worst).
What is one limitation of traditional linear arrays?
the inability to dynamically create thin slices.

What newer types of transducers have improved slice thickness resolution?
Newer transducers, called one and one-half dimensional arrays, create thinner beams with improved slice thickness resolution over a greater range of depths. The face of the 1 ½ - D array transducer has numerous crystals arranged side by side.

How are 1½-D arrays different from traditional linear, phased, and convex arrays?
Traditional arrays have one crystal occupying each position, while 1½-D arrays have multiple crystals stacked in the up-and-down direction.
Three-Dimensional Imaging
Three dimensional (3-D), or volume imaging, is a recent innovation in ultrasound. A 2-D array, with thousands of elements arranged in a checkerboard pattern, CREATES 3-D images.
How does a 2-D array improve elevational resolution
With 2-D array transducers, the sound beams are electronically focused in both the lateral and thickness planes. The beams are electronically steered, creating entire scan planes that sweep through anatomic structures.
How does a 2-D array create a 3-D image?
It electronically steers sound beams to create multiple scan planes that sweep through the anatomy.
4-D imaging
a trendy way of describing real-time 3-D imaging, 4-D imaging also uses 2-D array probes.

What are side lobes?
Created by single-element transducers, extra areas of sound energy that extend outside the main beam axis in the far zone

Why are strong side lobes a problem?
Structures inside the side lobes can create unwanted reflections that appear in the image as artifacts. Reflections arising from within the side lobes degrade lateral resolution.

What are grating lobes?
Extra sound beams that extend away from the main beam, created by array transducers.
Why are grating lobes undesirable?
They degrade lateral resolution, reduce image quality, and create artifacts.

Modern transducers are designed to reduce the strength of side and grating lobes, how is this accomplished?
Apodization? A process that changes the electrical spike voltages to reduce the strength of side and grating lobes.
What is subdicing?
Dividing one PZT crystal into several smaller crystals called sub elements These sub-elements are electrically joined and act as if they are a single crystal which reduces grating lobes.