SCREEN FILM RADIOGRAPHIC TECHNIQUE

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/92

encourage image

There's no tags or description

Looks like no tags are added yet.

93 Terms

1
New cards

Kilovolt peak, milliampere, exposure time, distance

Exposure factors

2
New cards

Exposure Factors

The factors that influence and determine the quantity and quality of x-radiation

3
New cards

Radiation quantity

refers to radiation intensity

measured in mGya or mGya/mAs (mR or mR/mAs)

4
New cards

Radiation quality

refers to x-ray beam penetrability

best measured by the half-value layer (HVL)

5
New cards

Half value layer

used to measure radiation quality

6
New cards

Kilovolt peak, current, time, distance

FOUR PRIME EXPOSURE FACTORS

7
New cards

Kilovolt peak

The primary control of x-ray beam quality and therefore beam penetrability

8
New cards

Kilovolt peak

controls screen-film radiographic contrast

9
New cards

Increasing kVp

• More x-rays are emitted and have higher energy and greater penetrability.

• Produce more scatter radiation due to the Compton effect.

• Less differential absorption occurs, therefore reducing image contrast

10
New cards

Milliamperes

Determines the number of x-rays produced and therefore the radiation quantity

11
New cards

One ampere

equal to 1 coulomb (C) of electrostatic charge flowing each second in a conductor.

12
New cards

Milliamperes

As more electrons flow through the x-ray tube, more x-rays are produced.


With a constant exposure time, mA controls the x-ray quantity and therefore the patient radiation dose.

13
New cards

Fixed

X-ray quality remains__with a change in mA

14
New cards

Exposure time

Usually are kept as short as possible.

Short ___ reduces motion blur.

The mA and __ are usually combined and used as mAs

15
New cards

mAs

what controls OD?

16
New cards

Indirect fashion

Time and mA can be used to compensate for each other in an

17
New cards

Distance

affects exposure of the IR according to the inverse square law.

18
New cards

SID

largely determines the intensity of the x-ray beam at the image receptor.

19
New cards

Distance

has no effect on radiation quality

20
New cards

Square Law

Derived from the Inverse Square Law


It allows a radiologic technologist to calculate the required change in mAs after a change in SID to maintain constant OD

21
New cards

Longer SID

results in less magnification, less focal spot blur, and improved spatial resolution

22
New cards

More mAs

must be used because of the effects of the direct square law

23
New cards

Focal Spot Size, Filtration, High-Voltage Generation

IMAGING SYSTEM CHARACTERISTICS

24
New cards

Two

Most x-ray tubes are equipped with___ focal-spot sizes

25
New cards

Small and large

• 0.5 mm/1.0 mm

• 0.6 mm/1.2 mm

• 1.0 mm/2.0 mm

26
New cards

Does not change xray quality or quantity

Changing the focal spot for a given kVp/mAs setting

27
New cards

Large focal spots

For general imaging.

Can be used to image thick or dense body parts.

Provides shorter exposure time.

Many more x-rays can be produced with the large focal spot because the anode heat capacity is higher

28
New cards

Small focal spots

Reserved for fine-detail radiography.

Normally used during extremity radiography and in examination of other thin body parts

29
New cards

Inherent, Added, Compensating

Three types of X-ray filtration

30
New cards

Inherent filters

Glass or metal envelope of the x-ray tube.

Has a value of 0.5 mm Al equivalent.

Variable-aperture light-localizing collimator usually provides an additional 1.0 mm Al equivalent.

Most of this is attributable to the reflective surface of the collimator's mirror.

To meet the required total filtration of 2.5 mm Al, an additional 1- mm Al filter is inserted between the x-ray tube housing and the collimator.

31
New cards

0.5 mm Al

inherent filter value

32
New cards

1.0 mm Al

Variable-aperture light-localizing collimator usually provides an additional __

33
New cards

Added filter

Intentional placement of filtering materials, such as aluminum or copper, in the path of the X-ray beam to absorb unwanted low-energy photons before they reach the patient

34
New cards

Compensating filter

These are shapes of aluminum mounted onto a transparent panel that slides in grooves beneath the collimator.


These filters balance the intensity of the x-ray beam to deliver a more uniform exposure to the image receptor

35
New cards

Wedge shaped, trough shaped

Types of compensating filters

36
New cards

Effects of Added Filtration

1. Removes low-energy photons.

2. Increases the average energy of the X-ray beam (increases beam quality).

3. Decreases the total number of photons (decreases beam quantity).

4. Reduces unnecessary radiation exposure to the patient.

37
New cards

Single phase, three phase, high frequency

Three basic types of high-voltage generators

38
New cards

Three-phase power

is more efficient than single-phase power

39
New cards

High Voltage Generation

The x-radiation emitted is nearly constant rather than pulsed.

40
New cards

High Frequency generation

results in even greater x-ray quantity and quality.

41
New cards

Body Habitus

The general size and shape of a patient

42
New cards

Thickness, composition, pathology

Patient Factors

43
New cards

Sthenic

Radiographic technique charts are based on

44
New cards

Thickness

the more x-radiation is required.

It should not be guessed.

45
New cards

Calipers

are used to measure the thickness of the anatomy being imaged.

46
New cards

Composition

Various tissues often are described by their degree of radiolucency or radiopacity.

47
New cards

Radiolucent tissue

attenuates few x-rays and appears black

48
New cards

Radiopaque tissue

absorbs x-rays and appears white

49
New cards

Destructive, Constructive

PATHOLOGY

50
New cards

Destructive Pathology

tissue became more radiolucent

51
New cards

Constructive Pathology

causes the tissue to become more radiopaque

52
New cards

Destructive

Active tuberculosis, atrophy, bowel obstruction, cancer, degenerative arthritis, emphysema, osteoporosis, pneumothorax

53
New cards

Constructive

Aortic aneurysm, ascites, atelectasis, cirrhosis, hypertrophy, metastases, pleural effusion, pneumonia, sclerosis

54
New cards

Optical Density, Contrast, Image Detail, Distortion

IMAGE-QUALITY FACTORS

55
New cards

mAs, SID

OD is controlled by?

56
New cards

directly proportional

mAs vs. OD

57
New cards

inversely proportional

SID vs. OD

58
New cards

directly proportional

kVp vs. OD

59
New cards

30%

change in mAs to produce a perceptible change in OD

60
New cards

4%

change in kVp to produce a perceptible change in OD

61
New cards

15% kVp rule

If the OD on the radiograph is to be increased with the use of kVp, an increase in kVp of 15% is equivalent to doubling the mAs.

62
New cards

kVp

the major factor used in controlling radiographic contrast

63
New cards

Image Detail

Describes the sharpness of appearance of small structures on the radiograph.

64
New cards

Distortion

Misrepresentation of object size and shape on the radiograph.


__ is reduced by positioning the anatomical part of interest in a plane parallel to that of the image receptor.

65
New cards

Artifacts

undesirable optical densities or blemishes on a radiograph or any other medical image.


can interfere with the visualization of anatomical structures and can lead to misdiagnoses.

66
New cards

cause is identified

Artifacts can be controlled when __

67
New cards

Exposure, Handling, Processing

Radiographic artifacts occur in three areas

68
New cards

Exposure artifacts

associated with the manner in which the radiographer conducts the examination.


are usually easy to detect and correct.

69
New cards

Unexpected foreign object

Improper patient preparation

70
New cards

Double exposure

reuse of cassettes already exposed

71
New cards

Blur

Improper patient movement, including breathing

72
New cards

Grid cutoff artifact

improper patient positioning

73
New cards

Obscured detail

poor screen-film contact

74
New cards

Processing Artifacts

Artifacts produced during film processing.

Most are pressure-type artifacts caused by the transport system

75
New cards

Pressure type artifacts

usually sensitize the emulsion and appear as higher optical density

76
New cards

Lower OD

Those that scrape or remove emulsion appear as

77
New cards

Proper processor QC program, frequent cleaning

Processing artifacts are eliminated with a

78
New cards

Guide shoe marks

Occur when the guide shoes in the turnaround assembly of the processor are sprung or improperly positioned.

79
New cards

Pi lines

Occur at 3.1416-inch (π) intervals because of dirt or a chemical stain on a roller.


3.1416 inches represents one revolution of a roller.


The artifact appears perpendicular to the film’s direction of travel through the processor

80
New cards

Dirty rollers

can cause emulsion pick-off and gelatin buildup, which result in sludge deposits on the film.


These artifacts usually appear as sharp areas of increased or reduced OD.

81
New cards

Chemical fog

looks like light or radiation fog and is usually a uniform dull gray.

82
New cards

Dichroic stain

Improper or inadequate processing chemistry can result in a special type of chemical fog

83
New cards

Curtain effect

The dichroic stain appears as a ___ on the radiograph

84
New cards

Yellow, green, blue, purple

Chemical stains on a radiograph can appear ___

85
New cards

Wet-Pressure Sensitization

Occurs when pressure is applied to film while it is still wet, particularly during film processing.


Wet-pressure sensitization is a common artifact that is produced in the developer tank.


Irregular or dirty rollers cause pressure during development and produce small circular patterns of increased OD

86
New cards

Developer tank

Wet-pressure sensitization is a common artifact that is produced in the __

87
New cards

Irregular or dirty rollers

cause pressure during development and produce small circular patterns of increased OD

88
New cards

Handling and Storage Artifacts

Proper facility design helps reduce handling and storage artifacts.

89
New cards

Light or Radiation fog

White light leaks in the darkroom or within the cassette cause streak-like artifacts of increased OD.


If the safelight has an improper filter, too bright, or the safelight is too close to the film processing tray, the image may be fogged.


Films left in the x-ray examination room during an exposure can become fogged by radiation.


Radiation fog and safelight fog look alike.

90
New cards

Pressure or Kink Marks

Rough handling before processing can cause scratches and kink marks. •

Although the kink mark may appear as a fingernail mark, it is not.

It is caused by the kinking or abrupt bending of film.

Both events usually appear as increased OD.

91
New cards

Static

probably the most obvious artifact.


It is caused by the buildup of electrons in the emulsion and is most noticeable during the winter and during periods of extremely low humidity.

92
New cards

Crown, tree, smudge

Three distinct patterns of static

93
New cards

Hypo Retention

The yellow-brown stain that slowly appears on a radiograph after a long storage time


Happens when not all of the residual thiosulfate from fixing was removed during washing, and silver sulfide slowly builds up and appears yellow in the stored radiograph.