primary factors
mA, kVp, exposure time, distance
mAs
measurement of tube current, quanitity
exposure time
controls duration of exposure, quantity
kVp
sped of electrons move in tube, quantity and quality
how can energy of electron affect quality
higher energy means higher penetrability
distance, inverse square law
distance of tube to image receptor, quanitity
secondary factors
contrast, receptor exposure, spatial resolution, distortion
contrast definition and its purpose
degree of differentiation between light and dark areas in radiograph, make anatomy more visible
scales of contrast
long/low and short/high
air gap technique
increased OID to decrease scatter
grids purpose
absorbed scatter and unwanted radiation
beam restriction (collimation)
reducing area leads to less scatter and better contrast
patient factors that affect contrast
pathology, patient habits, tissue type and density
destructive pathology affect on contrast
short/high
additive pathology affect on contrast
low/long, more technique
anode heel effect
intensity is greater on cathode side (120) compared to anode (75)
generator output on receptor exposure
single phase, triple phase, high frequency (less technique, better radiation efficiency)
spatial resolution
sharpness and quality of anatomical structural lines from attenuation
receptor exposure factors (8)
mAs, kVp, SID, grids, tube filtration, beam restriction, anode heel effect, patient factors
focal spot blur/ penumbra
area of UNsharpness
umbra
area of sharpness
contrast factors (6)
kVp, OID, grids, tube filtration, beam restriction, patient factors
affects of unsharpness and spatial resolution on a LARGE focal spot size
unsharpness increases, spatial resolution decreases
affects of unsharpness and spatial resolution on a SMALL focal spot size
unsharpness decreases, spatial resolution increases
spatial resolution factors (6)
OID, SID, focal spot size, motion, patient factors, angle
how can SID affect focal spot blur
large SID, decrease blur or small SID, increases blur
OID affect on focal spot blur
increase OID, increase blur or decrease OID, decrease blur
distortion factors (3)
magnification, foreshortening, elongation
mAs distance compensation formula
mAs1/mAs2= (d1)2/(d2)2
magnification factor
MF= SID/SOD
15 % rule
increasing kvp is the same thing as doubling mas or decreasing kvp is the same thing as dividing your mas
object size
object size = image size/MF
focal spot blur formula
focal spot size/SOD = focal spot blur/OID
double focus tube
tube with two filaments and focal spots
fractional focus tube
tube with smaller focal spot and anode angle less that 5 degrees
focal spot/track
region of the anode target where electrons interacts to produce x rays
properties of x ray (13)
no mass, invisible, electrically neutral, travel at speed of light, not optically focused, polyenergetic, range of energies, travel in straight lines, fluoresces substances, penetrate body, absorption or scatter in body, produce secondary radiation, cause damage to tissue
photon
bundle of energy
wavelength
distance between two corresponding points on a sine wave, measured in angstroms
frequency
number of wavelength that pass through a point, measured in Hertz
energy
measured in units of eV (10 kEV to 100 kEV)
short to long waveforms
high energy/frequency (short), low energy/frequency (long)
conditions necessary for x ray production (4)
separation, focusing, acceleration, sudden stop or break
components of x ray tube (5)
glass/metal enclosure, cathode, anode, induction motor, protective housing
cathode side and contains what?
negative side of tube, contains filament and focusing cup
anode
positive side of tube, stationary and rotating
purpose of anode (4)
receive and conduct electrons, mechanical support, dissipate heat
induction motor
powers rotating tube, stator on outside and rotor on inside of enclosure, works through electromagnetic induction
protective housing
guards against excessive radiation exposure and electric shock, contains oil
leakage radiation
any radiation that exits the protective housing (less than 1 mgy /hr at 1m)
glass/metal enclosure
withstands heat temperatures (glass), longer lifespan (metal), vacuum allows efficient x ray production and life
atomic number and melting point of Tungsten
#74, 3400 degrees C
motor that runs the tube
energy converted into head and x ray (%)
99% heat, 1% x ray
different focal spot size
thin areas use small filament (<200 mA), thick areas use large filament (>200 mA)
anode heel effect
intensity on cathode side is greater than anode, 120 to 75
causes of tube failure (9)
pitting, crazing, cracking, bearing damage, prolonged rotoring, filament breakage, arcing, oil overheating, physical abuse
different method to prevent tube failure (5)
tube/anode rating chart, warm x ray tube, high kvp, low mas, dont rotor long
different factors to consider when using a radiographic rating chart (6)
filament size, anode speed, target angle, number of exposures, tube design, cold vs hot tube, power supply, rectification
electromagnetic spectrum (EMS)
composed of different electromagnetic energies and radiations grouped by wavelength and frequency
how does the cathode work
low voltage circuit causes thermionic emission bunching up electrons to space charge, ejects to anode target
how does anode work
anode rotates during exposure, electron stream spread over a larger area
heat units
amount of heat produced from exposure, HU= mA x time x kVp x generator factor
generator factors from heat units
single phase (1), three phase (1.35), high frequency (1.40)
radiographic rating chart
shows unsafe for x ray tube, any combination of time mAs or kvp under mA curve is unsafe