Kilovolt peak, milliampere, exposure time, distance
Exposure factors
Exposure Factors
The factors that influence and determine the quantity and quality of x-radiation
Radiation quantity
refers to radiation intensity
measured in mGya or mGya/mAs (mR or mR/mAs)
Radiation quality
refers to x-ray beam penetrability
best measured by the half-value layer (HVL)
Half value layer
used to measure radiation quality
Kilovolt peak, current, time, distance
FOUR PRIME EXPOSURE FACTORS
Kilovolt peak
The primary control of x-ray beam quality and therefore beam penetrability
Kilovolt peak
controls screen-film radiographic contrast
Increasing kVp
• More x-rays are emitted and have higher energy and greater penetrability.
• Produce more scatter radiation due to the Compton effect.
• Less differential absorption occurs, therefore reducing image contrast
Milliamperes
Determines the number of x-rays produced and therefore the radiation quantity
One ampere
equal to 1 coulomb (C) of electrostatic charge flowing each second in a conductor.
Milliamperes
As more electrons flow through the x-ray tube, more x-rays are produced.
With a constant exposure time, mA controls the x-ray quantity and therefore the patient radiation dose.
Fixed
X-ray quality remains__with a change in mA
Exposure time
Usually are kept as short as possible.
Short ___ reduces motion blur.
The mA and __ are usually combined and used as mAs
mAs
what controls OD?
Indirect fashion
Time and mA can be used to compensate for each other in an
Distance
affects exposure of the IR according to the inverse square law.
SID
largely determines the intensity of the x-ray beam at the image receptor.
Distance
has no effect on radiation quality
Square Law
Derived from the Inverse Square Law
It allows a radiologic technologist to calculate the required change in mAs after a change in SID to maintain constant OD
Longer SID
results in less magnification, less focal spot blur, and improved spatial resolution
More mAs
must be used because of the effects of the direct square law
Focal Spot Size, Filtration, High-Voltage Generation
IMAGING SYSTEM CHARACTERISTICS
Two
Most x-ray tubes are equipped with___ focal-spot sizes
Small and large
• 0.5 mm/1.0 mm
• 0.6 mm/1.2 mm
• 1.0 mm/2.0 mm
Does not change xray quality or quantity
Changing the focal spot for a given kVp/mAs setting
Large focal spots
For general imaging.
Can be used to image thick or dense body parts.
Provides shorter exposure time.
Many more x-rays can be produced with the large focal spot because the anode heat capacity is higher
Small focal spots
Reserved for fine-detail radiography.
Normally used during extremity radiography and in examination of other thin body parts
Inherent, Added, Compensating
Three types of X-ray filtration
Inherent filters
Glass or metal envelope of the x-ray tube.
Has a value of 0.5 mm Al equivalent.
Variable-aperture light-localizing collimator usually provides an additional 1.0 mm Al equivalent.
Most of this is attributable to the reflective surface of the collimator's mirror.
To meet the required total filtration of 2.5 mm Al, an additional 1- mm Al filter is inserted between the x-ray tube housing and the collimator.
0.5 mm Al
inherent filter value
1.0 mm Al
Variable-aperture light-localizing collimator usually provides an additional __
Added filter
Intentional placement of filtering materials, such as aluminum or copper, in the path of the X-ray beam to absorb unwanted low-energy photons before they reach the patient
Compensating filter
These are shapes of aluminum mounted onto a transparent panel that slides in grooves beneath the collimator.
These filters balance the intensity of the x-ray beam to deliver a more uniform exposure to the image receptor
Wedge shaped, trough shaped
Types of compensating filters
Effects of Added Filtration
1. Removes low-energy photons.
2. Increases the average energy of the X-ray beam (increases beam quality).
3. Decreases the total number of photons (decreases beam quantity).
4. Reduces unnecessary radiation exposure to the patient.
Single phase, three phase, high frequency
Three basic types of high-voltage generators
Three-phase power
is more efficient than single-phase power
High Voltage Generation
The x-radiation emitted is nearly constant rather than pulsed.
High Frequency generation
results in even greater x-ray quantity and quality.
Body Habitus
The general size and shape of a patient
Thickness, composition, pathology
Patient Factors
Sthenic
Radiographic technique charts are based on
Thickness
the more x-radiation is required.
It should not be guessed.
Calipers
are used to measure the thickness of the anatomy being imaged.
Composition
Various tissues often are described by their degree of radiolucency or radiopacity.
Radiolucent tissue
attenuates few x-rays and appears black
Radiopaque tissue
absorbs x-rays and appears white
Destructive, Constructive
PATHOLOGY
Destructive Pathology
tissue became more radiolucent
Constructive Pathology
causes the tissue to become more radiopaque
Destructive
Active tuberculosis, atrophy, bowel obstruction, cancer, degenerative arthritis, emphysema, osteoporosis, pneumothorax
Constructive
Aortic aneurysm, ascites, atelectasis, cirrhosis, hypertrophy, metastases, pleural effusion, pneumonia, sclerosis
Optical Density, Contrast, Image Detail, Distortion
IMAGE-QUALITY FACTORS
mAs, SID
OD is controlled by?
directly proportional
mAs vs. OD
inversely proportional
SID vs. OD
directly proportional
kVp vs. OD
30%
change in mAs to produce a perceptible change in OD
4%
change in kVp to produce a perceptible change in OD
15% kVp rule
If the OD on the radiograph is to be increased with the use of kVp, an increase in kVp of 15% is equivalent to doubling the mAs.
kVp
the major factor used in controlling radiographic contrast
Image Detail
Describes the sharpness of appearance of small structures on the radiograph.
Distortion
Misrepresentation of object size and shape on the radiograph.
__ is reduced by positioning the anatomical part of interest in a plane parallel to that of the image receptor.
Artifacts
undesirable optical densities or blemishes on a radiograph or any other medical image.
can interfere with the visualization of anatomical structures and can lead to misdiagnoses.
cause is identified
Artifacts can be controlled when __
Exposure, Handling, Processing
Radiographic artifacts occur in three areas
Exposure artifacts
associated with the manner in which the radiographer conducts the examination.
are usually easy to detect and correct.
Unexpected foreign object
Improper patient preparation
Double exposure
reuse of cassettes already exposed
Blur
Improper patient movement, including breathing
Grid cutoff artifact
improper patient positioning
Obscured detail
poor screen-film contact
Processing Artifacts
Artifacts produced during film processing.
Most are pressure-type artifacts caused by the transport system
Pressure type artifacts
usually sensitize the emulsion and appear as higher optical density
Lower OD
Those that scrape or remove emulsion appear as
Proper processor QC program, frequent cleaning
Processing artifacts are eliminated with a
Guide shoe marks
Occur when the guide shoes in the turnaround assembly of the processor are sprung or improperly positioned.
Pi lines
Occur at 3.1416-inch (π) intervals because of dirt or a chemical stain on a roller.
3.1416 inches represents one revolution of a roller.
The artifact appears perpendicular to the film’s direction of travel through the processor
Dirty rollers
can cause emulsion pick-off and gelatin buildup, which result in sludge deposits on the film.
These artifacts usually appear as sharp areas of increased or reduced OD.
Chemical fog
looks like light or radiation fog and is usually a uniform dull gray.
Dichroic stain
Improper or inadequate processing chemistry can result in a special type of chemical fog
Curtain effect
The dichroic stain appears as a ___ on the radiograph
Yellow, green, blue, purple
Chemical stains on a radiograph can appear ___
Wet-Pressure Sensitization
Occurs when pressure is applied to film while it is still wet, particularly during film processing.
Wet-pressure sensitization is a common artifact that is produced in the developer tank.
Irregular or dirty rollers cause pressure during development and produce small circular patterns of increased OD
Developer tank
Wet-pressure sensitization is a common artifact that is produced in the __
Irregular or dirty rollers
cause pressure during development and produce small circular patterns of increased OD
Handling and Storage Artifacts
Proper facility design helps reduce handling and storage artifacts.
Light or Radiation fog
White light leaks in the darkroom or within the cassette cause streak-like artifacts of increased OD.
If the safelight has an improper filter, too bright, or the safelight is too close to the film processing tray, the image may be fogged.
Films left in the x-ray examination room during an exposure can become fogged by radiation.
Radiation fog and safelight fog look alike.
Pressure or Kink Marks
Rough handling before processing can cause scratches and kink marks. •
Although the kink mark may appear as a fingernail mark, it is not.
It is caused by the kinking or abrupt bending of film.
Both events usually appear as increased OD.
Static
probably the most obvious artifact.
It is caused by the buildup of electrons in the emulsion and is most noticeable during the winter and during periods of extremely low humidity.
Crown, tree, smudge
Three distinct patterns of static
Hypo Retention
The yellow-brown stain that slowly appears on a radiograph after a long storage time
Happens when not all of the residual thiosulfate from fixing was removed during washing, and silver sulfide slowly builds up and appears yellow in the stored radiograph.