Outer cells
When utilizing AEC, chest x-rays should utilize these cells to prevent over exposure of lung fields
150%
An AEC backup timer is set at this in regard to its expected mA
Time
AEC only controls this
600
Automatic termination of an exposure is required by law to occur at this mAs when more than 50 kVp is utilized
Overexposure
When an activated AEC chamber is collimated off this results
3
How many chambers do AEC chambers typically have?
Aluminum Filters
This is used to measure half value layers
Plus or minus 2%
An SID error of this is allowed with testing collimator accuracy
Repeat image analysis
This quality control measure should be performed on a continual basis
Quality control and quality assurance
These are the two types of quality management
Focal spot size estimation
The pinhole camera, star test pattern, and line pair resolution test tool are all tools used to perform quality control for this
Plus or minus 5%
This is the deviation allowed for exposure timer settings
Additive
Pneumonia is considered this type of pathology
Density setting
This setting regulates the density of an image and should not be used to compensate for part thickness
Anatomically programmed radiography (APR)
This is a feature found on modern digital x-ray systems that stores pre-set exposure techniques for specific examinations and patient sizes
Approximately 10 kVp
In a clinical setting, a 15% change in kVp is equal to this
Proportional anatomy groupings
This is the method of determining exposure techniques for one body part based on a known technique for another body part with similar thickness and tissue composition
mAs
When using a variable kVp technique chart, this remains nearly fixed
Focal plane
All recorded detail in objects outside of this will appear blurred
Time required to travel tomographic amplitude
Exposure time has to match this
Amplitude
The total distance traveled by the tube is known as this
Before exposure
To avoid motion artifacts in tomography, this is when tube motion should occur
Sharp
The anatomy located at the fulcrum of a tomographic image appears like this
10
In digital tomosynthesis, the image database is created from approximately this many exposures at different angles
Exposure angle, Inversely proportional
Section thickness is controlled by this and has this relationship to it
Destructive
Degenerative arthritis, emphysema, and pneumothorax are examples of this type of pathology
Increase
When going from no grid, to utilizing a grid, this is how exposure factors should be adjusted
Phantom image
Tube motion parallel to the long axis of a structure results in this
Directly above appropriate chamber
Anatomy should be positioned like this in regards to the ionization chamber(s) when utilizing AEC
2.5
For a half value layer test, the following exposures are measured by an ion chamber with increasing amounts of aluminum absorber placed in the beam as indicated below.
This is the HVL for this x-ray beam
0 mm Al: 200 mR
1 mm Al: 160 mR
2 mm Al: 120 mR
3 mm Al: 80 mR
Fulcrum
What determines focal plane location?
Exposure angle
What controls section thickness?
Thinner section
Larger tomographic angle results in what?
Parallel tube motion and blur overlap
What causes phantom images?
Exposure time
Which exposure factor does AEC directly control?
Outer cells
Which AEC cells are commonly used for chest radiography?
Additive pathology
What pathology requires increased exposure factors?
Destructive pathology
What pathology requires decreased exposure factors?
Ionization chamber
What is the most common modern AEC detector
Linear motion
What tomographic movement remains most common today?
QA and QC
What are the two components of quality management?
QA
Which focuses on people and processes?
QC
Which focuses on equipment?
Filtration required to reduce beam intensity by half
What does HVL measure?
DQE
What is the primary QA parameter for digital detectors?
+/- 5%
What is the acceptable kVp accuracy?
+/- 5%
What is the acceptable exposure reproducibility?
MQSA
Which law mandates FDA approval of mammography services?
The Joint Commission (TJC)
What organization is the largest accreditor?
1. Establish criteria
2. Monitor performance
3. Collect data
4. Analyze data
5. Evaluate data
What are the five steps of QA monitoring?