flow of internal components of linac (MV)
modular (magnetron/klystron) — pulse into waveguide (microwaves)
electron gun injects electrons into waveguides (electrons come from heating tungsten filaments)
microwaves accelerate electrons through the accelerating waveguide (@ near SOL), using steering magnets used to focus beam
theres a vacuum in the waveguide to avoid contamination with other particles
bending magnets deflect the beam to the tungsten xray target and when electrons interact, theyre converted into photons/xrays
photons/xrays focused thru primary collimator to rid of scatter
shape the beam thru the flattening filter — to uniformly distribute photon into cone shape of beam
beam pass thru 2 ion chambers (dual ionization chambers) — measuring dose thats delivered
pass thru treatment head (jaws: form x and y field size) and multi-leaf collimator (MLC: conforms beam to unique shapes thats specific to the treatment)
where do electrons come from in a linac
produced by heating a tungsten filament (↑ heat = ↑ electrons)
what prevents overheating of the linac
machine (accelerating waveguide areas) are bathed in water to cool down and prevent overheating
gantry
large, rotating arm-like structure that houses and moves the radiation-producing components around the patient during radiation therapy
Rotates 360 deg.
AKA of bed
bed, table, couch
how far does the gantry rotate vs collimator vs floor
gantry + collimator = both 360 deg.
floor = 180 deg.
how does the couch move?
move laterally. longitudinally, verticaly, rotation (along floor) (some beds allow for pitch and roll tilting ⟹ “6-degrees of freedom”)
6 degrees of freedom
bed moving laterally, longitudinally, vertically, rotation, + pitch (head/feet rise), and roll (R/L tilt)
onboard kV (CBCT) imaging + location on the linac
Onboard kV (CBCT) imaging: source panel/detector
Kv produces diagnostic level xrays for pictures (orthogonal pictures)
located — on the left and right side of the gantry (hands that come out as the source and/or detector
orthogonal pictures
two 2D X-ray projection images (can be taken in AP or lateral view, 90 deg. angle from each other) — moving machine
CBCT
cone beam computed tomography
taking pictures while rotating kV image 360deg.
MV imaging panel + location
AKA “portal imager” taking pictures of the treatment itself
located — south of the gantry
hand pendant
handheld control device to control linac/couch/positioning
mechanical isocentre
invisible point in the room where everything is centered and everything rotates relative to this point
collimator, gantry, bed etc all move relative to the mechanical isocentre
radiation isocentre
point in space where the radiation beam intersect when the gantry is rotated and the beam is on
isocentre
basically the same place of where the mechanical and radiation isocentre is located
CAX
central axis
SSD
source-skin distance
SAD
source-axis distance (axis = isocentre in pt) — ū ~100cm
SID
source-imager distance
TTH
tabletop height (isocenter to flat top of table)
separation
total thickness of pt’s body from entrance of radiation beam to the exit point of radiation
laser
thin beams of light that intersect at the isocentre
acting as visual reference for the isocentre
used to position pt and accurately align the isocentre in the body
what color are the lasers typically
red or green
where are the lasers in the linac room
Ceiling laser: used when gantry is rotated, bisects pt into R/L planes AND superior-inferior plans
Sagittal laser: shine from foot of bed and splits pt into R/L planes
right/left laser: bisect into SUP-INF AND ANT-POST
ceiling lasers
replaces the gantry’s crosshair when its rotated in different angles (thats not zero’d)
is the linac room light or dark
Ū light (esp during cleaning, marking tattoos entry/exit), but during treatment = dim lighting to see field and lasers for setup