1/87
Looks like no tags are added yet.
What primarily causes Scatter radiation?
Compton effect
Somewhat coherent scatter
What is contrast?
degree of difference in OD between areas of an image
What would only be in an ideal image?
Transmittance and absorption
What does scatter do?
Reduce image contrast by exposing parts of film
What are the main factors that contribute to scatter radiation?
kVp (energy)
Patient thickness
X-ray field size
What increases both Compton and photoelectric effect?
Energy (kVp)
T/F: Photoelectric decreases much faster than compton?
True
Is there more compton or photoelectric at higher energies?
Compton
Why would we ever raise the kVp?
Patient dose
Transmittance
What occurs to patient dose is kVp is to low?
Increases
What occurs to transmittance if kVp is to low?
Not enough x-rays will go through to create image
What happen to scatter radiation as the field size increases?
Increases
What is the most common beam-restricting device?
Light-localizing variable aperture collimator
What is the first part that a collimator controls?
off-focus radiation
What is the benefit of a adjustable collimator?
Variable x-ray beam size
How thick are the lead leaves in a collimator?
3mm
What are the devices used to reduce scatter?
Beam restricting devices
Radiographic grids
What are beam-restricting devices?
Effect what reaches the patient
What do radiographic grids effect?
Remnant beam (after patient)
How much scatter can a high-quality grid reduce?
80-90%
What is the interspace material in a grid made of?
Aluminum or plastic
What are the important dimensions of a grid?
Height of grid
Thickness of strips
Width of interspace material
What determines the grid ratio?
Height of grid by the interspace width
Which are the more effective grids?
High ratio
What is the downside to a higher ratio grid?
More radiation exposure necessary
What ratio do most systems use?
8:1 to 12:1
Which grid will clean up 85% of scatter?
5:1
Which grid will clean up near 97%?
16:1
What is the range of grid rations?
5:1 to 16:1
What is grid frequency?
Grid strips per centimeter
What range of frequencies do most grids have?
25-45 lines per centimeter
What are the results of using high-frequency grids?
Less distinct grid lines on radiograph
Requires high exposure, higher patient dose
What is the principal function of a grid?
Improve image contrast by absorbing scatter
What is the contrast improvement factor? (k)
image contrast with grid / image contrast without grid
What value in contrast improvement factor indicates no improvement?
1
How does a higher grid ratio and grid frequency affect contrast improvement factor?
Increase
What must be increased to maintain OD when using a grid?
mAs
What is the Bucky factor? (B)
The amount of increase is expressed
Higher the grid ratio or grid frequency, higher the B
What is another name for the Bucky factor?
Grid factor
What is the equation for the Bucky factor?
Patient dose grid / patient dose without grid
What is the simplest type of grid?
Parallel
What is a result of decreased SID?
More potential for grid cutoff
What can also influence grid cutoff?
X-ray field size
What occurs if there is short SID?
OD decreases towards the edge of image receptor
What is the goal of focused grids?
Minimize grid cutoff
How are strips aligned in a focused grid?
With divergence of x-ray beam
What is the purpose of moving grids?
Eliminate grid lines
What kinds of grids will have less noticeable lines?
Higher frequency
What is the name for the moving grid?
Potter-Bucky Diaphragm
What is the largest problem with grids?
Misalignment
What can cause misalignment of a grid?
Off center
Off level
Off focus
Upside down
What is the air-gap technique?
Increase OID (object to image receptor distance)
What grid is the air gap technique equivalent to?
8:1
What is the main disadvantage to air-gap technique?
Image magnification
What still needs to be increased with air-gap technique?
mAs
At what rate does mAs need to be increased with air-gap technique?
10% for every cm
When was computed radiography introduced?
1980s by Fujifilm Medical
What other names did the computed radiograph have?
“Photostimulable Phosphor Plate Radiography”
PSP plate radiography
What is photostimulable luminescence?
Receiving x-rays and storing the energy as trapped electrons, then releasing them as light when exposed to a different light source
How is Photostimulable luminescence accomplished?
Photostimulable phosphors (PSPs) in the plate
What are storage phosphor screens?
Latent image is stored for a period
What needs to occur once an image plate is exposed and the latent image is formed?
Stimulated to obtain the manifest image
What causes the PSP to emit light?
Finely focused laser beam
In what light region does the PSP emit light?
Blue
What is used to clear the image plate to prepare for a new image?
Intense white light
What contains the laser photomultiplier tubes and photodiodes?
Computed radiography reader
What does the Computed radiography reader do?
Removes PSP plate, scan it, erase it, and replace it
When was digital radiography introduced?
1980s by Fujifilm Medical after CR
What does digital radiography not have compared to computed radiography?
No film
No PSP
What does digital radiography use?
Flat panel detectors (FPDs)
What is direct digital radiography?
Converts x-ray photons into electrical signal
What is indirect digital radiography?
Converts x-ray photons into visible light, then into an electrical charge
What are charge-coupled device detectors? (CCDs)
Use indirect conversion of x-ray photons into visible light and then into electrical charge
What do direct flat panel detectors use?
Amorphous selenium photoconductor to convert x-rays into electrical signals
What processes the electrical signals in direct flat panel detectors?
Thin film transistor (TFT)
What kind of flat panel detectors are most commonly used in mammography where high resolution is required?
Direct
What does indirect flat panel detectors use?
Phosphor layer for producing visible light, which is read by a photodiode
What is a downside to CCD detectors?
Potential to produce high amount of image noise due to demagnification process
What is bit depth?
System is the number of discrete sampling levels used to interpret and convert analog signal into a digital one
What does a high bit depth allow for?
More accurate representation of the incoming signal
Requires more memory
What represents the number of bit depth levels?
2^n
n = image bit size
2^16 (bit) = 65, 536 levels
Why is a small bit ranges beneficial?
Accuracy
What is another way to make bits smaller?
Shrink total range of energy values
Reduces dynamic range
What is the dynamic range of detector?
Range of exposure values over which the signal will be properly digitized
What will result from a low dynamic range?
Limit range of exposure values able to be represented
What will result from a high dynamic range?
Each bit covering a greater exposure value range
What is a DICOM image file
Standardized image file type for medical imaging
Digital Imaging and Communications in Medicine
What is the filename extension for a DICOM file?
.dcm