How often should you check Exposure Timer Accuracy?
Annually
What is the Acceptable Range of Exposure Timer Accuracy?
or - 5%
How often should you check mA Linearity?
Annually
What is the Acceptable Range of mA Linearity?
How often should you check Exposure Reproducability?
Annually
What is the Acceptable Range of Exposure Reproducability?
How often should you check Kilovoltage Peak Calibration?
Annually
What is the Acceptable Range of Kilovoltage Peak Calibration?
How often should you check Half-Value Layer / Filtration?
Annually
What is the Acceptable Range of Half-Value Layer / Filtration?
Exceed required minimum (amount of thickness of absorber material "usually aluminum" needed to reduce the intensity of the x-ray beam to one-half of the original)
How often should you check Collimation?
Semi-Annually
What is the Acceptable Range of Collimation?
How often should you check Vertical Beam Alignment?
Semi-Annually
What is the Acceptable Range of Vertical Beam Alignment?
How often should you check Focal Spot Size / Condition?
Annually
What is the Acceptable Range of Focal Spot Size / Condition?
How often should you check the Image Receptor System?
Semi-Annually
What is the Acceptable Range of a Image Receptor System?
Zero visible defects
How often should you check the Automatic Exposure Control?
Annually
What is the Acceptable Range of the Automatic Exposure Control?
With HVL, the higher the HVL of an x-ray beam, the _____ penetrating the beam
more
Acceptance testing of x-ray equipment is performed by a ______ prior to daily use.
medical physicist
In medical imaging departments with a quality control program in place, the great majority of repeated exposures are due to improper:
Positioning
The device (provided by many manufacturers of display monitors) designed to directly measure luminance, or the light intensity emitted from a display monitor, is the:
Photometer
Determining how consistently radiologists' image interpretations match ultimate patient diagnosis is called ____.
outcome assessment/analysis
T/F: Quality control refers to the calibration and monitoring of equipment.
True
The minimum source-to-table distance for stationary fluoroscopy units is:
15 inches
The minimum source-to-table distance for mobile fluoroscopy units is:
12 inches
The ability to be given a negative diagnosis when no disease is present is termed:
specificity
The ability to detect disease is termed:
sensitivity