Flat-Panel Detector (FPD) Digital Systems are also known as:
Digital Radiography (DR)
2 Types of Flat Panel Detectors
Direct Conversion Detectors
Indirect Conversion Detectors
Direct Conversion Detectors
Use a photoconductor, which is capable of converting x-rays into an electronic signal.
Indirect Conversion Detectors
Use a scintillator, which converts incoming x-ray photons to light.
Indirect Conversion has 2 steps:
Convert incoming x-ray photons to light first, which is accomplished through the use of a scintillator.
Convert light photons to an electronic signal using a photodetector.
Indirect DR systems include either a:
Charged-coupled device (CCD)
Amorphous silicon with a thin film transistor (TFT) array
Direct DR systems without a scintillation/light conversion step have a _______________.
Higher DQE
DQE (Detective Quantum Efficiency)
% of incoming x-ray photons that are detected and absorbed by the IR for transformation to the x-ray image.
High DQE
Detector uses x-rays efficiently to produce a detailed image with less radiation (most desirable).
Low DQE
Detector “wastes” x-rays and requires more dose to get the same clarity.
Direct conversion uses _____________________ and a TFT.
Amorphous Selenium
Both types of indirect detectors need a __________ to change the x-ray photons to light.
Scintillator
The main difference between TFT detectors and CCD in indirect DR is the way the light is converted to ______________.
An electrical signal
Scintillators emit light _____________ which can reduce spatial resolution.
Isotropically
Because the "glow" of light spreads out, the detector can't always tell precisely where each x-ray landed. This ____________ the image sharpness
Softens/decreases
Indirect TFT requires a scintillator such as ___________________, or _________________ to emit light. Which is more commonly used today?
Cesium iodine (CsI)
Gadolinium oxysulfide (GdOD)
CsI
Once the light reaches the __________________ for detection, it is converted into electrical charges that are then deposited on a TFT.
Amorphous silicon (a-Si)
TFT collects the electric charges produced by either the selenium or silicon as an array or matrix of pixel-size ____________________.
Detector elements (DEL)
Detector Elements (DEL)
Receives exposure and collects signal. Smallest sensing unit in a digital x-ray detector.
Each DEL pixel is the same size and measures around _______, about the width of a human hair.
100-200 microns (μm)
DEL pixels and screen pixels are ______________.
Different
DEL pixel lives in the ___________ while screen pixels live in the _______________.
Detector
Display Monitor
TFTs act like buckets to _______________ electricity so the computer can make the picture.
Collect and move
The TFT acts like a ______. When open, the stored charge in the DEL flows out into the ____________. When closed, the charge _____________.
Switch
Readout electronics
Stay put
Each pixel detector element (DEL) includes the ______________, the _____________, and the ________________.
Sensing area
Capacitor
TFT
The _________ refers to the sensing area compared to the non-sensing electronics area.
Fill factor
______ is the fill factor used in modern DELs.
80/20
As fill factor % goes up, spatial resolution (sharpness) and contrast resolution (subtle differences in gray) will ___________.
Go up
Better resolution is achieved with a __________________ paired with a _________________.
Smaller DEL size
Greater fill factor
95% fill factor provides ________________________ than 75% fill factor.
More spatial resolution
Smaller DEL size of 100 µ (mammo) provides _____________________________.
More spatial resolution
Larger DEL size of 200μ (x-ray) provides _____ spatial resolution.
Less
CCD (Photodetector)
Sensor chip that is capable of converting visible light into an electric charge and storing it (no TFT involved).
The light produced (by a scintillator) from x-ray interactions must be optically coupled to the CCD sensor chip by ______ or _________.
Lenses
Fiber optics
Indirect TFT
“Bucky Brigade”
Detector itself collects the lgith and converts it to electrical charge right in each pixel
More common system in diagnostic radiography.
Indirect CCD
“Mini Camera”
Light is sent to a separate “camera chip” to capture the signal.
Older system - no common anymore
_______________________ may be used in place of the CCD as a photodetector.
CMOS (Complimentary Metal Oxide Semi-Conductors)
____ is slower, more expensive, and consumes more power but has better image quality and higher DQE.
CCD
_____ is faster, cheaper, and consumes less power but has lesser image quality and lower DQE.
CMOS
Direct Conversion Systems
Directly convert incoming x-ray photons to an electronic digital signal.
The photoconductor used in direct conversion systems is _________________.
Amorphous selenium (a-Se)
In a direct conversion system, ionization caused by the x-ray photons results in the selenium atoms _________ for collection.
Freeing electrons
Direct conversion is more common in ______ due to sharp detail capability.
Mammo
Image Acquisition
Both direct and indirect resulting data is the latent image that needs to be read out.
Processing
Only the exposed DELs are read. 2 Step process
Pre-processing
Post-Processing
The first step in DR processing is pre-processing, which _________________________.
Corrects for any faulty DELs
The second step in DR processing is post-processing, which _______________________ before it is displayed on the monitor.
Optimizes image contrast, brightness, and detail
DR detectors have 2 general classifications:
Portable detectors
Integral detectors (built-in to bucky)
Most of today's FPDs are _________ and use TFTs with non-tethered wi-fi technology.
Portable
Some FPDs can withstand weights of _____.
500 lbs
The average cost of a FPD is ______.
$25k-50k
Which of the following are used as scintillors in Indirect DR systems? (select all that apply)
Cesium Iodine
Amorphous Silicon
Amorphous Selenium
Gadolinium Oxysulfide
Cesium Iodine
Gadolinium Oxysulfide
Which DR system uses amorphous selenium as a photoconductor?
Direct DR - TFT
Image Resolution is improved with selection of: (6)
Smaller x-ray tube focal spots
Shorter OIDs
Longer SIDs
Shorter exposure times with higher mA selections
Grids (over 80 kVp)
Tighter collimation
In digital imaging, kVp and contrast are ____________, and mAs does not affect ___________.
Not directly related
Image brightness
Final digital image quality is essentially a function of ___________.
Total exposure to the detector.
Higher ____ values are now recommended with digital detectors.
kVp
mAs exposures that are too low will result in ________________.
Quantum mottle/Noise
IR exposure changes should be at least __________ to produce a visible change.
25-33%
How should you change exposure factors for an additive condition?
5-15% increase in kVp
How should you change exposure factors for an destructive condition?
25-50% decrease in mAs
Fixed kVp Systems
Use an optimal kVp that is not changed and variable mAs.
Variable kVp Systems
Use an optimal mAs that is not changed and variabel kVp.
Both kVp systems will work, but ______________ are better suited for use with digital image receptors.
Fixed kVp systems
S# (Fuji, Konica, Phillips)
Range 150-250
Higher = underexposed
Lower = overexposed
EI# (Carestream, Siemens)
Range 1500-1800
Higher = overexposed
Lower = underexposed
Exposure Index (EI#)
Acutal exposure
Target EI (EIt)
Ideal exposure
Deviation Index (DI)
% deviation from target. DI value of 2 or more needs to be evaluated for a repeat.
Contrast/brightness is not an issue when evaluating an image because __________________ will fix it
Rescaling and LUT
What are the two most common issues found when evaluating an image?
Poor positioning
Noise
Because noise is a common issue in digital radiography, we tend to overcompensate with ________ and allow the computer to correct the image. This leads to ________.
Higher mAs
Dose creep
A DI of 3 is ____% too high and indicates _____________. Is a repeat warranted?
100
Excessive overexposure
Repeat only if there is burn-out.
A DI of -3 is ___% too low and indicates _____________. Is a repeat warranted?
50
Excessive underexposure
Repeat required due to excessive noise.
EIt-EI#= ___________
Deviation Value
Deviation Value/EIt= _________
% Deviation
Data Drop
Extreme overexposure in which DELs are overwhelmed with photon energy and become incapable of recognizing high-energy values.
Images with data drop will demonstrate _____ contrast.
Poor
Detector Saturation
Data-drop throughout an entire region of the detector and loss of diagnositc info or mimic of destructive pathology.
Digital Processing Latitude
Range of x-ray exposures that a digital detecor can capture and still produce a usable image.
A wide digital processing latitude means the system is ______ forgiving and makes it ______ to get a good image.
More
Easier
Benefits of increased latitude include the ability to image structures of ____________ without loss of visibility and increased ____________ in under or overexposure.
Varying attenuation values
Margin for error
Increased latitude maintains a useful __________.
Image brightness
Be wary of ______________, ___________________, and ______________with wide latitude systems.
Underexposed quantum mottle
Overexposed flattened saturation
Dose creep
A deviation index value of 0.5 indicates:
Nearly optimum exposure.
In what situation are you most likely to see noise on your DR radiograph? (select all that apply)
When you did not decrease technique for a smaller patient.
When you did not increase technique for a larger patient.
Underexposure
Overexposure
When you did not increase technique for a larger patient.
Underexposure
Advantages of Digital X-Rays (3)
Digital images can be manipulated without re-exposure to patients.
Require less physical storage.
Can be sent to remote locations.
Computer system consists of:
Hardware
Software
Hardware
Physical component of the computer; consists of input and output devices to allow info to be directed outside the computer.
Software
Set of instructions or program to operate the computer.
CPU (Central Processing Unit)
Primary control center of the computer which includes memory.
2 Types of Memory
RAM (Random Access Memory)
ROM (Read Only Memory)
RAM (Random Access Memory)
Volatile, losing info when computer is off.
ROM (Read Only Memory)
Hardwired to computer, stays in computer even when off.
Any given radiology facility may perform __________ examinations per year, requiring ______ of computer memory. Long-term storage may require ______.
150,000
3.2 TB
10-16 TB
Network
2 or more objects sharing resources and information.
Server
A centralized computer managing network resources; stores, sends, and receives data.
Informatics
The science of processing data for storage and retrieval; information science.
Clinical Informatics (BMI)
Utilizes data and information technology to deliver healthcare services.
Electronic Health Record (EMR)
Electronic medical records.