151 Terms

1
New cards

swan-ganz (PA) catheter

flexible, radio-opaque 7 Fr, 110 cm PVC catheter marked in 10 cm increments from distal tip. PVC remains pliable at room temp and softens at body temp

2
New cards

swan-ganz (PA) catheter use

Obtains direct invasive hemodynamics - CVP, PAP, PCWP, SvO2, CO in high-risk cardiac surgical cases, cardiogenic shock, and MCS evaluation

3
New cards

PAC distal lumen (yellow)

measures pulmonary artery pressure (PAP) continuously and pulmonary capillary wedge pressure (PCWP) when wedged. Used to draw true mixed venous blood samples (SvO2). Terminates at catheter tip

4
New cards

PAC proximal lumen (blue)

measures central venous pressure (CVP), carries cold injectate for thermodilution CO, infusion site for vasoactive drugs. Terminates 30 cm above distal tip, positioning the opening in the RA when distal tip is in PA

5
New cards

PAC balloon port

Inflated with air then closed with stopcock during insertion to cushion tip and allow blood to “float” the catheter into the PA. Wedged to occlude antegrade flow and measure PCWP

6
New cards

PAC thermistor

Temperature-sensing electrical leads terminating 4 cm behind catheter tip in PA. Measures core blood temp in PA and detects change over time following cold injectate push to calculate CO via thermodilution

7
New cards

Multi-lumen access catheter (MAC/Cordis)

Large bore central venous access sheath (approx 9 Fr) with a hemostatic valve. High volume fluid resuscitation capability and houses the portal for PAC insertion

8
New cards

CVP/PAC insertion sites

Right internal jugular vein is preferred - short, direct shot into RA and lowest risk for pneumothorax and displacement. Subclavian vein carries higher risk of pneumothorax, femoral vein carries higher risk of infection and thrombosis.

9
New cards

seldinger technique

standard procedure for percutaneous access to blood vessels

10
New cards

seldinger steps

vessel cannulation with syringe needle. advance guidewire through needle. remove needle while maintaining wire position. dilate tissue tract over wire. advance catheter over guidewire. remove guidewire and secure line.

11
New cards

normal RAP

2-8 mmHg, low-pressure rolling hills waveform

12
New cards

normal RVP

15-25 / 3-8 mmHg, tall steep systolic towers with very low DBP and NO dicrotic notch

13
New cards

normal PAP

15-25 / 8-15 mmHg, Mean: 10-20 mmHg. Diastolic pressure step-up, dicrotic notch = pulmonic valve closure

14
New cards

normal PCWP/PAWP/LAP

Mean 8-15 mmHg; higher rolling hills waveform

15
New cards

pulmonary capillary wedge pressure (PCWP/PAWP)

Pressure measured when inflated balloon occludes antegrade flow in a small PA branch. Creates static fluid column through pulmonary capillaries to LA (Pascal’s Principle). Reflects LAP and LVEDP/LV preload.

16
New cards

a wave - CVP waveform

Atrial contraction (follows EKG P-wave)

17
New cards

c wave - CVP waveform

Tricuspid valve bulging during early systole.

18
New cards

x descent - CVP waveform

Atrial relaxation and downward pull of tricuspid valve during ventricular ejection/systole

19
New cards

v wave - CVP waveform

Passive atrial filling against closed tricuspid valve

20
New cards

y descent - CVP waveform

Atrial emptying into ventricles after tricuspid valve opens during ventricular diastole

21
New cards

absent a wave - CVP waveform

Loss of effective atrial contraction or kick, usually caused by A-fib or junctional rhythms

22
New cards

elevated v wave - CVP waveform

Hallmark of tricuspid regurgitation. Pressure from RV systole reads in the RA

23
New cards

mixed venous oxygen saturation (SvO2)

Blood drawn from PAC distal lumen provides a true mixed venous saturation because coronary sinus drainage has thoroughly mixed in the right heart with systemic venous drainage. Reflects global balance between DO2 and VO2. Normal = 60-80%

24
New cards

thermodilution CO

A known volume of cold saline is rapidly injected into proximal RA port. Distal thermistor records the resulting temperature drop curve in the PA, area underneath is inversely proportional to CO.

25
New cards

pacing catheters

specialized central venous or PACs with electrical pacing wires that contact RV endocardium to induce depolarization in bradycardia or heart block

26
New cards

PAC complications

Insertion: Ventricular arrhythmias (PVCs, Vtach), pneumothorax, bleeding/hematoma, arterial puncture. Indwelling: PA rupture or infarction, knotting, infection, endocardial or valve trauma

27
New cards

PAC monitoring system

PAC catheter, introducer sheath (cordis), non-compliant pressure tubing, 3 way stopcocks, pressure transducers, continuous flush system with heparinized saline and pressure bag (300 mmHg), transducer cable, pressure monitor, continuous EKG leads

28
New cards

elevated a wave - CVP waveform

Increased resistance to RA emptying, requiring greater pressure and force to push into RV. Tricuspid stenosis, reduced RV compliance, pulmonary HTN, increased RV afterload

29
New cards

cannon a wave - CVP waveform

Massive atrial pressure spike when RA contracts against a closed tricuspid valve. Caused by AV dissociation, AV block, junctional rhythms where atria and ventricles contract simultaneously.

30
New cards

Pascal’s Principle

Pressure applied to an enclosed fluid is transmitted undiminished to every portion of the fluid and to the walls of the containing vessel.

31
New cards

Poiseuille’s Law

Q = pi x pressure gradient x r^4 / 8 x viscosity x length. Flow is directly proportional to the pressure difference and r^4, and inversely proportional to the fluid’s viscosity and the tube’s length

32
New cards

Resistance in series

Cumulative. Eg. Adding a cannula to a line

33
New cards

Resistance in parallel

Reduces total resistance by providing more pathways. Eg. Adding a Y connector to split a line

34
New cards

Reynolds Number (Re)

Density x Velocity x Diameter / Viscosity. Dimensionless ratio predicting whether flow is laminar (<2000) or turbulent (>4000) with transition flow in between.

35
New cards

steps to zero a pressure transducer

Block off primed line by turning the stopcock off to the circuit and remove cap to expose to atmosphere

36
New cards

Zeroing

Exposing all arms of the Wheatstone bridge to atmosphere to establish room pressure as the 0 mmHg baseline reference point.

37
New cards

Phlebostatic axis

Positioning transducer at the level of the right atrium: 4th intercostal space, mid-axillary line to eliminate hydrostatic pressure errors when monitoring patient hemodynamics

38
New cards

Transducer placement 1in above pressure line placement

Displayed pressure reads 2 mmHg too low

39
New cards

Transducer placement 1 in below pressure line placement

Displayed pressure reads 2 mmHg too high (gravity)

40
New cards

Fidelity

Ability to reproduce the actual pressure waveform, preserving amplitude timing and shape

41
New cards

Natural frequency (Fn)

The frequency at which a mechanical system prefers to oscillate naturally. Lower in longer + more compliant systems and higher in shorter + stiffer systems

42
New cards

Dynamic response

How accurately the system follows rapidly changing pressure

43
New cards

Resonance (underdamping)

Signal is amplified when the incoming pressure frequency approaches the natural frequency of the monitoring system. Causes overshoot & exaggerated waveform amplitude

44
New cards

Damping

Loss of energy from an oscillating system. Too little = excessive oscillation, waveform amplitude, Appropriate = faithful waveform, Too much = loss of waveform detail

45
New cards

Underdamped readings

High SBP, low DBP, wide pulse pressure. Artifact, overshoot, ringing. Relatively preserved MAP due to opposing systolic and diastolic error

46
New cards

Overdamped readings

Low SBP, high DBP, narrow pulse pressure. No dicrotic notch

47
New cards

Square-Wave (or Fast-Flush or snap) test

Brief pigtail/plunger pulling to expose the transducer to 300 mmHg pressure bag fluid to evaluate dampening of the waveform following squaring

48
New cards

causes of overdamping

Air bubbles (compressible break in fluid), blood or clots, excessive tubing length or compliance, kinks, small-lumen catheter or one up against the vessel wall

49
New cards

causes of underdamping/resonance

Hyperdynamic signals (more frequent striking of catheter), tachycardia, sepsis, tubing vibration

50
New cards

catheter whip

Motion artifact from catheter movement

51
New cards

Calibration

Verification or reproduction of the pressure signal

52
New cards

Hydrodynamic pressure

Pressure produced by fluid in motion

53
New cards

Hydrostatic pressure

Pressure exerted by a stationary fluid column.

54
New cards

viscosity

A fluid’s internal friction or resistance to flow between adjacent layers. Promotes decreased Re, laminar flow

55
New cards

density

Mass per unit volume of a fluid, representing its inertia. Increased inertia increases Re making turbulent flow more likely

56
New cards

Laminar flow

Smooth streamlined flow in parallel layers. Re < 2000. Provides predictable flow profiles, lower resistance to movement, lower pressure drops across vessels or tubing

57
New cards

Turbulent flow

Disorganized chaotic fluid movement with swirls. Re > 4000. Significantly increases fluid resistance and places hydrodynamic stress on vessel walls (jet-blasting, endothelial dysfunction, plaque disruption)

58
New cards

Resistance-type transducer

Converts physical pressure against a diaphragm into a proportional change in electrical resistance and voltage output. Most common type using a wheatstone bridge containing a bonded strain gauge.

59
New cards

bonded strain gauge

Microscopic resistor chip on the back of the transducer diaphragm. Alters electrical resistance in direct proportion to applied force. Zeroing balances resistance at atmospheric pressure

60
New cards

wheatstone bridge

Electrical circuit configuration of resistors (including strain gauges) that converts tiny pressure-induced resistance changes into a measurable voltage output signal

61
New cards

capillaries

Highest cross-sectional area, lowest flow velocity

62
New cards

arterioles and venules

High cross-sectional area, lower flow velocity

63
New cards

arteries and veins

Lower cross-sectional area, higher flow velocity

64
New cards

aorta and vena cavae

Lowest cross-sectional area, highest flow velocity

65
New cards

components of a pressure monitoring system

catheter, non-compliant tubing, flush system/pressure bag/stopcocks, transducer cable, amplifier/monitor

66
New cards

Gain

Adjusts signal sensitivity from the transducer and waveform amplification

67
New cards

Scale

Adjusts only the visual height/grid range of the waveform, not the pressure reading itself

68
New cards

Amplifier

Multiplies tiny millivolt signals from the Wheatstone bridge into usable output voltages

69
New cards

Continuous Flush System

Maintained by a pressure bag at 300 mmHg delivering 3-4 mL/hr of continuous solution to keep the catheter tip patent. Not necessary for CPB pressure monitoring.

70
New cards

Steps for ABP with manual cuff

Select appropriate size. Secure cuff around upper arm at heart level. Inflate to 30 mmHg over normal ABP to occlude flow. Place bell over brachial artery while cuff deflates 3 mmHg/second. P at first sound = SBP, P at sound disappearance = DBP.

71
New cards

manual cuff advantages

noninvasive, easy to apply, low risk, widely available, appropriate for many stable patients

72
New cards

manual cuff disadvantages

Time between measurements misses rapid hemodynamic changes, accuracy depends on fit and positioning, no direct waveform or blood sampling, confounded by motion, arrhythmias, poor perfusion, vascular disease.

73
New cards

other non-invasive BP monitoring methods

manual auscultation, oscillometric cuff, doppler-assisted, finger cuffs

74
New cards

arterial catheter sites

Radial preferred, then femoral (central) and axillary. Brachial and dorsalis pedis in peds

75
New cards

Arterial catheter complications

Bleeding, hematoma, thrombosis/occlusion, emboli, distal ischemia, infection, disconnection/dislodgement

76
New cards

peripheral vs central arterial waveforms

Peripherals show widened pulse pressure and slid down dicrotic notch

77
New cards

pulse pressure

SBP - DBP

78
New cards

blood pressure by age

Vessels get less compliant, BP increases and pulse pressure widens

79
New cards

preload

volume of blood in both ventricles at the end of diastole. measured by CVP/RVEDV (R) and PCWP/LVEDV (L)

80
New cards

afterload

resistance the ventricles must overcome to eject their blood volume. measured by PVR (R) and SVR (L)

81
New cards

contractility

Inotropy - force with which the heart muscle contracts. Measured by SV & EF. Increased by inotrope drugs

82
New cards

preload modulation

Giving volume and blood increase, giving diuretics decreases (more fluid excreted)

83
New cards

afterload modulation

vasoconstrictors increase, vasodilators decrease

84
New cards

CO

HR x SV

85
New cards

CI

CO / BSA

86
New cards

SV

EDV - ESV. Normal = 50-100 ml/beat

87
New cards

stroke volume index (SVI)

SV / BSA

88
New cards

systemic vascular resistance (SVR)

(MAP - CVP) / CO x 80. Normal = 900-1440 dynes x s x cm^-5

89
New cards

Frank Starling curve of contractility

As preload (x- axis) increases, SV (y-axis) also increases up to a point. Past which the heart over-distends and can’t pump efficiently (SV decreases)

90
New cards

MAP

(1/3) systolic + (2/3) diastolic

91
New cards

pulmonary vascular resistance (PVR)

(mPAP - LAP) / CO x 80. Normal = 20-120 dynes x s x cm^-5

92
New cards

components of arterial pressure waveform

anacrotic limb, systolic peak, dicrotic limb, dicrotic notch, end diastolic pressure

93
New cards

anacrotic limb of arterial pressure waveform

rapid pressure rise during ventricular ejection

94
New cards

systolic peak of arterial pressure waveform

Anacrotic notch is the waveform’s highest peak and reflects SBP

95
New cards

dicrotic limb of arterial pressure waveform

pressure falls as blood moves into the peripheral circulation

96
New cards

dicrotic notch

Late dip in arterial pressure wavform associated with aortic valve closure. Marks the start of diastole

97
New cards

end diastolic pressure

Lowest trough in arterial pressure waveform immediately before the next systolic upstroke

98
New cards

peripheral arteries

High-resistance vessels with branch points create reflected pressure waves that travel back toward the heart. These combine with the forward wave during systole to amplify distal diastolic pulses.

99
New cards

aorta and vena cavae

Lowest cross-sectional area, highest flow velocity

100
New cards

Cardiac output (CO)

Volume of blood pumped per minute. HR x SV. Normal adult resting = 4-8 LPM