A technologist performs an examination using AEC. If patient thickness increases while kVp and mA remain unchanged, what will typically happen?

Exposure time will decrease

Exposure time will remain unchanged

Exposure time will increase

The system will automatically increase kVp

Exposure time will increase

2
New cards

A technologist increases kVp while using AEC, keeping all other selections unchanged. What is the expected result?

Increased exposure time

Decreased exposure time

Increased mAs

No change in exposure time

Decreased exposure time

3
New cards

A technologist increases mA from 200 to 400 while using AEC. Assuming all other factors remain constant and the system functions properly, what will typically occur?

Exposure time will decrease, while total mAs remains approximately the same

Exposure time will increase, and total mAs will double

Exposure time will remain unchanged

The AEC system will automatically decrease kVp

Exposure time will decrease, while total mAs remains approximately the same

4
New cards

What is the primary purpose of the AEC backup timer?

To ensure the patient receives the same dose during every examination

To automatically adjust the focal spot size

To increase exposure time for larger patients

To prevent an excessively long exposure if AEC fails to terminate it appropriately

To prevent an excessively long exposure if AEC fails to terminate it appropriately

5
New cards

A technologist performs an AEC exposure, but the selected detector is positioned beneath an air-filled region rather than the intended anatomy. What is the most likely result?

Increased exposure time

Increased mAs

Premature exposure termination

Automatic activation of the backup timer

Premature exposure termination

6
New cards

What is the purpose of exposure adjustment controls (density compensation) when using AEC?

To change the selected kVp automatically

To adjust the amount of radiation required at the AEC detector before exposure termination

To change the size of the selected AEC detector

To automatically adjust SID

To adjust the amount of radiation required at the AEC detector before exposure termination

7
New cards

A technologist notices that the exposure indicator consistently indicates insufficient detector exposure when using AEC with proper positioning and technique selection. Which adjustment may be appropriate after confirming the system's EI guidance?

Select a negative exposure adjustment setting

Decrease SID without evaluating the examination setup

Select a positive exposure adjustment setting, if supported by the equipment protocol

Increase image brightness using postprocessing

Select a positive exposure adjustment setting, if supported by the equipment protocol

8
New cards

A technologist uses AEC for an abdominal examination. A dense radiopaque object lies directly over the selected AEC detector. What is the most likely effect?

The exposure may continue longer, increasing mAs and patient exposure

The exposure will terminate earlier because the object absorbs more radiation

The AEC system will automatically lower kVp

The object will have no effect on AEC operation

The exposure may continue longer, increasing mAs and patient exposure

9
New cards

Which statement best explains why AEC does not eliminate the need for technologist judgment?

AEC controls all exposure factors, including kVp and SID

AEC automatically corrects positioning and collimation errors

AEC determines the appropriate anatomical projection

AEC depends on proper detector selection, positioning, and appropriate technique settings

AEC depends on proper detector selection, positioning, and appropriate technique settings

10
New cards

What is the primary purpose of Anatomically Programmed Radiography (APR)?

To automatically terminate exposures based on detector radiation levels

To provide preset exposure factors based on the selected examination and projection

To eliminate the need for patient positioning

To automatically correct motion artifacts

To provide preset exposure factors based on the selected examination and projection

11
New cards

A technologist selects a PA chest examination using APR. Which combination of settings might the system automatically suggest?

Low kVp, long exposure time, and center AEC detector only

High mAs and long exposure time for every patient

High kVp, short exposure time, and outer AEC detectors

The same exposure factors used for an AP abdomen

High kVp, short exposure time, and outer AEC detectors

12
New cards

How do APR and AEC commonly work together?

APR selects preset exposure settings and AEC terminates the exposure based on radiation reaching the selected detector

APR determines exposure time while AEC selects patient positioning

Both systems independently determine image brightness during postprocessing

APR eliminates the need for AEC

APR selects preset exposure settings and AEC terminates the exposure based on radiation reaching the selected detector

13
New cards

A technologist performs an AP abdomen using AEC. The exposure indicator shows substantially higher detector exposure than intended. The patient was properly positioned, the correct AEC detector was selected, and kVp was appropriate. The technologist discovers that the exposure adjustment control was set to +2 instead of 0. What is the most appropriate explanation?

The positive adjustment caused AEC to terminate the exposure sooner

The positive adjustment increased the radiation required at the detector before terminating the exposure

The positive adjustment automatically increased kVp

The positive adjustment decreased the total mAs

The positive adjustment increased the radiation required at the detector before terminating the exposure

14
New cards

A technologist performs an AP lumbar spine examination using AEC. The first exposure produces an acceptable exposure indicator. For the second exposure, the technologist increases mA from 200 to 400 while maintaining the same kVp, positioning, and detector selection. Which outcome should the technologist expect?

The exposure time will decrease by approximately half, and total mAs will remain approximately the same

The exposure time will remain unchanged, and mAs will double

The exposure time will double, and receptor exposure will increase

The system will automatically decrease kVp to compensate

The exposure time will decrease by approximately half, and total mAs will remain approximately the same

15
New cards

A technologist performs an AP abdomen examination using AEC. The patient is not properly centered, causing the selected AEC detector to be partially uncovered by the anatomy. What is the most likely result?

The exposure time will increase because the detector receives less radiation

The exposure may terminate prematurely, resulting in insufficient receptor exposure

The system will automatically adjust the patient's positioning

The AEC system will automatically increase kVp

The exposure may terminate prematurely, resulting in insufficient receptor exposure

16
New cards

The software function responsible for evaluating the collimated borders of an exposed imaging plate is called:

algorithmic application

exposure field recognition

image equalization

exposure field recognition

17
New cards

Ideally, with digital imaging, the field size must be large enough so that at least 30% of the image receptor is exposed.

True

False

True

18
New cards

To avoid image mottle, expose all digital images without a grid whenever possible.

True

False

False

19
New cards

When creating exposure factors for a digital imaging system, which of the following factors no longer needs to be considered?

body habitus of the patient

ALARA

optimal kVp settings

all of the above must still be considered

all of the above must still be considered

20
New cards

If a technologist fails to properly align the exposure field with a digital image receptor, what undesirable results are likely to occur?

Underexposure and mottled

overexposure and saturation

histogram errors and sub-optimal exposure index values

histogram errors and sub-optimal exposure index values

21
New cards

One of the problems identified that technologist face with digital imaging systems and their technique selection is that:

visual cues indicating over or underexposure are no longer present on the image

the image can be easily sent to remote locations

kVp manipulation is severely limited compared to film/screen systems

automatic exposure control systems cannot be used in tandem with digital systems

visual cues indicating over or underexposure are no longer present on the image

22
New cards

The 15% rule can be used to adjust technique for changed patient thickness.

True

False

True

23
New cards

Digital imaging systems are inherently mAs driven.

True

False

False

24
New cards

It is recommended that all exposures be made using a minimum of 70 kVp for digital imaging.

True

False

False

25
New cards

The practice of using higher exposure factors than needed to obtain a diagnostic radiograph is commonly referred to as:

ALARA

Rounding up

Dose creep

Saturation 

Dose creep

26
New cards

Which of the following statements is TRUE regarding kVp’s role when using digital imaging systems?

decreasing kVp significantly increases image contrast

kVp  ensures adequate penetration of the anatomy

decreasing kVp decreases image brightness

80 kVp is now the recommended setting for virtually all exams

kVp  ensures adequate penetration of the anatomy

27
New cards

The total range of exposure intensities (signals) that a digital detector can capture and display is:

creep

dynamic range

stabilization

compensation

dynamic range

28
New cards

When an image is said to be saturated, this means that:

it has had inadequate time to dry

image noise has exceeded an optimal level

excessive overexposure has occurred

the backup timer engaged during an AEC exposure

excessive overexposure has occurred

29
New cards

The factors responsible for the phenomena of dose creep when digital imaging systems are used include:

  1. fear of mottle on the image

  2. no visible consequences for overexposure

  3. easy ability to take a repeat image


1, 2 & 3

1 & 3

2 & 3

1, 2 & 3

1, 2 & 3

30
New cards

Exposure field recognition errors could occur in which circumstance outlined below?

using more than 30 kVp than recommended for the part

imaging anatomy diagonally or not aligned with the edges on an imaging plate

setting an inadequate backup time during AEC use

activating all three cells during an AEC exposure

imaging anatomy diagonally or not aligned with the edges on an imaging plate

31
New cards

Which of the following is considered best practice with regard to setting techniques and or evaluating images when digital imaging systems are used?

increase mAs by two times for the next exposure if the image appears to bright on the monitor

to avoid a repeat exposure due to image mottle, always increase mAs

do not repeat an overexposed image if post-processing can be used

avoid collimating to ensure that all parts of the IR or detector array received remnant radiation

do not repeat an overexposed image if post-processing can be used

32
New cards

In digital imaging, ALL technique corrections and mottle prevention should only be made by changing mAs.

True

False

False

33
New cards

Which of the following is not a pre-processing step?

Normalization

Exposure field recognition

Histogram analysis

Windowing

Windowing

34
New cards

A segmentation failure can occur if:

The anatomy is too small

The image is not centered or is asymmetrical

The SID is too long

The histogram is too narrow

The image is not centered or is asymmetrical

35
New cards

To prevent segmentation errors, technologists should:

Take multiple exposures per IR

Collimate to less than 10% of the IR

Expose only one centered image per IR

Use the lowest possible kVp

Expose only one centered image per IR

36
New cards

For best practice, exposures must cover at least what percentage of the IR?

10%

20%

30%

50%

30%

37
New cards

Exposure field recognition failure is most likely when:

Exposure is made without collimation

Collimation is too tight

SID is increased

The LUT is incorrect

Exposure is made without collimation

38
New cards

VOI (Values of Interest) refers to:

The lowest and highest intensity values used for image display

The entire histogram range

The background radiation levels

The number of images on a plate

The lowest and highest intensity values used for image display

39
New cards

The purpose of a Look-Up Table (LUT) is to:

Recalculate SID

Record patient dose

Eliminate scatter

Compare the exposure histogram to a reference and adjust pixel values

Compare the exposure histogram to a reference and adjust pixel values

40
New cards

Automatic rescaling occurs when:

Exposure field borders are not visible

The mAs is doubled

The computer adjusts pixel values to match the reference histogram

The histogram is deleted

The computer adjusts pixel values to match the reference histogram

41
New cards

Selecting the wrong body part or LUT may result in:

Increased SNR

Improper brightness and contrast

Reduced patient dose

Lower image resolution only

Improper brightness and contrast

42
New cards

Which factor can distort a histogram?

Proper collimation

Adequate centering

Presence of a prosthesis

Correct exposure

Presence of a prosthesis

43
New cards

To limit histogram errors, the technologist should:

Use AEC correctly and center anatomy

Use double exposures

Over-collimate

Reduce SID below 20”

Use AEC correctly and center anatomy

44
New cards

Which of the following is a post-processing tool?

Histogram analysis

Edge enhancement

Exposure field recognition

Segmentation

Edge enhancement

45
New cards

A digital radiograph is acquired with the anatomy positioned near the edge of the detector rather than centered. Which problem is most likely to occur during pre-processing?

The system may incorrectly identify the exposure field, resulting in improper image processing.

The detector will automatically compensate for the positioning error by shifting the anatomy

The image will demonstrate increased spatial resolution because less detector area was exposed.

The system will increase the exposure index to compensate for the off-center positioning.

The system may incorrectly identify the exposure field, resulting in improper image processing.

46
New cards

During digital image processing, which sequence best represents the major steps of pre-processing?

Histogram analysis → Segmentation → Normalization → Exposure field recognition

Segmentation → Exposure field recognition → Histogram analysis → Normalization

Exposure field recognition → Normalization → Segmentation → Histogram analysis

Normalization → Histogram analysis → Exposure field recognition → Segmentation

Segmentation → Exposure field recognition → Histogram analysis → Normalization

47
New cards

A technologist fails to collimate appropriately during a digital radiographic examination. How might this affect the final image?

The system will automatically remove all scatter without affecting the histogram.

The increased exposure field will improve the accuracy of the exposure index.

The computer may include unwanted pixel data, producing incorrect brightness, contrast, or exposure index values.

The detector will automatically decrease its sensitivity to prevent histogram errors

The computer may include unwanted pixel data, producing incorrect brightness, contrast, or exposure index values.

48
New cards

A radiographic image has adequate receptor exposure but demonstrates incorrect brightness and contrast after processing. Which explanation is most consistent with a histogram analysis error?

The exposure time was too short for the computer to create a histogram.

The detector was unable to convert x-ray energy into an electronic signal.

The system correctly identified the anatomy but permanently lost all pixel data.

The software incorrectly identified the values of interest and applied inappropriate grayscale processing.

The software incorrectly identified the values of interest and applied inappropriate grayscale processing.

49
New cards

Which technologist action would best reduce the likelihood of exposure field recognition errors?

Center the anatomy within the properly collimated exposure field.

Increase mAs to ensure the software receives sufficient exposure data.

Use the largest available field size to prevent the loss of anatomical information.

Rely on post-processing to correct improperly identified exposure borders.

Center the anatomy within the properly collimated exposure field.

50
New cards

A histogram displays a large spike caused by pixel values outside the anatomy of interest. If these values are incorrectly included in the analysis, what is the most likely result?

The system will increase the number of pixels in the detector matrix.

The histogram may become skewed, affecting the grayscale representation of the anatomy.

The spatial resolution will improve because a greater range of pixel values is analyzed.

The detector will automatically decrease the exposure received by the patient.

The histogram may become skewed, affecting the grayscale representation of the anatomy.

51
New cards

During histogram analysis, the computer identifies the values of interest (VOI). What is the primary purpose of this process?

To determine whether the radiograph was acquired using an appropriate SID.

To calculate the total radiation dose absorbed by the patient.

To identify the range of pixel values representing the anatomy of interest while excluding irrelevant data.

To modify the original x-ray exposure before image acquisition is completed.

To identify the range of pixel values representing the anatomy of interest while excluding irrelevant data.

52
New cards

A technologist notices that a digital image has an incorrect exposure index even though the selected exposure factors were appropriate. Which situation could explain this finding?

The software incorrectly identified the collimated exposure field and included inappropriate data in its analysis.

The technologist selected a higher kVp, which always causes an inaccurate exposure index.

The radiograph was displayed using magnification, permanently altering the detector exposure.

The technologist adjusted the window level after acquisition, changing the original detector signal.

The software incorrectly identified the collimated exposure field and included inappropriate data in its analysis.

53
New cards

Two radiographs of the same anatomy are acquired using slightly different exposure factors, yet both demonstrate similar displayed brightness and contrast. Which process best explains this result?

The detector automatically adjusts the radiation output before each exposure.

The computer modifies the original exposure factors after image acquisition.

Automatic rescaling and LUT processing adjust pixel values to produce a consistent image appearance.

The computer increases the number of pixels to compensate for differences in exposure.

Automatic rescaling and LUT processing adjust pixel values to produce a consistent image appearance.

54
New cards

A technologist selects the incorrect anatomical examination before acquiring a digital radiograph. Why might this affect the processed image?

The system will automatically reduce the x-ray beam intensity to match the selected anatomy.

The detector will change its physical pixel size to accommodate the incorrect exam.

The computer will be unable to record any exposure information from the detector.

An inappropriate examination-specific LUT may be applied, resulting in incorrect grayscale mapping.

An inappropriate examination-specific LUT may be applied, resulting in incorrect grayscale mapping.

55
New cards

During automatic rescaling, the acquired exposure histogram is compared with a reference histogram. What occurs when the two do not match?

Pixel values are adjusted to better match the expected grayscale representation of the selected examination.

The original x-ray exposure is repeated automatically using corrected exposure factors.

The detector adds anatomical structures that fall outside the reference histogram.

The computer increases spatial resolution until the acquired and reference histograms match.

Pixel values are adjusted to better match the expected grayscale representation of the selected examination.

56
New cards

Why can digital radiographs appear acceptable despite variations in the selected kVp and mAs?

Digital detectors eliminate the relationship between exposure factors and receptor exposure.

Image processing can maintain a consistent appearance when exposure remains within an acceptable range.

The computer automatically changes the amount of radiation that originally reached the patient.

The LUT eliminates quantum noise regardless of the amount of radiation reaching the detector.

Image processing can maintain a consistent appearance when exposure remains within an acceptable range.

57
New cards

Which statement best describes a limitation of automatic rescaling in digital radiography?

It can correct positioning errors as long as the anatomy remains within the detector.

It can recover anatomical information lost because of insufficient x-ray penetration.

It can maintain consistent brightness and contrast, but it cannot compensate for every exposure or acquisition error.

It prevents exposure index errors by eliminating the need for exposure field recognition.

It can maintain consistent brightness and contrast, but it cannot compensate for every exposure or acquisition error.

58
New cards

A technologist adjusts the window width of a digital radiograph. Which image characteristic is primarily affected?

The amount of radiation originally received by the detector.

The overall brightness of the displayed image.

The physical dimensions of the detector matrix.

The range of grayscale values displayed, affecting image contrast.

The range of grayscale values displayed, affecting image contrast.

59
New cards

A technologist applies edge enhancement to a digital radiograph. What is the intended effect?

To increase the exposure received by the detector and reduce quantum mottle.

To emphasize boundaries between structures, making edges appear sharper.

To increase the number of pixels within the image matrix.

To correct anatomical distortion caused by improper positioning.

To emphasize boundaries between structures, making edges appear sharper.

60
New cards

A digital image appears too bright immediately after processing, but the anatomy is adequately penetrated. Which action is most appropriate before considering a repeat exposure?

Increase mAs and repeat the exposure to produce a darker image.

Increase kVp and repeat the exposure to improve detector sensitivity.

Evaluate the image and use appropriate window settings rather than repeating solely because of its displayed brightness.

Repeat the image using a smaller focal spot to correct the brightness.

Evaluate the image and use appropriate window settings rather than repeating solely because of its displayed brightness.

61
New cards

A technologist routinely increases mAs because digital radiographs continue to appear acceptable even when higher-than-necessary exposures are used. Which concern does this practice illustrate?

Decreased detector dynamic range caused by repeated exposures.

Reduced patient dose resulting from increased exposure latitude.

Improved contrast resolution caused by increased receptor exposure.

Dose creep caused by unnecessarily high exposure factors being masked by digital processing.

Dose creep caused by unnecessarily high exposure factors being masked by digital processing.

62
New cards

A radiographic technique of 75 kVp at 24 mAs is adjusted using the 15% rule. Which new technique would approximately maintain receptor exposure?

86 kVp at 12 mAs.

86 kVp at 24 mAs.

75 kVp at 12 mAs.

64 kVp at 12 mAs.

86 kVp at 12 mAs.

63
New cards

A digital radiograph demonstrates areas of anatomy that appear dark and burned out because excessive exposure exceeded the detector's ability to preserve useful information. What is the most appropriate interpretation?

The image demonstrates quantum mottle caused by insufficient exposure.

The image demonstrates a positioning error that can be corrected through image inversion.

The image may demonstrate detector saturation, resulting in loss of anatomical information.

The image demonstrates normal automatic rescaling and requires no further evaluation.

The image may demonstrate detector saturation, resulting in loss of anatomical information.

64
New cards

When developing a digital technique chart, which approach is emphasized in the lecture as a method of reducing patient exposure while maintaining adequate penetration?

Use the lowest available kVp and increase mAs to compensate.

Use the same exposure factors for all anatomical regions to maintain consistency.

Increase mAs whenever possible because digital systems have wide exposure latitude.

Use an appropriately higher kVp with reduced mAs while considering the examination and patient.

Use an appropriately higher kVp with reduced mAs while considering the examination and patient.

65
New cards

DQE is used:

To calculate the median average value of a central pixel

To measure a phosphor’s ability to convert x-ray energy to an output signal

To determine an image’s exposure index – EI value after exposure

As a measure of the image receptor dose when DR detectors are used.

To measure a phosphor’s ability to convert x-ray energy to an output signal

66
New cards

In general, if an imaging receptor system has a higher DQE:

Higher kVp’s should be used with higher mAs

Image noise will be increased

More mAs should be used

Lower technical factors are required

Lower technical factors are required

67
New cards

The EI values used in all digital imaging systems are uniform among all manufacturers.

True

False

False

68
New cards

The term proposed by the International Electro-technical Commission(IEC) for use across all manufacturers as an exposure indicator system is:

Sensitivity number

DICOMM Value

Exposure quantum

Exposure Indices & Deviation Indices (EI and DI numbers)

Exposure Indices & Deviation Indices (EI and DI numbers)

69
New cards

Using the recommended deviation index system:  a DI value of +1represents an exposure to the imaging plate that was:

100% too high

50% too high

26% too high

25% too low

50% too low

26% too high

70
New cards
<p><span><strong>For this question consider the following:</strong></span></p><ul><li><p><span><strong>&nbsp;an IDEAL&nbsp; EI#&nbsp; is&nbsp; 800&nbsp; &nbsp; &nbsp;and&nbsp; &nbsp;</strong></span></p></li><li><p><span><strong>an OPTIMAL DI#&nbsp; is 0.0</strong></span></p></li><li><p><span><strong>DI# guidelines: any image with DI# below -3.0&nbsp; or above +3.0 must be repeated.</strong></span></p></li></ul><p>&nbsp;</p><p>Exposure: 90kVp @ 80mAs&nbsp; with EI# 1600 and DI#&nbsp; +3.0&nbsp;</p><p>What will be your new mAs to achieve an optimal exposure level on your next image?</p><p><strong>90kVp&nbsp; @ ?__ mAs</strong></p><p></p><p>40 mAs</p><p>60 mAs</p><p>80 mAs</p><p>160 mAs</p>

For this question consider the following:

  •  an IDEAL  EI#  is  800     and   

  • an OPTIMAL DI#  is 0.0

  • DI# guidelines: any image with DI# below -3.0  or above +3.0 must be repeated.

 

Exposure: 90kVp @ 80mAs  with EI# 1600 and DI#  +3.0 

What will be your new mAs to achieve an optimal exposure level on your next image?

90kVp  @ ?__ mAs


40 mAs

60 mAs

80 mAs

160 mAs

40 mAs

71
New cards
<p><span><strong>For this question consider the following:</strong></span></p><ul><li><p><span><strong>&nbsp;an IDEAL&nbsp; EI#&nbsp; is&nbsp; 800&nbsp; &nbsp; &nbsp;and&nbsp; &nbsp;</strong></span></p></li><li><p><span><strong>an OPTIMAL DI#&nbsp; is 0.0</strong></span></p></li><li><p><span><strong>DI# guidelines: any image with DI# below -3.0&nbsp; or above +3.0 must be repeated.</strong></span></p></li></ul><p>&nbsp;</p><p>Exposure: 60kVp @ 2.5mAs&nbsp; with EI# 550 and DI#&nbsp; -2.0&nbsp;</p><p>What will be your new mAs to achieve an optimal exposure level on your next image?</p><p><strong>60kVp&nbsp; @ ?__ mAs</strong></p><p>&nbsp;</p><p>1.4 mAs</p><p>2.5 mAs</p><p>3.5 mAs</p><p>5.0 mAs</p>

For this question consider the following:

  •  an IDEAL  EI#  is  800     and   

  • an OPTIMAL DI#  is 0.0

  • DI# guidelines: any image with DI# below -3.0  or above +3.0 must be repeated.

 

Exposure: 60kVp @ 2.5mAs  with EI# 550 and DI#  -2.0 

What will be your new mAs to achieve an optimal exposure level on your next image?

60kVp  @ ?__ mAs

 

1.4 mAs

2.5 mAs

3.5 mAs

5.0 mAs

3.5 mAs

72
New cards
<p><span><strong>For this question consider the following:</strong></span></p><ul><li><p><span><strong>&nbsp;an IDEAL&nbsp; EI#&nbsp; is&nbsp; 800&nbsp; &nbsp; &nbsp;and&nbsp; &nbsp;</strong></span></p></li><li><p><span><strong>an OPTIMAL DI#&nbsp; is 0.0</strong></span></p></li><li><p><span><strong>DI# guidelines: any image with DI# below -3.0&nbsp; or above +3.0 must be repeated.</strong></span></p></li></ul><p>&nbsp;</p><p>Situation:&nbsp; You completed a Lateral L-Spine using&nbsp; 90kVp @ 50 mAs.&nbsp; But the EI# is 250 and the DI# is -3.0.</p><p>What would be the appropriate change to mAs on the repeated exposure to reach an optimal exposure level?</p><p>90 kVp&nbsp; @<strong> [</strong><span><strong><u>___?_____]</u></strong></span><strong> mAs</strong></p><p><strong><em>Enter the numerical value only for your new mAs</em></strong></p>

For this question consider the following:

  •  an IDEAL  EI#  is  800     and   

  • an OPTIMAL DI#  is 0.0

  • DI# guidelines: any image with DI# below -3.0  or above +3.0 must be repeated.

 

Situation:  You completed a Lateral L-Spine using  90kVp @ 50 mAs.  But the EI# is 250 and the DI# is -3.0.

What would be the appropriate change to mAs on the repeated exposure to reach an optimal exposure level?

90 kVp  @ [___?_____] mAs

Enter the numerical value only for your new mAs

75 (with margin: 1)

73
New cards
<p><span><strong>For this question consider the following:</strong></span></p><ul><li><p><span><strong>&nbsp;an IDEAL&nbsp; EI#&nbsp; is&nbsp; 800&nbsp; &nbsp; &nbsp;and&nbsp; &nbsp;</strong></span></p></li><li><p><span><strong>an OPTIMAL DI#&nbsp; is 0.0</strong></span></p></li><li><p><span><strong>DI# guidelines: any image with DI# below -3.0&nbsp; or above +3.0 must be repeated.</strong></span></p></li></ul><p>&nbsp;</p><p>&nbsp;</p><p>Exposure: 70kVp @ 25mAs&nbsp; with EI# 1400 and DI# +1.0.&nbsp;</p><p>What will be your new mAs to achieve an optimal exposure level on your next image?</p><p><strong>70kVp&nbsp; @ [___?___] mAs</strong></p><p><strong><em>Enter the numerical value only for your new mAs</em></strong></p>

For this question consider the following:

  •  an IDEAL  EI#  is  800     and   

  • an OPTIMAL DI#  is 0.0

  • DI# guidelines: any image with DI# below -3.0  or above +3.0 must be repeated.

 

 

Exposure: 70kVp @ 25mAs  with EI# 1400 and DI# +1.0. 

What will be your new mAs to achieve an optimal exposure level on your next image?

70kVp  @ [___?___] mAs

Enter the numerical value only for your new mAs

17 or 17.5 (with margin: 1)

74
New cards
<p><span><strong>For this question consider the following:</strong></span></p><ul><li><p><span><strong>&nbsp;an IDEAL&nbsp; EI#&nbsp; is&nbsp; 800&nbsp; &nbsp; &nbsp;and&nbsp; &nbsp;</strong></span></p></li><li><p><span><strong>an OPTIMAL DI#&nbsp; is 0.0</strong></span></p></li><li><p><span><strong>DI# guidelines: any image with DI# below -3.0&nbsp; or above +3.0 must be repeated.</strong></span></p></li></ul><p>&nbsp;</p><p>Exposure: 90kVp @ 45mAs&nbsp; with EI# 1400 and DI#&nbsp; +2.0&nbsp;</p><p>What will be your new mAs to achieve an optimal exposure level on your next image?</p><p><strong>90kVp&nbsp; @ ______ mAs</strong></p>

For this question consider the following:

  •  an IDEAL  EI#  is  800     and   

  • an OPTIMAL DI#  is 0.0

  • DI# guidelines: any image with DI# below -3.0  or above +3.0 must be repeated.

 

Exposure: 90kVp @ 45mAs  with EI# 1400 and DI#  +2.0 

What will be your new mAs to achieve an optimal exposure level on your next image?

90kVp  @ ______ mAs

27 (with margin: 1)

75
New cards

Each digital imaging receptor manufacturer uses the SAME Exposure Indicator numeric system. 

So we can simply use 800 as the ideal EI# for imaging plates in all clinical sites.

True

False

False

76
New cards

The method of using the DI# to adjust your exposure is currently still in development, but as a general rule you are best to use this DI# as a starting point for correcting technique errors when repeating an image is needed.

True

False

True

77
New cards

Doubling the mAs for the DIGITAL image receptor will result in increased brightness of the displayed digital image after processing. 

True

False

False

78
New cards

Manipulating mAs for DIGITAL imaging affects the IR exposure.

True

False

True

79
New cards

A key advantage of digital imaging compared to film/screen is:

Lower resolution

No need for collimation

Ability to adjust image after exposure

Permanent image appearance

Ability to adjust image after exposure

80
New cards

The histogram represents:

Patient dose

Radiation exposure values received by the detector

Film density range

SID versus mAs relationship

Radiation exposure values received by the detector

81
New cards

Latitude in digital imaging refers to:

The detector’s geographic position

The ability to display shades of gray

The range of exposures that produce a diagnostic image

The degree of collimation

The range of exposures that produce a diagnostic image

82
New cards

Dose-Area Product (DAP) measures:

Detector sensitivity

Image brightness

Patient dose and irradiated field size

Exposure index calibration

Patient dose and irradiated field size

83
New cards

DAP increases when:

Collimation is tighter

mAs is reduced

SID is doubled

Field size increases at same technique

Field size increases at same technique

84
New cards

DAP is measured using:

A chamber at the image receptor

An ionization chamber near the collimator

A photodiode inside the monitor

A dosimeter on the technologist

An ionization chamber near the collimator

85
New cards

 In a logarithmic EI scale, a change of 300 EI corresponds to:

Doubling or halving mAs

Doubling kVp

15% change in kVp

Halving kVp

Doubling or halving mAs

86
New cards

In an inversely proportional EI system, a high “S” number means:

Proper exposure

Overexposure

Underexposure

Increased contrast

Underexposure

87
New cards

In a proportional EI system: 

Low EI means overexposure

High EI means underexposure

EI changes have no meaning

High EI means overexposure

High EI means overexposure

88
New cards

The Deviation Index (DI) represents:

Total patient dose

The difference between actual and target detector exposure

The brightness of the image

Histogram tail value

The difference between actual and target detector exposure