1/44
Looks like no tags are added yet.
Turbo factor is another name for
Echo train length (ETL)
What is the effect of increasing TR on image weighting?
It may change image weighting depending on other parameters
What defines a single-shot FSE sequence?
All k-space lines are acquired in one TR period
What type of pulse sequence uses a 180° rephasing pulse to generate an echo?
Spin-echo
What is the echo in MRI?
Signal detected at TE made of rephased magnetic moments
What is pathology weighting in MRI?
IR image with long TE showing bright pathology
The pulse sequence that uses multiple 180° RF rephasing pulses to fill multiple lines of k-space is
Fast spin echo (FSE)
What does phase reordering do in MRI?
Reorders gradient amplitudes and polarities for phase encoding
What effect does a higher echo train length (ETL) have on FSE images?
Increases T2 weighting
The long TR and long TE results in which type of image weighting?
T2
Which sequence is commonly used to suppress fat signal?
STIR
Compared to conventional spin echo sequences, fast spin echo (FSE) sequences
Always have short scan times
What regenerates the echo in a gradient-echo pulse sequence
Gradient reversal
What is the effective TE in a turbo spin echo?
The TE that best reflects image weighting
The time from the 180° RF inverting pulse to the 90° RF excitation pulse is called the
TI
Which sequence is typically used for brain imaging to see lesions adjacent to CSF?
FLAIR
in STIR sequences, "tau" refers to
Time between 90° and 180° pulses and from there to the echo
What is the primary purpose of FLAIR sequences in MRI?
Suppress cerebrospinal fluid (CSF) signal
The echo train length (ETL) is associated with which sequence?
Fast spin echo
What type of RF pulse is used to generate a spin echo?
180
What type of RF pulse does an inversion recovery sequence begin with?
180
The Larmor frequency is affected by the application of RF pulses.
False
A spin-echo is produced by
A 180° RF pulse following excitation
The point in time where there is no longitudinal magnetization of fat in STIR is called the __________
Null point
Inversion recovery sequences increase signal from
Contrast-enhanced structures
In Fast Inversion Recovery sequences, which tissue signal is often suppressed?
Fat
The scan time for a FSE sequence can be calculated as TR x NSA x (_________):
Number of phase encodings/ ETL
The time it takes for spins to dephase immediately after the 90° RF pulse is called:
Tau
What does TI represent in inversion recovery sequences?
Time from 180° inversion to 90° excitation
T2* decay is eliminated by the 180° RF pulse.
True
The effective TE in TSE sequences is used to determine _____________.
Image weighting
Which parameter affects the scan time of a sequence?
NSA (Number of Slices Acquired) , Number of phase encoding steps, TR (ALL OF THE ABOVE)
What is the purpose of a 180° RF rephasing pulse?
To rephase magnetic moments in the transverse plane
STIR sequences are particularly useful in musculoskeletal imaging because they:
Suppress fat signal effectively
What is the purpose of triple IR prep?
Null both fat and blood
In a Fast Spin Echo sequence, scan time is reduced by:
Increasing the number of phase encoding steps per TR
The method of filling k-space where all lines are filled at once is called:
Single shot turbo spin echo
What is an echo train in MRI?
Series of 180° pulses and echoes
T1-weighted images are typically produced using which combination of TR and TE?
Short TR and short TE
What is the benefit of using dual echo sequences?
Provide both PD and T2 weighted images
The phenomenon where fat remains bright on T2-weighting images due to reduced spin-spin interactions is called:
J coupling
A sequence that uses a 180-degree rephasing pulse to generate an echo is called a:
Spin echo
RARE is another name for:
Turbo spin-echo
Reducing ESP (echo spacing) in FSE sequences helps to minimize:
Image blurring
What is the primary function of a DRIVE sequence?
Achieve high signal from water with short TR