⭐*The position-indicating device (PID) is an extension of the a. tubehead. b. extension arm. с. control panel. d. exposure button,
a. tubehead.
*____ regulate(s) the manufacture and installation of dental x-ray equipment. a. The federal government b. State governments с. Local governments d. Municipal governments
a. The federal government
*The ___ allow(s) for positioning of the tubehead. a. control devices b. extension arm c. control panel d. exposure button
b. extension arm
⭐*You have your patient in the chair, you have placed the receptor in your patient's mouth, and you have positioned the position-indicating device. Which of the following should you do next to take the x-rays? a. Turn on the on-off switch. b. Push the exposure button. с. Turn on the exposure light. d. Adjust the control devices.
b. Push the exposure button.
⭐*Which of the following statements is true of the receptor holder? a. It is used to align an extraoral dental x-ray receptor. b. It requires the patient to stabilize the receptor in the mouth. C. It is required when using the intraoral paralleling technique. d. It is required when using the intraoral bisecting technique.
C. It is required when using the intraoral paralleling technique.
⭐*A beam alignment device can be used to help the dental radiographer position the PID in relation to the ___ and the receptor. a. control panel b. extension arm C. collimator d. tooth
d. tooth
*During an exposure a) The exposure button is briefly depressed and then released to initiate the exposure. b) The exposure button is firmly depressed until the preset exposure time is completed. c) A beep sounds to signal initiation of the exposure. d) The exposure light on the control panel is illuminated to signal the completion of the exposure.
b) The exposure button is firmly depressed until the preset exposure time is completed.
⭐*Infection control is critical in your dental operatory because pathogens can be transmitted easily through which of the following
your patient to you.
you to your patient.
from one patient to another patient. Question options: a) 1, 3 b) 1, 2 c) 2, 3 d) 1, 2, 3
d)1, 2, 3
⭐*(1) Highly resistant bacterial and mycotic (fungal) spores are not killed during disinfection procedures; (2) however, highly resistant bacterial and mycotic spores are killed during sterilization. Question options: a) The first statement is true; the second statement is false. b) Both statements are true. c) Both statements are false. d) The first statement is false; the second statement is true.
b) Both statements are true.
⭐*Critical instruments are defined as instruments a) or devices that do not come in contact with mucous membranes. b) that contact but do not penetrate soft tissue. c) that are used to penetrate soft tissue or bone. d) that contact but do not penetrate bone.
c) that are used to penetrate soft tissue or bone.
⭐*Which of the following are considered to be semicritical instruments? a) The x-ray control panel b) The exposure button c) The lead apron d) Receptor-holding devices
d) Receptor-holding devices
*Covering exposed surfaces with disposable materials ____________ adequate protection ____________ the need for surface cleaning and disinfection between patients. a) does not provide; but does eliminate b) neither provides; nor eliminates c) provides; while eliminating d) provides; but does not eliminate
c) provides; while eliminating
*Who among the following categories of dental professionals is qualified to mount dental radiographs?
Dentist
Dental hygienist
Dental assistant Question options: a)2, 3 b)1 only c)1, 2 d)1, 2, 3
d)1, 2, 3
*It is desirable that radiographs be identified with each of the following pieces of information, but it is critical that they be labeled with the Question options: a)radiographer's name. b)number of films. c)patient's name and date of exposure. d)dentist's name.
c)patient's name and date of exposure.
*In the labial mounting method, radiographs are placed in the film mount with the __________ side of the identification dot facing the viewer. They are then viewed from the _________ aspect. a)depressed; labial b)depressed; lingual c)raised; labial d)raised; lingual
c)raised; labial
*Two indicators that will help you in determining which way to mount bite-wing radiographs are (1) the curve of Spee curves _______________ toward the distal, and (2) mandibular molars have __________ distinct roots. Question options: a)upward; two b)downward; three c)downward; two d)upward; three
a)upward; two
*Which one of the following is the only member of the dental team who is legally allowed to interpret and diagnose radiographs? Question options: a)Dentist, dental hygienist or dental assistant b)Dentist c)Dentist or dental hygienist d)Dentist or dental assistant
b)Dentist
*Which of the following would you place in the patient's mouth in order to take dental x-rays? A. Image B. Image receptor C. Dental radiograph D. Radiograph
B. Image receptor
⭐*Recommendations concerning gloves would fall under which of the following categories of infection control practices that directly relate to dental imaging procedures? A.Personal protective equipment B.Hand washing and care of hands C.Sterilization or disinfection of instruments D.Cleaning and disinfection of dental unit and environmental surfaces
A.Personal protective equipment (PPE)
⭐*An oral examination limits the practitioner to knowledge of what is seen clinically. Dental images allow the practitioner to see many conditions that are not apparent clinically. A.Both statements are true. B.Both statements are false. C.The first statement is true; the second statement is false. D.The first statement is false; the second statement is true.
A.Both statements are true.
⭐*Added filtration in the dental x-ray tubehead A.refers to the placement of tungsten discs in the path of the x-ray beam between the collimator and the tubehead seal. B.filters out shorter-wavelength x-rays from the x-ray beam. C.results in a lower-energy beam. D.results in a more penetrating useful beam.
D.results in a more penetrating useful beam.
⭐*In order to avoid the primary beam while taking dental images, you should stand at a ____-degree to ____-degree angle to the beam. A.0; 45 B.45; 90 C.90; 135 D.135; 180
C.90; 135
*Through the use of dental images, the dental radiographer can detect ________ that cannot be detected clinically.
diseases
lesions
conditions of teeth and bones A.1, 2, 3 B.1, 2 C.2, 3 D.3 only
A.1, 2, 3
⭐*Which of the following should you do prior to taking x-rays to protect your patient from radiation exposure? A.Proper film processing B.Proper prescribing of radiographs C.A lead apron D.A thyroid collar
B.Proper prescribing of radiographs
*In dental imaging, no __________________ are used.
critical instruments
semicritical instruments
noncritical instruments A.1, 2, 3 B.1, 2 C.3 only D.1 only
D.1 only
⭐*On some dental x-ray machines, only the ____________ can be altered; the other parameters are preset by the manufacturer. A.kilovoltage B.exposure time C.PID length D.milliamperage
B.exposure time
⭐*According to the text, __________ is one of the most important uses of dental images. A.habitual rote frequency B.detection C.maintaining a routine D.insurance verification
B.detection
⭐*Dental images are a(n) _______________ comprehensive patient care. A.adjunct to B.necessary component of C.option for D.supplement to
B.necessary component of
*Preparation of supplies and equipment involves sterilizing which of the following items? A.Film B.Film-holding devices C.Lead apron D.PID
B.Film-holding devices
⭐*(1) You can use dental images to help educate your patient on his dental caries and periodontal disease. (2) You can also use dental images to help your patient see a picture of common conditions that cannot be seen clinically. These include cysts, extra teeth, and retained roots. A.Both statements are true. B.Both statements are false. C.The first statement is true, and the second statement is false. D.The first statement is false, and the second statement is true.
A.Both statements are true.
⭐*(1) It is your responsibility to maintain competence in taking dental images; taking continuing education courses is one example. (2) You should always do your best to properly position and expose receptors in a manner that is accurate, time efficient, and comfortable for your patient. A.Both statements are true. B.Both statements are false. C.The first statement is true, and the second statement is false. D.The first statement is false, and the second statement is true.
A.Both statements are true.
⭐*The thyroid collar A.is recommended for all intraoral exposures. B.is recommended for all extraoral exposures. C.is placed over the patient's chest and lap and extends to protect the thyroid. D.is recommended for both intraoral and extraoral exposures.
A.is recommended for all intraoral exposures.
⭐*Another term for infection control is __________, which means the absence of pathogens, or disease-causing microorganisms. A.antiseptic B.antibiotic C.antiinfective D.asepsis
D.asepsis
⭐*State and federal laws require that dental x-ray machines operating above 70 kVp have a minimum total (inherent plus added filtration) of ____ mm of aluminum filtration. A.1.5 B.2.0 C.2.5 D.5.0
C.2.5
⭐*There are three conditions that must be present for an infection to occur. Which of the following conditions is a necessary component for an infection to occur? A.Small amounts of the infectious disease with low infectivity B.A host that is not vulnerable to infection C.A way for the infectious disease to enter the host D.An aseptic environment
C.A way for the infectious disease to enter the host
⭐*The ALARA concept states that all radiation must be kept A.as long as readily achievable. B.as low as reasonably allowable. C.as low as reasonably achievable. D.as long as reliably achievable.
C.as low as reasonably achievable.
*Inherent filtration in the dental x-ray tubehead A.includes filtration that takes place when the primary beam passes through the glass window of the x-ray tube, the insulating oil, and the tubehead seal. B.includes the placement of aluminum discs in the path of the x-ray beam between the collimator and the tubehead seal. C.alone meets the standards regulated by state and federal law. D.is equivalent to approximately 5.0 mm of aluminum.
A. includes filtration that takes place when the primary beam passes through the glass window of the x-ray tube, the insulating oil, and the tubehead seal.
*A circular collimator produces a cone-shaped beam that is __________ inch(es) in diameter. A.<1 B.1.25 C.2.75 D.5.0
C.2.75
*A rectangular collimator A.restricts the beam 3 times more than a circular collimator. B.restricts the beam less than a circular collimator. C.decreases patient exposure considerably. D.is considerably larger than a size 2 intraoral receptor.
C.decreases patient exposure considerably.
*Which of the following types of position-indicating devices (PIDs) produces scatter radiation? A.Rectangular B.Conical C.Round D.Square
B.Conical
*Which type of PID would be most effective in reducing patient exposure? A.Conical B.A 16-inch round PID C.An 8-inch rectangular PID D.A 16-inch rectangular PID
D.A 16-inch rectangular PID
*Which of the following can you do after x-ray receptor exposure to reduce(s) the amount of x-radiation your patient receives? A.Proper placement of the lead apron B.Proper image retrieval C.Proper instrument technique D.Selecting proper exposure settings
B.Proper image retrieval
*According to the current recommendations (2003) of the National Council on Radiation Protection and Measurements, the current MPD for occupationally exposed persons is ___________/year. A.500 millirem B.5.0 rem C.50 rem D.5000 rem
B.5.0 rem
*You are 36 years old. What is your cumulative occupational dose recommended by the National Council on Radiation Protection and Measurements (NCRP)? A.360 millirems B.360 mSv C.180 millirems D.180 mSv
B.360 mSv
*Your patient is a 5-year-old boy who you have been seeing since he was 3 years old. He is always eating candy and drinking cola beverages. Six months ago he had A, J, and T restored. One year ago he had B, I, and L restored. Which of the following should you do regarding taking radiographs? A.Take bite-wings and occlusal radiograph today. B.Insurance covers bite-wings once a year, take the dental images in 6 months. C.Re-evaluate in 1 year for the need for dental images. D.Dental images are taken every 2 years.
A.Take bite-wings and occlusal radiograph today.
*Your next patient is a 35-year-old man who you have been seeing for the past 10 years. The only treatment he has needed since coming to the dental practice has been his 6-month "prophys." He has no periodontal issues and is on a 6-month recall cycle. You last took four bite-wing images 1 year ago. Which of the following should you do? A.Take bite-wings during this appointment. B.Take the images in 6 months. C.Take the images in 1 to 2 years. D.No images will be needed for 5 years.
C.Take the images in 1 to 2 years.
⭐*When should you avoid taking retakes? A.Whenever it is convenient because the dose is so low B.Only if the patient is uncomfortable with exposure C.Only if older (pre-1975) equipment is used D.At all times
D.At all times
*Which of the following should you do to prepare your treatment room prior to taking a dental image? A.Wipe with an antiseptic. B.Follow infection control procedures. C.Sterilize critical instruments. D.Wipe with a low-level disinfectant
B.Follow infection control procedures.
*Which of the following viruses may be present in oral secretions?
Cytomegalovirus
Hepatitis B virus
Herpes simplex virus
Human immunodeficiency virus 1, 2, 3 1, 2, 3, 4 2, 3, 4 2, 4
1, 2, 3, 4
*Which of the following intraoral receptor holders is a disposable Styro- foam bite-block? a. EEZEE-Grip b. Stabe bite-block с. EndoRay d. Uni-bite
B
*Features of the Stabe bite-block include a. a double-ended instrument. b. a slot for receptor retention. C. a molded plastic that can be sterilized. d. a receptor that is held between two serrated plastic grips.
b. a slot for receptor retention.
*Under which of the following conditions must you wear your mask and safety glasses? Question options: a)Only during surgical procedures b)Whenever spatter and aerosolized sprays of blood and saliva are likely c)Whenever the high-speed handpiece is used d)If desired
b)Whenever spatter and aerosolized sprays of blood and saliva are likely
⭐*Which of the following surfaces on the x-ray machine must be covered or disinfected?
Control panel
Exposure button
Tubehead
Position-indicating device (PID) Question options: a)2 only b)1, 2, 3, 4 c)2, 3, 4 d)1, 2, 3
b)1, 2, 3, 4
⭐*After seating the patient, the radiographer must complete which of the following procedures before washing the hands and putting on gloves?
Chair adjustment
Headrest adjustment
Placement of the lead apron Question options: a)1, 2, 3 b)2, 3 c)3 only d)1, 2
a)1, 2, 3
*Which of the following items should be removed by the patient during preparation for dental image procedures?
Eyeglasses
Dentures
Chewing gum Question options: a)1, 3 b)1, 2, 3 c)2, 3 d)2 only
b)1, 2, 3
*Radiographs are intended to be placed in a film mount/holder in Question options: a)anatomic order. b)the order in which they were processed. c)the order prescribed by the American Dental Association (ADA) nomenclature. d)the order in which they were exposed.
a)anatomic order.
⭐*When dental images are properly prescribed, exposed, and processed, the benefit of disease detection ____________ the risk of small doses of x-radiation. A.does not outweigh B.is considered equal C.slightly outweighs D.far outweighs
D.far outweighs
*The collimator ____________ A.is always round. B.restricts the size and shape of the x-ray beam. C.is a solid piece of aluminum. D.is fitted within the copper stem beneath the molybdenum cup.
B.restricts the size and shape of the x-ray beam.
⭐*The Rinn XCP is a A. beam alignment device. B. Styrofoam bite-block. C. snap-on metal collimating device. D. Uni-bite.
A. beam alignment device.
*A film mount may be made of
cardboard.
lead foil.
plastic.
vinyl.
1, 3, 4
*What does filtration result in?
A higher energy and more penetrating useful beam.
*What do machines operating at or below 70 kVp require?
A minimum total of 1.5mm aluminum filtration.
⭐*What is the fastest speed film on the market?
Fast speed or F-Speed (Kodak Insight)
*How far should the radiographer be from the tube head during exposure?
6 feet.
*What should the radiographer do while taking radiographs?
Leave the room or take a position behind a suitable barrier or wall.
*What does ALARA stand for?
As Low As Reasonably Achievable
⭐*What is a pathogen?
A microorganism that causes disease.
*For an infection to occur by one of the modes of transmission, the following conditions must be present:
Susceptible host
*Noncritical Instruments
Don't come in contact w/mucous membranes. Examples are PID, exposure button, tubehead.
*Intermediate-Level Disinfectants
EPA regulated chemical germicides labeled as both hospital disinfectants and tuberculocidals.
*Preparation of the patient before exposure include?
Chair Adjustment
⭐*What is the position indicating device or PID?
An extension of the x-ray tube head used to direct the x-ray beam.
*What does the beam alignment device do?
Stabilizes the receptor in the mouth and reduces the chance for movement. Also eliminates the need for the patient to hold the receptor in position.
*On most units the kVp and mA are what?
Preset by the manufacturer and cannot be adjusted. Some unites only allow for adjustment of the exposure time due to manufacturer presets.
*What is the primary benefit of dental x-rays?
The detection of disease, helps to identify and prevent problems as well as minimize future problems.
*What is a dental radiographer?
Any person who positions, exposes, and processes dental x-ray image receptors.
*What are the 3 modes of transmission for infection?
Patient to Dental Professional
*Preparation of the dental radiographer before exposure include?
Hand hygiene
*Where do you place barrier devices in the radiography operator?
On radiography equipment, control panel, tube head, tape on mouse, exposure buttons and a bag on headrest. Place plastic wrap on computer monitor and keyboard.
*What should you do with beam alignment devices when finished?
Sterilize them, they go in the autoclave.
*What should you do with the wired sensors?
Wipe with CaviWipe, and do not submerge in cold sterile solution. When in use you MUST cover with plastic sheath.
*T/F: Pointed cones should not be used because of increased scatter radiation.
TRUE
*what material is used in a collimator?
Lead
*which term describes the dose of radiation that the body can endure with little or no chance of injury?
Maximum permissable dose
*No federal standards existed for dental x-ray machines manufactured before the year …
1974
*T/F: the federal government dictates how dental x ray equipment is used.
FALSE: only state and local governement can dictate how it is used
*T/F State and local governments dictate codes that pertain to the use of dental x-ray equipment
TRUE
*T/F: with a portable dental x ray unit, operator exposure is limited by using a lead acrylic disk shield around P.I.D
TRUE
*T/F: a dental image contains less information than a written record
FALSE
*T/F: the dental radiographer may be assigned to monitor and implement quality control procedures
TRUE
*T/F Operator competence is maintained by repeatedly performing dental imaging duties.
False
*what is the primary purpose of infection prevention?
to prevent the transmission of infectious diseases
*list the three possible routes for disease transmission
1. direct contact with pathogen
2. indirect contact with pathogen
3. direct contact with airborne contaminants
*EPA registered chemical germicides labeled as both disinfectants and tuberculosidal agents are classified as what?
intermediate level disinfectants
*EPA registered chemical germicides labeled only as hospital disinfectants are classified as…
low level disinfectants
*When was the federal radiation control for safety and health act ordered?
1968
*When was the federal consumer patient radiation health and safety act mandated?
1981
*The dental x-ray tubehead must be equipped with
-Aluminum Filters
*total filtration
inherent + added
*What does added filtration do?
Filter out longer wavelengths, lower energy, produces a more useful, penetrating beam.