117 Terms

1
New cards

thermionic emission

process that occurs when a filament is becomes heated, and releases electrons

2
New cards

The amount of electricity that flows through the filament in the x-ray tube is determined by which CT parameter

mA

3
New cards

Ray definiton

part of x-ray beam that travels through patient and falls on one detector

4
New cards

ray sum definiton

total measurement of radiation that has traveled through a patient and fell on a single detector

5
New cards

view definiton

A collection of measurements that was calculated by the DAS

  • measurement is proportionate to # of rays that have struck the detector


6
New cards

algorithms definition

a set of rules or directions for getting a specific output from a specific input

  • should not be vauge, and should describe operation and must terminate after a certain number of steps


7
New cards

measurement data or scan data

comes from detectors

undergoes processing before the image can be further reconstructed using reconstruction algorithms

8
New cards

raw data

pre-processsed data that undergoes reconstction with image algrithms used by scanner

9
New cards

image data

image data that has undergone convolution and is projected onto a matrix

10
New cards

Convolution is

filtration algorithm that removes blurring in an image during back projection

11
New cards

Interpolation is

used in helical CT scans and is mathmatical process that uses data points above and below a slice to reconstruct an image

12
New cards

Mini IP is used to display

The minimum attenuation within a voxel

13
New cards

addaptive array detectors adjust the inner x-ray beam so that

only the inner half of the two detectors are exposed to the x-ray beam

14
New cards

A MIP image displays

the tissues with the greatest attenuation within the voxel

15
New cards

Which of the following transmit acquisition data to the computer?

DAS

16
New cards

The accuracy of the CT number compared to the linear attenuation coefficient is:

linearity

17
New cards

The voxel size depends on which of the following factors?

slice thickness

18
New cards

SSP can be used to describe which quality parameter?

logitudinal spatial resoltuion

19
New cards

It is common for the gantry of a CT Scanner to be able to tilt between:

12-30 degrees

20
New cards

Adjusting the filament size will have an impact on which of the following?

spatial resolution

21
New cards

Detectors that contain channels that are equal in all directions are known as:

uniform matrix array

22
New cards

pixel size is impacted by the

matrix

23
New cards

A reduction in scatter radiation will have what impact on image quality?

improve contrast resolution

24
New cards

The mathematical process that uses data points above and below a slice to reconstruct an image is known as:

interpolation

25
New cards

what are the three major sources of noise within a CT image

quantum mottle

component limitations

reconstruction algorithms

26
New cards

An acceptable range of variability when determining uniformity is:

5 HU

27
New cards

Utilization of ATCM will impact which of the following?

patient radiation dose

28
New cards

ATCM stands for

automatic tube current modulation

29
New cards

Which term is used to describe the rate that photons pass through an area in a set amount of time?

photon flux

30
New cards

Which of the following completes the majority of the processing of the image?

array processor

31
New cards

The MTF graph is used to represent a scanners ability for:

spatial resolution

32
New cards

Partial volume averaging can be compensated for through the use of:

acquiring thinner slices

33
New cards

The standards that ensure that medical data can easily be transferred and viewed from any device with viewing capabilities are called:

DICOM

34
New cards

Stability refers to the steadiness of a detectors response to which type of radiation?

transmitted

35
New cards

A volume of tissue or volume element is known as:

voxel

36
New cards

A contiguous data set means that:

there was no gap in the image data set

37
New cards

Assigning a value to the attenuation signal that is created by the detector is known as:

Linear attenuation coefficient

38
New cards

Decreasing the DFOV would do which of the following?

magnify the image

39
New cards

The inherent filtration of the x-ray tube used in CT consists of:

3mm of aluminum equivalency

40
New cards

Imaged areas with low variations in densities would have:

Low spatial frequency

41
New cards

Which of the following artifacts may be resolved through the calibration of the CT scanner?

ring

42
New cards

Increasing the rate of photons results in which of the following?

decrease noise

43
New cards

Gas ionization detectors are primarily used in:

third generation ct scanners

44
New cards

The limiting resolution of  MTF is:

10%

45
New cards

Which type of beam is utilized in CT (select two)?

polychromatic

heterogenous

46
New cards

Directing electrons to two different areas on the focal spot is referred to as a:

Flying focal spot

47
New cards

Filtered back projection uses what type of filter to remove the blurring that occurs with back projection?

Convolution

48
New cards

The campaign that was created to emphasize the importance of dose optimization for pediatrics is known as:

image gently

49
New cards

MSCT scanners utilize which type of detector?

Scintillation

50
New cards

A short circuit of the x-ray tube will result in:

Tube arcing

51
New cards

Artifacts on an image that are caused by the increase in the average energy of the x-ray beam is known as:

beam hardening

52
New cards

Metal artifacts can be compensated for through the use of:

Increasing the kV

53
New cards

To reduce the impact of metal artifacts on a CT scan, measures inlcude

remove metal from image when possible

utilizing thin sections

increasing kV

using software that can reduce the impacts of streaking

54
New cards

As the number of detector rows increases, the dose:

decreases

55
New cards

Setting the desired thickness of slices to be acquired is also known as:

Detector collimation

56
New cards

Cone beam CT are utilized in order to:

Expose the entire row of detectors evenly

57
New cards

Adjusting which of the following will have the greatest impact on dose?

kV

58
New cards

Which of the following is a result of beam hardening?

Cupping

59
New cards

cupping artifacts occur as a result of beam hardening, due to the beam becoming

harder in the center of the object being imaged, than it does at the edge

60
New cards

Placing several ROIs on a phantom image to measure the CT number is part of what QC test?

linearity

61
New cards

The term used to describe a filtration algorithm that removes blurring in an image during back-projection is:

convulution

62
New cards

The ability of the scanner to differentiate small differences in attenuation between closely spaced objects refers to which term?

Contrast resolution

63
New cards

Cone beam artifacts can be compensated for through the use of:

Using reconstruction algorithms

64
New cards

Which of the following is NOT a function of the RIS?

Store images

65
New cards

What impact does wide collimation have on a patient’s dose profile?

The dose is increased

66
New cards

characteristic x-rays are created when

an outer shell electron interacts with an inner shell electron and removes it

67
New cards

bremsstrahlung radiation is produced when

an electron is slowed as it passes the nucleus of an atom, causing it to slow down and lose some energy

68
New cards

When mA’s are doubled, the quanity of x-rays produced is

doubled

69
New cards

the quantity of x-rays is is directly proportional to

mAs

70
New cards

The quality of x-rasy is primarily controlled by the

kV

71
New cards

An increase in kV results in an increase of

beam quality - the x-ray beams ability to penetrate tissue

72
New cards

the inverse square law in ct states that

an increase in the distance of an object from the x-ray source = dose decrease

73
New cards

photoeletric effect

x-rays interacting with inner shell electrons and being absorbed. This results in complete absorption of the x-ray and the K shell electron being ejected from the atom.

74
New cards

Compton scattering

onsists of an incident x-ray interacting with an outer shell electron

75
New cards

Classical (Coherent) scattering

an incident x-ray interacting with an atom, which causes the atom to vibrate

76
New cards

as the x-ray tube gets closer to the iso-center of the gantry or closer to the detectors, radiation dose recieved by the patient

increases

77
New cards

filtration works to

absorb lower energy photons that only contribute to the patient dose

78
New cards

overbeaming occurs to

overcome the penumbra that occurs at the edge of the detectors

79
New cards

overranging occurs when

the radiation exposure begins prior to and after the image acquisition o

80
New cards

overranging is done to ensure

that there is enough data acquired from the beginning and end of the acquisition for reconstruction

81
New cards

wide collimation means that the patient’s dose

will be increased

82
New cards

narrow collimation of the x-ray beam means that the patient’s dose will be

decreased

83
New cards

When measuring radiation dose, exposure is

the measurement of the ionization of air

84
New cards

in radiation dose measurement, absorbed dose refers to

the amount of energy absorbed per unit of mass of material

85
New cards

in terms of radiation dose measurements, effective dose

quantifies an equivalent whole-body radiation dose from a partial body exposure

86
New cards

CTDI (CT dose Index) is

the measure of dose received in a single slice within an acquiisiton

87
New cards

CTDI100

a measurement taken with a 100mm long pencil ionzation chamber

88
New cards

CTDIw

accounts for the variations in density of an object being imaged

89
New cards

CTDIvol

Measures the radiation dose obtained during a helical scan and accounts for dose

does not take into consideration scan range; only estimates radiation dose within a given slice

90
New cards

DLP (Dose-Length product)

total amount of exposure for an exam, with consdieration of the scan length

91
New cards

several steps can be taken to reduce overall dose, especially in pediatric patients: steps include

Eliminating multiphase imaging

Reducing kV and mA

Ensuring clinical appropriateness

92
New cards

The image Gently campaign was created to increase awareness of dose optimization for

pediatric patients

93
New cards

The image wisely campaign was created to increase awareness of dose optimization for

adult patients

94
New cards

mA is ______ proportional to dose

directly

95
New cards

if mA doubles, dose ____

doubles

96
New cards

kV is ______ to dose

proportionalif

97
New cards

if kV increases, dose _____

increases

98
New cards

failure to iso-center the patient in the gantry can result in ___-

an under or over estimation of required dose for an exam

99
New cards

increase dose =

decrease noise

100
New cards

improve spatial resolution =

increase dose