3 ways to approach the CBC
Commercial laboratory analysis
In-clinic Automated Analyzer
Manual evaluation in-clinic
Most clinics will use a combination of all three
Often a strong preference for one over the others
Commercial Laboratories
Advantages
Excellent QC
Best equipment
High volume
Professional staff
Excellent quality assurance
Disadvantage:
Delayed results
Hematology is species-specific
Avoid human labs
Familiarity with many different species is required
Subjectivity of analysis
Automated Cell Counts in clinic
Technology includes:
the good, the bad, and the ugly
Several types on the market (-$$)
Can be more accurate than manual methods if used properly :)
More cells are evaluated :)
Machines require proper maintenance and calibration to produce consistent and accurate results :)
Requires less time and training :)
coulter counter
Dilute solution of cells in electrolyte solution
Drawn through aperture between 2 electrodes
Passing particles:
Change in electrical impedance (resistance)
Measurable voltage pulse
Proportional to cell size
Pulses are detected, analyzed and counted
the results
RBC count
Includes white cells
Proportion of WBCs is small; usually insignificant
RBCs must be lysed to obtain a WBC count
RBCs & platelets differentiated based on size
QBC Analyzer
Quatitative Buffy Coat Analysis
Differential centrifugation using a specialized float
Blood = plasma, buffy coat, RBCs
Buffy coat is further divided
Platelets: least dense (top of buffy coat)
Agranulocytes: middle (monocytes/lymphocytes)
Granulocytes: most dense (just above RBCs)
Lasercyte
flow cytometry
Recently more cost-accessible but harder to find (older model)
Cells pass through a laser beam
Absorb & scatter light
Each cell type produces a “signature”
Size
Nuclear configuration
Cytoplasmic inclusions
results:
Most generate a full CBC
5-part WBC differential
Platelet count
Reticulocyte count
RBC indices
Will flag(*)potential abnormalities or false readings
ProCyte one
Combination tech
“Newest” & most accurate
Combines 3 technologies:
Flow cytometry (Lasercyte)
Optical fluorescence
Enhances & improves detection and characterization of cells
Gold standard for reticulocyte & platelet assessment
Impedance
Lines up cells for individualization (counts)
Flags possible abnormalities or false readings with a key
whats missing
nRBCs
Platelet clumps
Precursor cells
Cannot distinguish:
RBC morphology (hints in MCHC; MCH)
Rouleaux
RBC inclusions
WBC toxicity
Döhle bodies
Disintegrated cells
Parasites (hints in WBC count)
Total Plasma Protein
Always calibrate refractometer with distilled water to the 1.000 line of the specific gravity scale
Read plasma on the TP/TS/SP (etc) scale!
Assessing the blood smear
Scan at 10x:
assess for general impressions, platelet clumps, microfilaria
Locate monolayer at 40x:
WBC differential
WBC abnormal morphology assessment
Use oil immersion 100x to assess:
RBC morphology in 10 random fields
Perform platelet estimate (Hema II)
RBC morphology:
Assess in 10 random oil immersion fields (OIF)
Interpret grading with chart
General categories: (assess entire field: mild/mod/ marked)
Anisocytosis: size variation
Poikilocytosis: shape variation
Rouleaux: decreased negative membrane charge - RBCs line up like stacks of coins