30 Terms

1
New cards

How often should you check Exposure Timer Accuracy?

Annually

2
New cards

What is the Acceptable Range of Exposure Timer Accuracy?

  • or - 5%


3
New cards

How often should you check mA Linearity?

Annually

4
New cards

What is the Acceptable Range of mA Linearity?

  • or - 10% of variation
5
New cards

How often should you check Exposure Reproducability?

Annually

6
New cards

What is the Acceptable Range of Exposure Reproducability?

  • or - 5%
7
New cards

How often should you check Kilovoltage Peak Calibration?

Annually

8
New cards

What is the Acceptable Range of Kilovoltage Peak Calibration?

  • or - 5%
9
New cards

How often should you check Half-Value Layer / Filtration?

Annually

10
New cards

What is the Acceptable Range of Half-Value Layer / Filtration?

Exceed required minimum (amount of thickness of absorber material "usually aluminum" needed to reduce the intensity of the x-ray beam to one-half of the original)

11
New cards

How often should you check Collimation?

Semi-Annually

12
New cards

What is the Acceptable Range of Collimation?

  • or - 2% of the SID
13
New cards

How often should you check Vertical Beam Alignment?

Semi-Annually

14
New cards

What is the Acceptable Range of Vertical Beam Alignment?

  • or - 2 degrees from vertical
15
New cards

How often should you check Focal Spot Size / Condition?

Annually

16
New cards

What is the Acceptable Range of Focal Spot Size / Condition?

  • or - 50% of nominal size
17
New cards

How often should you check the Image Receptor System?

Semi-Annually

18
New cards

What is the Acceptable Range of a Image Receptor System?

Zero visible defects

19
New cards

How often should you check the Automatic Exposure Control?

Annually

20
New cards

What is the Acceptable Range of the Automatic Exposure Control?

  • or - 10%
21
New cards

With HVL, the higher the HVL of an x-ray beam, the _____ penetrating the beam

more

22
New cards

Acceptance testing of x-ray equipment is performed by a ______ prior to daily use.

medical physicist

23
New cards

In medical imaging departments with a quality control program in place, the great majority of repeated exposures are due to improper:

Positioning

24
New cards

The device (provided by many manufacturers of display monitors) designed to directly measure luminance, or the light intensity emitted from a display monitor, is the:

Photometer

25
New cards

Determining how consistently radiologists' image interpretations match ultimate patient diagnosis is called ____.

outcome assessment/analysis

26
New cards

T/F: Quality control refers to the calibration and monitoring of equipment.

True

27
New cards

The minimum source-to-table distance for stationary fluoroscopy units is:

15 inches

28
New cards

The minimum source-to-table distance for mobile fluoroscopy units is:

12 inches

29
New cards

The ability to be given a negative diagnosis when no disease is present is termed:

specificity

30
New cards

The ability to detect disease is termed:

sensitivity