A technologist performs an examination using AEC. If patient thickness increases while kVp and mA remain unchanged, what will typically happen?
Exposure time will decrease
Exposure time will remain unchanged
Exposure time will increase
The system will automatically increase kVp
Exposure time will increase
A technologist increases kVp while using AEC, keeping all other selections unchanged. What is the expected result?
Increased exposure time
Decreased exposure time
Increased mAs
No change in exposure time
Decreased exposure time
A technologist increases mA from 200 to 400 while using AEC. Assuming all other factors remain constant and the system functions properly, what will typically occur?
Exposure time will decrease, while total mAs remains approximately the same
Exposure time will increase, and total mAs will double
Exposure time will remain unchanged
The AEC system will automatically decrease kVp
Exposure time will decrease, while total mAs remains approximately the same
What is the primary purpose of the AEC backup timer?
To ensure the patient receives the same dose during every examination
To automatically adjust the focal spot size
To increase exposure time for larger patients
To prevent an excessively long exposure if AEC fails to terminate it appropriately
To prevent an excessively long exposure if AEC fails to terminate it appropriately
A technologist performs an AEC exposure, but the selected detector is positioned beneath an air-filled region rather than the intended anatomy. What is the most likely result?
Increased exposure time
Increased mAs
Premature exposure termination
Automatic activation of the backup timer
Premature exposure termination
What is the purpose of exposure adjustment controls (density compensation) when using AEC?
To change the selected kVp automatically
To adjust the amount of radiation required at the AEC detector before exposure termination
To change the size of the selected AEC detector
To automatically adjust SID
To adjust the amount of radiation required at the AEC detector before exposure termination
A technologist notices that the exposure indicator consistently indicates insufficient detector exposure when using AEC with proper positioning and technique selection. Which adjustment may be appropriate after confirming the system's EI guidance?
Select a negative exposure adjustment setting
Decrease SID without evaluating the examination setup
Select a positive exposure adjustment setting, if supported by the equipment protocol
Increase image brightness using postprocessing
Select a positive exposure adjustment setting, if supported by the equipment protocol
A technologist uses AEC for an abdominal examination. A dense radiopaque object lies directly over the selected AEC detector. What is the most likely effect?
The exposure may continue longer, increasing mAs and patient exposure
The exposure will terminate earlier because the object absorbs more radiation
The AEC system will automatically lower kVp
The object will have no effect on AEC operation
The exposure may continue longer, increasing mAs and patient exposure
Which statement best explains why AEC does not eliminate the need for technologist judgment?
AEC controls all exposure factors, including kVp and SID
AEC automatically corrects positioning and collimation errors
AEC determines the appropriate anatomical projection
AEC depends on proper detector selection, positioning, and appropriate technique settings
AEC depends on proper detector selection, positioning, and appropriate technique settings
What is the primary purpose of Anatomically Programmed Radiography (APR)?
To automatically terminate exposures based on detector radiation levels
To provide preset exposure factors based on the selected examination and projection
To eliminate the need for patient positioning
To automatically correct motion artifacts
To provide preset exposure factors based on the selected examination and projection
A technologist selects a PA chest examination using APR. Which combination of settings might the system automatically suggest?
Low kVp, long exposure time, and center AEC detector only
High mAs and long exposure time for every patient
High kVp, short exposure time, and outer AEC detectors
The same exposure factors used for an AP abdomen
High kVp, short exposure time, and outer AEC detectors
How do APR and AEC commonly work together?
APR selects preset exposure settings and AEC terminates the exposure based on radiation reaching the selected detector
APR determines exposure time while AEC selects patient positioning
Both systems independently determine image brightness during postprocessing
APR eliminates the need for AEC
APR selects preset exposure settings and AEC terminates the exposure based on radiation reaching the selected detector
A technologist performs an AP abdomen using AEC. The exposure indicator shows substantially higher detector exposure than intended. The patient was properly positioned, the correct AEC detector was selected, and kVp was appropriate. The technologist discovers that the exposure adjustment control was set to +2 instead of 0. What is the most appropriate explanation?
The positive adjustment caused AEC to terminate the exposure sooner
The positive adjustment increased the radiation required at the detector before terminating the exposure
The positive adjustment automatically increased kVp
The positive adjustment decreased the total mAs
The positive adjustment increased the radiation required at the detector before terminating the exposure
A technologist performs an AP lumbar spine examination using AEC. The first exposure produces an acceptable exposure indicator. For the second exposure, the technologist increases mA from 200 to 400 while maintaining the same kVp, positioning, and detector selection. Which outcome should the technologist expect?
The exposure time will decrease by approximately half, and total mAs will remain approximately the same
The exposure time will remain unchanged, and mAs will double
The exposure time will double, and receptor exposure will increase
The system will automatically decrease kVp to compensate
The exposure time will decrease by approximately half, and total mAs will remain approximately the same
A technologist performs an AP abdomen examination using AEC. The patient is not properly centered, causing the selected AEC detector to be partially uncovered by the anatomy. What is the most likely result?
The exposure time will increase because the detector receives less radiation
The exposure may terminate prematurely, resulting in insufficient receptor exposure
The system will automatically adjust the patient's positioning
The AEC system will automatically increase kVp
The exposure may terminate prematurely, resulting in insufficient receptor exposure
The software function responsible for evaluating the collimated borders of an exposed imaging plate is called:
algorithmic application
exposure field recognition
image equalization
exposure field recognition
Ideally, with digital imaging, the field size must be large enough so that at least 30% of the image receptor is exposed.
True
False
True
To avoid image mottle, expose all digital images without a grid whenever possible.
True
False
False
When creating exposure factors for a digital imaging system, which of the following factors no longer needs to be considered?
body habitus of the patient
ALARA
optimal kVp settings
all of the above must still be considered
all of the above must still be considered
If a technologist fails to properly align the exposure field with a digital image receptor, what undesirable results are likely to occur?
Underexposure and mottled
overexposure and saturation
histogram errors and sub-optimal exposure index values
histogram errors and sub-optimal exposure index values
One of the problems identified that technologist face with digital imaging systems and their technique selection is that:
visual cues indicating over or underexposure are no longer present on the image
the image can be easily sent to remote locations
kVp manipulation is severely limited compared to film/screen systems
automatic exposure control systems cannot be used in tandem with digital systems
visual cues indicating over or underexposure are no longer present on the image
The 15% rule can be used to adjust technique for changed patient thickness.
True
False
True
Digital imaging systems are inherently mAs driven.
True
False
False
It is recommended that all exposures be made using a minimum of 70 kVp for digital imaging.
True
False
False
The practice of using higher exposure factors than needed to obtain a diagnostic radiograph is commonly referred to as:
ALARA
Rounding up
Dose creep
Saturation
Dose creep
Which of the following statements is TRUE regarding kVp’s role when using digital imaging systems?
decreasing kVp significantly increases image contrast
kVp ensures adequate penetration of the anatomy
decreasing kVp decreases image brightness
80 kVp is now the recommended setting for virtually all exams
kVp ensures adequate penetration of the anatomy
The total range of exposure intensities (signals) that a digital detector can capture and display is:
creep
dynamic range
stabilization
compensation
dynamic range
When an image is said to be saturated, this means that:
it has had inadequate time to dry
image noise has exceeded an optimal level
excessive overexposure has occurred
the backup timer engaged during an AEC exposure
excessive overexposure has occurred
The factors responsible for the phenomena of dose creep when digital imaging systems are used include:
fear of mottle on the image
no visible consequences for overexposure
easy ability to take a repeat image
1, 2 & 3
1 & 3
2 & 3
1, 2 & 3
1, 2 & 3
Exposure field recognition errors could occur in which circumstance outlined below?
using more than 30 kVp than recommended for the part
imaging anatomy diagonally or not aligned with the edges on an imaging plate
setting an inadequate backup time during AEC use
activating all three cells during an AEC exposure
imaging anatomy diagonally or not aligned with the edges on an imaging plate
Which of the following is considered best practice with regard to setting techniques and or evaluating images when digital imaging systems are used?
increase mAs by two times for the next exposure if the image appears to bright on the monitor
to avoid a repeat exposure due to image mottle, always increase mAs
do not repeat an overexposed image if post-processing can be used
avoid collimating to ensure that all parts of the IR or detector array received remnant radiation
do not repeat an overexposed image if post-processing can be used
In digital imaging, ALL technique corrections and mottle prevention should only be made by changing mAs.
True
False
False
Which of the following is not a pre-processing step?
Normalization
Exposure field recognition
Histogram analysis
Windowing
Windowing
A segmentation failure can occur if:
The anatomy is too small
The image is not centered or is asymmetrical
The SID is too long
The histogram is too narrow
The image is not centered or is asymmetrical
To prevent segmentation errors, technologists should:
Take multiple exposures per IR
Collimate to less than 10% of the IR
Expose only one centered image per IR
Use the lowest possible kVp
Expose only one centered image per IR
For best practice, exposures must cover at least what percentage of the IR?
10%
20%
30%
50%
30%
Exposure field recognition failure is most likely when:
Exposure is made without collimation
Collimation is too tight
SID is increased
The LUT is incorrect
Exposure is made without collimation
VOI (Values of Interest) refers to:
The lowest and highest intensity values used for image display
The entire histogram range
The background radiation levels
The number of images on a plate
The lowest and highest intensity values used for image display
The purpose of a Look-Up Table (LUT) is to:
Recalculate SID
Record patient dose
Eliminate scatter
Compare the exposure histogram to a reference and adjust pixel values
Compare the exposure histogram to a reference and adjust pixel values
Automatic rescaling occurs when:
Exposure field borders are not visible
The mAs is doubled
The computer adjusts pixel values to match the reference histogram
The histogram is deleted
The computer adjusts pixel values to match the reference histogram
Selecting the wrong body part or LUT may result in:
Increased SNR
Improper brightness and contrast
Reduced patient dose
Lower image resolution only
Improper brightness and contrast
Which factor can distort a histogram?
Proper collimation
Adequate centering
Presence of a prosthesis
Correct exposure
Presence of a prosthesis
To limit histogram errors, the technologist should:
Use AEC correctly and center anatomy
Use double exposures
Over-collimate
Reduce SID below 20”
Use AEC correctly and center anatomy
Which of the following is a post-processing tool?
Histogram analysis
Edge enhancement
Exposure field recognition
Segmentation
Edge enhancement
A digital radiograph is acquired with the anatomy positioned near the edge of the detector rather than centered. Which problem is most likely to occur during pre-processing?
The system may incorrectly identify the exposure field, resulting in improper image processing.
The detector will automatically compensate for the positioning error by shifting the anatomy
The image will demonstrate increased spatial resolution because less detector area was exposed.
The system will increase the exposure index to compensate for the off-center positioning.
The system may incorrectly identify the exposure field, resulting in improper image processing.
During digital image processing, which sequence best represents the major steps of pre-processing?
Histogram analysis → Segmentation → Normalization → Exposure field recognition
Segmentation → Exposure field recognition → Histogram analysis → Normalization
Exposure field recognition → Normalization → Segmentation → Histogram analysis
Normalization → Histogram analysis → Exposure field recognition → Segmentation
Segmentation → Exposure field recognition → Histogram analysis → Normalization
A technologist fails to collimate appropriately during a digital radiographic examination. How might this affect the final image?
The system will automatically remove all scatter without affecting the histogram.
The increased exposure field will improve the accuracy of the exposure index.
The computer may include unwanted pixel data, producing incorrect brightness, contrast, or exposure index values.
The detector will automatically decrease its sensitivity to prevent histogram errors
The computer may include unwanted pixel data, producing incorrect brightness, contrast, or exposure index values.
A radiographic image has adequate receptor exposure but demonstrates incorrect brightness and contrast after processing. Which explanation is most consistent with a histogram analysis error?
The exposure time was too short for the computer to create a histogram.
The detector was unable to convert x-ray energy into an electronic signal.
The system correctly identified the anatomy but permanently lost all pixel data.
The software incorrectly identified the values of interest and applied inappropriate grayscale processing.
The software incorrectly identified the values of interest and applied inappropriate grayscale processing.
Which technologist action would best reduce the likelihood of exposure field recognition errors?
Center the anatomy within the properly collimated exposure field.
Increase mAs to ensure the software receives sufficient exposure data.
Use the largest available field size to prevent the loss of anatomical information.
Rely on post-processing to correct improperly identified exposure borders.
Center the anatomy within the properly collimated exposure field.
A histogram displays a large spike caused by pixel values outside the anatomy of interest. If these values are incorrectly included in the analysis, what is the most likely result?
The system will increase the number of pixels in the detector matrix.
The histogram may become skewed, affecting the grayscale representation of the anatomy.
The spatial resolution will improve because a greater range of pixel values is analyzed.
The detector will automatically decrease the exposure received by the patient.
The histogram may become skewed, affecting the grayscale representation of the anatomy.
During histogram analysis, the computer identifies the values of interest (VOI). What is the primary purpose of this process?
To determine whether the radiograph was acquired using an appropriate SID.
To calculate the total radiation dose absorbed by the patient.
To identify the range of pixel values representing the anatomy of interest while excluding irrelevant data.
To modify the original x-ray exposure before image acquisition is completed.
To identify the range of pixel values representing the anatomy of interest while excluding irrelevant data.
A technologist notices that a digital image has an incorrect exposure index even though the selected exposure factors were appropriate. Which situation could explain this finding?
The software incorrectly identified the collimated exposure field and included inappropriate data in its analysis.
The technologist selected a higher kVp, which always causes an inaccurate exposure index.
The radiograph was displayed using magnification, permanently altering the detector exposure.
The technologist adjusted the window level after acquisition, changing the original detector signal.
The software incorrectly identified the collimated exposure field and included inappropriate data in its analysis.
Two radiographs of the same anatomy are acquired using slightly different exposure factors, yet both demonstrate similar displayed brightness and contrast. Which process best explains this result?
The detector automatically adjusts the radiation output before each exposure.
The computer modifies the original exposure factors after image acquisition.
Automatic rescaling and LUT processing adjust pixel values to produce a consistent image appearance.
The computer increases the number of pixels to compensate for differences in exposure.
Automatic rescaling and LUT processing adjust pixel values to produce a consistent image appearance.
A technologist selects the incorrect anatomical examination before acquiring a digital radiograph. Why might this affect the processed image?
The system will automatically reduce the x-ray beam intensity to match the selected anatomy.
The detector will change its physical pixel size to accommodate the incorrect exam.
The computer will be unable to record any exposure information from the detector.
An inappropriate examination-specific LUT may be applied, resulting in incorrect grayscale mapping.
An inappropriate examination-specific LUT may be applied, resulting in incorrect grayscale mapping.
During automatic rescaling, the acquired exposure histogram is compared with a reference histogram. What occurs when the two do not match?
Pixel values are adjusted to better match the expected grayscale representation of the selected examination.
The original x-ray exposure is repeated automatically using corrected exposure factors.
The detector adds anatomical structures that fall outside the reference histogram.
The computer increases spatial resolution until the acquired and reference histograms match.
Pixel values are adjusted to better match the expected grayscale representation of the selected examination.
Why can digital radiographs appear acceptable despite variations in the selected kVp and mAs?
Digital detectors eliminate the relationship between exposure factors and receptor exposure.
Image processing can maintain a consistent appearance when exposure remains within an acceptable range.
The computer automatically changes the amount of radiation that originally reached the patient.
The LUT eliminates quantum noise regardless of the amount of radiation reaching the detector.
Image processing can maintain a consistent appearance when exposure remains within an acceptable range.
Which statement best describes a limitation of automatic rescaling in digital radiography?
It can correct positioning errors as long as the anatomy remains within the detector.
It can recover anatomical information lost because of insufficient x-ray penetration.
It can maintain consistent brightness and contrast, but it cannot compensate for every exposure or acquisition error.
It prevents exposure index errors by eliminating the need for exposure field recognition.
It can maintain consistent brightness and contrast, but it cannot compensate for every exposure or acquisition error.
A technologist adjusts the window width of a digital radiograph. Which image characteristic is primarily affected?
The amount of radiation originally received by the detector.
The overall brightness of the displayed image.
The physical dimensions of the detector matrix.
The range of grayscale values displayed, affecting image contrast.
The range of grayscale values displayed, affecting image contrast.
A technologist applies edge enhancement to a digital radiograph. What is the intended effect?
To increase the exposure received by the detector and reduce quantum mottle.
To emphasize boundaries between structures, making edges appear sharper.
To increase the number of pixels within the image matrix.
To correct anatomical distortion caused by improper positioning.
To emphasize boundaries between structures, making edges appear sharper.
A digital image appears too bright immediately after processing, but the anatomy is adequately penetrated. Which action is most appropriate before considering a repeat exposure?
Increase mAs and repeat the exposure to produce a darker image.
Increase kVp and repeat the exposure to improve detector sensitivity.
Evaluate the image and use appropriate window settings rather than repeating solely because of its displayed brightness.
Repeat the image using a smaller focal spot to correct the brightness.
Evaluate the image and use appropriate window settings rather than repeating solely because of its displayed brightness.
A technologist routinely increases mAs because digital radiographs continue to appear acceptable even when higher-than-necessary exposures are used. Which concern does this practice illustrate?
Decreased detector dynamic range caused by repeated exposures.
Reduced patient dose resulting from increased exposure latitude.
Improved contrast resolution caused by increased receptor exposure.
Dose creep caused by unnecessarily high exposure factors being masked by digital processing.
Dose creep caused by unnecessarily high exposure factors being masked by digital processing.
A radiographic technique of 75 kVp at 24 mAs is adjusted using the 15% rule. Which new technique would approximately maintain receptor exposure?
86 kVp at 12 mAs.
86 kVp at 24 mAs.
75 kVp at 12 mAs.
64 kVp at 12 mAs.
86 kVp at 12 mAs.
A digital radiograph demonstrates areas of anatomy that appear dark and burned out because excessive exposure exceeded the detector's ability to preserve useful information. What is the most appropriate interpretation?
The image demonstrates quantum mottle caused by insufficient exposure.
The image demonstrates a positioning error that can be corrected through image inversion.
The image may demonstrate detector saturation, resulting in loss of anatomical information.
The image demonstrates normal automatic rescaling and requires no further evaluation.
The image may demonstrate detector saturation, resulting in loss of anatomical information.
When developing a digital technique chart, which approach is emphasized in the lecture as a method of reducing patient exposure while maintaining adequate penetration?
Use the lowest available kVp and increase mAs to compensate.
Use the same exposure factors for all anatomical regions to maintain consistency.
Increase mAs whenever possible because digital systems have wide exposure latitude.
Use an appropriately higher kVp with reduced mAs while considering the examination and patient.
Use an appropriately higher kVp with reduced mAs while considering the examination and patient.
DQE is used:
To calculate the median average value of a central pixel
To measure a phosphor’s ability to convert x-ray energy to an output signal
To determine an image’s exposure index – EI value after exposure
As a measure of the image receptor dose when DR detectors are used.
To measure a phosphor’s ability to convert x-ray energy to an output signal
In general, if an imaging receptor system has a higher DQE:
Higher kVp’s should be used with higher mAs
Image noise will be increased
More mAs should be used
Lower technical factors are required
Lower technical factors are required
The EI values used in all digital imaging systems are uniform among all manufacturers.
True
False
False
The term proposed by the International Electro-technical Commission(IEC) for use across all manufacturers as an exposure indicator system is:
Sensitivity number
DICOMM Value
Exposure quantum
Exposure Indices & Deviation Indices (EI and DI numbers)
Exposure Indices & Deviation Indices (EI and DI numbers)
Using the recommended deviation index system: a DI value of +1represents an exposure to the imaging plate that was:
100% too high
50% too high
26% too high
25% too low
50% too low
26% too high

For this question consider the following:
an IDEAL EI# is 800 and
an OPTIMAL DI# is 0.0
DI# guidelines: any image with DI# below -3.0 or above +3.0 must be repeated.
Exposure: 90kVp @ 80mAs with EI# 1600 and DI# +3.0
What will be your new mAs to achieve an optimal exposure level on your next image?
90kVp @ ?__ mAs
40 mAs
60 mAs
80 mAs
160 mAs
40 mAs

For this question consider the following:
an IDEAL EI# is 800 and
an OPTIMAL DI# is 0.0
DI# guidelines: any image with DI# below -3.0 or above +3.0 must be repeated.
Exposure: 60kVp @ 2.5mAs with EI# 550 and DI# -2.0
What will be your new mAs to achieve an optimal exposure level on your next image?
60kVp @ ?__ mAs
1.4 mAs
2.5 mAs
3.5 mAs
5.0 mAs
3.5 mAs
![<p><span><strong>For this question consider the following:</strong></span></p><ul><li><p><span><strong> an IDEAL EI# is 800 and </strong></span></p></li><li><p><span><strong>an OPTIMAL DI# is 0.0</strong></span></p></li><li><p><span><strong>DI# guidelines: any image with DI# below -3.0 or above +3.0 must be repeated.</strong></span></p></li></ul><p> </p><p>Situation: You completed a Lateral L-Spine using 90kVp @ 50 mAs. But the EI# is 250 and the DI# is -3.0.</p><p>What would be the appropriate change to mAs on the repeated exposure to reach an optimal exposure level?</p><p>90 kVp @<strong> [</strong><span><strong><u>___?_____]</u></strong></span><strong> mAs</strong></p><p><strong><em>Enter the numerical value only for your new mAs</em></strong></p>](https://assets.knowt.com/user-attachments/c983c1d9-0a6e-4dde-8027-eae6742fdc3a.jpg)
For this question consider the following:
an IDEAL EI# is 800 and
an OPTIMAL DI# is 0.0
DI# guidelines: any image with DI# below -3.0 or above +3.0 must be repeated.
Situation: You completed a Lateral L-Spine using 90kVp @ 50 mAs. But the EI# is 250 and the DI# is -3.0.
What would be the appropriate change to mAs on the repeated exposure to reach an optimal exposure level?
90 kVp @ [___?_____] mAs
Enter the numerical value only for your new mAs
75 (with margin: 1)
![<p><span><strong>For this question consider the following:</strong></span></p><ul><li><p><span><strong> an IDEAL EI# is 800 and </strong></span></p></li><li><p><span><strong>an OPTIMAL DI# is 0.0</strong></span></p></li><li><p><span><strong>DI# guidelines: any image with DI# below -3.0 or above +3.0 must be repeated.</strong></span></p></li></ul><p> </p><p> </p><p>Exposure: 70kVp @ 25mAs with EI# 1400 and DI# +1.0. </p><p>What will be your new mAs to achieve an optimal exposure level on your next image?</p><p><strong>70kVp @ [___?___] mAs</strong></p><p><strong><em>Enter the numerical value only for your new mAs</em></strong></p>](https://assets.knowt.com/user-attachments/6cf54247-c245-42ba-a6a7-a133077afb7b.jpg)
For this question consider the following:
an IDEAL EI# is 800 and
an OPTIMAL DI# is 0.0
DI# guidelines: any image with DI# below -3.0 or above +3.0 must be repeated.
Exposure: 70kVp @ 25mAs with EI# 1400 and DI# +1.0.
What will be your new mAs to achieve an optimal exposure level on your next image?
70kVp @ [___?___] mAs
Enter the numerical value only for your new mAs
17 or 17.5 (with margin: 1)

For this question consider the following:
an IDEAL EI# is 800 and
an OPTIMAL DI# is 0.0
DI# guidelines: any image with DI# below -3.0 or above +3.0 must be repeated.
Exposure: 90kVp @ 45mAs with EI# 1400 and DI# +2.0
What will be your new mAs to achieve an optimal exposure level on your next image?
90kVp @ ______ mAs
27 (with margin: 1)
Each digital imaging receptor manufacturer uses the SAME Exposure Indicator numeric system.
So we can simply use 800 as the ideal EI# for imaging plates in all clinical sites.
True
False
False
The method of using the DI# to adjust your exposure is currently still in development, but as a general rule you are best to use this DI# as a starting point for correcting technique errors when repeating an image is needed.
True
False
True
Doubling the mAs for the DIGITAL image receptor will result in increased brightness of the displayed digital image after processing.
True
False
False
Manipulating mAs for DIGITAL imaging affects the IR exposure.
True
False
True
A key advantage of digital imaging compared to film/screen is:
Lower resolution
No need for collimation
Ability to adjust image after exposure
Permanent image appearance
Ability to adjust image after exposure
The histogram represents:
Patient dose
Radiation exposure values received by the detector
Film density range
SID versus mAs relationship
Radiation exposure values received by the detector
Latitude in digital imaging refers to:
The detector’s geographic position
The ability to display shades of gray
The range of exposures that produce a diagnostic image
The degree of collimation
The range of exposures that produce a diagnostic image
Dose-Area Product (DAP) measures:
Detector sensitivity
Image brightness
Patient dose and irradiated field size
Exposure index calibration
Patient dose and irradiated field size
DAP increases when:
Collimation is tighter
mAs is reduced
SID is doubled
Field size increases at same technique
Field size increases at same technique
DAP is measured using:
A chamber at the image receptor
An ionization chamber near the collimator
A photodiode inside the monitor
A dosimeter on the technologist
An ionization chamber near the collimator
In a logarithmic EI scale, a change of 300 EI corresponds to:
Doubling or halving mAs
Doubling kVp
15% change in kVp
Halving kVp
Doubling or halving mAs
In an inversely proportional EI system, a high “S” number means:
Proper exposure
Overexposure
Underexposure
Increased contrast
Underexposure
In a proportional EI system:
Low EI means overexposure
High EI means underexposure
EI changes have no meaning
High EI means overexposure
High EI means overexposure
The Deviation Index (DI) represents:
Total patient dose
The difference between actual and target detector exposure
The brightness of the image
Histogram tail value
The difference between actual and target detector exposure