Chromatography
A separation method in which analytes move through or over a stationary phase at different rates.
Mobile phase
The gas or liquid that carries analytes through the column.
Stationary phase
The phase that interacts with and retains analytes inside the column.
Column
The tube where chromatographic separation takes place.
Analyte band or plug
The narrow region of sample occupying the column as it travels.
Elution
Movement of an analyte through and out of the column.
Retention
How strongly or how long an analyte is held by the stationary phase.
Resolution
How well two analytes are separated into distinct peaks.
Band broadening
Expansion of an analyte zone as it travels through the column.
Adsorption
Attraction of an analyte to a solid surface.
Absorption or partition
Movement of an analyte into a second phase, such as a liquid stationary phase.
Distribution coefficient, Kc
The concentration ratio Kc=[A]mp[A]sp between stationary and mobile phases.
Partition coefficient, Kx
The molar ratio of analyte in the stationary phase to analyte in the mobile phase; it accounts for phase volumes.
Selectivity factor, α
The ratio of partition coefficients for two analytes, α=KAKB; α=1 means no separation.
Dead time, t0
Time required for an unretained compound or mobile phase to travel from injector to detector.
Retention time, tR
Time from injection until the analyte peak reaches maximum response.
Retention factor, k
A normalized measure of retention: k=t0tR−t0.
Retention volume, VR
The mobile-phase volume needed to elute an analyte: VR=tR×mobile-phase velocity.
Chromatogram
A plot of detector response versus elution time or mobile-phase volume.
Theoretical plates, N
A measure of column efficiency; larger N generally means narrower peaks.
Plate height, H
Column length divided by theoretical plates: H=NL. Smaller H means better efficiency.
Peak baseline width, w
Width of a chromatographic peak measured at the baseline using tangent lines.
Resolution equation
For two peaks, Rs=w1+w22(tR2−tR1). Higher Rs means better separation.
Gaussian peak
An approximately bell-shaped peak produced when band spreading is statistically distributed.
Main causes of band broadening
Multiple paths, longitudinal diffusion, mass-transfer limits, and injection or detector effects.
Gas chromatography, GC
A technique that uses a gas mobile phase to separate volatile analytes in a column.
GC carrier gas
High-purity He, H2, or N2 that pushes analytes through the column.
Pressure regulator
Controls gas pressure and can provide approximately constant pressure flow.
Flow controller
Controls carrier-gas flow, commonly in mL/min.
GC injection volume
A small liquid sample volume, commonly about 0.2 to 20μL.
GC injection port
A heated inlet that vaporizes the sample before it enters the column.
On-column injection
Injection directly into the column without first vaporizing in a separate inlet.
Split injection
Only a small fraction of the vaporized sample enters the column; useful for concentrated samples.
Splitless injection
Most of the sample enters the column; useful for low-concentration samples.
GC stationary phase
A thin, nonvolatile liquid coating that analytes diffuse into and out of.
Main basis of GC separation
Analyte volatility or boiling point; stationary-phase interactions become more important when boiling points are similar.
Packed GC column
A column filled with particles; higher sample capacity but lower efficiency than capillary columns.
Capillary GC column
A narrow, long column with very high efficiency but lower sample capacity.
WCOT column
Wall-coated open tubular column with stationary phase directly on the capillary wall.
SCOT column
Support-coated open tubular column with stationary-phase-coated particles on the wall.
Temperature programming
Changing column temperature over time to improve separation and control analysis time.
Isothermal GC
Operating the column at one constant temperature.
Why start GC at a cooler temperature?
A cooler starting temperature helps concentrate or stack the sample at the head of the column.
Flame ionization detector, FID
A detector that burns analytes in an H2-air flame and measures ions produced.
FID response
Approximately proportional to the amount of reduced carbon; it is mass-sensitive.
FID limitations
It responds poorly to noncombustible gases such as H2O, CO2, SO2, CO, noble gases, and NOx.
GC-MS
A GC coupled to mass spectrometry for separation plus structural or mass-to-charge information.
GC elution order
Generally, lower-boiling analytes elute earlier, although stationary-phase interactions also matter.
Liquid chromatography, LC
A separation technique using a liquid mobile phase pumped through a stationary-phase column.
HPLC
High-performance liquid chromatography; high-pressure pumping gives controlled flow through small particles.
UHPLC
An advanced HPLC system using higher pressure and smaller particles for improved efficiency.
HPLC mobile phase
The liquid solvent or solvent mixture that carries analytes through the column.
Why mobile-phase composition matters
It strongly controls analyte affinity, retention, selectivity, and elution order.
HPLC degassing
Removal of dissolved gases to prevent bubbles and flow or detector problems.
Six-port injection valve
Loads a sample loop while mobile phase flows, then switches flow through the loop to inject the sample.
Guard column
A short protective column placed before the analytical column to trap contaminants.
Analytical column
The column used to perform the separation; it is not generally designed for sample recovery.
Size-exclusion chromatography
Separates molecules by how well they enter pores in the stationary-phase particles.
Size-exclusion elution order
Large molecules elute first because they enter fewer pores; small molecules take longer paths.
Ion chromatography
Separates ions by reversible exchange with charged groups on a stationary phase.
Cation exchange
A negatively charged stationary phase retains positively charged analytes.
Anion exchange
A positively charged stationary phase retains negatively charged analytes.
Normal-phase chromatography
Uses a polar stationary phase and nonpolar mobile phase; nonpolar analytes elute first.
Reversed-phase chromatography
Uses a nonpolar stationary phase, commonly C18 or C8, and a polar mobile phase.
Reversed-phase elution order
More polar analytes generally elute first; nonpolar analytes are retained longer.
Strong solvent in reversed phase
A less-polar organic solvent, such as methanol or acetonitrile, elutes nonpolar analytes faster.
Common reversed-phase mobile phases
Water, methanol, and acetonitrile.
Mobile-phase polarity estimate
PAB′=ϕAPA′+ϕBPB′, where ϕ is solvent fraction and P′ is solvent polarity.
HPLC detectors
Common detectors include UV/Vis, mass spectrometry, fluorescence, conductivity, and refractive index.
Gradient elution
Changing mobile-phase composition during a run to alter analyte retention and improve the separation.
Mass spectrometry, MS
An analytical method that creates gas-phase ions and measures their mass-to-charge ratios, m/z.
Mass spectrum
A plot of ion abundance or intensity versus mass-to-charge ratio, m/z.
Molecular ion
A radical ion containing the intact analyte molecule; often labeled M+∙ or M+.
Daughter ion
A fragment ion produced from the original molecular ion.
Base peak
The most intense peak in a mass spectrum; it is assigned 100% intensity and may not be M+.
Mass-spectrum fingerprint
The reproducible pattern of ions characteristic of an analyte under a given ionization method.
m/z
The mass-to-charge ratio of an ion; it is the quantity used to separate and report ions in MS.
Isotope peak
An additional peak caused by a naturally occurring heavier isotope, such as 13C or 37Cl.
13C isotope abundance
About 1% of carbon is 13C, so isotope peaks increase with the number of carbon atoms.
37Cl isotope abundance
About 33% of chlorine is 37Cl, producing a strong M+2 pattern when chlorine is present.
MS inlet system
Introduces the sample into the mass spectrometer, often from GC or HPLC.
Ion source
The component that converts analyte molecules into gas-phase ions.
Mass analyzer
Separates or selects ions according to their m/z values.
Ion detector
Senses ions and converts their arrival into an electronic signal.
Ion optics
Electrical or magnetic elements that accelerate, focus, direct, or stop ions.
MS vacuum system
Maintains very low pressure so ions have fewer collisions and the signal has less noise.
Why MS needs a vacuum
Collisions with other gas molecules can deflect ions, reduce transmission, and increase background noise.
Total ion chromatogram, TIC
A chromatogram showing the summed signal from all detected ion fragments versus time.
Electron ionization, EI
A relatively energetic ionization method that often produces many reproducible fragments.
Soft ionization
A gentler ionization approach that tends to preserve the molecular ion, especially for large biomolecules.
Qualitative use of MS
Identifies compounds from accurate mass, molecular ions, isotope patterns, and fragmentation fingerprints.
Quantitative use of MS
Determines the amount of an analyte from ion signal calibrated against standards.
MS isotope-ratio use
Measures isotope abundances for applications such as carbon dating or anti-doping analysis.
Accurate mass
A precise measured ion mass that helps determine elemental composition.
Nuclear magnetic resonance, NMR
A technique that uses nuclear spin and radiofrequency absorption in a magnetic field.
NMR-active nucleus
A nucleus with nonzero spin quantum number, such as 1H or 13C.
Nuclear spin quantum number, I
A property of a nucleus that determines its possible spin states: number of states=2I+1.
1H spin
The proton has I=21 and therefore two allowed spin states.
13C spin
Carbon-13 has I=21 and is NMR active; most carbon is carbon-12, which is NMR inactive.
NMR magnet, B0
The strong external magnetic field that causes nuclear spin states to have different energies.