digital radiography (DR)
cassette-less system
direct and indirect conversion
direct digital radiography DDR
x-rays —> electronic signals to digital
indirect digital radiography IDR
xrays —> visible light —> electronic signals to digital
is direct or indirect more common?
indirect because it is more cost effective (lower spatial resolution though because extra conversion)
difference between direct vs indirect?
direct does not use scintillator (converts xr to light)
indirect conversion can use which two devices?
CCD charge-coupled device (scintillation layer)
TFT thin film transistor (scintillation and photodiode layer)
thin film transistor used for?
angiography and FL
CCD
charge coupling device
used in indirect digital radiography
light sensing element
made of silicon
why is silicon used for CCD?
very sensitive to light
wide dynamic range
small size
fill factor
percentage of the actual pixel size that is sensitive to xrs
bigger fill factor desired
INC fill factor = ? resolution
INC fill factor = INC resolution
direct digital radiography layers
photoconductor (amorphous selenium)
thin film transistor TFT
analog to digital converter
does indirect or direct have higher resolution?
direct has higher resolution bc less steps but costs more money
digital images are a ?
matrix of picture element (pixels) values
each pixel is assigned what?
a shade of gray
bone
small signal
radiopaque
soft tissue
medium signal
gray
air
large signal
radiolucent
pixel
picture element
smallest visible unit of a digital image that determine spatial detail of an xray image
bit
smallest unit of digital data
How many brightness values each pixel can have OR how many shades of gray each pixel can take on
bit depth
pixel’s range of possible shades of gray (10, 12, 16 are common)
bit depth formula
2^n
INC bit depth = ? shades of gray = ? contrast resolution
INC bit depth = INC shades of gray = INC contrast resolution
INC pixels = ? matrix = ? resolution
INC pixel amount = INC matrix size = INC resolution
bit depth and contrast resolution related how?
directly proportional
No DR imaging system can resolve an object ?
No DR imaging system can resolve an object smaller than the pixel size
pixel size too big, object size too small
only one shade of gray per pixel
matrix size and spatial resolution relationship?
directly proportional
common matrix sizes
1024x1024, 2048x2048, 4096x4096
spatial frequency
describes how often image details repeat per unit distance in an image / how often image details change from light and dark over a given distance
spatial frequency measured how?
line pairs / mm
when do we want low spatial frequency?
larger structures, soft tissue outlines
when do we want high spatial frequency?
finer detail; looking for trabeculae detail, extremities, smaller bones
mtf
modulation transfer function
describes how well an imaging system preserves detail or contrast at different spatial frequencies
Ability of an imaging system to display objects of different sizes onto an image
mtf scale 0-1
1 = perfect representation
0 = no representation
dynamic range
range of values over which a system can respond
AKA gray scale
larger dynamic range
More shades of gray (long scale)
Low contrast
Better contrast resolution
Wide image latitude (margin of error for an image)
smaller dynamic range
Less shades of gray (short scale)
High contrast
Less contrast resolution
Narrow image latitude
image quality factors
brightness
contrast
resolution
distortion
exposure index
noise
brightness
intensity of light that represents the individual pixels in the image on the monitor
brightness replaces density
contrast
difference in brightness between light and dark areas of an image
scatter radiation control
Closely collimate (hardens the beam, filters out low energy xr)
Use grids (absorb scatter radiation before reaching IR)
Optimal kV
what determines spatial resolution in both film and digital ?
film = focal spot size
digital = pixel size
SNR
signal to noise ratio
high snr desired
mAs and SNR have direct relationship
INC mAs = ? SNR
INC mas = INC snr (direct relationship)
signal
number of xr photons that strike the detector
determined by mAs
noise
Other factors that negatively affect the final image
low snr makes what kind of image?
grainy image - quantum mottle bc not enough xr photons reaching IR
INC mas to fix !
electronic noise
inherent noise in the electronic system, power fluctuations, non-uniformity of detector
detective quantum efficiency
the ability of a material to absorb xr energy ; how efficient a IR absorbs xr signal and produces signal
DQE and pt dose relationship?
inverse
inc dqe = dec pt dose
30% rule:
A minimum of 30% of the IR must be exposed for an accurate exposure index
pt dose from film to DDR?
pt dose from film to IDR?
film to DDR = reduce dose 20-50%
film to IDR = reduce dose 30-50%
highest DQE of electronic conductors?
selenium