PULSE Week 4: Instrumentation of MRI

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/32

flashcard set

Earn XP

Description and Tags

Comprehensive vocabulary flashcards covering MRI instrumentation concepts from Week 4, including electromagnetism, magnet types, RF transmission/reception systems, coil configurations, gradient hardware, shimming, and safety.

33 Terms

1
New cards

Faraday's Law of Electromagnetism

A physical law stating that passing an electric current through a coiled wire generates a magnetic field, whose direction is determined by the direction of current flow.

2
New cards
<p>Right Hand Rule</p>

Right Hand Rule

A visual rule demonstrating the relationship between current flow and magnetic field direction, where the thumb points along the electric current and curled fingers (II–V) indicate the magnetic field.

3
New cards

Permanent Magnet (MRI)

An MRI system constructed of many small permanent magnets surrounding a solid iron core, producing a vertical magnetic field with low field strength (~0.4 T0.4\,\text{T}) that cannot be turned off via quenching.

4
New cards

Electromagnet (MRI)

A magnet system built from loops of copper wire powered by a continuous electric power supply, generating a horizontal magnetic field with high field strength that can be quenched in emergencies.

5
New cards

Resistive Magnet

An electromagnet constructed using an iron core encased in copper wiring that can be switched off directly by a power switch, but suffers from high operational power costs and strict water cooling requirements.

6
New cards

Superconducting Magnet

An electromagnet composed of low-resistance coils immersed in a liquid helium cryogen bath at −269∘C-269^\circ\text{C} (4∘K4^\circ\text{K}), delivering high magnetic fields (1.5 T+1.5\,\text{T}+) and high field homogeneity with low power consumption.

7
New cards

Fringe Field

The zone of external magnetic attraction extending beyond the physical bore of the MRI scanner, which increases with higher magnet strength and closer proximity to the magnet.

8
New cards

Gauss

The unit of measurement used to quantify fringe field strength, where 10,000 Gauss=1 Tesla10{,}000\,\text{Gauss} = 1\,\text{Tesla}.

9
New cards

Passive Shielding

A magnetic containment method installed during scanner construction that encases the magnet in a heavy steel shell to absorb the majority of the fringe field.

10
New cards

Active Shielding

Modern magnetic shielding employing opposing bucking coils that generate a counter-current to restrict and minimize the peripheral range of the fringe field.

11
New cards

Transmitter RF Coil

The RF coil located deepest within the MRI structural layers perpendicular to B0B_0, responsible for transmitting B1B_1 radiofrequency excitation pulses to tissue at its specific precessional frequency.

12
New cards

Specific Absorption Rate (SAR) Limit

The FDA regulatory limit restricting RF energy absorption in tissue to 4 W/kg4\,\text{W/kg} and limiting core body temperature elevation to 1∘C1^\circ\text{C}.

13
New cards

Receive RF Coil

A coil selected and attached to the exam table by the MRI technologist to detect and capture the radiofrequency MR echo emitted by relaxing patient tissues.

14
New cards

Transmit-Receive (T/R) Coil

A dual-purpose coil designed to both transmit the RF pulse and receive the MR signal, exciting smaller anatomical areas to significantly lower patient SAR and B1 RMSB_1\,\text{RMS} deposition.

15
New cards

Phase Array Coil

A dedicated receiver coil constructed of two separate jointed sections positioned anterior and posterior to the patient like a sandwich, supporting parallel imaging to increase SNR and shorten scan time.

16
New cards

Volumetric Coil

A dedicated coil engineered to closely encircle specific 3D anatomy (e.g., head, knee, wrist), providing high signal-to-noise ratio and enabling small FOV scanning and motion reduction.

17
New cards

Surface Coil

A flat, multi-element coil suited for large, linear fields of view such as the spine, allowing multiple spinal segments to be scanned without re-landmarking.

18
New cards

Annefact

A glaring image artifact caused when surface coil elements located outside the active FOV remain switched on during scanning.

19
New cards

Radio Frequency Tuning

The pre-scan calibration process adjusting bandwidth and transmit gain (TG, typically 140–180140\text{--}180) to match patient morphology and tissue dielectric load for optimal signal transmission.

20
New cards

Gradient Coils

Conductive copper wire loops through which electric currents are pulsed to produce intentional, linear variations across the main magnetic field for spatial encoding.

21
New cards

Time-Varying Magnetic Field (TVMF) Limit

The FDA safety boundary defined not by a numerical value, but by the physical threshold at which the patient begins to experience peripheral nerve stimulation.

22
New cards

X-Gradient

The physical gradient coil that produces a linear field slope along the patient's left-to-right axis, coding for sagittal slice selection.

23
New cards

Y-Gradient

The physical gradient coil that creates a linear field variation along the patient's anterior-to-posterior axis, selecting the coronal imaging plane.

24
New cards

Z-Gradient

The physical gradient coil that establishes a linear magnetic slope from inferior to superior along the bore, selecting the axial imaging plane.

25
New cards

Gradient Amplitude

The strength or magnitude of electric current applied to a gradient coil to generate the linear field variation necessary for spatial localization.

26
New cards
<p>Rise Time and Fall Time</p>

Rise Time and Fall Time

The duration required for a gradient to ramp up to its full maximum amplitude (Rise Time) and the duration it takes to ramp down completely back to zero (Fall Time).

27
New cards

Slew Rate

A measure of gradient performance defined as the gradient amplitude multiplied by rise time, expressed in millitesla per meter per second (mT/m/s\text{mT/m/s}).

28
New cards

Duty Cycle

The percentage (%\%) of time during one repetition period (TR\text{TR}) that a gradient coil is actively energized at amplitude.

29
New cards

Magnetic Field Homogeneity

The degree of static magnetic field (B0B_0) uniformity across the isocenter of the magnet, measured in parts per million (ppm).

30
New cards

Shimming

The adjustment process using mechanical iron plates or active electrical coils to eliminate magnetic field inhomogeneities caused by scanner materials or room surroundings.

31
New cards

Passive Shimming

A permanent shimming technique performed during magnet installation by mechanically inserting ferromagnetic steel trays to compensate for static field defects.

32
New cards

Active Shimming

Dynamic shimming carried out by the MRI technologist via the operator console, energizing electromagnetic shim coils with current to balance the field prior to scanning.

33
New cards

Magnet Quench

The rapid boiling off and venting of liquid cryogen (helium) from a superconducting magnet, carrying serious risks of chamber asphyxiation and eardrum rupture from room pressurization.