1/81
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
Unipolar
Large sensing window between device and electrode in the heart
Bipolar
small, more specific sensing window between 2 electrodes at the tip
Can pacemakers function unipolar and bipolar?
Yes
Unipolar is _____ _______ when bipolar fails
backup option
Measure sensing in
mV
Measure pacing in
Volts
Minimum P wave value
greater than or equal to 2 mV
Minimum R wave value
greater than or equal to 5 mV
Sensing
When a device sees natural/intrinsic depolarization
Devices sense cardiac depolarizations by
measuring change in electrical potential of myocardial cells
Fence in sensing
tells us less or more sensitive and has an inverse relationship
Less sensitive fence means
device sees less, fence is higher
More sensitive fence means
device sees more, fence is lower
Oversensing
seeing too much (like r and t waves), fence needs to be raised
Undersensing
not seeing enough (won't sense r wave), fence needs to be lowered
Oversensing causes
Under pacing
- skipped/missed beats
Undersensing causes
overpacing
- faster/unnecessary paces and pro arrhythmic
Do we ever want to pace atrium in A flutter or A fib?
NO
Oversensing on RV lead can cause
pacing inhibition and device misinterpretation of VF
How to reproduce noise?
isometric exercises
Noise Oversensing
- can cause underpacing and misinterpretted VF
What happens if we undersense AF or VF?
could continue to pace during these fast rates
Near and farfield over sensing
sensing an extra signal that disrupts timing cycle
comes from same chamber or different chambers/outside the heart
Nearfield Oversensing
Oversensing in same chamber
- V channel senses R and T waves, double counting
Farfield Oversensing
- one channel senses signals from another chamber or from outside the heart
- A channel sensing V signals
Atrial lead shape
J shape in RAA
Farfield R wave
Oversensing
- A channel senses R waves from V channel
- can cause misinterpretation of atrial arrhythmias and false mode switch
Why would A channel see farfield R waves?
- A channel fence too low
- V output too high
T wave Oversensing
Nearfield
- counts T wave as R wave in V channel
- can cause false positive VT
Can device discern R v. T wave?
NO
Why might T waves be large?
Hyperkalemia
Atrial Lead Dislodgement
- usually happens in first 24h
- A lead falls into RV or dislodges from RAA
What do we need before discharging patient with new device?
X-ray for lead placement
What does atrial lead dislodgment look like on strip?
- atrial pacing impulse captures the V (lead has fallen into RV)
Polarization
Layers of negative and positive ions surrounding the electrode during pulse stimulus. It can slow the movement of a charge.
Capture
Successful depolarization of cardiac tissue from pacing stimulus
Stimulation threshold
minimum amount of energy required to cause depolarization
Conduction
Movement of charge under influence of electrical field
Voltage
V
- force that drives current (amplitude)
Pulse Width
ms
- duration of time in a pacing stimulus
Current
mA
- flow of electrons measured in milliamps, not programmable; is calculated
Output Capacitor
Stores and delivers pacing impulses
Impedence
Ohms
- resistance to the flow of current within an electrical circuit
T/F Thresholds can change
True
T/F Autocapture algorithms can be tricked
true
T/F A threshold can be checked on pacemaker dependent patient
True
T/F Afib always needs threshold check
false: can't outrun AF
Pacemaker battery made out of
lithium/iodine
Beginning of Life (BOL)
when the device is first put in
Elective Replacement Interval (ERI)
- device needs to be changed out
- 3-6 months left usually
End of Life (EOL)
pacemaker stops working and has to be changed out
Battery cathode
positive
Lead cathode
negative
Battery Anode
negative
Lead Anode
positive
Lead tip is always
cathode and negative
Depolarization with device
Electrons flow from pulse generator to negative cathode tip
- influence ions for big Na+ influx
In CHB what phase of AP are we in?
Phase 4
PM goal in AP
get to phase 0 to depolarize
Threshold purpose
to have enough energy to cause depolarization
Pacing is
programmable depolarization through amplitude and pulse width
Amplitude
how many volts are delivered per pacing impulse
Pulse width in pacing
duration of impulse, usually 0.4 or 0.5 ms
Goal of pacing
effective and safe output setting and maximizing battery longevity
Safety margin
2:1
- sensitivity: half
- amplitude: double
Threshold test
Control/outrun the heart rate
- may decrease AVD for ventricular threshold
Never run threshold when
patient is in VT
Rate limit for threshold test
120 bpm
Pacemaker dependent threshold test
Run in DDD, need ventricular support
How to know if patient is pacemaker dependent?
Run a sensing test
Can you run atrial threshold in AF?
No
In dual chamber device testing in DDD, how many chambers decremented at a time?
one
What to know before running threshold test
Presenting and Underlying rhythm
Overdrive Pacing requires
Power (adequate starting amplitude) and speed (faster than underlying rhythm)
Decrement power
decrease amplitude incrementally and watch for loss of capture
What is the threshold?
Not output where test stops
- It is previous output with successful depolarization
Factors affecting threshold
- Lead Integrity
- Lead location
- medications
- arrhythmias
- electrolyte imbalances
Causes of failure to capture
- lead dislodgment (physician fix)
- poor connection at connector block (least likely) (physician fix)
- lead maturation
- lead failure
- low output (change amplitude)
Fusion Beat
1/2 intrinsic beat, 1/2 paced
- timing cycle issue, AVD too short
Pseudofusion beat
- pace falls in the middle of intrinsic beat during absolute refractory and does not capture
- increase sensitivity (V) if d/t undersensing
Pseudo Pseudofusion beat
A pace in intrinsic R wave and V pace in T wave
- undersensing P and R waves
Reasons for Automatic Capture Algorithms
- more frequent testing
- prolong battery life