thermionic emission
process that occurs when a filament is becomes heated, and releases electrons
The amount of electricity that flows through the filament in the x-ray tube is determined by which CT parameter
mA
Ray definiton
part of x-ray beam that travels through patient and falls on one detector
ray sum definiton
total measurement of radiation that has traveled through a patient and fell on a single detector
view definiton
A collection of measurements that was calculated by the DAS
measurement is proportionate to # of rays that have struck the detector
algorithms definition
a set of rules or directions for getting a specific output from a specific input
should not be vauge, and should describe operation and must terminate after a certain number of steps
measurement data or scan data
comes from detectors
undergoes processing before the image can be further reconstructed using reconstruction algorithms
raw data
pre-processsed data that undergoes reconstction with image algrithms used by scanner
image data
image data that has undergone convolution and is projected onto a matrix
Convolution is
filtration algorithm that removes blurring in an image during back projection
Interpolation is
used in helical CT scans and is mathmatical process that uses data points above and below a slice to reconstruct an image
Mini IP is used to display
The minimum attenuation within a voxel
addaptive array detectors adjust the inner x-ray beam so that
only the inner half of the two detectors are exposed to the x-ray beam
A MIP image displays
the tissues with the greatest attenuation within the voxel
Which of the following transmit acquisition data to the computer?
DAS
The accuracy of the CT number compared to the linear attenuation coefficient is:
linearity
The voxel size depends on which of the following factors?
slice thickness
SSP can be used to describe which quality parameter?
logitudinal spatial resoltuion
It is common for the gantry of a CT Scanner to be able to tilt between:
12-30 degrees
Adjusting the filament size will have an impact on which of the following?
spatial resolution
Detectors that contain channels that are equal in all directions are known as:
uniform matrix array
pixel size is impacted by the
matrix
A reduction in scatter radiation will have what impact on image quality?
improve contrast resolution
The mathematical process that uses data points above and below a slice to reconstruct an image is known as:
interpolation
what are the three major sources of noise within a CT image
quantum mottle
component limitations
reconstruction algorithms
An acceptable range of variability when determining uniformity is:
5 HU
Utilization of ATCM will impact which of the following?
patient radiation dose
ATCM stands for
automatic tube current modulation
Which term is used to describe the rate that photons pass through an area in a set amount of time?
photon flux
Which of the following completes the majority of the processing of the image?
array processor
The MTF graph is used to represent a scanners ability for:
spatial resolution
Partial volume averaging can be compensated for through the use of:
acquiring thinner slices
The standards that ensure that medical data can easily be transferred and viewed from any device with viewing capabilities are called:
DICOM
Stability refers to the steadiness of a detectors response to which type of radiation?
transmitted
A volume of tissue or volume element is known as:
voxel
A contiguous data set means that:
there was no gap in the image data set
Assigning a value to the attenuation signal that is created by the detector is known as:
Linear attenuation coefficient
Decreasing the DFOV would do which of the following?
magnify the image
The inherent filtration of the x-ray tube used in CT consists of:
3mm of aluminum equivalency
Imaged areas with low variations in densities would have:
Low spatial frequency
Which of the following artifacts may be resolved through the calibration of the CT scanner?
ring
Increasing the rate of photons results in which of the following?
decrease noise
Gas ionization detectors are primarily used in:
third generation ct scanners
The limiting resolution of MTF is:
10%
Which type of beam is utilized in CT (select two)?
polychromatic
heterogenous
Directing electrons to two different areas on the focal spot is referred to as a:
Flying focal spot
Filtered back projection uses what type of filter to remove the blurring that occurs with back projection?
Convolution
The campaign that was created to emphasize the importance of dose optimization for pediatrics is known as:
image gently
MSCT scanners utilize which type of detector?
Scintillation
A short circuit of the x-ray tube will result in:
Tube arcing
Artifacts on an image that are caused by the increase in the average energy of the x-ray beam is known as:
beam hardening
Metal artifacts can be compensated for through the use of:
Increasing the kV
To reduce the impact of metal artifacts on a CT scan, measures inlcude
remove metal from image when possible
utilizing thin sections
increasing kV
using software that can reduce the impacts of streaking
As the number of detector rows increases, the dose:
decreases
Setting the desired thickness of slices to be acquired is also known as:
Detector collimation
Cone beam CT are utilized in order to:
Expose the entire row of detectors evenly
Adjusting which of the following will have the greatest impact on dose?
kV
Which of the following is a result of beam hardening?
Cupping
cupping artifacts occur as a result of beam hardening, due to the beam becoming
harder in the center of the object being imaged, than it does at the edge
Placing several ROIs on a phantom image to measure the CT number is part of what QC test?
linearity
The term used to describe a filtration algorithm that removes blurring in an image during back-projection is:
convulution
The ability of the scanner to differentiate small differences in attenuation between closely spaced objects refers to which term?
Contrast resolution
Cone beam artifacts can be compensated for through the use of:
Using reconstruction algorithms
Which of the following is NOT a function of the RIS?
Store images
What impact does wide collimation have on a patient’s dose profile?
The dose is increased
characteristic x-rays are created when
an outer shell electron interacts with an inner shell electron and removes it
bremsstrahlung radiation is produced when
an electron is slowed as it passes the nucleus of an atom, causing it to slow down and lose some energy
When mA’s are doubled, the quanity of x-rays produced is
doubled
the quantity of x-rays is is directly proportional to
mAs
The quality of x-rasy is primarily controlled by the
kV
An increase in kV results in an increase of
beam quality - the x-ray beams ability to penetrate tissue
the inverse square law in ct states that
an increase in the distance of an object from the x-ray source = dose decrease
photoeletric effect
x-rays interacting with inner shell electrons and being absorbed. This results in complete absorption of the x-ray and the K shell electron being ejected from the atom.
Compton scattering
onsists of an incident x-ray interacting with an outer shell electron
Classical (Coherent) scattering
an incident x-ray interacting with an atom, which causes the atom to vibrate
as the x-ray tube gets closer to the iso-center of the gantry or closer to the detectors, radiation dose recieved by the patient
increases
filtration works to
absorb lower energy photons that only contribute to the patient dose
overbeaming occurs to
overcome the penumbra that occurs at the edge of the detectors
overranging occurs when
the radiation exposure begins prior to and after the image acquisition o
overranging is done to ensure
that there is enough data acquired from the beginning and end of the acquisition for reconstruction
wide collimation means that the patient’s dose
will be increased
narrow collimation of the x-ray beam means that the patient’s dose will be
decreased
When measuring radiation dose, exposure is
the measurement of the ionization of air
in radiation dose measurement, absorbed dose refers to
the amount of energy absorbed per unit of mass of material
in terms of radiation dose measurements, effective dose
quantifies an equivalent whole-body radiation dose from a partial body exposure
CTDI (CT dose Index) is
the measure of dose received in a single slice within an acquiisiton
CTDI100
a measurement taken with a 100mm long pencil ionzation chamber
CTDIw
accounts for the variations in density of an object being imaged
CTDIvol
Measures the radiation dose obtained during a helical scan and accounts for dose
does not take into consideration scan range; only estimates radiation dose within a given slice
DLP (Dose-Length product)
total amount of exposure for an exam, with consdieration of the scan length
several steps can be taken to reduce overall dose, especially in pediatric patients: steps include
Eliminating multiphase imaging
Reducing kV and mA
Ensuring clinical appropriateness
The image Gently campaign was created to increase awareness of dose optimization for
pediatric patients
The image wisely campaign was created to increase awareness of dose optimization for
adult patients
mA is ______ proportional to dose
directly
if mA doubles, dose ____
doubles
kV is ______ to dose
proportionalif
if kV increases, dose _____
increases
failure to iso-center the patient in the gantry can result in ___-
an under or over estimation of required dose for an exam
increase dose =
decrease noise
improve spatial resolution =
increase dose