What does electrophoresis separate?
Charged molecules in an electric field
Where do cations migrate towards?
Cathodes
Where do anions migrate towards?
Anode
What is migration affected by?
Net charge
Molecular size/shape
Strength of the electric field
Properties of the support medium
Buffer pH/ionic strength
Temp
Greater charge =
Generally greater migration
Large molecule =
Generally slower migration
Proteins are ampholytes
They can carry positive or negative depending on pH
If pI is greater than pH then
Protein becomes more positively charged and will migrate toward the cathode
If pH is greater than pI then
Protein becomes more negatively charged, migrates toward the anode
If pH = pI, then
Net charge = 0
Electrophoretic migration stops
What does the buffer do?
Carries current and maintains pH, pH determines the charge on the proteins and therefore affects migration
What is the purpose of the support/separation medium?
Provides the environment in which molecules separate.
Electrodes and power supply do what in electrophoresis?
Create the electric field that drives migration
Detection and quantification system
Separated proteins can be stained or detected optically
A densitometer converts separated proteins zones into peaks that can be quantified
Agarose is what?
Predominantly charge-based separation
Polyacrylamide is what?
Based on charge and molecular size
Agarose - what should MLS students know?
Commonly used for serum protein electrophoresis; large pores, so proteins separate primarily according to charge
Polyacrylamide - what should MLS students know?
Smaller/tunable pores; separates based on charge and size b/c of molecular sieving
Cellulose acetate - what should MLS students know?
Older/historical support medium; largely replaced by agarose for many applications
Capillary - what should MLS students know?
Separation occurs inside a narrow capillary; rapid, automated, high-resolution method
Explain the process of detection and densitometry:
After conventional gel electrophoresis, separated proteins form bands or zones
Bands may be stained so they can be visualized
A densitometer measures absorbance across the separated zones
The resulting graph is an electropherogram
Area under each peak is proportional to the relative amount of protein in that fraction
Monoclonal gammopathy pattern
Narrow spike in the gamma region
Polyclonal inflammation pattern
Broad gamma-region increase
Nephrotic syndrome pattern
low albumin w/ increased a2 region
Cirrhosis pattern
Beta-gamma bridging
Hypogammaglobulinemia
Decreased gamma region
Where does capillary electrophoresis (CE) perform electrophoresis separation?
A very narrow capillary
Why does capillary electrophoresis use a high voltage?
The small capillaries dissipate heat efficiently
What are the advantages of capillary separation?
Rapid separation, high resolution, very small sample volume, automation, online detection
Electroosmotic flow
Capillary zone electrophoresis, buffer movement through the capillary helps carry analytes toward the detector
Analytes still separate b/c they have different electrophoresis mobilities
IFE separates proteins according to what?
Their isoelectric points
A pH gradient is established across what?
A pH gradient is established across the MEDIUM
Each protein migrates until what?
pH = protein's pI
When the pH = pI the net charge becomes what? And what happens to migration?
Net charge becomes 0
Migration stops
What will happen to proteins once they stop moving and net charge is zero?
Proteins become focused into sharp bands
What are Southern blots used to detect?
DNA
What are Northern blots used to detect?
RNA
What are Western blots used to detect?
Protein
How are separated molecules transferred in blotting techniques?
Transferred from a gel to a membrane
Which blotting techniques are uncommon? Why are they uncommon?
Southern and Northern blotting, they have been replaced by PCR, sequencing, and other molecular techniques
Western blotting has been replaced by what?
Newer immunoassays and molecular tests
Which blotting technique is historically important?
Western
Too much sample?
Overloaded bands or poor resolution
Too little sample
Faint bands or poor sensitivity
Wrong sample type
Plasma fibrinogen can add an extra band
Heat/current problems
Band broadening or distorted migration
Buffer problem?
pH/ionic strength changes migration
Uneven wetting/electrodes?
Unequal migration across gel
Stain/detection problem?
Weak or unreliable quantification
Point-of-application artifact
Band at the origin can mimic abnormally
What should you ask before interpreting a strange band?
Is it artifact or true abnormal protein?